K Number
K132225
Device Name
EXTRICARE
Manufacturer
Date Cleared
2014-05-15

(302 days)

Product Code
Regulation Number
878.4780
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The extriCARE® 3600 Negative Pressure Wound Therapy System is indicated for wound management via the application of negative pressure to the wound by the removal of wound exudate, infectious materials, and tissue debris from the wound bed. The extriCARE® 3600 Negative Pressure Wound Therapy System is indicated for the following wound types: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafis.

Device Description

The system consists of a vacuum pump, canister, tubing, NPWT bandage or NPWT Foam Kits. In operation, the device is attached via the tubing to a Negative Pressure Wound bandage or foam kits. Maximum pressure and continuous mode or intermittent mode selection are digitally programmable. The extriCARE 3600 pump can be sold alone or as a part of the extriCARE 3600 system with extriCARE bandages or foam kits. The extriCARE bandages are an all-in-one wound dressing; with no wound packing required which is cleared in the predicate K110078. The foam kit consists of Polyurethane foam, transparent film drape, paper ruler, Suction Bell with connecting (drainage) tube and clamp, which is included in this submission for review.

AI/ML Overview

The provided 510(k) summary describes a Negative Pressure Wound Therapy (NPWT) system, which is a medical device and not an AI/ML algorithm. Therefore, many of the requested categories related to AI/ML device performance (e.g., sample size for training, expert ground truth, MRMC study, AI improvement effect size) are not applicable to this document.

However, I can extract the information provided regarding the device's acceptance criteria and the studies conducted to demonstrate its performance as a medical device.

Acceptance Criteria and Device Performance for extriCARE® 3600 Negative Pressure Wound Therapy System

Since this is a medical device (Negative Pressure Wound Therapy System) and not an AI/ML algorithm, the acceptance criteria are not typically expressed in terms of metrics like sensitivity, specificity, or AUC, but rather in terms of compliance with functional, safety, and performance standards. The "reported device performance" in this context refers to the successful completion of these tests.

Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance
Functional PerformanceEnsure device meets design input when used with NPWT bandage and foam kits, including:Series of bench tests conducted to ensure device functions successfully.
- Sensors functionSuccessfully met.
- Hook functionSuccessfully met.
- Canister functionSuccessfully met.
- Clamp functionSuccessfully met.
- Air filter functionSuccessfully met.
- Simulated wound performance testSuccessfully met.
Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1:2005 and IEC 60601-1-2:2007 standardsEMC testing performed on representative samples; device met the standards.
Electrical SafetyCompliance with AAMI ES 60601-1:2005 standardsElectrical safety testing performed on representative samples; device met the standards.
BiocompatibilityFor parts with direct body contact (bandage and foam kits), compliance with ISO-10993 for:All tests successfully passed for bandage and foam kit samples.
- In-vitro CytotoxicitySuccessfully passed.
- Skin irritationSuccessfully passed.
- Skin SensitizationSuccessfully passed.

Study Details:

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Test Set Sample Size: The document does not specify a numerical sample size for "test sets" in the context of clinical trials or data-driven algorithm validation. Instead, it mentions "representative samples of the device" for functional, mechanical, EMC, and electrical safety testing. For biocompatibility, "bandage samples and foam kits samples" were tested. No specific number of devices or clinical cases is provided.
    • Data Provenance: The document does not provide information on the country of origin of data or whether tests were retrospective or prospective, as these are not relevant for the type of bench and biocompatibility testing described.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This question is not applicable as the document describes a physical medical device undergoing bench and biocompatibility testing, not an AI/ML algorithm requiring expert-labeled ground truth for a test set.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This question is not applicable for the same reasons as above. Adjudication methods are relevant for resolving discrepancies in expert-labeled ground truth in AI/ML performance studies.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This question is not applicable. An MRMC comparative effectiveness study is used for evaluating the impact of AI on human reader performance, which is not relevant for a physical NPWT system.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable as the document concerns a physical medical device, not a standalone AI algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For functional and safety testing, the "ground truth" is adherence to established engineering specifications, international standards (e.g., IEC 60601 series), and biocompatibility standards (ISO-10993). There isn't a "ground truth" in the sense of clinical diagnoses or outcomes data for this type of device submission.
  7. The sample size for the training set:

    • This question is not applicable as this is a physical medical device, not an AI/ML algorithm that requires a training set.
  8. How the ground truth for the training set was established:

    • This question is not applicable for the same reason mentioned above.

{0}------------------------------------------------

510K Summary

page 1 of 3

c
Cara Cara Cara Ca

510(K) SUMMARY

Submitter:

Devon Medical Inc.

Contact Person:

Ruth Wu, COO 1100 First Avenue, Suite 202 King of Prussia, PA 19406 866.446.0092 Phone: 484.636.0211 Fax:

Common Classification & Proprietary Names:

Negative Pressure Wound Therapy System Common Names: Proprietary Names: extriCARE 3600 extriCARE Foam Kit

Date Prepared:

January 16th, 2013

Classification

The classification name, 21 CFR Part and Paragraph number, product code and classification of the extriCARE 3600:

FDAClassificationProductCodeClass
21 CFR878.4780Powered SuctionPumpOMPII

Predicate Devices:

The extriCARE 3600 Negative Pressure Wound Therapy System and extriCARE Foam Kit is substantially equivalent to the following:

Predicate DeviceManufacturer510(k)#
extriCARE 2400Devon MedicalK110078
V.A.C.® Freedom™Kinetic Concepts, Inc.K032310
Wound Pro ApexAccuro Medical ProductsLLCK100823
NPWT FoamDressing KitsSmith & NephewK082211

Device Description

The system consists of a vacuum pump, canister, tubing, NPWT bandage or NPWT Foam Kits. In operation, the device is attached via the tubing to a

{1}------------------------------------------------

510K Summary

Negative Pressure Wound bandage or foam kits. Maximum pressure and continuous mode or intermittent mode selection are digitally programmable. The extriCARE 3600 pump can be sold alone or as a part of the extriCARE 3600 system with extriCARE bandages or foam kits. The extriCARE bandages are an all-in-one wound dressing; with no wound packing required which is cleared in the predicate K110078. The foam kit consists of Polyurethane foam, transparent film drape, paper ruler, Suction Bell with connecting (drainage) tube and clamp, which is included in this submission for review.

Intended Use:

The extriCARE 3600 Negative Pressure Wound Therapy System is indicated for wound management via the application of negative pressure to the wound by the removal of wound exudate, infectious materials, and tissue debris from the wound bed. The extriCARE 3600 Negative Pressure Wound Therapy System is indicated for the following wound types: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.

Technological Characteristics:

The manufacturer believes that the technological characteristic of the extriCARE 3600 are substantially similar to those of the predicate devices. The extriCARE 3600 has very similar components to its predicate devices and very similar principles of operation. The device consists of an electrically generated source of vacuum and a canister. Like the predicates, vacuum, is applied for a specified period of time and intensity, according to the physician's prescription.

Both extriCARE bandages and foam dressing kits are alternatively used together with the extriCARE 3600 pumps. The extriCARE bandages are the same as stated in the previously cleared submission K110078.

The extriCARE foam kits use similar components and similar principles of operation to the predicate device. Like the predicate, the foam is placed into the wound and attached to a vacuum device, which may provide negative pressure to the wound site which may promote healing by removing wound exudates.

Performance Testing

To ensure the performance of extriCARE® 3600 device meet the design input when conjunction with NPWT bandage and foam kits, a series of bench tests were conducted to ensure device's functions, which include the sensors, hook, canister, clamp, air filter and a simulated wound performance test.

{2}------------------------------------------------

510K Summary

EMC Testing

To verify that the device design met it's functional and performance requirements, representative samples of the device underwent function and mechanical testing, EMC testing in accordance with IEC 60601-1:2005, IEC 60601-1-2:2007 and electrical safety testing in accordance with AAMI ES 60601-1:2005 standards.

K13225 page 3 of 3

Biocompatibility

The extriCARE 3600 system consists of the pump, bandages or alternative foam kits. While the pump has no direct body contact when used as indicated, the bandage and foam kits do have direct body contact. Per ISO-10993 requirements, the bandage samples and foam kits samples were tested for the following items and all tests were successfully passed.

  • In-vitro Cytotoxicity .
  • Skin irritation .
  • Skin Sensitization .

Statement of Substantial Equivalence

The extriCARE 3600 system is substantially equivalent in technology, function, operating parameters, and intended use to predicate devices that are currently commercially available and in distribution, and does not raise any new questions of safety or effectiveness.

Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Devon Medical Products, believes that the extriCARE 3600 system (pump, bandage, foam, and accessories) is safe and effective and substantially equivalent to the predicate devices as described herein.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the border of the circle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

May 15, 2014

Devin Medical, Inc. Ruth Wu, Chief Operating Officer 1100 First Avenue, Suite 202 King of Prussia, Pennsylvania 19406

Re: K132225

Trade/Device Name: extriCARE® 3600 Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: February 11, 2014 Received: February 14, 2014

Dear Ms. Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dr has master and regulations administered by other Federal agencies. You must or any it all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ruth Wu, Chief Operating Officer

.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

for

Binita Ashar, MD, MBA, FACS Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

cc: DMC 510(k) Staff Division D.O.

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K132225

Device Name

extriCARE 3600 Negative Pressure Wound Therapy System

Indications for Use (Describe)

The extriCARE® 3600 Negative Pressure Wound Therapy System is indicated for wound management via the application of negative pressure to the wound by the removal of wound exudate, infectious materials, and tissue debris from the wound bed. The extriCARE® 3600 Negative Pressure Wound Therapy System is indicated for the following wound types: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafis.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

ound Da

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the The burden time for this Colloon of information to recess, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.