(207 days)
The MD300W4 wrist pulse oximeter is a portable non-invasive device intended for spot checking ,data collection and recording of arterial oxygen saturation (SpO2) and pulse rate of adult and pediatric patient at home and hospital( including clinical use in internist/ surgery, Anesthesia, intensive care and etc).
The proposed device MD300W4 consists of MCU circuit, power supply circuit, SpO2 module circuit, display circuit, Flash memory circuit, GPRS module circuit, charging circuit, real-time clock circuit, button circuit.
It can measure, store, review and display the SpO2% and pulse rate value, time, ID number, pulse bar and battery power status, the connection of probe, and transmit data by GPRS or USB cable.
The power supply is 4.2V Li-battery with capacity 1250mAh. The device can not be used to measure when it charges for the Li- battery.
The Pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and a photodetector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is infrared light.
Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in probe collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The proposed device has one model detachable sensor as the accessory of which the skin-contacting material is silicon. It is listed below:
M-50G (510(k) number: K082487)
The proposed device is not for life-supporting or life-sustaining, not for implant. The device or probe is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological products.
The device is software-driven and the software validation is provided in Section 14 Software on page 14-1.
Here's an analysis of the provided text to fulfill your request:
Acceptance Criteria and Device Performance Study for Wrist Pulse Oximeter MD300W4 (K122046)
Based on the provided 510(k) summary, the device is a modified Wrist Pulse Oximeter, MD300W4, and its performance is evaluated against the predicate device, MD300W (K081125), and established standards.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the MD300W4 are largely aligned with its predicate device and industry standards for pulse oximeters. The reported device performance is presented as compliance with these specifications.
| Characteristic | Acceptance Criteria (Target/Standard) | Reported Device Performance (MD300W4 Results) |
|---|---|---|
| SpO2 Accuracy | 70%-100% ±3% (General) | 70%-100% ±3% (Claimed by manufacturer) |
| 90%-100% ±2%, 80%-90% ±3%, 70%-80% ±3% (Specific ranges) | 90%-100% ±2%, 80%-90% ±3%, 70%-80% ±3%, <70% unspecified (Matches predicate and alleged performance) | |
| PR Accuracy | ±2% or 2 bpm, whichever is greater (General) | ±2% or 2 bpm, whichever is greater (Matches predicate and alleged performance) |
| SpO2 Display Range | 0%-100% | 0%-100% |
| SpO2 Measuring Range | 70%-100% | 70%-100% |
| SpO2 Resolution | 1% | 1% |
| PR Measuring Range | 30~235bpm | 30~235bpm |
| PR Resolution | 1bpm | 1bpm |
| Biocompatibility | Compliance with ISO 10993 "Biological Evaluation of Medical Devices" | Meets the requirement of Biocompatibility, in compliance with ISO 10993. |
| Software Validation | Compliance with FDA Guidance for "Content of Premarket Submissions for Software Contained in Medical Devices" | In compliance with FDA Guidance. |
| Electrical Safety | IEC 60601-1:1988+A1:1991+A2:1995 | Designed, tested, and will be manufactured in accordance with IEC 60601-1. |
| EMC | IEC 60601-1-2:2007 | Designed, tested, and will be manufactured in accordance with IEC 60601-1-2. |
| Performance (Pulse Oximeters) | ISO 9919:2005 | Designed, tested, and will be manufactured in accordance with ISO 9919:2005. Clinical test conducted according to ISO9919:2005 Annex EE. |
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size for Test Set: 12 health adults.
- Data Provenance: Prospective clinical test conducted in Yue Bei People's Hospital. The country of origin can be inferred as China, given the manufacturer's location and the hospital name.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that SpO2 values were compared to "arterial blood sample oxygen saturation (functional SaO2) as measured by CO-Oximetry," referring to this as the "Golden Standard Co-Oximeter." This implies a technical measurement-based ground truth rather than expert consensus on subjective interpretations.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method. The ground truth is established directly by a "Golden Standard" Co-Oximeter for arterial blood sample oxygen saturation, rather than through subjective expert review requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is a pulse oximeter, which directly measures physiological parameters (SpO2 and pulse rate) and does not involve "human readers" interpreting images or data in the context of an "AI" system that would enhance human performance. The study described is a clinical accuracy study comparing the device's readings against a reference method.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, a standalone performance evaluation was done. The clinical test described involved comparing the device's SpO2 measurements directly against arterial blood sample oxygen saturation measured by a CO-Oximeter. This assesses the device's algorithmic performance in isolation (determining SpO2 from its sensor input) without human intervention in the measurement process itself, beyond the initial setup.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The type of ground truth used was objective measurement data from a "Golden Standard" Co-Oximeter measuring functional SaO2 from arterial blood samples. This is considered a highly accurate and direct physiological measurement.
8. The Sample Size for the Training Set
The document does not mention a separate training set or its sample size. The clinical test described appears to be for validation/testing, not for training a machine learning model. For a traditional medical device like a pulse oximeter, the core algorithm for converting optical signals to SpO2 and PR is typically based on established physiological models and calibration, rather than iterative machine learning training on a large dataset in the way an AI diagnostic algorithm would be.
9. How the Ground Truth for the Training Set Was Established
As no separate training set is mentioned (or implied as relevant for this type of device's core function), the method for establishing ground truth for a training set is not applicable based on the provided information.
{0}------------------------------------------------
Exhibit F 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) number is: ___K122046
Statement
Wrist Pulse Oximeters MD300W4 is a modification device to Wrist Pulse Oximeter
| MD300W4 | MD300W | |
|---|---|---|
| Display Creen | OLED | Segment LCD |
| Transmission mode | USB and GPRS | USB |
| Power supply | lithium-ion rechargeable battery | 1 AAA |
| Enclosure | 133mm×63mm×33mm | 60mm×50mm×20mm |
MD300W (K081125). Their differences between them are listed below:
1. 510(k) Owner's Information
Establishment Registration Number: 3005569927
Beijing Choice Electronic Technology Co., Ltd.
Room 1127-1128 Building B, Bailangyuan
Fuxing Road, No. A36,
100039 Beijing, China
Contact Person: Mr. Lei Chen /Quality Director
Phone: +86-10-88790480
Fax: 215-4052545
Email: cc@choicemmed.com
2. Prepared Date: Aug. 8, 2012
3. Proposed Device Information
Device Common Name: Pulse Oximeter -
Device Trade Name: Wrist Pulse Oximeter
Model: MD300W4
{1}------------------------------------------------
Classification Name: Oximeter
Class: II
Regulation Number: 21 CFR 870.2700
Product Code: DQA
Panel: Anesthesiology
4. Predicate Device
Wrist Pulse Oximeter MD300W (K081125)
Beijing Choice Electronic Technology Co., Ltd.
Room 1127-1128 Building B, Bailangyuan
Fuxing Road, No. A36
100039 Beijing, China
5. Device Description
The proposed device MD300W4 consists of MCU circuit, power supply circuit, SpO2 module circuit, display circuit, Flash memory circuit, GPRS module circuit, charging circuit, real-time clock circuit, button circuit.
It can measure, store, review and display the SpO2% and pulse rate value, time, ID number, pulse bar and battery power status, the connection of probe, and transmit data by GPRS or USB cable.
The power supply is 4.2V Li-battery with capacity 1250mAh. The device can not be used to measure when it charges for the Li- battery.
The Pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and a photodetector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is infrared light.
Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in probe collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The proposed device has one model detachable sensor as the accessory of which the skin-contacting material is silicon. It is listed below:
M-50G (510(k) number: K082487)
The proposed device is not for life-supporting or life-sustaining, not for implant. The device
2
{2}------------------------------------------------
or probe is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological products.
The device is software-driven and the software validation is provided in Section 14 Software on page 14-1.
| Proposed Device | Predicate Device | |
|---|---|---|
| ITEM | Wrist Pulse OximeterMD300W4 | Wrist Pulse Oximeter MD300W(K081125) |
| Composition | A main unit, SpO2 sensor | A main unit, SpO2 sensor |
| Design principle | Skin, bone, tissue, and venousvessels normally absorb a constantamount of light over time. Thephotodetector in probe collects andconverts the light (660nm red lightand 940nm near infrared light)into electronic signal which isproportional to the light intensity.The arteriolar bed normallypulsates and absorbs variableamounts of light during systole anddiastole, as blood volume increasesand decreases. The ratio of lightabsorbed at systole and diastole istranslated into an oxygensaturation measurement. Thismeasurement is referred to asSpO2. | Skin, bone, tissue, and venousvessels normally absorb a constantamount of light over time. Thephotodetector in probe collects andconverts the light (660nm red lightand 940nm near infrared light) intoelectronic signal which isproportional to the light intensity.The arteriolar bed normally pulsatesand absorbs variable amounts oflight during systole and diastole, asblood volume increases anddecreases. The ratio of lightabsorbed at systole and diastole istranslated into an oxygen saturationmeasurement. This measurement isreferred to as SpO2 |
| Patient ContactMaterial | Medical Silica gel and Nylon | Medical Silica gel and Nylon |
| Displayunitspecification | OLED | Segment LCD |
| Working time | Work for 10h continuously | Work for 12h continuously |
| Power | 4.2V Li-battery | 1 *AAA |
| Display Parameter | SpO2, pulse rate | SpO2, pulse rate |
| SpO2 displayrange | 0%-100% | 0%-100% |
| SpO2 measuringrange | 70%-100% | 70%-100% |
| SpO2 Resolution | 1% | 1% |
| SpO2 Accuracy | 70%-100%+3% | 70%-100%+3% |
| 90%-100%±2%, 80%-90%±3%, 70%-80%±3%, <70% unspecified | 90%-100%±2%, 80%-90%±3%, 70%-80%±3%, <70% unspecified | |
| PR Measuring | 30~235bpm | 30~235bpm |
| Range | ||
| PR resolution | 1bpm | 1bpm |
| PR accuracy | ±2% or 2 bpm, whichever isgreater | 30-99bpm, ± 2 bpm;100-235: ±2% |
| Data record | 72-hour | 72-hour |
| Transmissionmode | USBGPRS | USB |
| Operatingtemperature | 5°C~40°C | 5°C~40°C |
| Relativehumidity | ≤80% | ≤80% |
| Sensor | M-50G | M-50G |
| SpO2 module | CSN604 | CSN604 |
6. Comparison list of the technological characteristics
3
{3}------------------------------------------------
7. Intended use
The MD300W4 wrist pulse oximeter is a portable non-invasive device intended for spot checking ,data collection and recording of arterial oxygen saturation (SpO2) and pulse rate of adult and pediatric patient at home and hospital( including clinical use in internist/ surgery, Anesthesia, intensive care and etc).
8. Non-clinical Test
The MD300W4 wrist pulse oximeter is designed, tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
- IEC 60601-1:1988+A1:1991+A2:1995, Medical Electrical Equipment Part1: . General requirements for safety.
- IEC 60601-1-2:2007, Medical Electrical Equipment Part 1-2: General . requirements for basic safety and essential performance, Collateral standard: Electromagnetic compatibility - Requirements and tests.
- ISO 9919:2005, Medical electrical equipment Particular requirements for the basic . safety and essential performance of pulse oximeter equipment for medical use
- FCC 47 CFR Part 15B, Part 22 Subpart H and Part 24 Subpart E .
- ANSI/IEEE Std. C95.1-1992 in accordance with the requirements of FCC Report . and Order: ET Docket 93-62, and OET Bulletin 65 Supplement C
4
{4}------------------------------------------------
The Software Validation is in compliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices".
The test results indicate that the safety and effectiveness of the proposed devices is identical to that of the predicate device.
9. Clinical Test
The clinical test is conducted according to ISO9919:2005 Annex EE in Yue Bei Pepople's Hospital. 12 health adults are selected as the subjects. The system is evaluated during steady state / non-motion conditions with various levels of induced hypoxia resulting in stable oxygen saturation levels between 100% and 70% SaO2. The SpO2 are compared to arterial blood sample oxygen saturation (functional SaO2) as measured by CO-Oximetry.
There are no adverse events during the study. It can be determined from the study results that the SpO2 accuracy of the proposed device is compliance to the specification claimed by the manufacturer compared with "Golden Standard" Co-Oximeter. So the proposed device is safe and effective as well as the predicate device.
The accuracy claimed by the manufacturer is identical to that of the predicate device. And the test result indicates that the safety and effectiveness of the proposed devices is identical to that of the predicate device.
10. Substantial Equivalence Determination
According to the non-clinical and clinical test results, the proposed devices are as safe, as effective and perform as well as the predicate device. So the proposed devices are Substantially Equivalent (SE) to the predicate device which is US legally market device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 4, 2013
Mr. Lei Chen Quality Director Beijing Choice Electronic Technology Company, Limited No. 9 Shuangyuan Road, Badachu Hi-Teeh Zone, Shijingshan District Beijing, China 100041
Re: K122046
Trade/Device Name: Wrist Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: January 28, 2013 Received: January 28, 2013
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Chen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony Lo m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Exhibit A Indications for Use
510(k) Number (if known): K122046 Device Name: Wrist Pulse Oximeter
Indications for Use:
The MD300W4 wrist pulse oximeter is a portable non-invasive device intended for spot checking ,data collection and recording of arterial oxygen saturation (SpO2) and pulse rate of adult and pediatric patient at home and hospital( including clinical use in internist/ surgery, Anesthesia, intensive care and etc).
V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lester W. Schultheis Jr 2013.02.01 15:33:43 -05'00'
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology. General Hospital Infection Control, Dental Devices
510(k) Number: K122046
A-1
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).