(504 days)
The Vital Signs Patch (VSP) system is intended to be used on patients in a clinical environment for the continuous, non-invasive monitoring of ECG, Heart Rate (HR), respiration rate, surface temperature, and arterial blood oxygen saturation (intended use group adults 21 and above), when prescribed by a physician or other qualified healthcare professional.
The Vital Signs Patch (VSP) system is designed to monitor vital signs of patients. The internal LifeWatch Technologies Ltd. (LWT in short) Part Number of the VSP is CG-1101B. The VSP system consists of the following components: Vital Signs Patch, VSP transmitter ("Brain"), Gateway, Clinical Backend.
Here's a breakdown of the acceptance criteria and study information for the Vital Signs Patch system, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it states that the device was subjected to extensive verification/validation testing to ensure it meets all functional and performance requirements and is fit for its intended use. The performance details are primarily linked to the adherence to voluntary performance standards and a specific clinical trial for SpO2.
Here's an attempt to infer and summarize based on the provided text, linking performance to the mentioned standards:
| Parameter Monitored | Acceptance Criteria (Inferred from Standards) | Reported Device Performance |
|---|---|---|
| ECG, Heart Rate | ANSI/AAMI/ISO EC57:1998 (R) 2008: Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms | "Software Functional Unit Verification", "System Level Software Validation", "Arrhythmia Detection Algorithm Performance Validation" all performed to ensure the device meets functional and performance requirements and is fit for its intended use, conforming to EC57 and EC38 standards. |
| Respiration Rate | Implicitly covered by general functional and performance requirements for vital signs monitoring. | "Software Functional Unit Verification", "System Level Software Validation" to ensure the device meets functional and performance requirements and is fit for its intended use. ECG electrodes used for impedance measurement. |
| Surface Temperature | ISO 80601-2-56:2009: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature | "Software Functional Unit Verification", "System Level Software Validation" to ensure the device meets functional and performance requirements and is fit for its intended use, conforming to ISO 80601-2-56. |
| SpO2 | ISO 80601-2-61:2011: Particular requirements for basic safety and essential performance of pulse oximeter equipment | "Clinical trial of the SPO2 function" was performed. Conforms to ISO 80601-2-61. Specific results are not provided in this summary. The device's SpO2 function is stated to be substantially equivalent to the cleared MD300W4 Beijing Wrist Pulse Oximeter. |
| General Safety | IEC 60601-1:2005 (Safety), IEC 60601-1-2:2007 (EMC), ISO 14971:2007 (Risk Mgt), ISO 10993-1:2009 (Biocompatibility) | Electrical Safety, EMC, and FCC testing by an independent, certified, external test laboratory. Conforms to IEC 60601-1, IEC 60601-1-2, ISO 14971, and ISO 10993-1. |
| Software | IEC 62304:2006: Medical Device Software Software Lifecycle Processes, FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11 2005" | "Software Verification and Validation", "Software Functional Unit Verification", "System Level Software Validation", "Arrhythmia Detection Algorithm Performance Validation". Conforms to IEC 62304 and FDA software guidance. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document explicitly mentions a "Clinical trial of the SPO2 function." However, it does not specify the sample size for this clinical trial or for any other performance test.
- Data Provenance: The document does not specify the country of origin of the data. It also does not specify if the studies were retrospective or prospective, although "Clinical trial of the SpO2 function" typically implies a prospective study.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not provide information on the number of experts used or their qualifications for establishing ground truth for any of the tests, including the SpO2 clinical trial.
4. Adjudication Method
The document does not specify any adjudication methods (e.g., 2+1, 3+1, none) used for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being performed, nor any effect size of human readers improving with or without AI assistance. The device is for monitoring vital signs, not for assisting human interpretation of complex medical images or data that would typically involve a MRMC study.
6. Standalone Performance Study
The document indicates that studies were performed on the device's functional and performance requirements. The Arrhythmia Detection Algorithm Performance Validation and the Clinical trial of the SPO2 function suggest standalone performance evaluations of these specific features, focusing on the algorithm's accuracy in detecting arrhythmias and the SpO2 sensor's performance against a reference. While not explicitly termed "standalone performance study," the objective of these tests, especially against recognized standards, inherently assesses the algorithm/device performance without constant human intervention in data collection during the measurement phase. The device then presents data to a human for diagnosis.
7. Type of Ground Truth Used
- For SpO2: A "Clinical trial" was conducted. In such trials for pulse oximetry, the ground truth is typically established by comparing the device's readings against arterial blood gas (ABG) measurements from a co-oximeter, which is considered the gold standard for SpO2. This is implied by the nature of such a clinical trial conforming to ISO 80601-2-61.
- For ECG/Arrhythmia Detection: The ground truth for algorithm performance validation against standards like ANSI/AAMI/ISO EC57 often involves using truth-labeled ECG databases (e.g., MIT-BIH Arrhythmia Database) or synchronously recorded and extensively reviewed ECGs by expert cardiologists.
- For other vital signs (HR, Respiration, Temperature): Ground truth would typically be established by comparing the device's readings against established reference methods or calibrated medical devices known for their accuracy.
8. Sample Size for the Training Set
The document does not provide any information regarding the sample size used for the training set for any of the algorithms or features.
9. How the Ground Truth for the Training Set Was Established
The document does not provide any information on how the ground truth for the training set was established.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2014
Lifewatch Technologies Ltd. Asher Kassel Director of RA and QA 2 Pekeris Street Rehovot, 7670202, Israel
Re: K132407
Trade/Device Name: Vital Signs Patch System (VSP) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: DSI, MHX, DQA, FLL Dated: December 10, 2014
Received: December 11, 2014
Dear Asher Kassel,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Traditional 510(k) for Vital Signs Patch Section 002: Indications for Use Statement
Indications for Use
510(k) Number (if known): K132407
Device Name: Vital Signs Patch
Indications for Use
The Vital Signs Patch (VSP) system is intended to be used on patients in a clinical environment for the continuous, non-invasive monitoring of ECG, Heart Rate (HR), respiration rate, surface temperature, and arterial blood oxygen saturation (intended use group adults 21 and above), when prescribed by a physician or other qualified healthcare professional.
Contraindications:
- o The VSP is not intended for use by persons with any type of defibrillator, external or internal (ICD); the VSP must be detached from the patient before using a defibrillator on the patient
- The VSP is not to be used in a magnetic resonance imaging (MRI) environment. The VSP device must be removed from the patient's skin before he/she undergoes MRI procedure
- 0 The VSP is not a "life-saving" or therapeutic device; the VSP supplies vital signs data to a doctor or technician for the purpose of diagnosis by such (or other qualified) personnel
- 0 The VSP is not intended for use on patients with unhealed surgical incision/dressings on the thoracic or abdominal regions
- The VSP is not intended for use on patients with skin or soft tissue damage on the area where the o VSP is placed (such as burns, irritation, infections, wounds, etc.)
- o The VSP is not intended for use in the diagnosis of myocardial infarction or for chest pain monitoring.
Warning: The temperature function of the Vital Signs Patch measures and reports surface temperature on the upper chest where direct measurements of body core temperature are required, it is recommended to utilize appropriate core temperature monitoring devices for this purpose.
| Prescription Use X | AND/OR | Over-The-Counter Use _ |
|---|---|---|
| (Part 21 CFR 801 Subpart D) | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for LifeWatch Technologies. The logo consists of a pattern of dots on the left, followed by the text "LifeWatch" in a stylized font. Below "LifeWatch" is the word "Technologies" in a smaller font. There is a small star to the right of the word "LifeWatch".
510(k) Summary: Vital Signs Patch system
Introduction
This document contains the 510(k) summary for the Vital Signs Patch system, Part Number CG-1101B. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).
| Submitter | LifeWatch Technologies Ltd. | ||
|---|---|---|---|
| EstablishmentRegistration Number | 9681879 | ||
| Address | 2 Pekeris St., P.O.B. 527, Rehovot, 7610303, Israel | ||
| Contact person: | Asher Kassel, Director of RA & QA, LifeWatch Technologies Ltd. | ||
| Phone: | 972-8-9484010(direct) | Fax: | 972-8-9484044 |
| E-mail: | asherk@lifewatch.com | ||
| Date Prepared: | June 14, 2014 | ||
| Primary predicatedevice | PROPAQ LT Vital Signs Monitor, manufactured by Welch AllynProtocol, Inc. and cleared in K033378 on March 1st, 2004. | ||
| Secondary predicatedevice | VS200 Aingeal Physiological Patient Monitor manufactured byIntelesens Ltd.; cleared in K110015 on May 20, 2011. | ||
| SP02 predicatedevice | MD300W4 Beijing Wrist Pulse OximeterChoice Electronic Technology Co., Ltd. cleared in K122046 | ||
| Trade Name: | Vital Signs Patch system | ||
| Regulation No: | 21 CFR 870.1025; 870.2700 and 880.2910 | ||
| Classification: | Class II | ||
| Product Codes: | DSI; MHX; DQA; FLL |
1.1. Device Description and Summary of the Technological Characteristics / Principles of Operation
The Vital Signs Patch (VSP) system is designed to monitor vital signs of patients. The internal LifeWatch Technologies Ltd. (LWT in short) Part Number of the VSP is CG-1101B. The VSP system consists of the following components:
Vital Signs Patch
A Vital Signs Patch (VSP Patch) – A Polyethylene substrate coated with biocompatible adhesive with an ECG sensor with four electrodes (3 leads), a temperature sensor, a blood oxygen saturation sensor, a respiration rate sensor (utilizes the ECG electrodes for impedance measurement), a 3.0V Li-MnO2 battery to activate the Brain and a cradle to hold the Brain.
VSP transmitter ("Brain")
The VSP transmitter (also called "Brain") is the main Processing & Transmitting Unit of the VSP device; the Brain has a 16-pin connector that is plugged into the cradle of the Patch. Once the Brain is inserted into the cradle of the Patch, it receives power from the battery installed in the Patch and starts to operate. The Brain
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for LifeWatch Technologies. The logo consists of a pattern of dots on the left, followed by a curved line and the words "LifeWatch" in a serif font. Below "LifeWatch" is the word "Technologies" in a smaller, sans-serif font. There is a small star to the right of the word "LifeWatch".
acquires all signals coming from the sensors embedded in the Patch.
The Brain integrates a low power RF component (Nordic) for short-range communication with the Gateway. The Brain is also equipped with Red and Green LED's and a buzzer to visually and audibly indicate system status.
Gateway
The Gateway is a dedicated SW application designed by LifeWatch Technologies (LWT) that runs on a LifeWatch Technologies approved mobile device running Android OS v 2.3.5 or above, and equipped with Nordic nRF24L series 2.4 GHz RF communication chip. The Gateway includes a user interface for alarms (such as low arterial oxygen saturation) and monitoring. Communication with the Clinical Backend is performed using wireless (3G cellular or Wi-Fi) technology.
Clinical Backend
The VSP Clinical Backend (VSP CBE or CBE) receives vital signs data from the VSP Gateway, processes them and presents the data on a Clinical console. The VSP system is capable of processing and storing patient vital signs in a centralized database; the length of the recording is limited by the available storage capacity.
1.2. Indications for Use
The Vital Signs Patch system is intended to be used by patients for the continuous, noninvasive monitoring of ECG, Heart Rate (HR), respiration, surface temperature and arterial blood oxygen saturation (intended use group adults 21 and above), when prescribed by a physician or other qualified healthcare professional.
1.3. Contraindications
Contraindications for Use:
- The VSP is not intended for use by persons with any type of defibrillator, external or ● internal (ICD); the VSP must be detached from the patient before using a defibrillator on the patient
- . The VSP is not to be used in a magnetic resonance imaging (MRI) environment. The VSP device must be removed from the patient's skin before he/she undergoes MRI procedure
- The VSP is not a "life-saving" or therapeutic device: the VSP supplies vital signs data . to a doctor or technician for the purpose of diagnosis by such (or other qualified) personnel
- The VSP is not intended for use on patients with unhealed surgical incision/dressings . on the thoracic or abdominal regions
- . The VSP is not intended for use on patients with skin or soft tissue damage on the area where the VSP is placed (such as burns, irritation, infections, wounds, etc.)
- . The VSP is not intended for use in the diagnosis of myocardial infarction or for chest pain monitoring.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for LifeWatch Technologies. The logo consists of a triangular shape made up of small dots on the left, followed by a curved line resembling an ear. To the right of the ear-like shape is the text "LifeWatch" in a bold, sans-serif font, with a small star symbol next to it. Below "LifeWatch" is the word "Technologies" in a smaller, lighter font.
Warning: The temperature function of the Vital Signs Patch measures and reports surface temperature on the upper chest where direct measurements of body core temperature are required, it is recommended to utilize appropriate core temperature monitoring devices for this purpose.
1.4. Clinical and non-clinical performance data for the Vital Signs Patch svstem
The Vital Signs Patch system has been subjected to extensive verification / validation testing. Final testing of the system included various performance tests and software validation tests designed to ensure that the device meets all of its functional and performance requirements and is fit for its intended use. The following list summarizes the testing performed on the device:
- Software Verification and Validation .
- Software Functional Unit Verification O
- System Level Software Validation O
- O Arrhythmia Detection Algorithm Performance Validation
- Hardware Verification ●
- Clinical trial of the SPO2 function ●
- Electrical Safety, EMC, SPO2 and FCC testing by an independent, certified, external test laboratory.
1.5. Voluntary Performance Standards:
This 510(k) submission was written in accordance with the FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11 2005". The design of the Vital Signs Patch system conforms to the following voluntary standards:
-
- ANSI/AAMI/ISO EC57:1998 (R) 2008: Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms (FDA Recognition Number: 3-73 )
-
- ANSI/AAMI EC38:2007: Ambulatory Electrocardiograph (FDA Recognition Number: 3-65 )
-
- ISO 14971:2007: Medical devices application of risk management to medical devices (FDA Recognition Number: 5-40)
-
- IEC 60601-1:2005, 300 edition: Medical electrical equipment; Part 1: General requirements for safety (FDA Recognition Number: 5-4)
-
- ISO 10993-1:2009: Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process (FDA Recognition Number: 2-179)
-
- IEC 60601-1-2:2007, Medical electrical equipment; Part 1: 2. Collateral Std.: EMC; requirements and tests (FDA Recognition Number: 5-53)
-
- ISO 80601-2-61:2011: Medical electrical equipment -- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment (FDA Recognition Number: 1-85)
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for LifeWatch Technologies. The logo consists of a blue, stylized image of a radar or sonar-like wave on the left, followed by the words "LifeWatch" in a larger, bolder blue font. Below "LifeWatch" is the word "Technologies" in a smaller, lighter blue font. A small star is to the right of the word "LifeWatch".
-
- ISO 80601-2-56:2009, Medical electrical equipment Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature (FDA Recognition Number: 6-232)
-
- IEC 62304:2006, Medical Device Software Software Lifecycle Processes (FDA Recognition Number: 13-8)
Substantial Equivalence:
The Vital Signs Patch system is substantially equivalent with respect to indications for use, technological characteristics and performance characteristics to the identified legally marketed predicate devices.
§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).
(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.