K Number
K132356
Manufacturer
Date Cleared
2014-01-31

(186 days)

Product Code
Regulation Number
872.4200
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This medical device is only intended for dental treatment in the area of dentistry. Is is intended to be used for the following applications:

  • SANAO 200 L / 40 ST: The removal of decayed matter, cavity and crown preparations, the removal of filings and surface finishing of tooth and restoration surfaces.
  • SANAO 40 / 40 L. / 10 / 10 L.: Cavity preparations, caries excavation, endodontics, surface finishing of tooth and restoration surfaces.
  • SANAO PSI / PSO: prophylaxis treatment.
Device Description

All SANAO Contra-Angle and Straight dental handpieces can be powered by an air-motor or an electronic micro-motor and are intended for use by trained professionals in the field of general dentistry. They are ergonomically shaped, reusable and some models are optionally available with a glass rod lighting system. Depending on the model the devices possess a 1- or 3- port spray or no spray. For different applications and tools there is a range of devices available with different transmissions for maximum output rotations ranging from 8,000 to 200,000 rpm. The devices are color-coded according to their respective gear ratio. The shanks of the instruments are in conformance with the respective ISO standards.

AI/ML Overview

The SciCan SANAO Dental Handpiece, Contra- and Right-Angle Attachment, Dental (K132356) underwent a non-clinical evaluation to demonstrate substantial equivalence to its predicate device, the STATIS Handpieces (K071891). This evaluation primarily relied on literature research and comparison of technological characteristics to existing marketed devices.

1. Table of Acceptance Criteria and Reported Device Performance

The device's acceptance criteria were based on adherence to recognized international standards for dental handpieces. The reported performance is based on the claim of compliance with these standards and comparison to the predicate devices.

Acceptance Criteria (Standard)Reported Device Performance
ISO 7785-2: Dental Handpieces - Part 2 Straight and Geared Angled HandpiecesAll SANAO Handpieces samples were compliant with ISO 7785-2.
ISO 1797: Dentistry – Shanks for rotary instrumentsShanks of SANAO instruments conform to respective ISO standards.
ISO 3964: Dental Handpieces Coupling dimensionsCoupling for SANAO motor conforms to ISO 3964.
ISO 10993-1: Biological evaluation of medical devicesRisk analysis conducted. Materials (Stainless Steel, PEEK) are biocompatible.
ISO 14971: Medical devices - Application of risk management to medical devicesRisk analysis was conducted, showing minor risk level.
ISO 17664: Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices(Implicitly met as reusable device, documentation provided - not explicitly stated but part of compliance)
ISO 17665-1: Sterilization of health care products - Moist heat - Part 1(Implicitly met as reusable device, documentation provided - not explicitly stated but part of compliance)
ISO 9687: Dental equipment - Graphical symbols(Implicitly met as device is appropriately labeled/marked)
ISO 21531: Dentistry - Graphical symbols for dental instruments(Implicitly met as device is appropriately labeled/marked)

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not explicitly state the specific number of SANAO Handpieces tested in the bench tests. It mentions "All samples were compliant with ISO 7785-2."
  • Data Provenance: The evaluation was non-clinical, based on literature research and bench testing. The data provenance would therefore be from internal SciCan testing (bench testing) and publicly available information related to existing medical devices (literature research). No specific country of origin for this data is mentioned, but SciCan GmbH is based in Germany. The study was retrospective in the sense that it relied on existing standards and market data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This was a non-clinical evaluation based on engineering and performance standards, not a clinical study requiring expert interpretation of medical images or patient outcomes.

4. Adjudication Method for the Test Set

Not applicable. As a non-clinical evaluation, there was no adjudication method involving human interpretation of results in the traditional sense of a clinical trial. Compliance with ISO standards was likely determined by engineers and quality assurance personnel.

5. Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader, multi-case comparative effectiveness study was not done. The evaluation was non-clinical and focused on device performance specifications.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This device is a mechanical dental handpiece, not an algorithm or AI system.

7. Type of Ground Truth Used

The "ground truth" for this device's evaluation was primarily compliance with established international performance standards (e.g., ISO 7785-2, ISO 1797, ISO 3964) and the demonstrated substantial equivalence to predicate devices based on their known performance and safety profiles.

8. Sample Size for the Training Set

Not applicable. There was no "training set" as this is a mechanical device, not an AI/ML algorithm.

9. How the Ground Truth for the Training Set Was Established

Not applicable.

{0}------------------------------------------------

K132DS6

001_510k SUMMARY

Owner/Company name, address

SciCan GmbH Wangener Str. 78 88299 Leutkirch Germany

U.S. Contact

John Westermeier SciCan Inc. 701 Technology Drive Canonsburg, PA 15317 USA Phone: 949.887.6547

Application Correspondent Josef Sillet

Manager of Regulatory Affairs SciCan GmbH Wangener Str. 78 88299 Leutkirch Germany

Phone: +49-756198343-660 Fax: +49-756198343-699 E-Mail: jsillet@scican.com

Date prepared April 15th, 2013

  • Name of device
    Trade Name: Classification Name: Regulation Name: Regulatory Class: Product Code:

SANAO Handpiece, Contra- And Right-Angle Attachment, Dental Dental handpiece and accessories Class I EGS

JAN 31 2014

{1}------------------------------------------------

Predicate devices

The SANAO is substantially equivalent to the following legally marketed devices:

K071891 510(k): Trade name: STATIS Models 1.5L / 1.1 / 1.1L / 4.1 / 1.1ST / 6.1 EFB Product code: Classification Name: Handpiece, Air-powered, Dental

The STATIS Handpieces Slow-Speed Models STATIS 1.5L / 1.1 / 1.1L / 4.1 / 1.1ST / 6.1 are the exact predicate devices for SANAO Handpieces 200L / 40L / 40 / 40ST/ 10L / 10 / PSI / PSO.

Description of the device

All SANAO Contra-Angle and Straight dental handpieces can be powered by an air-motor or an electronic micro-motor and are intended for use by trained professionals in the field of general dentistry. They are ergonomically shaped, reusable and some models are optionally available with a glass rod lighting system. Depending on the model the devices possess a 1- or 3- port spray or no spray. For different applications and tools there is a range of devices available with different transmissions for maximum output rotations ranging from 8,000 to 200,000 rpm. The devices are color-coded according to their respective gear ratio. The shanks of the instruments are in conformance with the respective ISO standards.

、Statement of intended use

The SANAO instruments are intended to be used for general dentistry work. The intended use of the respective instruments is depending on the gear ratio.

  • SANAO 200 L / 40 ST: The removal of decayed matter, cavity and crown preparations, the removal of fillings and surface finishing of tooth and restoration surfaces.
  • SANAO 40 / 40 L / 10 / 10 L: Cavity preparations, caries excavation, endodontics, surface finishing of tooth and restoration surfaces.
  • SANAO PSI / PSO: prophylaxis treatment. -

Summary of Technological Characteristics

The devices have been designed according to the following standards:

PerformanceISO - 7785-2 Straight and Geared Angle Handpieces
Coupling for instrumentsISO - 1797 Dentistry – Shanks for rotary instruments
Coupling for motorISO - 3964 Dental Handpieces Coupling dimensions

For user comfort, the SANAO handpieces are ergonomically shaped. All contact surfaces with the patient are made out of stainless steel. The devices can be powered by an air-motor or an electronic micro-motor. Some models are available with a glass rod lighting system. Depending on the model the devices either possess a 1- or a 3- port spray or no spray. The different transmissions allow for different maximum output rotations ranging from 8,000 to 200,000 rpm.

SciCan SANAO 510(k)

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "SciCan" in a stylized font. The letters are in a dark gray color, and the "i" has a dot above it that is slightly offset to the right. The font appears to be a serif font, and the letters are slightly rounded.

Clinical and non-clinical tests

A non-clinical evaluation, based on literature research, has been done. The evaluation of the applicable market data showed that SANAO handpieces do not pose more known or new clinical risks than similar medical devices currently on the market. Based on those results clinical test have not been executed.

Statement of substantial equivalence

The SANAO dental handpieces are similar in design, function and intended use to other dental handpieces currently distributed in the U.S. An example for substantially equivalent devices are the STATIS Contra-Angle and Straight Handpieces.

The following tables show a comparison between the different SANAO dental handpieces and the predicates.

Comparison SANAO 200L / STATIS 1.5L

SANAO 200LSTATIS 1.5L(K071891)
PictureImage: SANAO 200LImage: STATIS 1.5L
Indications for UseThis medical device is:• intended for dental treatments only• suggested for the followingapplications:The removal of decayed matter, cavityand crown preparations, the removal offillings and surface finishing of toothand restoration surfaces.This medical device is:• intended for dental treatments only• suggested for the followingapplications:Crown preparation and all other tasksthat require the highest level ofcontrolled power
SANAO 200LSTATIS 1.5L(K071891)
DesignContra-AngleContra-Angle
Ratio1:51:5
identificationRedRed
CoolingWater/AirWater/Air
SprayYesYes
Air water ports3 port3 port
Operational modesPowered by air motor or electronicmicro-motorPowered by air motor or electronicmicro-motor
LightYesYes
FiberopticsYesYes
Dimensions (mm)Head Height: 13.9Head Ø: 10.2Overall length:97.7Head Height: 15.2Head Ø: 9.8Overall length:95.8
Type of chuckPush buttonPush button
Coupling with motorISO 3964ISO 3964
Coupling instrumentISO 1797 type 3 (1,60mm)ISO 1797 type 3 (1,60mm)
Composition of materials
Patient contact surfacesStainless Steel covered with ChromePEEKStainless Steel covered with Chrome
Water LinesStainless steel 304LStainless steel 304L
Technical specification
ChuckPush buttonPush button
Light intensity7,500 LUX5,000 Lux
Bur Extraction force> 22N>22N
Recommended settings -Air cooling8 Nl/mn at 2.5bar/36.3psi8-10 Nl/mn
Recommended settings -air/water pressureSpray air 2.7 bar/39.2 psi Water spray2 bar/29 psiSpray air 36 psiWater spray 36 psi
SANAO 200LSTATIS 1.5L(K071891)
Speed input max.40,000 rpm40,000 rpm
Speed output max.200,000 rpm200,000 rpm
Chuck designISO 1797 type 3Shaft Ø: 1.6mmLength 25 mmISO 1797 type 3Shaft Ø: 1.6mmLength 21mm
SANAO 40L / 40STATIS 1.1 / 1.1L(K071891)
Picture
Indications for UseThis medical device is:• intended for dental treatment only• suggested for the followingapplications: preparation of cavities,caries excavation, endodonticapplications, processing of tooth andrestoration surfaces.This medical device is:• intended for dental treatment only• suggested for the followingapplications:polishing, preparing, reducing andexcavation
Device Design
DesignContra-AngleContra-Angle
Ratio40L / 40 1:11.1/ 1.1L 1:1
identificationBlueBlue
CoolingWater/AirWater/Air
SprayYesYes
Air water ports1 port3 port
Operational modesPowered by air motor or electronicmicro-motorPowered by air motor or electronicmicro-motor
LightYes, for L modelYes, for L model
FiberopticsYesYes
Dimensions (mm)Head Height: 13.9Head Ø: 10.2Overall length:97.7Head Height: 15.2Head Ø: 9.8Overall length:95.8
Type of chuckPush buttonPush button
Coupling with motorISO 3964ISO 3964
Coupling instrumentISO 1797 type 1 (2,35mm)ISO 1797 type 1 (2,35mm)
Composition of materials
Patient contact surfacesStainless Steel covered with ChromePEEKStainless Steel covered with Chrome
Water LinesStainless steel 304LStainless steel 304L
Technical specification
ChuckPush buttonPush button
SANAO 40L / 40STATIS 1.1 / 1.1L(K071891)
Light intensity11,300 LUX (for L model)6,300 Lux (for L model)
Bur Extraction force> 45 N>45N
Recommended settings -air cooling8 NI/mn at 2.5bar/36.3psi8-10 NI/mn
Recommended settings -air/water pressureSpray air 2.7 bar/39.2 psi Water spray2 bar/29 psiSpray air 36 psiWater spray 36 psi
Speed input max.40,000 rpm40,000 rpm
Speed output max.40,000 rpm40,000 rpm
Chuck designISO 1797 type 1Shaft Ø: 2.334 - 2.350mmLength 22 mmISO 1797 type 1Shaft Ø: 2.350mmLength: Unknown
SANAO 10L / 10STATIS 4.1(K071891)
PictureImage: SANAO 10L / 10Image: STATIS 4.1
Indications for UseThis medical device is:• intended for dental treatment only• suggested for the followingapplications:preparation of cavities, cariesexcavation, endodontic applications,processing of tooth and restorationsurfaces.This medical device is:• intended for dental treatment only• suggested for the followingapplications:4.1: removing caries, endodontics,plaque and polishing of the toothsurface after oral hygiene treatment
Device Design
DesignContra-AngleContra-Angle
Ratio5:14:1
identificationGreenGreen
CoolingWater/AirWater/Air
SprayYesYes
Air water ports1 port3 port
Operational modesPowered by air motor or electronicmicro-motorPowered by air motor or electronicmicro-motor
LightYes, for L modelNo
FiberopticsYes, for L modelNo
Dimensions (mm)Head Height: 13.9Head Ø: 10.2Overall length:97.7Head Height: 15.2Head Ø: 9.8Overall length:95.8
Type of chuckPush buttonPush button
Coupling with motorISO 3964ISO 3964
SANAO 10L / 10STATIS 4.1(K071891)
Patient contact surfacesStainless Steel covered with ChromePEEKStainless Steel covered with Chrome
Water LinesStainless steel 304LStainless steel 304L
Technical specification
ChuckPush buttonPush button
Light intensity10,300 LUX (for L model)Not applicable
Bur Extraction force> 45N> 45N
Recommended settings -air cooling8 Nl/mn at 2.5bar/36.3psi8-10 Nl/mn
Recommended settings -air/water pressureSpray air 2.7 bar/39.2 psi Water spray 2bar/29 psiSpray air 36 psiWater spray 36 psi
Speed input max.40,000 rpm40,000 rpm
Speed output max.8,000 rpm10,000 rpm
ISO 1797 type 1ISO 1797 type 1
Chuck designShaft Ø: 2.334 - 2.350 mmShaft Ø: 2.350 mm
Length 25 mmLength: unknown
SANAO 40 STSTATIS 1.1 ST(K071891)
PictureImage: SANAO 40 STImage: STATIS 1.1 ST
Indications for UseThis medical device is:• only intended for dental treatment.• suggested for the followingapplications:Removal of carious material, cavitiesand crown preparations, removal offillings, processing of tooth andrestoration surfaces.This medical device is:• only intended for dental treatment.• suggested for the followingapplications:Preparation in the anterior region,osteotomy, root resection or extra-oralprosthesis correction.
Device Design
DesignStraight handpieceStraight handpiece
Ratio1:11:1
identificationBlueBlue
CoolingWater/AirWater/Air
SprayYesYes
Air water ports1 port1 port
Operational modesPowered by air motor or electronicmicro-motorPowered by air motor or electronicmicro-motor
LightNoNo
FiberopticsNoNo
Dimensions (mm)$Ø$ : 19.9Overall length:83.7$Ø$ : 19.6Overall length:84.9
Type of chuckMechanical chuckMechanical chuck
SANAO 40 STSTATIS 1.1 ST(K071891)
Coupling with motorISO 3964ISO 3964
Coupling instrumentISO 1797 type 2 (2,35mm)ISO 1797 type 2 (2,35mm)
Composition of materials
Patient contact surfacesStainless Steel covered with ChromeStainless Steel covered with Chrome
Water LinesStainless steel 304LStainless steel 304L
Technical specification
ChuckMechanical chuckMechanical chuck
Light intensityNot applicableNot applicable
Bur Extraction force> 45N>45N
Recommended settings -air cooling8 NI/mn at 2.5bar/36.3psi8-10 Nl/mn
Recommended settings -air/water pressureSpray air 2.7 bar/39.2 psi Water spraySpray air 36 psi
air/water pressure2 bar/29 psiWater spray 36 psi
Speed input max.40,000 rpm40,000 rpm
Speed output max.40,000 rpm40,000 rpm
ISO 1797 type 2ISO 1797 type 2
Chuck designShaft Ø: 2.334 - 2.350 mmShaft Ø: 2.350
Length 25 mmLength: unknown
SANAO PSI / PSOSTATIS 6.1(K071891)
PictureImage: SANAO PSI / PSOImage: STATIS 6.1
Indications for UseThis medical device is:· only intended for dental treatment.· suggested for the followingapplication: prophylaxis treatmentThis medical device is:· only intended for dental treatment.· suggested for the followingapplications:prophylaxis
Device Design
DesignContra-AngleContra-Angle
Ratio5:16:1
identificationGreenGreen
CoolingNot applicableNot applicable
SprayNoNo
Air water portsNoneNone
Operational modesPowered by air motor or electronicmicro-motorPowered by air motor or electronicmicro-motor
LightNo lightNo light
FiberopticsNoneNone
Dimensions (mm)Head Height: PSI 10.8mm PSO13.8mmHead Ø: 8.2Overall length:96.7Head Height: 15.2Head Ø: 7.3Overall length:88.7
Type of chuckNot applicableNot applicable
Coupling with motorISO 3964ISO 3964
Coupling instrumentPSI with Screw-In-Function (with ISO13295 type 3)For mandrels with Snap-On-Function(ISO 13295 type 5) or mandrels withScrew-In function (ISO 13295 type 3)
SANAO PSI / PSOSTATIS 6.1(K071891)
Function (ISO 13295 type 5)
Composition of materials
Patient contact surfacesStainless Steel covered with ChromePEEKStainless Steel covered with Chrome
Water LinesNo water lineNo water line
Technical specification
ChuckNo chuckNo chuck
Light intensityNot applicableNot applicable
Bur Extraction forceNot applicableNot applicable
Recommended settings -air coolingNot applicableNot applicable
Recommended settings -air/water pressureNot applicableNot applicable
Speed input max.40,000 rpm40,000 rpm
Speed output max.8,000 rpm6,666 rpm
Chuck designNo chuckNo chuck

{3}------------------------------------------------

SciCan

: ·

. . . . .

SciCan SANAO 510(k)

{4}------------------------------------------------

the state the state of the states of the states

.

.

. . . . .

.

.

SciCan SANAO 510(k) ،

:

:

:

{5}------------------------------------------------

Comparison SANAO 40L / 40 STATIS vs. 1.1 / 1.1L

{6}------------------------------------------------

{7}------------------------------------------------

Comparison SANAO 10L / 10 vs. STATIS 4.1

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "SciCan" in a bold, sans-serif font. The letters are closely spaced together, giving the word a compact appearance. The overall impression is clean and professional.

{9}------------------------------------------------

Comparison SANAO 40 STvs. STATIS 1.1 ST

{10}------------------------------------------------

SciCan

.

:

.

{11}------------------------------------------------

Comparison SANAO PSI / PSO vs. STATIS 6.1

{12}------------------------------------------------

As shown in the comparison tables, the indications for use of the SANAO dental handpieces are substantially equivalent to those of the SciCan STATIS Handpieces (K071891). As described above, the principle of operation and design features of the SANAO Handpieces are identical to those of the predicates. The gear ratio of SANAO PSI / PSO / 10 / 10L is different to that of the predicates. Therefore, the maximum rotation speed at the chuck is different between the SANAO PSI / PSO / 10 / 10L and the predicates. Also some differences can be found in the light intensity, dimensions and the external material from SANAO to STATIS. In order to evaluate that those differences do not affect safety and effectiveness of the SANAO Handpieces bench testing was performed. Bench testing demonstrated that the SANAO Handpieces met the requirement of the ISO 7785-2 which was used for evaluation of the predicate devices. Based on the above results, the SANAO Handpieces are substantially equivalent to the predicates and do not raise any new safety and effectiveness concern.

Bench Testing

The Bench tests were performed in order to ensure the safety and effectiveness of the SANAO Handpieces, verify conformity to ISO 7785-2 and demonstrate substantial equivalence to the predicates. All samples were compliant with ISO 7785-2 and demonstrated substantial equivalance to the predicates.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "SciCan" in a bold, sans-serif font. The letters are closely spaced, and the overall appearance is clean and modern. The image is simple and straightforward, with the focus entirely on the text.

Risk analysis method & actions

A Risk Management Plan to identify the risks and activities for managing those risks was carried out during design and development of the SANAO Dental handpieces, accessories and associated devices.

The Risk Analysis of the SANAO line of dental handpieces was conducted in accordance with ISO 14971: 2007. The analysis showed that the level of risk associated with the SANAO handpieces is minor and not any greater as the level of risks identified for similar medical devices currently on the market.

Applicable Standards

StandardsNo.StandardsOrganizationStandards TitleVersion
1ISO 10993-1ISOBiological evaluation of medical devices2009
2ISO 14971ISOMedical devices - Application of riskmanagement to medical devices2007
3ISO 7785-2ISODental Handpieces - Part 2Straight and Geared Angled Handpieces1995
4ISO 17664ISOSterilization of medical devices - Informationto be provided by the manufacturer for theprocessing of resterilizable medical devices2004
5ISO 17665-1ISOSterilization of health care products - Moistheat - Part 1: Requirements for thedevelopment, validation and routine controlof a sterilization process for medical devices2006
6ISO 9687ISODental equipment - Graphical symbols1995
7ISO 21531ISODentistry - Graphical symbols for dentalinstruments2009
8ISO 3964ISODental Handpieces - Coupling Dimensions1982
9ISO 1797-1ISODental Rotary Instruments - Shanks2011

Conclusion

The indications for Use of the SANAO Handpieces are substantially equivalent to those of the SciCan STATIS (K071891). In addition, the principle of operation and design features of the SANAO Handpieces are identical to those of the predicates. The gear ratio of the SANAO PSI / PSO / 10 / 10L is different to that of the predicates. Therefore, the maximum rotation speed at the chuck is different between the SANAO PSI / PSO / 10 / 10L and the predicates. There are also some

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "SciCan" in a bold, sans-serif font. The letters are closely spaced together, giving the word a compact appearance. The overall impression is one of a logo or brand name.

differences in the light intensions and the external material from SANAO to the predicates. Although there are those differences in characteristics between the SANAO Handpieces and the predicates, a number of performance testing indicated that the SANAO Handpieces met the requirements of the recognized consensus or voluntary standard. Based on the above results, we conclude that the SANAO-Handpieces are substantially equivalent to the predicates and do not raise any new safety and effectiveness concem.

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 31,2014

SciCan GmbH C/O Mr. Alexander Schapovalov TUV SUD American, Incorporated 1775 Old Highway 8 NW, Suite 104 New Brighton, MN 56112-1891

Re: K132356

Trade/Device Name: SANAO Dental Handpieces Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EGS Dated: January 12, 2014 Received: January 16, 2014

Dear Mr. Schapovalov:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{16}------------------------------------------------

Page 2 – Mr. Schapovalov

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/16/Picture/8 description: The image contains the words "Kwame Ulmer" in a bold, sans-serif font. The letters are arranged vertically, with "Kwame" on top and "Ulmer" below. To the right of the words is a geometric design that resembles a triangle with horizontal lines inside. The design is outlined in black and white.

for

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{17}------------------------------------------------

001_INDICATION FOR USE STATEMENT

510(k) Number (if known): K 132356

Device Name:

SANAO Dental Handpieces

Indications for Use

This medical device is only intended for dental treatment in the area of dentistry. Is is intended to be used for the following applications:

  • SANAO 200 L / 40 ST: The removal of decayed matter, cavity and crown preparations, the " removal of filings and surface finishing of tooth and restoration surfaces.
  • SANAO 40 / 40 L. / 10 / 10 L.: Cavity preparations, caries excavation, endodontics, surface . finishing of tooth and restoration surfaces.
  • SANAO PSI / PSO: prophylaxis treatment. t

X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use __ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runner DDS, AA Mary S. Runner -S
2014.01.31
08:22:05 -05'00'

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.