K Number
K131366
Date Cleared
2013-10-11

(151 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACCU-CHEK Aviva Expert System is indicated as an aid in the treatment of Insulin-requiring diabetes. The ACCU-CHEK Aviva Expert System consists of the ACCU-CHEK Aviva Expert Meter, ACCU-CHEK Aviva Plus test strips, ACCU-CHEK Aviva control solutions, and ACCU-CHEK Bolus Advisor. The ACCU-CHEK Aviva Expert System is Intended to facilitate the optimization of glycemic control in patients who are trained in multiple daily insulin injection therapy and are under the supervision of healthcare professionals experienced in managing insulin treated patients.

The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used for the quantitative measurement of giucose in fresh capillary whole blood samples drawn from the fingertips. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended for selftesting outside the body (in vitro diagnostic use) by people with diabetes. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used by a single person and should not be shared. The ACCU-CHEK Aviva Expert blood giucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Altemative site testing should NOT be used with the ACCU-CHEK Aviva Expert blood glucose monitoring system. The ACCU-CHEK Aviva Expert System is intended for prescription home use only.

The ACCU-CHEK Aviva Expert meter is also indicated for the calculation of an insulin dose or carbohydrate intake based on user-entered data. The ACCU-CHEK Bolus Advisor, as a component of the Accu-Chek Aviva Expert meter, is intended for use in providing insulin dose recommendations in response to blood glucase, health events, and carbohydrate input. The ACCU-CHEK Bolus Advisor is intended to provide direction for insulin adjustment within the scope of a preplanned treatment program from a healthcare professional. Before its use, a physician or healthcare professional must prescribe the ACCU-CHEK Aviva Expert System and provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the ACCU-CHEK Bolus Advisor. Once programmed, a patient must consult with his/her physician or healthcare professional before making any changes to these ACCU-CHEK Bolus Advisor settings.

Device Description

The ACCU-CHEK Aviva Expert System consists of the following:

  • ACCU-CHEK Aviva Expert meter .
  • ACCU-CHEK Bolus Advisor (a component of the Aviva Expert meter) .
  • . ACCU-CHEK Aviva Plus test strips (K101299)
  • ACCU-CHEK Aviva control solutions (K043474) .

The ACCU-CHEK Aviva Expert system is a blood glucose monitoring system that makes use of the ACCU-CHEK Aviva Plus test strips (cleared under K101299) and the ACCU-CHEK Aviva control solutions (cleared under K043474). The industrial design of the ACCU-CHEK Aviva Expert meter is nearly identical to that of the ACCU-CHEK Aviva Combo meter (cleared under kl 11353); the only difference between the two devices is that the ACCU-CHEK Aviva Expert meter does not have Bluetooth wireless capability and therefore cannot be connected to an insulin pump, whereas the ACCU-CHEK Aviva Combo meter does have Bluetooth capability and can be connected to an insulin pump.

The ACCU-CHEK Aviva Expert system provides the user with the ability to measure capillary blood glucose levels when a sample of capillary blood is applied to the test strip. The meter also provides an optional insulin bolus calculator (the ACCU-CHEK Bolus Advisor) designed for use by individuals with diabetes who require insulin. This feature is optional in that a user can simply obtain a blood glucose value through capillary blood testing and does not need to use the insulin bolus calculator portion of the system if it is not desired. The insulin bolus calculator algorithm is identical to the bolus calculator algorithm that was cleared during the Aviva Combo 510(k) submission (K111353). For the ACCU-CHEK Aviva Expert system, this bolus calculator is meant to be used by patients with diabetes on multiple daily insulin injection (MDI) therapy. In order to calculate the appropriate bolus of insulin, the ACCU-CHEK Bolus Advisor takes the measured bG, the target bG, the carbohydrate intake, the insulin-to-carbohydrate ratio, the insulin sensitivity, health events (such as exercise), the time of day, and the active insulin into account. Before using the ACCU-CHEK Aviva Expert system, a physician or healthcare professional must provide the patient-specific target blood glucose, insulinto-carbohydrate ration, and insulin sensitivity parameters.

AI/ML Overview

Here's a detailed breakdown of the acceptance criteria and study information for the ACCU-CHEK Aviva Expert System, based on the provided 510(k) Summary:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for blood glucose monitoring systems typically align with ISO 15197 (or similar standards) for accuracy. The provided document details the accuracy of the device against a reference method. It also outlines precision performance.

Performance MetricAcceptance Criteria (Implied by Predicate/General Standards)Reported Device Performance
Accuracy (Method Comparison)
Glucose < 75 mg/dL>95% of results within ±15 mg/dL of laboratory reference (a common standard for this range)41/48 (85.4%) within ±5 mg/dL 48/48 (100%) within ±10 mg/dL 48/48 (100%) within ±15 mg/dL
Glucose ≥ 75 mg/dL>95% of results within ±15% of laboratory reference (a common standard for this range)147/252 (58.3%) within ±5% 222/252 (88.1%) within ±10% 246/252 (97.6%) within ±15% 250/252 (99.2%) within ±20%
Precision (Repeatability - Within Lot)For targets < 75 mg/dL, SD ≤ 5.0 mg/dL; for targets ≥ 75 mg/dL, CV ≤ 5.0% (from Similarities Table)Mean [mg/dL]: 42.1, 84.5, 137.8, 208.2, 345.0 SD [mg/dL]: 1.2, 2.2, 3.3, 5.6, 7.9 CV [%]: 2.9, 2.6, 2.4, 2.7, 2.3
Precision (Reproducibility - Day-to-Day)Not explicitly stated as acceptance criteria in the provided text, but data is given.Control Solutions Mean [mg/dL]: Low: 45.1, Mid: 117.6, High: 303.0 SD [mg/dL]: Low: 1.1, Mid: 2.4, High: 5.1 CV [%]: Low: 2.4, Mid: 2.0, High: 1.7
Measuring Range20 - 600 mg/dL (from Similarities Table of predicate)20 - 600 mg/dL
Sample Hematocrit10 to 65% (from Similarities Table of predicate)10 to 65%
Maximum Altitude10,000 feet (from Similarities Table of predicate)10,000 feet
Operating Temperature/Humidity14 to 38°C (57 to 100°F), 10 to 80% r.h. (from Similarities Table of predicate)14 to 38°C (57 to 100°F), 10 to 80% r.h.

2. Sample Size Used for the Test Set and Data Provenance

  • Accuracy Test Set (Method Comparison):
    • For glucose concentrations less than 75 mg/dL: 48 samples
    • For glucose concentrations greater than or equal to 75 mg/dL: 252 samples
  • Precision Test Set (Repeatability - Within Lot): 100 samples for each of 5 glucose levels.
  • Precision Test Set (Reproducibility - Day-to-Day): 100 samples for each of 3 control solution levels.
  • Data Provenance: Not explicitly stated in the provided text, but it’s typically part of a controlled clinical or laboratory study. It can be assumed to be from a prospective study, likely conducted internally or by a contracted lab, designed to evaluate the device's performance. The country of origin is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the given 510(k) summary. For glucose monitoring systems, ground truth is typically established by a laboratory reference method (e.g., YSI analyzer) operated by trained laboratory professionals, rather than by medical experts establishing a diagnosis. The summary mentions "laboratory reference" implicitly by comparing results to specified mg/dL or percentage deviations.

4. Adjudication Method for the Test Set

This information is not provided and is generally not applicable in the context of blood glucose meter accuracy studies. The ground truth (reference method glucose value) serves as the definitive comparator, and there is no "adjudication" necessary between multiple interpretations of this ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this device. This type of study is used for diagnostic systems where human readers interpret medical images or data, and an AI might assist in that interpretation. The ACCU-CHEK Aviva Expert System is a blood glucose meter with an optional bolus calculator, not an imaging or interpretive AI system that a human "reads." While it has an "Expert System" and "Bolus Advisor" component, this is an algorithm that provides a calculation rather than an AI assisting human interpretation of complex medical data.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the performance data presented (accuracy and precision) represents the standalone performance of the ACCU-CHEK Aviva Expert System as a blood glucose meter. The bolus advisor component is an algorithm that provides a calculation based on input, and its performance would be assessed algorithmically (e.g., accuracy of dosage calculation based on given parameters), which would inherently be a "standalone" algorithmic evaluation. The provided data focuses on the glucose measurement accuracy, which is the primary standalone function.

7. The Type of Ground Truth Used

The ground truth used for the accuracy study is a laboratory reference method for blood glucose measurement. This is implied by the phrasing "Results for glucose concentrations less/greater than or equal to 75 mg/dL" and the comparison to numeric deviations (± mg/dL or ± %). For such devices, a highly accurate laboratory analyzer (e.g., YSI Glucose Analyzer) is the standard for establishing ground truth.

8. The Sample Size for the Training Set

The provided 510(k) summary does not specify the sample size for any training set. Given that this device is presenting itself as substantially equivalent based on identical core technology (test strip, bG measurement algorithm, bolus calculator algorithm) to a previously cleared predicate device, extensive new training data for foundational algorithms like glucose measurement might not have been required for this specific submission if the underlying algorithms were already established and validated in the predicate. If any specific machine learning or adaptable components were newly introduced, their training data would typically be described. However, the bolus calculator algorithm is stated to be "identical" to the predicate.

9. How the Ground Truth for the Training Set Was Established

Since no training set details are provided, information on how its ground truth was established is not available in this document. If new training data were used for any component, the ground truth would likely also be established via a laboratory reference method for blood glucose values and clinical scenarios for bolus calculations.

{0}------------------------------------------------

510(k) Summary

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

1. Submitter Name, Address, Contact

Roche Diagnostic Corporation 9115 Hague Rd. Indianapolis, IN 46250 (317) 521-4793 Contact Person: Nate Carrington Date Prepared: February 12, 2013

OCT
11 2013

2. Device Name

Proprietary names:

ACCU-CHEK Aviva Expert System ACCU-CHEK Aviva Plus Test Strip ACCU-CHEK Bolus Advisor ACCU-CHEK Aviva Expert Meter

Classification name: Glucose dehydrogenase, glucose test system (21 C.F.R. § 862.1345); Class II

Classification name: Drug dosing calculation (21 C.F.R. § 868.1890); Class II

LFR, Glucose Dehydrogenase NDC, Drug Dosing Calculator

3. Predicate Device

ACCU-CHEK Aviva Combo meter, cleared as a component of the ACCU-CHEK Combo System in #K111353

The ACCU-CHEK Aviva Combo blood glucose system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitoring the effectiveness of diabetes control. The ACCU-CHEK Aviva Combo meter is also indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data.

{1}------------------------------------------------

4. Device Description

The ACCU-CHEK Aviva Expert System consists of the following:

  • ACCU-CHEK Aviva Expert meter .
  • ACCU-CHEK Bolus Advisor (a component of the Aviva Expert meter) .
  • . ACCU-CHEK Aviva Plus test strips (K101299)
  • ACCU-CHEK Aviva control solutions (K043474) .

The ACCU-CHEK Aviva Expert system is a blood glucose monitoring system that makes use of the ACCU-CHEK Aviva Plus test strips (cleared under K101299) and the ACCU-CHEK Aviva control solutions (cleared under K043474). The industrial design of the ACCU-CHEK Aviva Expert meter is nearly identical to that of the ACCU-CHEK Aviva Combo meter (cleared under kl 11353); the only difference between the two devices is that the ACCU-CHEK Aviva Expert meter does not have Bluetooth wireless capability and therefore cannot be connected to an insulin pump, whereas the ACCU-CHEK Aviva Combo meter does have Bluetooth capability and can be connected to an insulin pump.

The ACCU-CHEK Aviva Expert system provides the user with the ability to measure capillary blood glucose levels when a sample of capillary blood is applied to the test strip. The meter also provides an optional insulin bolus calculator (the ACCU-CHEK Bolus Advisor) designed for use by individuals with diabetes who require insulin. This feature is optional in that a user can simply obtain a blood glucose value through capillary blood testing and does not need to use the insulin bolus calculator portion of the system if it is not desired. The insulin bolus calculator algorithm is identical to the bolus calculator algorithm that was cleared during the Aviva Combo 510(k) submission (K111353). For the ACCU-CHEK Aviva Expert system, this bolus calculator is meant to be used by patients with diabetes on multiple daily insulin injection (MDI) therapy. In order to calculate the appropriate bolus of insulin, the ACCU-CHEK Bolus Advisor takes the measured bG, the target bG, the carbohydrate intake, the insulin-to-carbohydrate ratio, the insulin sensitivity, health events (such as exercise), the time of day, and the active insulin into account. Before using the ACCU-CHEK Aviva Expert system, a physician or healthcare professional must provide the patient-specific target blood glucose, insulinto-carbohydrate ration, and insulin sensitivity parameters.

5. Intended Use

The ACCU-CHEK Aviva Expert System is indicated as an aid in the treatment of insulinrequiring diabetes. The ACCU-CHEK Aviva Expert System consists of the ACCU-CHEK Aviva Expert Meter, ACCU-CHEK Aviva Plus test strips, ACCU-CHEK Aviva control solutions, and ACCU-CHEK Bolus Advisor. The ACCU-CHEK Aviva Expert System is intended to facilitate the optimization of glycemic control in patients who are trained in multiple daily insulin injection therapy and are under the supervision of healthcare professionals experienced in managing insulin treated patients.

510(k) Summary

{2}------------------------------------------------

The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used by a single person and should not be shared. The ACCU-CHEK Aviva Expert blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the ACCU-CHEK Aviva Expert blood glucose monitoring system. The ACCU-CHEK Aviva Expert System is intended for prescription home use only.

The ACCU-CHEK Aviva Expert meter is also indicated for the calculation of an insulin dose or carbohydrate intake based on user-entered data. The ACCU-CHEK Bolus Advisor, as a component of the Accu-Chek Aviva Expert meter, is intended for use in providing insulin dose recommendations in response to blood glucose, health events, and carbohydrate input. The ACCU-CHEK Bolus Advisor is intended to provide direction for insulin adjustment within the scope of a pre-planned treatment program from a healthcare professional. Before its use, a physician or healthcare professional must prescribe the ACCU-CHEK Aviva Expert System and provide the patient-specific target blood glucose, insulin-to-carboyhydrate ratio, and insulin sensitivity parameters to be programmed into the ACCU-CHEK Bolus Advisor. Once programmed, a patient must consult with his/her physician or healthcare professional before making any changes to these ACCU-CHEK Bolus Advisor settings.

6. Substantial Equivalence

The ACCU-CHEK Aviva Expert System is substantially equivalent to the ACCU-CHEK Aviva Combo System. Below is a table that provides a comparison between the ACCU-CHEK Aviva Expert System and its predicate device.

510(k) Summary

{3}------------------------------------------------

Similarities Table

System Feature/ClaimDetail
Test StripIdentical: Both systems make use of the Aviva Plus teststrip.
Test Strip ProductionProcesses and Lot-ReleaseCriteriaIdentical: Both systems make use of the Aviva Plus teststrip.
bG MeasurementAlgorithmIdentical: The fundamental scientific technology for themeasurement of blood glucose has not changed from thepredicate.
Meter Main IntegratedCircuit BoardIdentical: The firmware components and layout areidentical between the two meters.
Meter Display Module,Display Frame, and ButtonModuleIdentical: The same display unit, display frame, and buttonunit are used by the two meters.
Meter Housing and BatteryContactsIdentical: For both meters, the plastic parts are producedby the same molding tool using the same material, and thebattery contacts are the same.
Meter ProductionEnvironmentIdentical: Both meters are produced on one productionline/process.
Underdose Detection andMeter FailsafesIdentical: The fundamental scientific technology for themeasurement of blood glucose has not changed from thepredicate.
Integrity Check for StripIdentical: Early in the measurement sequence, the metermeasures the resistance of the gold on the un-dosed strip toassure that it has been properly inserted and that the qualityis not compromised. The meter measures the backgroundconductivity and electrical current prior to dosing to assurethat the reagent quality is not compromised or that the stripwas not prematurely dosed.
Insulin Bolus CalculatorAlgorithmIdentical: Both systems use the exact same boluscalculator, the ACCU-CHEK Bolus Advisor, for insulindosing calculations.

.

:

·

{4}------------------------------------------------

Similarities Table (continued)

SystemFeature/ClaimACCU-CHEK Aviva ExpertMeter with Aviva PlusTest StripACCU-CHEK Combo Meterwith Aviva Plus Test StripPredicate(K111353)
Indications forUseQuantitative measurement ofglucose (sugar) in capillary bloodfrom the finger tip, indicated fordiabetes management bycalculating an insulin dose orcarbohydrate intakeQuantitative measurement ofglucose (sugar) in capillary bloodfrom the finger tip, indicated fordiabetes management bycalculating an insulin dose orcarbohydrate intake
Test PrincipleAmperometric DetectionAmperometric Detection
EnzymeMut. Q-GDHMut. Q-GDH
SampleHematocrit10 to 65%10 to 65%
MaximumAltitude10,000 feet10,000 feet
Measuring Range20 - 600 mg/dL20 - 600 mg/dL
Sample Volume0.6 µL0.6 µL
Test Time5 seconds5 seconds
OperatingTemperature andRelativeHumidity14 to 38°C(57 to 100°F)10 to 80% r.h.14 to 38°C(57 to 100°F)10-80% r.h.
CodingCode key insertion; lot-specificcode key provided with each boxof Aviva Plus test stripsCode key insertion; lot-specificcode key provided with each boxof Aviva Plus test strips
PrecisionFor response targets below 75mg/dL, the SD is ≤ 5.0 mg/dL,and for response targets ≥ 75mg/dL, the CV is ≤ 5.0%.For response targets below 75mg/dL, the SD is ≤ 5.0 mg/dL,and for response targets ≥ 75mg/dL, the CV is ≤ 5.0%.
Double DosingNoNo
Alternate SiteTestingNoNo
SystemFeature/ClaimACCU-CHEK Aviva ExpertMeter with Aviva PlusTest StripACCU-CHEK Combo Meterwith Aviva Plus Test StripPredicate(K111353)
Closed and OpenVial Shelf LifeStability18 months18 months
Control SolutionsAqueous, 2 levels, uses ACCU-CHEK Aviva Control SolutionsAqueous, 2 levels, uses ACCU-CHEK Aviva Control Solutions
PrimaryPackagingStandard flip top vialStandard flip top vial
HandlingAutomatic on/off with stripinsertion or by pressing buttonAutomatic on/off with stripinsertion or by pressing button
Dimensions3.7 x 2.1 x 1 in LWH;approximately 3.6 oz. withbatteries inserted3.7 x 2.1 x 1 in LWH;approximately 3.6 oz. withbatteries inserted
Data TransferInfraredInfrared
Date RemindersYesYes
bG TestRemindersYesYes
Alarm ClockRemindersYesYes
Target bG LevelsYesYes
Health EventsYesYes
Electronic DiaryYesYes
Limitations ofProcedureGalactose >15 mg/dL will causeoverestimation of blood glucoseresults.Galactose >15 mg/dL will causeoverestimation of blood glucoseresults.
Lipemic Samples >1800 mg/dLLipemic Samples >1800 mg/dL
Intravenous administration ofascorbic acid which results inblood concentrations of ascorbicacid >3 mg/dL will causeoverestimation of blood glucoseresultsIntravenous administration ofascorbic acid which results inblood concentrations of ascorbicacid >3 mg/dL will causeoverestimation of blood glucoseresults
SystemFeature/ClaimACCU-CHEK Aviva ExpertMeter with Aviva PlusTest StripACCU-CHEK Combo Meterwith Aviva Plus Test StripPredicate(K111353)
RF WirelessCapabilityNoYes; additional components tosatellite board for Bluetoothcommunication with ACCU-CHEK Spirit Combo insulininfusion pumps
MeasurementUnits of InsulinBolus ResultCalculations0.5 units (appropriate forsyringe/pen administration ofinsulin)0.1 units (appropriate for pumpadministration of insulin)
Basal Insulin*Diary feature for tracking basalinsulinBasal insulin is controlled viapump
User GroupDiabetes patients treated withmultiple daily insulin injection(MDI) therapyDiabetes patients treated withinsulin pump therapy or multipledaily insulin injection (MDI)therapy

.

:

{5}------------------------------------------------

Similarities Table (continued)

б

{6}------------------------------------------------

Differences Table

  • The basal insulin values that are recorded do not influence the bolus advice.

7. Data demonstrating substantial equivalence

Performance testing on the ACCU-CHEK Aviva Expert System demonstrated that the device meets the performance requirements for its intended use. The data demonstrate that the system is substantially equivalent to the predicate device.

Below is the method comparison data for the system:

Results for glucose concentrations less than 75 mg/dL

Within +5 mg/dLWithin +10 mg/dLWithin + 15 mg/dL
41/48 (85.4%)48/48 (100%)48/48 (100%)

510(k) Summary

{7}------------------------------------------------

Results for glucose concentrations greater than or equal to 75 mg/dL

Within +5 %Within +10 %Within + 15%Within + 20 %
147/252 (58.3%)222/252 (88.1%)246/252 (97.6%)250/252 (99.2%)

Below is the repeatability (wtihin lot) precision for the system:

Blood12345
N100100100100100
Mean [mg/dL]42.184.5137.8208.2345.0
SD [mg/dL]1.22.23.35.67.9
CV [%]2.92.62.42.72.3

Below is the reproducibility (intermediate or day-to-day) precision for the system:

Control solutionsLowMidHigh
N100100100
Mean [mg/dL]45.1117.6303.0
SD [mg/dL]1.12.45.1
CV [%]2.42.01.7

510(k) Summary

.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an abstract human figure with three flowing lines extending from its body, resembling a bird in flight or a flowing ribbon.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 11, 2013

Roche Diagnostics Corporation C/O Nathan A. Carrington, Ph.D. 9115 Hague Rd INDIANAPOLIS IN 46250-0457

Re: K131366

Trade/Device Name: ACCU-CHEK Aviva Expert Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: LFR, NDC Dated: August 26, 2013 Received: August 29, 2013

Dear Dr. Carrington:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements conceming your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{9}------------------------------------------------

Page 2-Mr. Carrington

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincercly yours,

Carol C. Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K131366

Device Name: ACCU-CHEK Aviva Expert Blood Glucose Monitoring System

Indications for Use:

The ACCU-CHEK Aviva Expert System is indicated as an aid in the treatment of Insulin-requiring diabetes. The ACCU-CHEK Aviva Expert System consists of the ACCU-CHEK Aviva Expert Meter, ACCU-CHEK Aviva Plus test strips, ACCU-CHEK Aviva control solutions, and ACCU-CHEK Bolus Advisor. The ACCU-CHEK Aviva Expert System is Intended to facilitate the optimization of glycemic control in patients who are trained in multiple daily insulin injection therapy and are under the supervision of healthcare professionals experienced in managing insulin treated patients.

The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used for the quantitative measurement of giucose in fresh capillary whole blood samples drawn from the fingertips. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended for selftesting outside the body (in vitro diagnostic use) by people with diabetes. The ACCU-CHEK Aviva Expert blood glucose monitoring system is intended to be used by a single person and should not be shared. The ACCU-CHEK Aviva Expert blood giucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Altemative site testing should NOT be used with the ACCU-CHEK Aviva Expert blood glucose monitoring system. The ACCU-CHEK Aviva Expert System is intended for prescription home use only.

The ACCU-CHEK Aviva Expert meter is also indicated for the calculation of an insulin dose or carbohydrate intake based on user-entered data. The ACCU-CHEK Bolus Advisor, as a component of the Accu-Chek Aviva Expert meter, is intended for use in providing insulin dose recommendations in response to blood glucase, health events, and carbohydrate input. The ACCU-CHEK Bolus Advisor is intended to provide direction for insulin adjustment within the scope of a preplanned treatment program from a healthcare professional. Before its use, a physician or healthcare professional must prescribe the ACCU-CHEK Aviva Expert System and provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the ACCU-CHEK Bolus Advisor. Once programmed, a patient must consult with his/her physician or healthcare professional before making any changes to these ACCU-CHEK Bolus Advisor settings.

Prescription Use XX
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEEDED)

Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR)

Stayce Beck Division Sign-Off Office of In Vitro Devices and Radiologic Health

510(k)

Page 1 of _

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.