K Number
K132999
Manufacturer
Date Cleared
2014-04-14

(202 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cervical Stand Alone System is indicated for spinal fusion procedures at one level in the cervical spine (C2-T1), in skeletally mature patients with degenerative disc disease (DDD), DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.

The Cervical Stand Alone System is used with autograft bone material and the two titanium alloy screws which accompany the implant.

Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the Cervical Stand Alone System in the cervical spine.

Device Description

The Cervical Stand Alone device is a stand-alone spacer system designed to provide structural stability in skeletally mature individuals following discectomy. The spacers are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. Screws are inserted through the anterior titanium portion of the implant into adjacent vertebral bodies for bony fixation. The spacer is to be filled with autogenous bone graft material.

The Cervical Stand Alone spacer is manufactured from PEEK and Titanium with Titanium bone screws that allow intradiscal fixation to the vertebral body. The superior and inferior suffaces of the implant have a pattern of ridges that provide increased stability and help prevent migration of the device.

AI/ML Overview

Here's an analysis of the provided text regarding the Cervical Stand Alone System, focusing on acceptance criteria and the supporting study:

Acceptance Criteria and Device Performance for Cervical Stand Alone System

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from ASTM Standards)Reported Device Performance (from Non-Clinical Tests)
Static Axial Compression (ASTM F2077-11)Conducted
Dynamic Axial Compression (ASTM F2077-11)Conducted
Static Torsion (ASTM F2077-11)Conducted
Dynamic Torsion (ASTM F2077-11)Conducted
Static Compression Shear (ASTM F2077-11)Conducted
Dynamic Compression Shear (ASTM F2077-11)Conducted
Subsidence (ASTM F2267-04)Conducted
Expulsion (ASTM Draft Standard F-04.25.02.02)Conducted

Note: The provided document states that these tests were "conducted" in accordance with the specified ASTM standards. It does not provide specific numerical performance results or explicit pre-defined pass/fail criteria beyond the adherence to the standards. The basis of substantial equivalence is that the new device has "the same intended use, similar indications for use, the same technological characteristics and design, same materials and the same principles of operation as the predicate devices." Implied acceptance criteria are that the device's mechanical performance is equivalent to, or better than, the predicate devices when tested to these standards.

2. Sample Size for Test Set and Data Provenance

  • Sample Size: Not explicitly stated in the provided text. The document mentions "mechanical testing" but does not detail the number of units or repetitions used for each test.
  • Data Provenance: Not applicable in the context of clinical data. This study is described as "Non-Clinical Test Conducted." Therefore, there is no country of origin for clinical data, nor is it retrospective or prospective clinical data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not applicable. This was a non-clinical, mechanical performance study. There was no "ground truth" derived from expert interpretation of images or patient outcomes. The "ground truth" for these tests would be the established scientific principles of mechanical engineering and the specifications of the ASTM standards themselves.

4. Adjudication Method for the Test Set

  • Not applicable. As this was a non-clinical mechanical test, there was no need for expert adjudication. The results were likely compared against predefined limits or against the performance of predicate devices as per the ASTM standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. The document describes non-clinical mechanical testing, not a clinical MRMC study involving human readers or AI assistance.

6. Standalone (Algorithm Only) Performance Study

  • No. This device is a physical medical implant (intervertebral body fusion device), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

7. Type of Ground Truth Used

  • The "ground truth" for this study was the mechanical performance specifications and methodologies outlined in the ASTM standards (ASTM F2077-11, ASTM F2267-04, and ASTM Draft Standard F-04.25.02.02). The device's performance was evaluated against the requirements, methods, and expected behaviors established by these engineering standards.

8. Sample Size for the Training Set

  • Not applicable. Since this is a physical medical device undergoing mechanical testing, there is no "training set" in the context of machine learning or AI models. The design and manufacturing processes are informed by engineering principles and standards, but not by a data-driven training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set. The "ground truth" for the design and manufacturing of such a device is rooted in:
    • Biomechanical principles of the spine.
    • Material science properties of PEEK and Titanium.
    • Clinical experience with similar devices (as inferred by the predicate devices).
    • Engineering design specifications aiming to meet the performance requirements of the relevant ASTM standards.

{0}------------------------------------------------

510(k) SUMMARY

Cervical Stand Alone System

APR 1 4 2014

510(k) Owner Information

Name:Address:Orthofix Inc.3451 Plano ParkwayLewisville, TX 75056
Telephone Number:Fax Number:Email:214.937.2145214-937-3322nataliavolosen@orthofix.com
Registration Number:3008524126
Contact Person:Natalia VolosenSenior Regulatory Affairs Specialist
Date Prepared:April, 10, 2014
Name of DeviceTrade Name / ProprietaryName:Cervical Stand Alone System
Common Name:Intervertebral body fusion device
Product Code:OVE - Cervical intervertebral fusion device with integrated fixation
Regulatory Classification:Class II - 21 CFR § 888.3080 - Intervertebral body fusion device
Review Panel:Orthopedic Device Panel
Predicate Devices:K101812 - CONSTRUX Mini PEEK Spacer System, SE 9/27/10K091088 - LDR Spine Cervical Interbody Fusion System, SE07/14/09K083389 - COALITION Spacer, SE 03/26/09K102606 - AVS Anchor-C Cervical Cage System, SE 04/22/11

Reason for 510(k) Submission: New product offering

Device Description

The Cervical Stand Alone device is a stand-alone spacer system designed to provide structural stability in skeletally mature individuals following discectomy. The spacers are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. Screws are inserted through the anterior titanium portion of the implant into adjacent vertebral bodies for bony fixation. The spacer is to be filled with autogenous bone graft material.

The Cervical Stand Alone spacer is manufactured from PEEK and Titanium with Titanium bone screws that allow intradiscal fixation to the vertebral body. The superior and inferior suffaces of the implant have a pattern of ridges that provide increased stability and help prevent migration of the device.

{1}------------------------------------------------

Intended Use / Indications for Use

The Cervical Stand Alone System is indicated for spinal fusion procedures at one level in the cervical spine (C2-T1), in skeletally mature patients with degenerative disc disease (DDD), DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.

The Cervical Stand Alone System is used with autograft bone material and the two titanium alloy screws which accompany the implant.

Patients must have undergone a regimen of at least six (6) weeks of non-operative treatment prior to being treated with the Cervical Stand Alone System in the cervical spine.

Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices

The technological characteristics of the Cervical Stand Alone System are similar to the predicate devices in terms of design, dimensions, intended use, materials, and performance characteristics

PERFORMANCE DATA - Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence

Mechanical testing consisting of Static and Dynamic Axial Compression Test, Static and Dynamic Torsion Test, Static and Dynamic Compression Shear Test, Subsidence Test and Expulsion Test were conducted in accordance to ASTM F2077-11 standard for Test Method for Intervertebral Body Fusion Devices, ASTM F2267-04 standard for Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device under Static Axial Compression and in accordance with ASTM Draft Standard F-04.25.02.02, "Static Push-out Test Method for Intervertebral Body Fusion Devices.

Basis of Substantial Equivalence

The new Cervical Stand Alone system has the same intended use, similar indications for use. the same technological characteristics and design, same materials and the same principles of operation as the to the AVS Anchor-C Cervical Cage System. CONSTRUX Mini PEEK Spacer svstem. COALITION Spacer and LDR Spine Cervical Interbody Fusion System device.

{2}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 14, 2014

Orthofix, Incorporated Ms. Natalia Volosen Senior Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056

Re: K132999

Trade/Device Name: Cervical Stand Alone System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II · Product Code: OVE Dated: March 6, 2014 Received: March 7, 2014

Dear Ms. Volosen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Ms. Natalia Volosen

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDeviccs/Resourcesfor You/Industrv/default.htm.

Sincerely yours.

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K132999

Device Name

Cervical Stand Alone System

Indications for Use (Describe)

The Cervical Stand Alone System is indicated for spinal fusion procedures at one level in the cervical spine (C2-11), in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.

The Cervical Stand Alone System is used with autograft bone material and the two titanium alloy screws which accompany the implant.

Patients must have undergone a regiment of at 16) weeks of non-operative treatment prior to being treated with the Cervical Stand Alone System in the cervical spine.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY 46. . . ..... :"

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Anton E. Dmitriev. Ph

Division

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration

  • Office of Chief Information Officer
  • Paperwork Reduction Act (PRA) Staff
  • PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.