(128 days)
The Titan powder free nitrile patient examination glove, blue color, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It has blue color and is sold as non sterile.
Powder free nitrile examination gloves are made of synthetic nitrile rubber, and are non sterile that meets all of the requirements of ASTM standard D 6319-10, except for sterility requirements. They have blue color.
Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided text:
Device: Powder Free Patient Examination Gloves, Blue Color (Brand Name: Titans)
1. Table of Acceptance Criteria and Reported Device Performance
| Description | Acceptance Criteria (Standard) | Reported Device Performance (Brand Name: Titans) |
|---|---|---|
| Dimensions | ASTM D 6319-10 | Meets ASTM D 6319-10 (Published 03/18/2011) |
| Physical Property | ASTM D 6319-10 | Meets ASTM D 6319-10 (Published 03/18/2011) |
| Free of Pinhole | ASTM D5151 (AQL 2.5) | Meets ASTM D5151 (AQL 2.5) (Published 08/20/2012) |
| Residue Powder | ASTM D6124 | Meets ASTM D6124 (Published 08/20/2012) |
| Primary Skin Irritation | ISO 10993-10: 2010 | Not a primary skin irritant under the conditions of the study |
| Dermal Sensitization | ISO 10993-10: 2010 | Not a dermal sensitizer under the conditions of the study |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for the individual tests (e.g., for pinhole detection, physical properties, or biocompatibility). It only indicates that "Tests were conducted following the recommended procedures outlined in the respective consensus standards." The data provenance (country of origin, retrospective/prospective) and specific sample sizes for these tests are not provided in the summary.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Not applicable. This device is a patient examination glove, and the testing involves adherence to established ASTM and ISO standards for physical properties, performance, and biocompatibility. Ground truth in this context is defined by the objective measurements and pass/fail criteria of these standards, not by expert interpretation or consensus on images or clinical cases.
4. Adjudication Method for the Test Set
Not applicable. As the "ground truth" is based on objective measurements against consensus standards, an adjudication method for reconciling expert opinions is not necessary or mentioned.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic devices where human readers interpret data (e.g., medical images) and AI might assist them. For patient examination gloves, the evaluation centers around predetermined physical, chemical, and biological performance standards.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is a physical medical device (gloves), not an algorithm or AI. Performance is assessed through standardized laboratory tests, not by an algorithm.
7. The Type of Ground Truth Used
The ground truth used is based on established consensus standards (ASTM D6319-10, ASTM D5151, ASTM D6124, ISO 10993-10) with specific, objective pass/fail criteria for dimensions, physical properties, absence of pinholes, residue powder levels, and biocompatibility endpoints (skin irritation and sensitization).
8. The Sample Size for the Training Set
Not applicable. This device is a manufactured product (gloves), not an AI algorithm requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no AI algorithm or training set, this question is not relevant to the described device.
{0}------------------------------------------------
Section 5: 510k Summary
Section 5: 510(k) Summary:
This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92
Submitter & Foreign Manufacture Identification 5.1
Hebei HongSen Plastics Technology Co., Ltd The Eastern Industrial Accumulation Area Nangong City, Xingtai City, Hebei Province, China Tel: 86-311-85656588 Submitter's FDA Registration Number: N/A
SEP 2 5 2013
5.2 US Agent and Contact Person
Charles Shen Manton Business and Technology Services 5 Carey Street Pennington, NJ 08534 Tel: 608-217-9358 Email: cyshen@aol.com
Date of Summary: April 15, 2013 5.3
5.4 Device Name: Proprietary Name:
Common Name: Classification Name: Device Classification: Regulation Number: Panel: General Product Code:
Powder Free Patient Examination Gloves, Blue Color (Brand Name: Titans) Patient examination glove Patient examination glove I 21 CFR 880.6250 Hospital LZA
Predicate Device Information: ર્ડ
:
K121992, "Patient Nitrile Examination Gloves, Powder Free, Non-Sterile, Blue (1) Color", manufactured by "Xinwei (Shandong) Plastic and Rubber Products Co., Ltd."
ર્સ્વર્સ Device description:
Powder free nitrile examination gloves are made of synthetic nitrile rubber, and are non sterile that meets all of the requirements of ASTM standard D 6319-10, except for sterility requirements. They have blue color.
{1}------------------------------------------------
5.7 Intended Use:
The Titans powder free nitrile patient examination glove, blue color, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It has blue color and is sold as non sterile.
5.8 Comparison to Predicate Devices
The powder free nitrile examination gloves, blue, non sterile are compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.
K121992, "Patient Nitrile Examination Gloves, Powder Free, Non-Sterile, Blue (1) Color", manufactured by "Xinwei (Shandong) Plastic and Rubber Products Co., Ltd."
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
| Description | Our Device | Predicate Device (K121992) |
|---|---|---|
| Indication forUse | Disposable device intended for medicalpurposes that is worn on the examiner's handor finger to prevent contamination betweenpatient and examiner. It has blue color and issold as non sterile. | A patient examination glove is adisposable device intended formedical purposes that isworn upon the examiner's hands orfinger to prevent contaminationbetween patient and examiner(2ICFR 880.6250) |
| Basic Design | A garment covering the hand and waistarea. Gloves have separate sheaths oropenings for each finger and the thumb. | Same |
| Materials | Nitrile rubber | Same |
| Size | XS, S, M, L, XL, XXL | S, M, L,XL |
| Single Use | Yes | Yes |
| Color | blue | Blue |
| Sterile | Non sterile | None Sterile |
Table 5.1: Comparison of Intended Use, Design, and Material
Our device is essentially identical to the predicate device in terms of indications for use, design, and material between our device and the predicate devices.
{2}------------------------------------------------
Section 5: 510k Summary
The following table shows similarities and differences of the performance between our device and the predicate devices. Tests were conducted following the recommended procedures outlined in the respective consensus standards, and results for Powder Free Patient Examination Gloves, Blue Color (Brand Name: Titans), manufactured by "Hebei HongSen Plastics Technology Co., Ltd. " met all relevant requirements in the test standards, and are comparable to the predicate device.
| Description | Our Device | Predicate Device (K121992) |
|---|---|---|
| Dimension | Meets ASTM D 6319-10(Published 03/18/2011) | Meets ASTM D 6319-10(Publication date unknown) |
| Physical Property | Meets ASTM D 6319-10(Published 03/18/2011) | Meets ASTM D 6319-10(Publication date unknown) |
| Free of Pinhole | Meets ASTM D5151 (AQL2.5)(Published 08/20/2012) | Meets ASTM D5151(Publication date unknown) |
| Residue Powder | Meets ASTM D6124(Published 08/20/2012) | Meets ASTM D6124(Publication date unknown) |
| Primary Skin Irritation(ISO 10993-10: 2010) | Not a primary skin irritant underthe conditions of the study | Same |
| Dermal sensitization(ISO 10993-10: 2010) | Not a dermal sensitizer under theconditions of the study | Same |
| Table 5.2: Comparison of Physical, Biocompatibility and Performance Testing | |||
|---|---|---|---|
A brief discussion of the non-clinical testing data that was submitted, 5.9 referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its user-defined acceptance criteria and intended uses:
Powder Free Patient Examination Gloves, Blue Color (Brand Name: Titans) meet requirements per ASTM D6319-10, ASTM D6124-06, ASTM D 5151-06, and ISO 10993-10. It is safe and effective, and its performance meets the requirements of its predefined acceptance criteria and intended uses.
{3}------------------------------------------------
A brief discussion of the clinical submitted, reference, or relied on in the 5.10 premarket notification submission for a determination of substantial equivalence.
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
Substantial Equivalent Conclusions 5.11
Based on the comparison of intended use, design, materials, and performance, our Powder Free Patient Examination Gloves, Blue Color (Brand Name: Titans) are substantial equivalent to its predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2013
Hebei HongSen Plastics Technology Company Limited C/O Charles Shen Manton Business and Technology Services 5 Carey Street PENNINGTON NJ 08534 US
- Re: K131440
Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue Color (Brand Name Titans) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: July 11, 2013 Received: July 1, 2013
Dear Mr. Shen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
FDA Richard C.
Chapman
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
for
Enclosure
{6}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K131440
Device Name
Powder Free Nitrile Patient Examination Gloves, Blue color (Brand Name: Titans)
Indications for Use (Describe)
The Titan powder free nitrile patient examination glove, blue color, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It has blue color and is sold as non sterile.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
EF
PSC | | | shindang Services (301) | 443-6740
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
NE ON MIN FOR THE FOR FOR FOR FOR FOR FOR FOR FOR FOR Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Concurrence
FORM FDA 3881 (6/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Elizabeth F. Claverie 2013.09.25 14:04:24
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.