(113 days)
The LenSx 550 Laser System is indicated for anterior capsulotomy and laser phacofragmentation during cataract surgery. The anterior capsulotomy and phacofragmentation procedures may be performed either individually or consecutively during the same surgery.
The LenSx 550 Laser is an ophthalmic surgical laser that has previously been cleared for use in anterior capsulotomy during cataract surgery (K082947). The LenSx 550 generates femtosecond laser pulses that are scanned in a three-dimensional pattern in the eye. Localization of these laser pulses is accomplished by an aiming beam that identifies the surfaces of the lens, and an off-the-shelf, disposable contact lens and suction ring assembly that fixates the eye relative to the delivery system. As described in K082947, anterior capsulotomy is performed by delivering a cylindrical pattern of laser pulses to intersect the anterior lens capsule.
The same surgical device, with only minor software changes, is now also being proposed for use in phacofragmentation during cataract surgery. As described in this premarket notification, laser phacofragmentation divides the lens into quadrants, in the same way that division of the lens into quadrants is performed with standard ultrasound phacofragmentation in the initial step of phacoemulsification.
To perform phacofragmentation, the LenSx system delivers a series of laser pulses to form two intersecting ellipsoidal planes. Photodisruption of the lens tissue in this pattern, which appears as a cross from a surgical (top) view, segments the nucleus into four sections before the eye is physically entered by any instruments or devices.
Following entry into the eye and then into lens, the fragmented nuclear quadrants created by the LenSx 550 are further fragmented and removed using a standard phacoemulsification probe.
Here's a breakdown of the acceptance criteria and study information for the LenSx 550 Laser System, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria (e.g., a specific percentage reduction in ultrasound power, or a target accuracy for laser pulse placement). Instead, the performance is described in comparative and qualitative terms.
| Performance Metric | Acceptance Criteria (Implicit from study design) | Reported Device Performance |
|---|---|---|
| Phacofragmentation Accuracy & Reproducibility | High degree of accuracy and reproducibility for laser pulse placement in plastic substrates. Confirmed accuracy and reproducibility of localized laser tissue effects in ex vivo porcine eyes. | Demonstrated a high degree of accuracy and reproducibility for laser pulse placement in plastic substrates. Further confirmed the accuracy and reproducibility of the localized laser tissue effects in ex vivo porcine eyes. |
| Reduction in Ultrasound Power | Significantly reduced ultrasound power required to complete phacofragmentation compared to standard methods with a predicate device (Alcon Series 20000 Legacy). | The study group (using laser phacofragmentation) required significantly less manual manipulation and ultrasound power to achieve the desired phacofragmentation effect compared to the control group (standard ultrasonic phacoemulsification). |
| Clinical Outcomes | Excellent clinical results with no significant adverse events, comparable to advanced phacoemulsification techniques reported in previous studies. | Clinical results were excellent in both the laser phacofragmentation and control groups, with no significant adverse events. Results in both groups were consistent with those of advanced phacoemulsification techniques reported in previous studies. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Clinical Study - Test Set): A total of 92 subjects, divided into two groups of 46 eyes each.
- Study Group: 46 eyes
- Control Group: 46 eyes
- Data Provenance: Prospective clinical study performed outside the U.S. (specific country not mentioned).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts or their qualifications for establishing the ground truth of the clinical outcomes or the success of phacofragmentation. The clinical study evaluated performance, and "clinical results were excellent in both groups, with no significant adverse events," implying expert assessment by the surgical team or researchers involved, but details are not provided.
4. Adjudication Method for the Test Set
The document does not describe any explicit adjudication method (e.g., 2+1, 3+1). The "clinical results were excellent" implies a direct observation and assessment by the surgical team or lead investigators, rather than a formal adjudication process post-facto by a separate panel.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted in the traditional sense of evaluating multiple human readers' diagnostic performance. This was a clinical comparative effectiveness study comparing a new surgical technique (LenSx laser phacofragmentation) with a standard technique (ultrasonic phacoemulsification).
- Effect Size: The study found that the laser phacofragmentation group required "significantly less manual manipulation and ultrasound power." The document states "significantly reduced" but does not provide a specific quantitative effect size (e.g., a percentage reduction or a statistical measure like Cohen's d) for the improvement in ultrasound power or manual manipulation.
6. Standalone (Algorithm Only) Performance Study
Yes, aspects of standalone performance were evaluated through pre-clinical studies:
- Plastic substrates: Demonstrated "a high degree of accuracy and reproducibility for laser pulse placement."
- Ex vivo porcine eyes: Further confirmed "the accuracy and reproducibility of the localized laser tissue effects."
These pre-clinical studies assessed the device's inherent capability to perform the intended physical action (laser phacofragmentation) without human intervention in the fragmentation process itself (though a human would operate the device).
7. Type of Ground Truth Used
- Pre-clinical studies (plastic substrates, ex vivo porcine eyes): The ground truth was based on physical observation and measurement of laser pulse placement accuracy, reproducibility, and localized tissue effects.
- Clinical study:
- Effectiveness of phacofragmentation: Ground truth implicitly based on surgical observation of the extent of fragmentation, and measured reduction in ultrasound power needed for subsequent phacoemulsification.
- Safety: Ground truth based on clinical observation of adverse events.
- Clinical outcomes: Ground truth based on overall clinical assessment by the medical professionals involved, and comparison to "advanced phacoemulsification techniques reported in previous studies."
8. Sample Size for the Training Set
The document does not specify a separate training set in the context of machine learning. The studies described are performance evaluations of the device's hardware and software logic. The device itself (LenSx 550 Laser System) was previously cleared (K082947) for anterior capsulotomy, implying its core technology was already established. The current submission focuses on a new indication (phacofragmentation) with "only minor software changes." The pre-clinical studies and clinical studies described are for validation of the new indication, not for training a specific AI model.
9. How the Ground Truth for the Training Set Was Established
As no specific AI training set or process is described, this information is not applicable based on the provided text. The device's operation appears to be based on pre-programmed patterns and optical localization rather than a continuously learning AI model that requires a ground-truthed training set for its development.
{0}------------------------------------------------
APR 2 3 2010
510(K) SUMMARY
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).
SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED
| a. | Applicant: | LenSx Lasers, Inc.33 Journey, Suite 175Aliso Viejo, CA 92656Tel: (949) 360-6010Fax: (949) 360-6028 |
|---|---|---|
| ---- | ------------ | -------------------------------------------------------------------------------------------------------------------- |
Judy Gordon, D.V.M. Contact Person: b. ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, CA 92651 judy@clinregconsulting.com Tel: (949) 715-0609 Fax: (949) 715-0610
NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME
| a. Trade/Proprietary Name: | LenSx 550 Laser System |
|---|---|
| b. Common/Usual Name: | LenSx 550 Laser System |
| c. Classification Name: | Laser Instrument, Surgical, Powered |
| d. Classification Code(s): | 21 CFR 886.4390 ; 79 OOE |
PREDICATE DEVICES
| 510(K) # | TRADE NAME | MANUFACTURER |
|---|---|---|
| K082947 | LenSx 550 Laser System | LenSx Lasers, Inc. |
| K993154 | Dodick Laser PhotolysisSystem | ARC Laser Corporation |
| K952213 | Alcon Series 20000 LegacyPhacoemulsifier | Alcon Laboratories |
{1}------------------------------------------------
DEVICE DESCRIPTION
The LenSx 550 Laser is an ophthalmic surgical laser that has previously been cleared for use in anterior capsulotomy during cataract surgery (K082947). The LenSx 550 generates femtosecond laser pulses that are scanned in a three-dimensional pattern in the eye. Localization of these laser pulses is accomplished by an aiming beam that identifies the surfaces of the lens, and an off-the-shelf, disposable contact lens and suction ring assembly that fixates the eye relative to the delivery system. As described in K082947, anterior capsulotomy is performed by delivering a cylindrical pattern of laser pulses to intersect the anterior lens capsule.
The same surgical device, with only minor software changes, is now also being proposed for use in phacofragmentation during cataract surgery. As described in this premarket notification, laser phacofragmentation divides the lens into quadrants, in the same way that division of the lens into quadrants is performed with standard ultrasound phacofragmentation in the initial step of phacoemulsification.
To perform phacofragmentation, the LenSx system delivers a series of laser pulses to form two intersecting ellipsoidal planes. Photodisruption of the lens tissue in this pattern, which appears as a cross from a surgical (top) view, segments the nucleus into four sections before the eye is physically entered by any instruments or devices.
Following entry into the eye and then into lens, the fragmented nuclear quadrants created by the LenSx 550 are further fragmented and removed using a standard phacoemulsification probe.
STATEMENT OF INTENDED USE
The LenSx 550 Laser System is indicated for anterior capsulotomy and laser phacofragmentation during cataract surgery. The anterior capsulotomy and phacofragmentation procedures may be performed either individually or consecutively during the same surgery.
TECHNOLOGICAL CHARACTERISTICS COMPARISON
The LenSx 550 Laser System proposed for use in phacoframentation is identical with respect to all technological characteristics to the previously cleared LenSx 550 for use in anterior capsulotomy (K082947). Substantial equivalence is also drawn to previously cleared phacofragmentation systems.
The LenSx 550 Laser System delivers femtosecond laser pulses to produce a pattern of photodisruption in the lens nucleus. Highly localized, laser mediated acoustic and cavitation effects associated with photodisruption precisely fragment the lens nucleus.
Predicate devices include laser and ultrasound phacoemulsification devices cleared for phacofragmentation, such as the Alcon Series 20000 Legacy (Alcon Laboratories, Inc. K952213) and the Dodick Photolysis system (ARC Laser Corp., K993154). Ultrasound
{2}------------------------------------------------
phacofragmentation devices utilize a piezoelectric ultrasonic source that generates similar acoustic and cavitation effects that, along with mechanical effects from the probe, result in phacofragmentation.
Laser devices cleared for phacofragmentation, including the Dodick Photolysis System, use high energy nanosecond laser pulses delivered via a probe to create the mechanical, cavitation and acoustic effects that mediate phacofragmentation.
Since the femtosecond pulses are delivered optically through the cornea, the LenSx Laser performs phacofragmentation without the mechanical effects and ultrasonic shockwaves/energy transmitted to the tissue by handheld phacofragmentation probes.
BRIEF SUMMARY OF PRECLINICAL AND CLINICAL PERFORMANCE TEST RESULTS
The performance data supporting substantial equivalence of the LenSx 550 femtosecond laser system to the predicate devices are summarized as follows:
- Pre-clinical studies applying the LenSx phacofragmentation pattern in plastic . substrates demonstrated a high degree of accuracy and reproducibility for laser pulse placement.
- LenSx phacofragmentation performed in ex vivo porcine eyes further confirmed . the accuracy and reproducibility of the localized laser tissue effects. This testing also demonstrated that the ultrasound power required to complete the phacofragmentation procedure was significantly reduced in comparison to procedures performed using standard phacofragmentation and phacoemulsification techniques with the Alcon Series 20000 Legacy, the most common such device.
- A prospective clinical study was performed outside the U.S. to evaluate clinical . performance, with a total of 92 subjects randomly assigned to one of the following treatment groups:
-
- A study group of 46 eyes underwent initial laser phacofragmentation using the LenSx 550 Laser, followed by phacoemulsification using the Alcon Series 20000 Legacy.
-
- A control group of 46 eyes underwent phacofragmentation and phacoemulsification with the Alcon Series 20000 Legacy (predicate device), using a surgical technique previously demonstrated to significantly reduce ultrasound power requirements.
-
Compared to the control group where only standard ultrasonic phacoemulsification was used, the study group (using laser phacofragmentation) required significantly less manual manipulation and ultrasound power to achieve the desired phacofragmentation effect. Clinical results were excellent in both groups, with no significant adverse events. Results in both groups were consistent with those of advanced phacoemulsification techniques reported in previous studies (Can, 2004).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an emblem of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is stylized with thick, black lines, giving it a bold and recognizable appearance.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
APR 2 3 2010
LenSx Lasers, Incorporated c/o Judy F. Gordon, D.V.M. ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, California 92651
Re: K094052
Trade/Device Name: LenSx 550 Laser System, Model 550 Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: OOE, HQC Dated: March 23, 2010 Received: March 25, 2010
Dear Dr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Edith m.h
Malvina B. Eydelman, M.Ø Director Division of Ophthalmic, Neurological, and Ear, Nose, Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
Device Name(s): LenSx 550 Laser System
Indications for Use:
The LenSx 550 Laser System is indicated for anterior capsulotomy and laser phacofragmentation during cataract surgery. The anterior capsulotomy and phacofragmentation procedures may be performed either individually or consecutively during the same surgery.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _ 1 _ of
Daryl Kaufman M.A.
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices .
510(k) Number K0-94052
§ 886.4390 Ophthalmic laser.
(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.