(50 days)
Not Found
No
The summary describes a standard in vitro diagnostic assay using chemical reagents and a chemistry analyzer. There is no mention of AI/ML in the device description, intended use, or performance studies.
No
This device is an in vitro diagnostic (IVD) device used to quantitatively measure albumin in human serum or plasma. It is used for diagnostic purposes by providing information for the diagnosis and treatment of diseases, not for therapy itself.
Yes
The "Intended Use / Indications for Use" section explicitly states "For in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma on ADVIA Chemistry systems. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys." This directly indicates the device's role in diagnosing diseases.
No
The device description explicitly states it is a reagent and calibrator kit, which are physical components used in an in vitro diagnostic assay. It is not solely software.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The "Intended Use / Indications for Use" section explicitly states "For in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma".
- Device Description: The device is a reagent kit and calibrator designed to be used on automated chemistry systems for testing biological samples (serum or plasma).
- Performance Studies: The document describes performance studies like precision, method comparison, and analytical specificity, which are typical evaluations for IVD devices.
N/A
Intended Use / Indications for Use
For in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma on ADVIA Chemistry systems. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys.
For in vitro diagnostic use in the calibration of the ADVIA Chemistry Albumin BCP Assay (ALBP) on ADVIA Chemistry systems.
Product codes (comma separated list FDA assigned to the subject device)
CJW, JIX
Device Description
The Albumin BCP reagents are ready-to-use liquid reagents packaged for use on the automated ADVIA 1650 Chemistry systems. Reagents are supplied in two configurations: fill volume of 18 mL in a 20 mL wedge or 35 mL in a 40 mL wedge, 4 wedges/kit.
The calibrator is a multi-analyte human serum based product containing albumin derived from human serum. The kit consists of 3 vials of one-level calibrator which are lyophilized. The target concentration of this calibrator is 4.3 g/dL. The volume per vial (after reconstitution with deionized water) is 2.0 mL.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The following data represent typical performance for the ADVIA Chemistry Albumin BCP Assay and were collected on ADVIA 1650 Chemistry system. Substantial equivalence was demonstrated by testing several performance characteristics including precision, method comparison, interfering substances and analytical range. All of the evaluation studies gave acceptable results when compared to the predicate device. These studies support that the ADVIA Chemistry Albumin BCP Assay is substantially equivalent to the predicate device.
1. Precision
Within run and Total Precision were established by Assaying serum sample pools and serum based controls. Each sample was Assaved 2 replicates per run, 2 runs per day, for at least 20 days. Precision estimates were computed according to CLSI document EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline.
Product | N | MEAN g/dL | Within Run SD | Within Run CV | Between Run SD | Between Run CV | Between Day SD | Between Day CV | Total SD | Total CV |
---|---|---|---|---|---|---|---|---|---|---|
Serum Control | 80 | 2.7 | 0.03 | 0.9 | 0.00 | 0.0 | 0.03 | 1.2 | 0.04 | 1.5 |
Serum Control | 80 | 4.0 | 0.03 | 0.7 | 0.01 | 0.3 | 0.03 | 0.6 | 0.04 | 1.0 |
Serum Pool | 80 | 3.5 | 0.00 | 0.0 | 0.00 | 0.0 | 0.00 | 0.0 | 0.00 | 0.0 |
Serum Pool | 80 | 5.2 | 0.04 | 0.7 | 0.02 | 0.3 | 0.02 | 0.4 | 0.05 | 0.9 |
2. Linearity/Assay reportable range
Linearity of the Assay on ADVIA 1650 Chemistry system was assessed by evaluating eleven dilutions across the measuring range. All samples were tested on the ADVIA 1650 Chemistry analyzer. Observed values were compared vs. expected values using linear regression analysis. The low end of the assay range is calculated based on the Limit of Quantitation. The high end of the assay range is based on the linearity calculations.
The linear/measuring range of the assay is 0.6 to 8.0 g/dL.
3. Limit of Blank, Limit of Detection, Limit of Quantitation
The estimations of the Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) were performed according to CLSI document EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, Approved Guideline. The LoB for the ADVIA Chemistry Albumin BCP Assay on the ADVIA 1650 Chemistry system is 0.1 g/dL. Limit of detection (LoD) is the smallest amount that this Assay can reliably detect to determine presence of an analyte. The LoD for for the ADVIA Chemistry Albumin BCP Assay on the ADVIA 1650 Chemistry system is 0.6 g/dL. LoB and LoD values are determined with proportions of false positives (a) less than 5% and false negatives (B) less than 5%, based on 720 determinations with 240 blank and 480 low-level sample replicates. . Limit of Quantitation (LoQ) is 0.6 g/dL based on an inter-assay precision of 10% bias. Bias is the difference in the results between the control sample (without the interferent) and the test sample (contains the interferent) expressed in percent. Bias exceeding 10% is considered interference.
Bilirubin (conjugated and unconjugated) tested up to 60 mg/dL, lipemia up to 525 mg/dL and hemoglobin up to 750 g/dL were shown not to cause significant interference with this assay.
7. Reference Interval (Expected Values)
Expected values are within 3.4 - 5.0 g/L **
** Willey DA, Savory J, Lasky F. An Evaluation of a Revised Albumin Method for the aca@ discrete clinical analyzer, Du Pont Company, Wilmington, DE, August 1982
Siemens provides this information for reference. As with all in vitro diagnostic Assays each laboratory should determine its own reference ranges for the diagnostic evaluation of patient results.
8. Reagent and Calibrator Stability:
Reagent: for opened products, once placed on the system reagents are stable for 60 days. The shelf life of the ADVIA Chemistry Albumin BCP Reagent is 12 months at 2-8°C. For unopened product, see the expiration date on the reagent carton.
Calibrator: for opened products, once the cap is removed, assigned values are stable for 8 hours when recapped immediately after use and stored at 2-8℃. The shelf life of the ADVIA Chemistry Albumin BCP Calibrator is 18 months at 2-8°C. For unopened product, see the expiration date on the calibrator carton.
The calibrator which is used for the ADVIA Chemistry Albumin BCP assay is the same calibrator (one level only) which was previously cleared as Siemens Dimension Clinical Chemistry System TP/ALB calibrator under K861700 and is relabeled for ADVIA Chemistry use
9. Traceability
The ADVIA Chemistry Albumin BCP Assay is traceable to ERM-DA470k Reference Material. Assigned values of the ADVIA Chemistry Albumin BCP Calibrator are traceable to this standardization
10. Value Assignment
The value assignment protocol for the ADVIA Chemistry Albumin BCP calibrator is the same as for the predicate device which was cleared under K861700.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1035 Albumin test system.
(a)
Identification. An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.(b)
Classification. Class II.
0
510(k) Summary of Safety and Effectiveness for the ADVIA® 1650 Chemistry Albumin BCP Assay (ALBP)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
-
- 510(k) Number: _k132664
-
- Applicant:
Address: Phone:
Contact: Kira Gordon, PhD Sr. Regulatory Affairs Specialist Siemens Healthcare Diagnostics, Inc 511 Benedict Ave, Tarrytown, NY 10591 (914) 524-2996 (914) 524-2500 (FAX)
OCT 1 6 2013
-
- Date: August 26, 2013
4. Proprietary and Established Names:
ADVIA Chemistry Albumin BCP Assay (ALBP) ADVIA Chemistry Albumin BCP Calibrator
5. Regulatory Information:
- Reagent
Regulation section: 21 CFR §862.1035, bromcresol purple dye-binding, albumin Classification: Class II Product Code: CJW · Panel: Clinical Chemistry (75)
Calibrator
Regulation section: 21 CFR §862.1150, calibrator, multi-analyte mixture Classification: Class II Product Code: JIX Panel: Clinical Chemistry (75)
6. Predicate Device:
Reagent
Device Name: Dimension Clinical Chemistry System Albumin Flex reagent cartridge Common Name: Dimension Clinical Chemistry System Albumin Assay 510(k) Number: K861700 Manufacturer: Siemens Healthcare Diagnostics, Inc.
Page 1 of 7
1
Calibrator
Device Name: Dimension Clinical Chemistry System TP/ALB Calibrator Common Name: Dimension Clinical Chemistry System TP/ALB Calibrator 510(k) Number: K861700 Manufacturer: Siemens Healthcare Diagnostics, Inc.
7. Intended Use:
See Indications for Use
8. Indications for Use:
For in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma on ADVIA Chemistry systems. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys.
For in vitro diagnostic use in the calibration of the ADVIA Chemistry Albumin BCP Assay (ALBP) on ADVIA Chemistry systems.
9. Device Description:
The Albumin BCP reagents are ready-to-use liquid reagents packaged for use on the automated ADVIA 1650 Chemistry systems. Reagents are supplied in two configurations: fill volume of 18 mL in a 20 mL wedge or 35 mL in a 40 mL wedge, 4 wedges/kit.
The calibrator is a multi-analyte human serum based product containing albumin derived from human serum. The kit consists of 3 vials of one-level calibrator which are lyophilized. The target concentration of this calibrator is 4.3 g/dL. The volume per vial (after reconstitution with deionized water) is 2.0 mL.
10. Test Principle:
The ADVIA Chemistry Albumin BCP Assay is an adaptation of the bromocresol purple (BCP) dye-binding method reported by Carter and Louderback, et al. In the ADVIA Chemistry ALBP Assay, serum or plasma albumin quantitatively binds to BCP to form an albumin-BCP complex that is measured as an endpoint reaction at 596/694 nm.
11. Substantial Equivalence Information:
Predicate device name: Dimension Clinical Chemistry System Albumin Flex® reagent cartridge (ALB) Dimension Clinical Chemistry System TP/ALB Calibrator 2
Predicate K number: K861700
Page 2 of 7
2
| Reagent:
Item | New Device:
ADVIA Chemistry Albumin BCP
Assay (ALPB) | Predicate Device: Siemens
Dimension Clinical
Chemistry System
Albumin Flex® reagent
cartridge (ALB) |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Analyte | Albumin | Same |
| Intended Use/
Indications for Use | For in vitro diagnostic use in the
quantitative measurement of albumin in
human serum or plasma on ADVIA
Chemistry systems.
Albumin measurements are used in the
diagnosis and treatment of numerous
diseases primarily involving the liver or
kidneys. | For in vitro diagnostic use in
the quantitative measurement
of albumin in human serum or
plasma. |
| Instrument to be
used | ADVIA 1650 Chemistry System | Dimension Clinical
Chemistry System |
| Measurement | quantitative | Same |
| Sample type | Serum, Plasma | Same |
| Reference interval | 3.4 - 5.0 g/dL | Same |
| Format | Liquid | Same |
| Use of Calibrators | Yes | Same |
| Analytical
measuring interval | 0.6 - 8.0 g/dL | Same |
| Method Principle | bromocresol purple (BCP) dye-binding
method | Same |
| Reagents | Single reagent | Same |
| Standardization | ERM-DA470k Reference Material | Same |
Calibrator
.
| Item | New Device:
ADVIA Chemistry Albumin BCP
Calibrator | Predicate Device: Siemens
Dimension Clinical
Chemistry System
TP/ALB calibrator |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Intended Use | For in vitro diagnostic use in the
calibration of the ADVIA Chemistry
Albumin BCP Assay (ALBP) on ADVIA
Chemistry systems | for in vitro diagnostic use in
the
calibration of Albumin BCP
Assay |
| Instrument | ADVIA 1650 Chemistry System | Dimension Clinical
Chemistry System |
| Measured Analytes
(value assigned) | Albumin | Albumin, Total Protein |
.
3
Form | Lyophilized | Same |
---|---|---|
Matrix | Human serum | Same |
Analyte source | Human source | Same |
Number of levels | One | Three |
Fill Volume | Reconstitute with 2 mL DI water | Same |
12. Standard/Guidance Document Reference
- Interference Testing in Clinical Chemistry: Approved Guideline Second Edition (CLSI . EP7-A2)
- Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; . Approved Guideline - Second Edition (CLSI EP17-A2)
- Evaluation of Precision Performance of Quantitative Measurement Methods: Approved . Guideline-Second Edition (CLSI EP5-A2)
13. Performance Characteristics
The following data represent typical performance for the ADVIA Chemistry Albumin BCP Assay and were collected on ADVIA 1650 Chemistry system. Substantial equivalence was demonstrated by testing several performance characteristics including precision, method comparison, interfering substances and analytical range. All of the evaluation studies gave acceptable results when compared to the predicate device. These studies support that the ADVIA Chemistry Albumin BCP Assay is substantially equivalent to the predicate device.
1. Precision
Within run and Total Precision were established by Assaying serum sample pools and serum based controls. Each sample was Assaved 2 replicates per run, 2 runs per day, for at least 20 days. Precision estimates were computed according to CLSI document EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline.
MEAN | Within Run | Between Run | Between Day | Total | ||||||
---|---|---|---|---|---|---|---|---|---|---|
Product | N | g/dL | SD | CV | SD | CV | SD | CV | SD | CV |
Serum Control | 80 | 2.7 | 0.03 | 0.9 | 0.00 | 0.0 | 0.03 | 1.2 | 0.04 | 1.5 |
Serum Control | 80 | 4.0 | 0.03 | 0.7 | 0.01 | 0.3 | 0.03 | 0.6 | 0.04 | 1.0 |
Serum Pool | 80 | 3.5 | 0.00 | 0.0 | 0.00 | 0.0 | 0.00 | 0.0 | 0.00 | 0.0 |
Serum Pool | 80 | 5.2 | 0.04 | 0.7 | 0.02 | 0.3 | 0.02 | 0.4 | 0.05 | 0.9 |
2. Linearity/Assay reportable range
Linearity of the Assay on ADVIA 1650 Chemistry system was assessed by evaluating eleven dilutions across the measuring range.. All samples were tested on the ADVIA 1650 Chemistry analyzer. Observed values were compared vs. expected values using linear regression analysis. The low end of the assay range is calculated based on the Limit of Quantitation. The high end of the assay range is based on the linearity calculations.
4
The linear/measuring range of the assay is 0.6 to 8.0 g/dL.
Image /page/4/Figure/1 description: The image is a linear regression plot titled "ADVIA 1650 Albumin BCP - Linearity, Linear Regression (Observed vs. Expected)". The x-axis is labeled "Expected g/dL" and ranges from 0.0 to 10.0, while the y-axis is labeled "Observed g/dL" and also ranges from 0.0 to 10.0. A linear regression line is plotted on the graph, and the equation of the line is given as Y = 1.016 x - 0.05, with R = 1.000 and N = 11.
3. Limit of Blank, Limit of Detection, Limit of Quantitation
The estimations of the Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) were performed according to CLSI document EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, Approved Guideline. The LoB for the ADVIA Chemistry Albumin BCP Assay on the ADVIA 1650 Chemistry system is 0.1 g/dL. Limit of detection (LoD) is the smallest amount that this Assay can reliably detect to determine presence of an analyte. The LoD for for the ADVIA Chemistry Albumin BCP Assay on the ADVIA 1650 Chemistry system is 0.6 g/dL. LoB and LoD values are determined with proportions of false positives (a) less than 5% and false negatives (B) less than 5%, based on 720 determinations with 240 blank and 480 low-level sample replicates. . Limit of Quantitation (LoQ) is 0.6 g/dL based on an inter-assay precision of 10% bias. Bias is the difference in the results between the control sample (without the interferent) and the test sample (contains the interferent) expressed in percent. Bias exceeding 10% is considered interference.
Bilirubin (conjugated and unconjugated) tested up to 60 mg/dL, lipemia up to 525 mg/dL and hemoglobin up to 750 g/dL were shown not to cause significant interference with this assay.
- Reference Interval (Expected Values)
Expected values are within 3.4 - 5.0 g/L **
** Willey DA, Savory J, Lasky F. An Evaluation of a Revised Albumin Method for the aca@ discrete clinical analyzer, Du Pont Company, Wilmington, DE, August 1982
Siemens provides this information for reference. As with all in vitro diagnostic Assays each laboratory should determine its own reference ranges for the diagnostic evaluation of patient results.
8. Reagent and Calibrator Stability:
Reagent: for opened products, once placed on the system reagents are stable for 60 days. The shelf life of the ADVIA Chemistry Albumin BCP Reagent is 12 months at 2-8°C. For unopened product, see the expiration date on the reagent carton.
6
Calibrator: for opened products, once the cap is removed, assigned values are stable for 8 hours when recapped immediately after use and stored at 2-8℃. The shelf life of the ADVIA Chemistry Albumin BCP Calibrator is 18 months at 2-8°C. For unopened product, see the expiration date on the calibrator carton.
The calibrator which is used for the ADVIA Chemistry Albumin BCP assay is the same calibrator (one level only) which was previously cleared as Siemens Dimension Clinical Chemistry System TP/ALB calibrator under K861700 and is relabeled for ADVIA Chemistry use
9. Traceability
The ADVIA Chemistry Albumin BCP Assay is traceable to ERM-DA470k Reference Material. Assigned values of the ADVIA Chemistry Albumin BCP Calibrator are traceable to this standardization
10. Value Assignment
The value assignment protocol for the ADVIA Chemistry Albumin BCP calibrator is the same as for the predicate device which was cleared under K861700.
Target, g/dL | 4.3 |
---|---|
Target Range, g/dL | 4.1 to 4.5 |
14. Conclusions
The ADVIA Chemistry Albumin BCP Assay with associated calibrator on ADVIA 1650 Chemistry system is substantially equivalent in principle and performance to the Siemens Healthcare Diagnostics Dimension Albumin Assay reagent and Dimension TP/ALB calibrator respectively, cleared under K861700.
7
Image /page/7/Picture/0 description: The image shows the logo for the Department of Health & Human Services (USA). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged around the perimeter. Inside the circle is a stylized emblem featuring three angled shapes resembling a bird in flight or a series of waves.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 16, 2013
Siemens Healthcare Diagnostics, Inc. C/O Kira Gordon, Ph.D. 511 Benedict Ave. TARRYTOWN NY 10591
Re: K132664
Trade/Device Name: ADVIA Chemistry Albumin BCP Assay (ALBP) ADVIA Chemistry Albumin BCP Calibrator Regulation Number: 21 CFR 862.1035 Regulation Name: Albumin test system Regulatory Class: II Product Code: CJW, JIX Dated: August 26, 2013 Received: August 27, 2013
Dear Dr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
8
Page 2-Dr. Gordon
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincercly yours,
Carol C. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
9
Indications for Use
510(k) Number (if known): K132664
Device Name:
ADVIA Chemistry Albumin BCP Assay (ALBP) ADVIA Chemistry Albumin BCP Calibrator
Indications for Use:
For in vitro diagnostic use in the quantitative measurement of albumin in human serum or plasma on ADVIA Chemistry systems. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys.
For in vitro diagnostic use in the calibration of the ADVIA Chemistry Albumin BCP Assay (ALBP) on ADVIA Chemistry Systems.
Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Image /page/9/Picture/12 description: The image shows the name "Yung W. Chan -S" in a bold, sans-serif font. The letters are large and clearly legible. The name appears to be a signature or heading, possibly indicating the author or subject of a document.
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K132664