K Number
K132553
Device Name
U&U STERILE PISTON SYRINGE WITHOUT NEEDLE
Date Cleared
2014-02-27

(197 days)

Product Code
Regulation Number
880.5860
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
U&U Sterile Piston Syringes is intended for use by health care professionals for general purpose fluid aspiration/ injection. U&U Sterile Piston Syringes is intended for use by health care professionals for general purpose fluid aspiration.
Device Description
The piston syringe is a device intended for medical purposes, consisting of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male Luer Slip/Lock connector (nozzle) for attaching the female Luer connector (hub) of a hypodermic single lumen needle, or for attaching other devices with a female Luer. The syringe is sterilized by EtO gas. And it is a Non-Pyrogenic and single use device. The mainly raw materials are PP, PE and rubber.
More Information

Not Found

No
The device description and intended use are for a standard, mechanical piston syringe. There is no mention of any software, algorithms, or data processing that would indicate the use of AI or ML. The "Mentions AI, DNN, or ML" and "Mentions image processing" sections are explicitly marked as "Not Found".

No
The device is described as a general-purpose fluid aspiration/injection device, which is not considered a therapeutic function. Therapeutic devices are typically used for treatment or therapy.

No.

The device description clearly states its intended use is for "general purpose fluid aspiration/injection," which is a procedural function, not a diagnostic one. It does not mention any role in identifying, measuring, or analyzing medical conditions or disease states.

No

The device description clearly describes a physical medical device (syringe) made of materials like PP, PE, and rubber, and mentions sterilization and physical components like a barrel and plunger. It does not describe software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for "general purpose fluid aspiration/injection" and "general purpose fluid aspiration." This describes a physical action of moving fluids into or out of the body, not the testing of samples in vitro (outside the body) to diagnose a condition.
  • Device Description: The description focuses on the physical components of a syringe (barrel, plunger, Luer connector) and its sterilization method. It doesn't mention any reagents, test strips, or other components typically associated with IVD devices used for analyzing biological samples.
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), detecting specific analytes, or providing diagnostic information.

IVD devices are specifically designed to perform tests on samples taken from the human body to provide information for the diagnosis, monitoring, or treatment of diseases or conditions. This syringe is a tool for administering or withdrawing fluids, which is a different function.

N/A

Intended Use / Indications for Use

U&U Sterile Piston Syringes is intended for use by health care professionals for general purpose fluid aspiration/ injection

Product codes

FMF

Device Description

The piston syringe is a device intended for medical purposes, consisting of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male Luer Slip/Lock connector (nozzle) for attaching the female Luer connector (hub) of a hypodermic single lumen needle, or for attaching other devices with a female Luer. The syringe is sterilized by EtO gas. And it is a Non-Pyrogenic and single use device. The mainly raw materials are PP, PE and rubber.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

health care professionals

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K110771

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).

0

132ss3

U&U Medical Technology Co., Ltd Dongzhou Village, Hengshangiao, Changzhou, Jiangsu, China U&U (HONGKONG) Medical Technology Co., Limited RM C1-D SF WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG [U&U Piston Syringe]

510(k) Submission

FEB 2 7 2014

12/08/13 Rev 0.00

Section_005 510(K) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92

Date Prepared: 12, 08.2013

1. Submitter Name and Address:

Owner Name:U&U Medical Technology Co., Ltd
Address:Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China
RM C1-D 6/F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN
HONG KONG
Contactor Name:Xuebo Wang
TEL:+86-13564751751
E-mail:Blackwang@tkmedical.com

Contract Manufacturer Name: ANHUI TIANKANG MEDICAL PRODUCTS CO., LTD. No 20 south renhe road tianchang, CHINA 239300 Address: Web: www.tkmedical.com

US Agent:

US Agent:Pan Angels Corp.
Address:3330 Fowler Street, Los Angeles, CA 90063,U.S.A
TEL:(323)422-8581
Contact person :Mr. Michael Kim

2. Submission Devices Information:

Trade/Proprietary Name: U&U Sterile Piston Syringe without needle Common Name: Piston Syringe Classification name:Piston Syringe. Class: II. Panel: 80. Procodes: FMF - Piston Syringe

3. Predicate Devices Information:

    1. Piston Syringe; Trade Name: 510(K) Number:
      BD Single Use, Hypodermic Syringe K110771

4. Devices Description:

Sterile Piston Syringes

The piston syringe is a device intended for medical purposes, consisting of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male Luer Slip/Lock connector (nozzle) for attaching the female Luer connector (hub) of a hypodermic single lumen needle, or for attaching other devices with a female Luer. The syringe is sterilized by EtO gas. And it is a Non-Pyrogenic and single use device. The mainly raw materials are PP, PE and rubber

00016

Page 1 of 3

[PMN-510(K) SUBMISSION]

.

1

U&U Medical Technology Co., Ltd Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China U&U (HONGKONG) Medical Technology Co., Limited RM C1-D SF WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG [U&U Piston Syringe]

| Ref
Number | Model
Number | Description | Size |
|---------------|-----------------|----------------------------|---------|
| TKSLS001 | TKSLS | Piston syringe (LUER SLIP) | 1cc/ml |
| TKSLS002 | TKSLS | Piston syringe (LUER SLIP) | 2cc/ml |
| TKSLS003 | TKSLS | Piston syringe (LUER SLIP) | 3cc/ml |
| TKSLS004 | TKSLS | Piston syringe (LUER SLIP) | 5cc/ml |
| TKSLS005 | TKSLS | Piston syringe (LUER SLIP) | 10cc/ml |
| TKSLS006 | TKSLS | Piston syringe (LUER SLIP) | 20cc/ml |
| TKSLS007 | TKSLS | Piston syringe (LUER SLIP) | 30cc/ml |
| TKSLS008 | TKSLS | Piston syringe (LUER SLIP) | 50cc/ml |
| TKSLS009 | TKSLS | Piston syringe (LUER SLIP) | 60cc/ml |
| TKSLL001 | TKSLL | Piston syringe (LUER LOCK) | 1cc/ml |
| TKSLL002 | TKSLL | Piston syringe (LUER LOCK) | 2cc/ml |
| TKSLL003 | TKSLL | Piston syringe (LUER LOCK) | 3cc/ml |
| TKSLL004 | TKSLL | Piston syringe (LUER LOCK) | 5cc/ml |
| TKSLL005 | TKSLL | Piston syringe (LUER LOCK) | 10cc/ml |
| TKSLL006 | TKSLL | Piston syringe (LUER LOCK) | 20cc/ml |
| TKSLL007 | TKSLL | Piston syringe (LUER LOCK) | 30cc/ml |
| TKSLL008 | TKSLL | Piston syringe (LUER LOCK) | 50cc/ml |
| TKSLL009 | TKSLL | Piston syringe (LUER LOCK) | 60cc/ml |

5. Intended Use:

Sterile Piston Syringes

U&U Sterile Piston Syringes is intended for use by health care professionals for general purpose fluid aspiration/ injection

6. Technological Characteristics:

Through compansons between the submitted devices with the predicate devices as follows tables. We believe the applicant devices are substantially equivalent with the predicate devices.

Sterile Piston Hypodermic Syringes Comparison Table

| Element of Comparison | Submission Device | Predicate Device
K110771 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | U&U Sterile Piston Syringes
is intended for use by health
care professionals for
general purpose fluid
aspiration/ injection | The BD Single Use,
Hypodermic Syringe is
intended for use by health
care professionals for general
purpose fluid aspiration/
injection. |
| Principle of Operation | Normal | Normal |
| Syringe Capacity | Various Sizes | Various Sizes |
| Nozzle Type | Luer Slip & Luer Lock | Luer Slip & Luer Lock |
| Lubricant for Barrel | Silicone Oil | Silicone Oil |
| Barrel Transparency | Transparent and Clear | Transparent and Clear |
| Gradations Legibility | Legible | Legible |

[PMN-510(K) SUBMISSION]

Page 2 of 3

000017

2

U&U Medical Technology Co., Ltd Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China U&U (HONGKONG) Medical Technology Co., Limited RM C1-D 8/F WING HING IND BLDG 14 HING YIP ST KWUN TONG KLN HONG KONG [U&U Piston Syringe]

Rev 0.00 12/08/13

Materials
BarrelPPPP
PlungerPEPE
PistonRubberRubber
PerformancesConforms to ISO7886-1Conforms to ISO7886-1
BiocompatibilityConforms to ISO10993Conforms to ISO10993
LabelingMeet the requirements of 21 CFR Part 801Meet the requirements of 21 CFR Part 801

7. Conclusion:

The materials, performance, and operational features of both the submitted device and the predicate device are substantially equivalent.

END

Kubo usac

Aug. 12. 2013

[PMN-510(K) SUBMISSION]

Pege 3 of 3

3

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 27, 2014

U&U Medical Technology Company, Limited Xuebo Wang Regulatory Affairs Manager Dongzhou Village, Hengshanqiao Town, Changzhou Jiangsu, China 213119

Re: K132553

Trade/Device Name: U&U Sterile Piston Syringe without Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: December 8, 2013 Received: January 29, 2014

Dear Mr. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

4

Page 2 - Mr. Wang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Image /page/4/Picture/6 description: The image shows the name "Kwame Ulmer's" in a stylized font. The letters are black and bold, and they are arranged in two rows. The first row contains the word "Kwame", and the second row contains the word "Ulmer's". The letters are partially obscured by a geometric design that is made up of lines, circles, and triangles. The design is also black and bold, and it is arranged in a way that it appears to be behind the letters.

for

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K132553

Device Name

U&U Sterile Piston Syringes Without Needle

Indications for Use (Describe)

U&U Sterile Piston Syringes is intended for use by health care professionals for general purpose fluid aspiration.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

Image /page/5/Figure/11 description: The image shows a document with the heading "FOR FDA USE ONLY" and the title "Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)". It includes the FDA logo and indicates that the document was digitally signed by Richard C. Chapman on February 25, 2014, at 09:02:57 -05'00'. The document is identified as FORM FDA 3881 (9/13) and is labeled as Page 1 of 2.

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

6

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."