K Number
K083389

Validate with FDA (Live)

Device Name
COALITION SPACER
Date Cleared
2009-03-26

(129 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The COALITION™ Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION™ Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant.

Device Description

The COALITION™ Spacer is a stand-alone cervical interbody fusion device used to provide structural stability in skeletally mature individuals following discectorny. The spacers are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to aid in expulsion resistance. Screws are inserted through the anterior titanium portion of the implant into adjacent vertebral bodies for bony fixation. The spacer is to be filled with autogenous bone graft material.

The COALITION™ Spacer is made from radiolucent polymer, with titanium alloy or tantalum markers, as specified in ASTM F2026, F136, F1295, and F560. The anterior portion of the implant and the mating screws are manufactured from titanium alloy, as specified in ASTM F136 and F1295.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called the COALITION™ Spacer, an intervertebral body fusion device. This document describes the device, its intended use, and its substantial equivalence to predicate devices. It does not contain information about acceptance criteria, device performance studies, sample sizes, expert qualifications, adjudication methods, multi-reader multi-case studies, standalone performance, or training/ground truth details typically associated with AI/ML device evaluations.

Therefore, I cannot provide the requested information. The document focuses on regulatory approval for a physical medical implant, not a software or AI/ML diagnostic tool.

{0}------------------------------------------------

K083389 Page$\frac{1}{1}$

510(k) SUMMARY: COALITION™ SPACER

MAR 2 6 2009

  • Company: Globus Medical Inc. 2560 General Armistead Ave. Audubon. PA 19403 (610) 415-9000
    Kelly J. Baker, Ph.D Contact: Director, Clinical Affairs & Regulatory

  • Device Name: COALITION™ Spacer
    Classification: Product Code ODP. Class II. 21 CFR §888.3080 Intervertebral body fusion device.

  • PATRIOT™ Predicate(s): Cervical Spacer (Colonial™ ACDF Spacer) K072991 and other legally marketed devices.

Device Description:

The COALITION™ Spacer is a stand-alone cervical interbody fusion device used to provide structural stability in skeletally mature individuals following discectorny. The spacers are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to aid in expulsion resistance. Screws are inserted through the anterior titanium portion of the implant into adjacent vertebral bodies for bony fixation. The spacer is to be filled with autogenous bone graft material.

The COALITION™ Spacer is made from radiolucent polymer, with titanium alloy or tantalum markers, as specified in ASTM F2026, F136, F1295, and F560. The anterior portion of the implant and the mating screws are manufactured from titanium alloy, as specified in ASTM F136 and F1295.

Intended Use:

The COALITION™ Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION™ Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant.

Basis for Substantial Equivalence:

The COALITION™ Spacer has been evaluated in accordance with the "Class II Special Controls Guidance Document: Intervertebral Fusion Device", June 12, 2007 and have been found to meet the criteria set forth in the quidance document in terms of indications, design, and performance.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is written around the perimeter of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Globus Medical, Inc. % Kelly J. Baker, Ph.D. Director, Clinical Affairs & Regulatory 2560 General Armistead Avenue Audubon, Pennsylvania 19403

SEP 12 2011

Re: K083389

Trade/Device Name: COALITION™ Spacer Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: March 23, 2009 Received: March 24, 2009

Dear Dr. Baker:

This letter corrects our substantially equivalent letter of March 26, 2009.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other

{2}------------------------------------------------

Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Mark A. Milken

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use Statement

X083389 510(k) Number: Device Name: COALITION™ Spacer

Indications:

The COALITION™ Spacer is a stand-alone interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-T1) at one level. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of nonoperative treatment. The COALITION™ Spacer is to be filled with autogenous bone graft material, and is to be used with two titanium alloy screws which accompany the implant.

OR Prescription Use × (Per 21 CFR §801.109)

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

D. Thm

Division Division of General, Restorative, and Neurological D

510(k) Number K68339

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.