(112 days)
Intended Use: AS-XXXX Suction Pump is intended to be used to remove bodily fluids from a patient's airvay or respiratory system.
The AS-XXXX (FAMILY SERIES) suction pump is a portable AC-powered suction pump. The AS-XXXX (FAMILY SERIES) suction pump creates a negative pressure (vacuum) that draws fluids through a disposable suction catheter that is connected by suction tubing to a collection container. The suctioned fluids are then trapped in the collection container for proper disposal. The device must only be used on the order of a physician. If practiced outside of the hospital setting, the care giver must be trained by a certified healthcare professional and the training must be recorded and documented. The AS-XXXX (FAMILY SERIES) suction pump consists of an on/off switch, a pump unit, a power cord, a collection container, relief valve, pressure adjustment knob, pressure gauge, microbial filter, long intermediate tubing, short intermediate tubing, and suction tubing. Disposable suction catheters are not packaged with this device and must be purchased separately.
The provided text describes the acceptance criteria and the study for the Ding Hwa AS-XXXX (Family Series) Suction Pump.
Here's the breakdown of the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (from predicate device) | Reported Device Performance (Ding Hwa AS-XXXX Suction Pump) |
|---|---|
| Substantially equivalent to and meets the same acceptance criteria as the predicate device (EMG Suction Unit SUA01-AXX 510(k)112421). | The Ding Hwa AS-XXXX Suction Pump is stated to be "substantially equivalent and meets the same acceptance criteria as the predicate device." |
| Compliance with IEC 60601-1, 3rd Ed. (General Requirements for Basic Safety and Essential Performance) | "All results of the testing met acceptance criteria." |
| Compliance with IEC 60601-1-2 (Collateral standard: Electromagnetic compatibility) | "All results of the testing met acceptance criteria." |
| Compliance with ISO 10079-1 (2009): Medical suction equipment -- Part 1: Electrically powered suction equipment -- Safety requirements. | "All results of the testing met acceptance criteria." The text explicitly states: "Ding Hwa AS-XXXX Suction Pump is substantially equivalent and meets the same acceptance criteria as the predicate device, EMG Suction Unit SUA01-AXX 510(k)112421. Non-clinical performance testing includes IEC 60601-1, 3rd Ed. MEDICAL ELECTRICAL EQUIPMENT - PART 1: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE, IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility and ISO 10079-1 (2009): Medical suction equipment -- Part 1: Electrically powered suction equipment -- Safety requirements. All results of the testing met acceptance criteria." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document refers to "non-clinical performance testing" and the results meeting "acceptance criteria," but does not detail the methodology, sample sizes, or data provenance for these tests beyond citing the standards. This device is a hardware product (suction pump), not an AI device, so typical AI study parameters like "test set" and "data provenance" often don't apply in the same way.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable for this type of medical device (a suction pump) and is not provided in the document. The "ground truth" for a suction pump would be its adherence to established performance and safety standards, not expert consensus on diagnostic images or other data typically associated with AI.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable for this type of medical device and is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable for this type of medical device (suction pump). MRMC studies are typically for medical imaging diagnostics or similar AI-assisted interpretation, which is not relevant here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable for this type of medical device. The device is a physical pump, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is established through compliance with recognized international and national standards for medical electrical equipment and suction equipment, specifically:
- IEC 60601-1, 3rd Ed. (General Requirements for Basic Safety and Essential Performance)
- IEC 60601-1-2 (Electromagnetic compatibility)
- ISO 10079-1 (2009): Medical suction equipment -- Part 1: Electrically powered suction equipment -- Safety requirements.
The device also based its "ground truth" on substantial equivalence to a predicate device (EMG Suction Unit, Model SUA01-AXX Series 510(k)112421), implying its performance matches that of an already cleared device.
8. The sample size for the training set
This information is not applicable for this type of medical device. "Training set" is a concept related to machine learning and AI, not mechanical medical devices like a suction pump.
9. How the ground truth for the training set was established
This information is not applicable for this type of medical device.
{0}------------------------------------------------
CO., LTD. IING H
510(k) Summary – Ding Hwa AS-XXXX (Family Series) Suction Pump Sec. 5:
| Date Summary wasPrepared | December 25, 2013 |
|---|---|
| 510(k) Submitter | Ding Hwa Co. Ltd.#121 Sec. 3 Zhongshan RoadDacun, Chang-Hwa, Taiwan 51542 |
| Primary Contact for this510(k) Submission | Robert O. DeanCompliance Systems International, LLc.1083 Delaware Rd.Buffalo, NY 14223716.440.7362compliancesystems@yahoo.com |
| Device Common Name | Powered suction pump |
| Trade Name | AS-XXXX Suction Pump (family series) |
| Device Product Codes andClassification Name | JCX, 21CFR878.4780, Powered suction pump |
| Predicate Device | EMG Suction Unit, Model SUA01-AXX Series 510(k)112421 |
| Device Description | The AS-XXXX (FAMILY SERIES) suction pump is a portable AC-powered suction pump.The AS-XXXX (FAMILY SERIES) suction pump creates a negative pressure (vacuum)that draws fluids through a disposable suction catheter that is connected by suctiontubing to a collection container. The suctioned fluids are then trapped in the collectioncontainer for proper disposal. The device must only be used on the order of a physician.If practiced outside of the hospital setting, the care giver must be trained by a certifiedhealthcare professional and the training must be recorded and documented. The AS-XXXX (FAMILY SERIES) suction pump consists of an on/off switch, a pump unit, a powercord, a collection container, relief valve, pressure adjustment knob, pressure gauge,microbial filter, long intermediate tubing, short intermediate tubing, and suction tubing.Disposable suction catheters are not packaged with this device and must be purchasedseparately. |
| Intended Use | AS-XXXX Suction Pump is intended to be used to remove body fluids from a patient'sairway or respiratory system. |
| TechnologicalCharacteristics | Ding Hwa AS-XXXX Suction Pump is of the same design, material and performancecharacteristics and intended use as the predicate device. |
| Summary of Testing | Ding Hwa AS-XXXX Suction Pump is substantially equivalent and meets the sameacceptance criteria as the predicate device, EMG Suction Unit SUA01-AXX510(k)112421. Non-clinical performance testing includes IEC 60601-1, 3rd Ed.MEDICAL ELECTRICAL EQUIPMENT - PART 1: GENERAL REQUIREMENTS FORBASIC SAFETY AND ESSENTIAL PERFORMANCE,IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basicsafety and essential performance - Collateral standard: Electromagnetic compatibilityand ISO 10079-1 (2009): Medical suction equipment -- Part 1: Electrically poweredsuction equipment -- Safety requirements.All results of the testing met acceptance criteria. |
| Substantial Equivalence | The AS-XXXX Suction Pump described in this 510(k) submission are substantiallyequivalent in all specifications and performance compared to the predicate deviceidentified in K112421 and do not raise any new types of safety and effectivenessissues when compared to the predicate device. |
EXPORT DEPT:Rm. 4B-02/03, Taipei World Trade Center, 4/Fl.,#5, Hsin-Yi Road, Section 5, Taipei,Taiwan 11011 TEL:+886- 2- 2345- 1868 FAX : +886- 2- 2345- 3162 FACTORY : # 121, Sec. 3, Zhongshan Road, Dacun, Chang-Hwa, Taiwan 51542 TEL : +886- 4- 888- 1486 FAX : +886- 4- 888- 1499
- FACTORY : # 43, Shin- Kung 1st Road, Pei- Tou Industrial Park, Pei- Tou, Chang- Hwa, Taiwan 52149
- TEL : +886- 4- 888- 1486 FAX : +886- 4- 888- 1499
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 28, 2014
Ding Hwa Company, Ltd. Mr. Robert O. Dean Compliance Systems International, LLC. 1083 Delaware Road Buffalo, New York 14223
Re: K140031
Trade/Device Name: Ding Hwa Co., Ltd - AS-XXXX Suction Pump (Family Series) Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: JCX Dated: March 26, 2014 Received: April 1, 2014
Dear Mr. Dean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
Image /page/1/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, composed of three curved lines.
{2}------------------------------------------------
Page 2 - Mr. Robert O. Dean
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices · Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) 140031
Device Name
Ding Hwa Co., Ltd - AS-XXXX Suction Pump (Family Series)
Indications for Use (Describe)
Intended Use: AS-XXX Suction Pump is intended to be used to remove bodily fluids from a patient's airvay or respiratory system.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
(
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
ध PSC Particles Samices (10) | 43-4211)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
g-et
§ 878.4780 Powered suction pump.
(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.