K Number
K131774
Device Name
HAMILTON-G5
Date Cleared
2013-10-09

(114 days)

Product Code
Regulation Number
868.5895
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care. The HAMILTON-G5 ventilator is intended for use by properly trained personnel under the direct supervision of a licensed physician. The HAMILTON-G5 ventilator may be used for transport within a hospital or hospital type facility provided compressed gas is supplied. The device is not to be used in the presence of flammable anesthetic agents or other iqnition sources. The ventilator is not to be used in an environment with magnetic resonance imaging (MRI) equipment. The device is not intended for transportation outside the hospital or for use in the home environment.
Device Description
The HAMILTON-G5 is an electronically controlled pneumatic intensive care ventilator ventilation system. It uses oxygen and air or heliox to ventilate adults, pediatrics, infants, and neonates. It is powered by AC, and with battery backup in order to protect against power failure or unstable power and to facilitate intra-hospital transport. The HAMILTON-G5's pneumatics deliver gas -- while its electrical systems controls pneumatics, monitors alarms and distributes power. The user interface consists of a LCD-display with touch screen, keys, and a press-and- turn knob. The HAMILTON-G5's new software version 2.30, includes a pulse oximetry function for continuous, non-invasive oxygen saturation monitoring (SpO2). Volume targeting in the HAMILTON-G5 is now supported by a new ventilation mode, called the Volume Support mode (VS). It is a flow-cycled, volume targeted, and pressure-regulated mode. The Volume Support (VS) mode is designed for spontaneously breathing patients. It provides support to patient-initiated breaths so as to deliver the desired tidal volume (V+), at a level appropriate to the patient's efforts. This mode allows the ventilator to change the support in response to changing patient conditions and inspiratory effort levels. To achieve this volume, the device decreases support when the patient's breathing activity increases or, conversely, increases support when the patient's inspiratory effort decreases.
More Information

No
The document describes a ventilator with a new ventilation mode (Volume Support) that adjusts support based on patient effort. While this involves dynamic response, the description does not mention the use of AI or ML algorithms for this adjustment. The mechanism described appears to be rule-based or algorithmic rather than learned from data.

Yes
The device is a ventilator, which delivers gas to patients to support breathing and is used therapeutically to treat conditions affecting respiration.

No

Explanation: The HAMILTON-G5 is described as a ventilator, a life support device, and its primary function is to provide respiratory support. While it has a pulse oximetry function for monitoring oxygen saturation, this monitoring feature is part of its overall function to manage ventilation rather than its primary purpose being diagnostic. Diagnostics fundamentally involve identifying or determining the nature of a disease or condition. The device's core intent is therapeutic ventilation.

No

The device description clearly states it is an "electronically controlled pneumatic intensive care ventilator ventilation system" that uses oxygen and air, is powered by AC with battery backup, and has a user interface with a physical display, keys, and a knob. While it includes new software features, the core device is a hardware ventilator.

Based on the provided text, the HAMILTON-G5 ventilator is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • Intended Use: The intended use clearly states that the device is for "intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients." This describes a life support and respiratory therapy function, not a diagnostic test performed on samples taken from the body.
  • Device Description: The description focuses on the mechanical and electronic aspects of delivering gas for ventilation, monitoring alarms, and providing power. While it includes a pulse oximetry function (SpO2), which measures oxygen saturation in the blood, this is a non-invasive measurement taken directly from the patient, not a test performed on a sample outside the body.
  • Lack of IVD Characteristics: The text does not mention any of the typical characteristics of an IVD, such as:
    • Analyzing biological samples (blood, urine, tissue, etc.)
    • Detecting or measuring substances in biological samples
    • Providing information for diagnosis, monitoring, or screening based on laboratory tests

The HAMILTON-G5 is a medical device used for direct patient care and treatment (ventilation), not for performing diagnostic tests on samples.

N/A

Intended Use / Indications for Use

The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care. The HAMILTON-G5 ventilator is intended for use by properly trained personnel under the direct supervision of a licensed physician. The HAMILTON-G5 ventilator may be used for transport within a hospital or hospital type facility provided compressed gas is supplied.

The device is not to be used in the presence of flammable anesthetic agents or other iqnition sources. The ventilator is not to be used in an environment with magnetic resonance imaging (MRI) equipment. The device is not intended for transportation outside the hospital or for use in the home environment.

Product codes (comma separated list FDA assigned to the subject device)

CBK, DQA

Device Description

The HAMILTON-G5 is an electronically controlled pneumatic intensive care ventilator ventilation system. It uses oxygen and air or heliox to ventilate adults, pediatrics, infants, and neonates. It is powered by AC, and with battery backup in order to protect against power failure or unstable power and to facilitate intra-hospital transport. The HAMILTON-G5's pneumatics deliver gas -- while its electrical systems controls pneumatics, monitors alarms and distributes power. The user interface consists of a LCD-display with touch screen, keys, and a press-and- turn knob.

The HAMILTON-G5's new software version 2.30, includes a pulse oximetry function for continuous, non-invasive oxygen saturation monitoring (SpO2).

Volume targeting in the HAMILTON-G5 is now supported by a new ventilation mode, called the Volume Support mode (VS). It is a flow-cycled, volume targeted, and pressure-regulated mode.

The Volume Support (VS) mode is designed for spontaneously breathing patients. It provides support to patient-initiated breaths so as to deliver the desired tidal volume (V+), at a level appropriate to the patient's efforts. This mode allows the ventilator to change the support in response to changing patient conditions and inspiratory effort levels. To achieve this volume, the device decreases support when the patient's breathing activity increases or, conversely, increases support when the patient's inspiratory effort decreases.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

adult and pediatric patients, and optionally infant and neonatal patients.

Intended User / Care Setting

properly trained personnel under the direct supervision of a licensed physician.

hospital and institutional environment where healthcare professionals provide patient care.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Safety testing of the HAMILTON-G5, with the new options, was conducted and shows that the device is substantially equivalent to the predicate devices for its intended use. The new Volume Support mode was further subjected to waveform performance testing, as described by the standard ASTM F1100-90. The data provided from these tests were shown to be substantially equivalent to a legally marketed device.

Verification and Validation testing also demonstrated that the SpO2 and pulse rate values, calculated by the OEM system, are not corrupted during communication to the HAMILTON-G5 host device. No modifications were made to the previously cleared oximeter systems. The results of the software verification and validation testing demonstrate that all specified requirements have been implemented correctly and completely.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K103803, K073179

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K100428, K974292, K011918, K032749, K043517

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the words "HAMILTON MEDICAL" in a bold, sans-serif font. The word "HAMILTON" is on the top line, and the word "MEDICAL" is on the bottom line. The text is black and the background is white. The image has some noise and artifacts, which may be due to the image being scanned or photographed.

HamilTon.G5

510(k) SUMMARY

| SUBMITTER: | HAMILTON MEDICAL AG
Via Crusch 8
Bonaduz, 7402
SWITZERLAND |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Ralph Aguila
Regulatory Affairs / Quality Engineer
Phone: +41 81 660 6845
Fax: +41 81 660 6020
e-mail: raaguila@hamilton-medical.ch |
| ESTABLISHMENT
REGISTRATION NUMBER: | 3001421318 |
| PREPARATION DATE: | September 4th, 2013 |
| TRADE NAME: | HAMILTON-G5 |
| CLASSIFICATION NAME: | CLASS II Ventilator, Continuous (CBK)
CLASS II Oximeter (DQA) |
| REGULATION NUMBER &
PRODUCT CODE | 21 CFR 868.5895 (CBK)
21 CFR 870.2700 (DQA) |
| PREDICATE DEVICE:
(PRIMARY) | HAMILTON-G5 (K103803) |
| PREDICATE DEVICE:
(SECONDARY) | MAQUET Servo-i (K073179)
For the Volume-Support mode
MASIMO RAINBOW SET RADICAL
PULSE CO-OXIMETER (K100428)
Nihon Kohden SpO2 OXIMETER,
PROBES, AND ACCESSORIES
(K974292, K011918, K032749, K043517) |

1

Image /page/1/Picture/0 description: The image shows the words "HAMILTON MEDICAL" in bold, black letters. The word "HAMILTON" is on the top line, and the word "MEDICAL" is on the bottom line. The letters are slightly distorted, and the image has a grainy texture.

Image /page/1/Picture/1 description: The image shows the text "HAMILTON-G5" in a bold, sans-serif font. The letters are black, and the text is slightly distressed, giving it a vintage or worn appearance. The letters "A", "M", "I", "L", "T", "O", and "N" have a white dashed line above them.

DEVICE DESCRIPTION

The HAMILTON-G5 is an electronically controlled pneumatic intensive care ventilator ventilation system. It uses oxygen and air or heliox to ventilate adults, pediatrics, infants, and neonates. It is powered by AC, and with battery backup in order to protect against power failure or unstable power and to facilitate intra-hospital transport. The HAMILTON-G5's pneumatics deliver gas -- while its electrical systems controls pneumatics, monitors alarms and distributes power. The user interface consists of a LCD-display with touch screen, keys, and a press-and- turn knob.

The HAMILTON-G5's new software version 2.30, includes a pulse oximetry function for continuous, non-invasive oxygen saturation monitoring (SpO2).

Volume targeting in the HAMILTON-G5 is now supported by a new ventilation mode, called the Volume Support mode (VS). It is a flow-cycled, volume targeted, and pressure-regulated mode.

The Volume Support (VS) mode is designed for spontaneously breathing patients. It provides support to patient-initiated breaths so as to deliver the desired tidal volume (V+), at a level appropriate to the patient's efforts. This mode allows the ventilator to change the support in response to changing patient conditions and inspiratory effort levels. To achieve this volume, the device decreases support when the patient's breathing activity increases or, conversely, increases support when the patient's inspiratory effort decreases.

INTENDED USE

The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where healthcare professionals provide patient care. The HAMILTON-G5 ventilator is intended for use by properly trained personnel under the direct supervision of a licensed physician. The HAMILTON-G5 ventilator may be used for transport within a hospital or hospital-type facility -- provided compressed gas is supplied.

The device is not to be used in the presence of flammable anesthetic agents or other ignition sources. The ventilator is not to be used in an environment with magnetic resonance imaging (MRI) equipment. The device is not intended for transportation outside the hospital or for use in the home environment.

2

Image /page/2/Picture/0 description: The image shows the words "HAMILTON" and "MEDICAL" stacked on top of each other. The text is in a bold, sans-serif font. The word "HAMILTON" is on the top line, and the word "MEDICAL" is on the bottom line. The text is black against a white background.

Image /page/2/Picture/1 description: The image shows the text "HAMILTON.G5" in a bold, sans-serif font. The text is black and appears to be a logo or brand name. The letters are closely spaced, and the dot between "HAMILTON" and "G5" is also in bold.

SUMMARY OF THE TECHNOLOGY AND PERFORMANCE SPECIFICATIONS COMPARISON WITH THE PREDICATED DEVICES

The indication-of-use statement for the proposed HAMILTON-G5 ventilator is the same as the predicate HAMILTON-G5. In addition, the technological characteristics and performance specifications of the probosed HAMILTON-G5 ventilator are substantially equivalent to those of the predicate devices.

The new pulse oximetry function uses previously cleared devices, like the Masimo Rainbow Set Radical Pulse co-oximeter & accessories, as well as the Nihon Kohden SpO2 oximeter & accessories. The new Volume Support mode in the HAMILTON-G5 is substantially equivalent to the MAQUET Servo-i's Volume Support mode.

HAMILTON MEDICAL has demonstrated the modified HAMILTON-G5 ventilator to be substantially equivalent. The proposed ventilator is considered to be substantially equivalent to the currently marketed predicate devices, which have been previously cleared by FDA.

COMPARISON TABLE BETWEEN PROPOSED AND PREDICATE DEVICE

The following table compares the proposed HAMILTON-G5 with its predicate device, the HAMILTON-G5 (K103803).

| | Proposed device:
HAMILTON-G5 | Predicate device:
HAMILTON-G5
(K103803) | Equivalence |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Software
version | SW 2.30 | SW 2.0 | Substantially
Equivalent |
| Ventilator
modes | (S)CMV, SIMV, SPONT, ASV,
P-CMV, P-SIMV, APVcmv,
APVsimv, DuoPAP, APRV,
NIV, NIV-ST, nCPAP-PS,
Volume Support | (S)CMV, SIMV, SPONT, ASV,
P-CMV, P-SIMV, APVcmv,
APVsimv, DuoPAP, APRV,
NIV, NIV-ST, nCPAP-PS | Substantially
Equivalent.
The Volume Support
mode is new in the
proposed
HAMILTON-G5 |
| Pulse Oximetry | Yes. | No | N/A

The proposed
HAMILTON-G5
Includes both the
Masimo (K100428)
and Nihon Kohden
Pulse Oximetry
(K974292, K011918,
K032749, K043517) |
| | Proposed device:
HAMILTON-G5 | Predicate device:
HAMILTON-G5
(K103803) | Equivalence |
| Indications of
Use | The HAMILTON-G5 ventilator
is designed for intensive care
ventilation of adult and
pediatric patients, and
optionally infant and neonatal
patients. The device is
intended for use in the hospital
and institutional environment
where health care
professionals provide patient
care. The HAMILTON-G5
ventilator is intended for use by
properly trained personnel
under the direct supervision of
a licensed physician. The
HAMILTON-G5 ventilator may
be used for transport within a
hospital or hospital type facility
provided compressed gas is
supplied.
The device is not to be used in
the presence of flammable
anesthetic agents or other
ignition sources. The ventilator
is not to be used in an
environment with magnetic
resonance imaging (MRI)
equipment. The device is not
intended for transportation
outside the hospital or for use
in the home environment. | The HAMILTON-GS ventilator
is designed for intensive care
ventilation of adult and pediatric
patients, and optionally infant
and neonatal patients. The
device is intended for use in the
hospital and institutional
environment where healthcare
professionals provide patient
care. The HAMILTON-GS
ventilator is intended for use by
properly trained personnel
under the direct supervision of
a licensed physician. The
HAMILTON-GS ventilator may
be used for transport within a
hospital or hospital-type facility
provided compressed gas is
supplied.
The device is not to be used in
the presence of flammable
anesthetic agents or other
ignition sources. The ventilator
is not to be used in an
environment with magnetic
resonance imaging (MRI)
equipment.
The device is not intended for
transportation outside the
hospital or for use in the home
environment. In the USA,
federal law restricts this device
to sale by or on the order of a
physician. | Substantially
Equivalent |
| Operation
environmental
requirements | •5 to 40 °C (41 to 104 °F)
•10 to 95%, non-condensing
•1013 to 600 hPa | •5 to 40 °C (41 to 104 °F)
•10 to 95%, non-condensing
•1013 to 600 hPa | Equivalent |
| Input Power | AC: 100 to 240 V, 50/60 Hz | AC: 100 to 240 V, 50/60 Hz | Equivalent |
| Power
Consumption | 50 VA typical | 50 VA typical | Equivalent |
| | Proposed device:
HAMILTON-G5 | Predicate device:
HAMILTON-G5
(K103803) | Equivalence |
| Battery | One batteries, Li-ion, sealed,
maintenance-free | One batteries, Li-ion, sealed,
maintenance-free | Equivalent |
| Battery
operating time
(typical) | 2 hours | 2 hours | Equivalent |
| Safety features | • Apnea backup ventilation
• Automatic self-tests
• Alarms (operator-adjustable /
non-adjustable)
• Alarm backup buzzer
• External flow sensor failure
mode
• Safety mode in case of
technical failures
• Air inlet HEPA filter
monitoring
• Monitored fan
• Event log | • Apnea backup ventilation
• Automatic self-tests
• Alarms (operator-adjustable /
non-adjustable)
• Alarm backup buzzer
• External flow sensor failure
mode
• Safety mode in case of
technical failures
• Air inlet HEPA filter
monitoring
• Monitored fan
• Event log | Equivalent |
| Emergency air
intake | In case of a power supply,
technical, or pneumatics failure
the ambient valve allows
spontaneous breathing. | In case of a power supply,
technical, or pneumatics failure
the ambient valve allows
spontaneous breathing. | Equivalent |
| Maximum
working
pressure
limitation | 60 cmH20 | 60 cmH20 | Equivalent |
| Maximum
inspiratory flow | 210 L/min | 210 L/min | Equivalent |

K131774

3

HAMILTON.G5

4

Image /page/4/Picture/0 description: The image shows the words "HAMILTON" and "MEDICAL" stacked on top of each other. The word "HAMILTON" is in a larger font than the word "MEDICAL". The text is in black and the background is white. The font is bold.

HzzvilLTOrA-G5

5

Image /page/5/Picture/0 description: The image shows the words "HAMILTON" and "MEDICAL" stacked on top of each other. The text is in a bold, sans-serif font and appears to be part of a logo or heading. The overall impression is of a professional or corporate identity.

NON-CLINICAL PERFORMANCE TESTS

Safety testing of the HAMILTON-G5, with the new options, was conducted and shows that the device is substantially equivalent to the predicate devices for its intended use. The new Volume Support mode was further subjected to waveform performance testing, as described by the standard ASTM F1100-90. The data provided from these tests were shown to be substantially equivalent to a legally marketed device.

Verification and Validation testing also demonstrated that the SpO2 and pulse rate values, calculated by the OEM system, are not corrupted during communication to the HAMILTON-G5 host device. No modifications were made to the previously cleared oximeter systems. The results of the software verification and validation testing demonstrate that all specified requirements have been implemented correctly and completely.

Draft Reviewer Guidance for Ventilators.1995
IEC 60601-1General Requirements for Safety
IEC 60601-1-2Electromagnetic Compatibility
IEC 60601-2-12Critical Care Ventilators
IEC 60601-2-49Essential performance of multi-function monitoring equipment.

CONCLUSION

The results of verification, validation, and testing activities demonstrate that the modified HAMILTON-G5 ventilator is substantially equivalent to the legally marketed devices identified above.

6

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an abstract human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002

October 9, 2013

Hamilton Medical AG Mr. Ralph Aguila Regulatory Affairs/Quality Engineer Via Crusch 8 BONADUZ 7402 SWITZERLAND

Re: K131774

Trade/Device Name: Hamilton-G5 Regulation Number: 21 CFR 868.5895 Regulation Name: Ventilator, Continuous Regulatory Class: II Product Code: CBK, DQA Dated: September 5, 2013 Received: September 9, 2013

Dear Mr. Aguila:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

7

Page 2 - Mr. Aguila

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part

807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/7/Picture/7 description: The image contains a combination of text and graphic elements. The text includes the name "Tejashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director." There is also the acronym "DAGRID" present. The word "FOR" is located in the bottom right corner of the image.

Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

8

INDICATIONS FOR USE STATEMENT

K131774

510(k) Number:

HAMILTON-G5 Device Name:

Indication for Use:

The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care. The HAMILTON-G5 ventilator is intended for use by properly trained personnel under the direct supervision of a licensed physician. The HAMILTON-G5 ventilator may be used for transport within a hospital or hospital type facility provided compressed gas is supplied.

The device is not to be used in the presence of flammable anesthetic agents or other iqnition sources. The ventilator is not to be used in an environment with magnetic resonance imaging (MRI) equipment. The device is not intended for transportation outside the hospital or for use in the home environment.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by Anya C. Harry -S V: c=US, o=U.S. Government, ou=HHS, Anva C. DA ou=People, cn=Anya C. Harry -S, 2342.19200300.100.1.1=0011315590 Date: 2013:10.09 15:28:27 -04'00'