K Number
K102606

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-04-22

(224 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stryker Spine AVS® Anchor-C Cervical Cage System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The AVS® Anchor-C Cervical Cage is to be used with autogenous bone graft and implanted via an open, anterior approach.

The AVS® Anchor-C Cervical Cage must be used with the internal screw fixation provided by AVS® Anchor-C Fixation Screws. This cervical device is to be used in patients who have had six weeks of non-operative treatment.

Device Description

The AVS® Anchor-C Cervical Cage is a hollow, rectangular-shaped PEEK Optima® LT1 (per ASTM F2026) cage assembled to a titanium alloy Ti6Al4V (per ASTM F136 and ISO 5832-3) plate and has one tantalum marker (per ASTM F560). It is intended for use as an interbody fusion device and is offered in a variety of heights, footprints, and lordotic angles to adapt to varying patient anatomies. The PEEK Optima® LTI cage portion consists of one closed pocket for graft containment and has serrations on the superior and inferior surfaces of the cage. The implant is designed to be used exclusively with the internal supplemental fixation provided (AVS® Anchor-C Fixation Screws). The AVS® Anchor-C Fixation Screws are constructed from titanium alloy Ti6Al4V (per ASTM F136 and ISO 5832-3) and possess clips (also constructed from titanium alloy Ti6Al4V per ASTM F136 and ISO 5832-3) that mate with internal features located within the AVS® Anchor-C Cervical Cage. Once fully seated into the holes, the screws are designed to lock into the titanium plate.

AI/ML Overview

The provided document describes the Stryker Spine AVS® Anchor-C Cervical Cage System, a medical device. It focuses on the mechanical testing performed to demonstrate substantial equivalence to predicate devices, rather than the performance of an AI system. Therefore, many of the requested fields regarding AI system performance, ground truth, and expert evaluation are not applicable or cannot be extracted from this document.

Here's an analysis of the available information:

Acceptance Criteria and Device Performance

The document states that "Testing in compliance with FDA's June 12, 2007 'Class II Special Controls Guidance Document: Intervertebral Body Fusion Device' was performed for the AVS® Anchor-C implant system and demonstrated substantially equivalent performance to the identified predicate device systems."

The acceptance criteria are implicitly defined by the successful demonstration of "substantially equivalent performance" to predicate devices based on the mechanical tests listed. The document does not provide specific numerical acceptance thresholds or reported performance values for the device itself. Instead, it indicates that the device met the criteria by performing comparably to already approved devices.

Table of Acceptance Criteria and Reported Device Performance:

Test Method (per ASTM/Draft Standard)Acceptance Criteria (Implicit)Reported Device Performance
Static Compression (ASTM F2077)Substantially equivalent to predicate devices (e.g., LDR MC+, Surgicraft STALIF C, etc.)Demonstrated substantially equivalent performance.
Dynamic Compression (ASTM F2077)Substantially equivalent to predicate devicesDemonstrated substantially equivalent performance.
Static Compression Shear (ASTM F2077)Substantially equivalent to predicate devicesDemonstrated substantially equivalent performance.
Dynamic Compression Shear (ASTM F2077)Substantially equivalent to predicate devicesDemonstrated substantially equivalent performance.
Static Torsion (ASTM F2077)Substantially equivalent to predicate devicesDemonstrated substantially equivalent performance.
Dynamic Torsion (ASTM F2077)Substantially equivalent to predicate devicesDemonstrated substantially equivalent performance.
Expulsion (ASTM F04-25-02-02 Draft)Substantially equivalent to predicate devicesDemonstrated substantially equivalent performance.
Subsidence (ASTM F2267)Substantially equivalent to predicate devicesDemonstrated substantially equivalent performance.

Additional Requested Information:

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify the sample size for each mechanical test. Mechanical tests typically use a small number of physical samples (e.g., 3-6) per test condition to represent the device.
  • Data Provenance: The data is from mechanical testing of the device, performed in compliance with FDA guidance. The country of origin of the data is not specified but is implicitly associated with the manufacturer (Stryker Spine, located in New Jersey, USA). The data is prospective, generated specifically for this 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This device is a physical implant, and the "ground truth" for mechanical testing is based on the physical properties and performance measured against established engineering standards (ASTM and ISO). There are no human "experts" establishing ground truth in the context of diagnostic interpretation for this type of submission.

4. Adjudication method for the test set:

  • Not Applicable. As this is mechanical testing of a physical device, there is no adjudication method in the context of human interpretation or AI output. The evaluation is based on engineering measurements and comparisons to predicate device performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI-enabled device or a diagnostic imaging device requiring human reader studies.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not Applicable. This is a physical medical implant, not an algorithm or AI system.

7. The type of ground truth used:

  • For the mechanical tests, the "ground truth" is defined by the objective measurements obtained from the various tests (e.g., compression strength, shear strength, expulsion resistance) and their comparison to the performance of predicate devices as per the relevant ASTM and ISO standards and FDA guidance.

8. The sample size for the training set:

  • Not Applicable. This is a physical medical implant. There is no training set in the context of machine learning or AI.

9. How the ground truth for the training set was established:

  • Not Applicable. As there is no training set, this question is not relevant.

{0}------------------------------------------------

12102606

Stryker Spine AVS® Anchor-C Cervical Cage System

Traditional S10(k) Premarket Notification

APR 2 2 2011

・j

510(k) Summary: AVS® Anchor-C Cervical Cage System
Submitter:Stryker Spine
2 Pearl Court
Allendale, New Jersey 07401
Contact PersonMs. Kimberly Lane
Sr. Regulatory Affairs Specialist
Phone: 201-760-8215
Fax: 201-760-8415
Email: kimberly.lane@stryker.com
Date PreparedApril 19, 2011
Trade NameAVS® Anchor-C Cervical Cage System
Proposed ClassClass II
Classification Nameand NumberIntervertebral body fusion device, 21 CFR 888.3080
Product CodeODP
Predicate DevicesThe AVS® Anchor-C Cervical Cage System was shown to besubstantially equivalent to the devices listed below:● LDR MC+, 510(k) # K091088● Surgicraft STALIF C, 510(k) #K072415● Depuy Bengal #K081917● Spinal Elements Crystal #K073351● Zimmer BAK/C # P980048● Medtronic AFFINITY #P000028
Device DescriptionThe AVS® Anchor-C Cervical Cage is a hollow, rectangular-shaped PEEK Optima® LT1 (per ASTM F2026) cage assembledto a titanium alloy Ti6Al4V (per ASTM F136 and ISO 5832-3) plate and has one tantalum marker (per ASTM F560). It isintended for use as an interbody fusion device and is offered in avariety of heights, footprints, and lordotic angles to adapt tovarying patient anatomies. The PEEK Optima® LTI cage
510(k) Summary: AVS® Anchor-C Cervical Cage Systemportion consists of one closed pocket for graft containment andhas serrations on the superior and inferior surfaces of the cage.The implant is designed to be used exclusively with the internalsupplemental fixation provided (AVS® Anchor-C FixationScrews). The AVS® Anchor-C Fixation Screws are constructedfrom titanium alloy Ti6Al4V (per ASTM F136 and ISO 5832-3) and possess clips (also constructed from titanium alloyTi6Al4V per ASTM F136 and ISO 5832-3) that matewith internal features located within the AVS® Anchor-CCervical Cage. Once fully seated into the holes, the screws aredesigned to lock into the titanium plate.
Intended UseThe Stryker Spine AVS® Anchor-C Cervical Cage System isindicated for anterior cervical interbody fusion procedures inskeletally mature patients with cervical disc disease at one levelfrom the C2-C3 disc to the C7-T1 disc. Cervical disc disease isdefined as intractable radiculopathy and/or myelopathy withherniated disc and/or osteophyte formation on posterior vertebralendplates producing symptomatic nerve root and/or spinal cordcompression confirmed by radiographic studies. The AVS®Anchor-C Cervical Cage is to be used with autogenous bonegraft and implanted via an open, anterior approach.The AVS® Anchor-C Cervical Cage must be used with theinternal screw fixation provided by AVS® Anchor-C FixationScrews. This cervical device is to be used in patients who havehad six weeks of non-operative treatment.
Summary of theTechnologicalCharacteristicsThe subject AVS® Anchor-C implant system and the predicatesshare similar design features:• Graft windows for packing autogenous bone
510(k) Summary: AVS® Anchor-C Cervical Cage System
Serrations on the superior and inferior surfaces
Comparable heights, widths, depths, and lordotic angles
Testing in compliance with FDA's June 12, 2007 "Class IISpecial Controls Guidance Document: Intervertebral BodyFusion Device" was performed for the AVS® Anchor-C implantsystem and demonstrated substantially equivalent performanceto the identified predicate device systems.
The following mechanical tests were performed:
Static Compression (per ASTM F2077)
Dynamic Compression (per ASTM F2077)
Static Compression Shear (per ASTM F2077)
Dynamic Compression Shear (per ASTM F2077)
Static Torsion (per ASTM F2077)
Dynamic Torsion (per ASTM F2077)
Expulsion (per ASTM F04-25-02-02 Draft)
Subsidence (per ASTM F2267)

Page 2 of 4

/"

sheet lot 3

{1}------------------------------------------------

K102606

Stryker Spine AVS® Anchor-C Cervical Cage System

ئيس

Page 3 of 4

sheet 2 of 3

{2}------------------------------------------------

K102606

Stryker Spine AVS® Anchor-C Cervical Cage System

Page 4 of 4

sheet 3 of 3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird symbol with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Stryker Spine % Ms. Kimberly Lane Senior Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401

ട്ട്‌ -

K102606 Trade/Device Name: Stryker Spine AVS® Anchor-C Cervical Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: April 4, 2011 Received: April 5, 2011

Dear Ms. Lane:

Re:

This letter corrects our substantially equivalent letter of April 22, 2011.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other

{4}------------------------------------------------

Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions. (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark A. Milbersen

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Stryker Spine AVS® Anchor-C Cervical Cage System

Traditional 510(k) Premarket Notification

Indications for Use

K102606 510(k) Number (if known): K

Device Name: Stryker Spine AVS® Anchor-C Cervical Cage System

Indications For Use:

The Stryker Spine AVS® Anchor-C Cervical Cage System is indicated for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C disc to the C7-T1 disc. Cervical disc disease is defined as intractable radioulopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographics producting synoding synador-C Cervical Cage is to be used with autogenous bone graft and implanted via an open, anterior annonce

The AVS Anchor-C Cervical Cage must be used with the internal screw fixation provided by AVS® Anchor-C Fixation Screws. This cervical device is to be used in patients who have had six weeks of nonoperative treatment.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

KIO2666 510(k) Number.

Page 2 of 2

N/A