K Number
K132069
Date Cleared
2013-09-23

(82 days)

Product Code
Regulation Number
878.3300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Desara® and Desara® Blue devices are intended to be used in females to position a mesh for the treatment of Genuine Stress Urinary Incontinence (SUI), mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.

Device Description

Desara and Desara® Blue devices are sterile, single-use mid-urethral slings used to provide support in the pelvic region to treat stress urinary or mixed incontinence. The devices are manufactured out of monofilament polypropylene yarn, which is knitted into a mesh. The devices have integral sleeves, tips and sutures to assist the surgeon in placement of the mesh, which are removed after placement of the device.

Desara® and Desara® Blue are designed to work with Caldera Medical's reusable introducers, which utilize a Universal Connection System offering surgeons the flexibility to choose their preferred surgical approach.

AI/ML Overview

The provided document is a 510(k) summary for the Desara® and Desara® Blue surgical mesh devices, which are used to treat Stress Urinary Incontinence (SUI) in females. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a standalone clinical study for a new device with novel technology that would typically involve the comprehensive details requested in your prompt.

Therefore, many of the requested details (like sample size for a test set, number of experts for ground truth, adjudication methods, MRMC studies, standalone algorithm performance, training set details) are not applicable to this type of submission. This 510(k) relies on demonstrating that the new devices are substantially equivalent to already approved predicate devices by showing similar technological characteristics and performance in bench and validation testing.

Here's an breakdown based on the information available:

Acceptance Criteria and Reported Device Performance

The document states that the Desara® and Desara® Blue devices were assessed against various mesh characteristics, and the results demonstrated "equivalent device function based upon its intended use to the predicate device, Desara® Mesh, (#K112609)." The acceptance criteria were therefore implicitly that the new devices perform comparably to the predicate and reference devices across these specified characteristics.

Acceptance Criteria (Measured Characteristics)Reported Device Performance (Relative to Predicate/Reference)
Material CompositionComprised of the same polypropylene mesh resin as the reference device, Ascend® Blue (#K101462). The same mesh knit and limited contact raw materials as Desara® Mesh (#K112609) were used.
Mesh ThicknessAssessed (Specific values not provided, but evaluated for equivalence).
Mesh Knit CharacteristicsAssessed (Specific values not provided, but evaluated for equivalence).
Pore SizeAssessed (Specific values not provided, but evaluated for equivalence).
Mesh DensityAssessed (Specific values not provided, but evaluated for equivalence).
Tensile StrengthAssessed (Specific values not provided, but evaluated for equivalence).
Mesh StiffnessAssessed (Specific values not provided, but evaluated for equivalence).
Flexural RigidityAssessed (Specific values not provided, but evaluated for equivalence).
Tear ResistanceAssessed (Specific values not provided, but evaluated for equivalence).
Burst StrengthAssessed (Specific values not provided, but evaluated for equivalence).
Suture PulloutAssessed (Specific values not provided, but evaluated for equivalence).
Pyrogen LevelsAssessed (Specific values not provided, but evaluated for equivalence).
BiocompatibilitySupported by the predicate devices Desara® Mesh (#K112609) and Ascend® Blue (#K101462), which passed all biocompatibility testing (same materials).
Aging, Shelf Life, Transportation, SterilizationSupported by the predicate device Desara® Mesh (#K112609) and Ascend® Blue (#K101462) documentation (same or similar materials, packaging, manufacturing, and sterilization processes).
Device Function"Results of both mechanical bench and validation testing demonstrate equivalent device function based upon its intended use to the predicate device, Desara® Mesh, (#K112609)."

Study Details (Based on available 510(k) information)

  1. Sample size used for the test set and the data provenance:

    • The document describes mechanical bench and validation testing, not typically a "test set" in the context of an AI/algorithm study with patient data. The sample sizes for these engineering tests are not explicitly provided in this summary but would be part of the underlying test reports.
    • Data Provenance: The testing appears to be primarily laboratory-based (mechanical bench and validation testing). There is no mention of patient data or its country of origin.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as there is no "ground truth" derived from expert review of patient data. The "truth" or reference for performance is established by comparing to the predicate device and relevant standards.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This is not applicable as there is no "test set" involving expert review of patient data that would require adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The device is a surgical mesh for SUI treatment, not an AI diagnostic or interpretive tool that would involve human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. This is a medical device (surgical mesh), not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" or reference for comparison in this substantial equivalence determination is the performance characteristics of the predicate device (Desara® Mesh #K112609) and reference device (Ascend® Blue #K101462), as well as compliance with established FDA guidance documents and international standards for mesh characteristics and biocompatibility.
  7. The sample size for the training set:

    • This is not applicable. There is no "training set" as this is a physical medical device, not a machine learning model.
  8. How the ground truth for the training set was established:

    • This is not applicable.

{0}------------------------------------------------

DERA MEDICAL

K132069 Paqe 1 of 3

SEP 2 3 2013

510(k) Summarv

Date of Summary: June 27, 2013

Applicant:

Contact:

F

Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 P (818) 879-6555 F (818) 879-6556

Vicki Gail QA/RA and Operations Manager P (818) 483-7602 F (818) 736-9083 E ygail@calderamedical.com

Desara®

Desara® Blue

Trade Name:

Device Class: Product Code: C.F.R Section: Common name:

Class II · OTN 21 CFR 878.3300 Surgical Mesh, Gynecologic, For Stress Urinary Incontinence, Female Obstetrics/Gynecology Panel

Desara® Mesh , K112609 Predicate Device:

Description of Device:

Classification Panel:

Desara and Desara® Blue devices are sterile, single-use mid-urethral slings used to provide support in the pelvic region to treat stress urinary or mixed incontinence. The devices are manufactured out of monofilament polypropylene yarn, which is knitted into a mesh. The devices have integral sleeves, tips and sutures to assist the surgeon in placement of the mesh, which are removed after placement of the device.

Desara® and Desara® Blue are designed to work with Caldera Medical's reusable introducers, which utilize a Universal Connection System offering surgeons the flexibility to choose their preferred surgical approach.

Intended Use of Device:

Desara® and Desara® Blue devices are intended to be used in females to position a mesh for the treatment of Genuine Stress Urinary Incontinence (SUI), mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.

Page 12

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Caldera Medical. The logo consists of a circular graphic on the left and the text "CALDERA MEDICAL" on the right. The graphic appears to be a stylized representation of a medical symbol. The text is in a bold, sans-serif font.

Summary of Technological Characteristics

Desara® and Desara® Blue devices submitted herein contain changes in polypropylene mesh resin, a decrease to the size of the junction tip and an optional design which contains blue colorant. Desara® and Desara® Blue are the same shape and size, utilize the same mesh knit and limited contact raw materials as the predicate device, Desara® Mesh (#K112609).

Desara® and Desara® Blue have a more specific intended use than that of the predicate device, Desara® Mesh (#K112609) and the same fundamental scientific technology.

Performance Summary

Desara® and Desara® Blue submitted herein are comprised of the same mesh resin as the reference device, Ascend® Blue, (#K101462), also a product of Caldera Medical. In accordance with the FDA's Guidance for the Preparation of a Premarket Notification Application for Surgical Mesh, the following mesh characteristics were assessed: mesh thickness, mesh knit characteristics, pore size, mesh density, tensile strength, mesh stiffness, flexural rigidity, tear resistance, burst strength, suture pullout and pyrogen levels.

Desara® Mesh (#K112609) and Ascend® Blue, (#K101462) were used as a reference for Desara® and Desara® Blue devices for biocompatibility since thev are comprised of the same materials. Desara® Mesh (#K112609) and Ascend® Blue, (#K101462) have passed all biocompatibility testing as indicated per the FDA guidance documents, FDA Device Advice, Guidance Documents (Medical Device and Radiation Emitting Product), 3. Biocompatibility and FDA Draft Guidance Entitled "Use of International Standard ISO-10993. "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing", dated April 23, 2013.

Results of both mechanical bench and validation testing demonstrate equivalent device function based upon its intended use to the predicate device, Desara® Mesh, (#K112609).

Testing documentation from Desara® Mesh (#K112609) and Ascend® Blue, (#K101462) were used to support the aging, sheif life, transportation and sterilization of Desara® and Desara® Blue submitted herein in, was performed in accordance with the FDA Guidance, Updated 510(k) Sterility Review Guidance K90-1; Final Guidance for Industry, FDA, FDA Device Advice, Guidance Documents (Medical Device and Radiation Emitting Product), 4. Labeling and FDA Consensus standard, ASTM F-1980-07, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices as the devices are comprised of the same or similar materials, and utilized the same packaging, manufacturing and sterilization processes as that of Desara® Mesh (#K112609).

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Caldera Medical. The logo consists of a circular graphic to the left of the company name. The company name, "CALDERA MEDICAL", is written in all capital letters and in a bold font.

Summary of Substantial Equivalence

The conclusions drawn from the non-clinical and clinical tests demonstrate that Desara® and Desara® Blue submitted herein are substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. The logo is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 23, 2013

Caldera Medical, Inc. % Vicki Gail QA/RA and Operations Manager 5171 Clareton Drive Agoura Hills, CA 91301

Re: K132069

Trade/Device Name: Desara and Desara Blue Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OTN Dated: June 27, 2013 Received: July 3, 2013

Dear Vicki Gail,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA),

You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 - Vicki Gail

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Caldera Medical. The logo consists of a circular graphic on the left, followed by the text "CALDERA MEDICAL" in bold, uppercase letters. The graphic appears to be a stylized representation of a medical symbol or emblem.

Indications for Use

510(k) Number (if known): K132069

Device Name: Desara® Desara® Blue

Indications for Use:

Desara® and Desara® Blue are intended to be used in females to position a mesh for the treatment of Genuine Stress Urinary Incontinence (SUI), mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.

Prescription Use(21 CFR 801 Subpart D)X
-----------------------------------------------

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Benjamin R. Fisher -S 2013.09.23 16:52:16 -04'00'

Page 1 of 1

Caldera Medical, Inc. Desam and Desam Blue Page 15

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.