K Number
K073179
Device Name
SERVO-I VENTILATOR SYSTEM MODEL 64 87 800 E4073, HELIOX OPTION, MODEL 6675585
Date Cleared
2008-06-09

(209 days)

Product Code
Regulation Number
868.5895
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The SERVO-i Ventilator System with Heliox option is intended for treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency. SERVO-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities. The SERVO-i Ventilator System with Heliox option is also MR conditional. The SERVO-i Ventilator System with Heliox option is indicated for use with the delivery of Air, Oxygen, or Heliox (a mixture of Helium and Oxygen).
Device Description
The predicate device SERVO-i Ventilator System is a ventilator, which gives, ventilation (Invasive and Non Invasive) to critical care patients in the weight range 0.5 to 250 Kg's. The modified device adds an option, called "Heliox option", which is a mechanical adaptor along with software for SERVO-i Ventilator System integration.
More Information

Not Found

No
The summary describes a ventilator with a mechanical and software modification for Heliox delivery, with no mention of AI or ML terms or related concepts like training/test sets or performance metrics typically associated with AI/ML.

Yes
The device is intended for "treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency," indicating a therapeutic purpose.

No

The document describes the device as a ventilator system intended for treatment and monitoring of patients with respiratory failure or insufficiency, indicating a therapeutic rather than diagnostic purpose.

No

The device description explicitly states that the modification includes a "mechanical adaptor along with software," indicating the presence of hardware components.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for the treatment and monitoring of patients with respiratory failure or insufficiency by providing ventilation. This is a therapeutic and monitoring function performed directly on the patient.
  • Device Description: The device is described as a ventilator system that provides ventilation. This is a life support and respiratory support device.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue, etc.) to provide information for diagnosis, monitoring, or treatment.

IVD devices are used to perform tests in vitro (outside the body) on biological samples to gain diagnostic or other health-related information. This device operates in vivo (on the patient's body) to support respiration.

N/A

Intended Use / Indications for Use

The SERVO-i Ventilator System with Heliox option is intended for treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency. SERVO-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities. The SERVO-i Ventilator System with Heliox option is also MR conditional.

The SERVO-i Ventilator System with Heliox option is indicated for use with the delivery of Air, Oxygen, or Heliox (a mixture of Helium and Oxygen).

Product codes (comma separated list FDA assigned to the subject device)

CBK

Device Description

The predicate device SERVO-i Ventilator System is a ventilator, which gives, ventilation (Invasive and Non Invasive) to critical care patients in the weight range 0.5 to 250 Kg's. The modified device adds an option, called "Heliox option", which is a mechanical adaptor along with software for SERVO-i Ventilator System integration.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

neonates, infants, and adults

Intended User / Care Setting

healthcare providers in hospitals or healthcare facilities.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K063404, K022674, K041524

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).

0

K073179
page 1 of 2

510(K) SUMMARY

As required by section 807.92(c)

JUN - 9 2008

Subscribers Name & Address

Maquet Critical Care AB Röntgenvägen 2 SE-171 95 Solna, Sweden Tel: (011) 46 8 730 73 00 Fax: (011) 46 8 730 78 38

Contact Persons for this submission:

Mr. Micael Johansson Tele. Direct; (011) 46 8 730 79 58 Email: micael.johansson@maquet.se

Mr. Jamie Yieh USA Contact : Director, Regulatory Affairs Maquet, Inc. 1140 Route 22 East, Suite 202 Bridgewater, NJ 08807 Phone: 908-947-2311 Fax: 908-947-2301 Email: jamie.yieh@maquet-inc.com

Trade Names

SERVO-i Ventilator System with Heliox option

Preparation date 2007-11-09

Device Classification

| Common Name | Classification
Number | Class | Regulation Number |
|-------------------------------------|--------------------------|-------|-------------------|
| Ventilator, continuous (Respirator) | 73 CBK | II | 21 CFR 868.5895 |

Predicate Device Identification

Legally marketed devices to which equivalence is being claimed510(k) #
Maquet, SERVO-i Ventilator SystemK063404
Viasys, Avea with Heliox optionK022674
GE Datex-Ohmeda, Aptaer Heliox Delivery SystemK041524

1

Device Description

Summary of technological characteristics of modified Device and Predicate Device: The predicate device SERVO-i Ventilator System is a ventilator, which gives, ventilation (Invasive and Non Invasive) to critical care patients in the weight range 0.5 to 250 Kg's. The modified device adds an option, called "Heliox option", which is a mechanical adaptor along with software for SERVO-i Ventilator System integration.

Intended Use of the Device:

The SERVO-i Ventilator System with Heliox option is intended for treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency. SERVO-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities. The SERVO-i Ventilator System with Heliox option is also MR conditional.

The SERVO-i Ventilator System with Heliox option is indicated for use with the delivery of Air, Oxygen, or Heliox (a mixture of Helium and Oxygen).

Comparison to predicate devices.

The subject device is comparable to the predicate devices in that the intended use is for delivery of Heliox, for the patient ranges of neonatal to adult, and in invasive and non-invasive applications. All devices utilize gas connectors or adapters and software enhancements.

2

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" arranged around the perimeter. Inside the circle is an abstract emblem resembling an eagle or bird-like figure, constructed from thick, stylized lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 9 2008

Maquet Critical Care AB C/O Mr. Jamie Yieh Director, Regulatory Affairs Maquet, Incorporated 1140 Route 22 East, Suite 202 Bridgewater, New Jersey 08807

Re: K073179

Trade/Device Name: SERVO-i Ventilator System with Heliox Option Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: May 12, 2008 Received: May 13, 2008

Dear Mr. Yieh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

3

Page 2 - Mr. Yieh

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Ches

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K073179

Device Name: SERVO-i Ventilator System with Heliox option

Indications for Use:

The SERVO-i Ventilator System with Heliox option is intended for treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency. SERVO-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities. The SERVO-i Ventilator System with Heliox option is also MR conditional.

The SERVO-i Ventilator System with Heliox option is indicated for use with the delivery of Air, Oxygen, or Heliox (a mixture of Helium and Oxygen).

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

hnd

Page 1 of 1

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K073179