K Number
K111353
Manufacturer
Date Cleared
2012-07-17

(428 days)

Product Code
Regulation Number
880.5725
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.

The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician. The ACCU-CHEK 360° Insulin Pump Configuration software facilitates the monitoring and programming of the pump settings.

The ACCU-CHEK Aviva Combo blood glucose system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitoring the effectiveness of diabetes control. The ACCU-CHEK Aviva Combo blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the ACCU-CHEK Aviva Combo blood glucose monitoring system. The ACCU-CHEK Aviva Plus test strips are for use with the ACCU-CHEK Aviva Combo blood glucose meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The ACCU-CHEK Aviva Combo meter can also be used to interface with, and remotely control, the ACCU-CHEK Spirit Combo insulin infusion pumps via radio frequency communication. The ACCU-CHEK Aviva Combo meter is also indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data.

For in vitro diagnostic use

The ACCU-CHEK Aviva Combo system is intended for single patient use and should not be shared.

Device Description

The ACCU-CHEK Combo System includes:

  • An ACCU-CHEK Spirit Combo insulin infusion pump and optional . accessory software (ACCU-CHEK 360° Insulin Pump Configuration software), and
  • . An ACCU-CHEK Aviva Combo blood glucose monitor. Both devices utilize disposables that are already legally marketed.

The ACCU-CHEK Spirit Combo utilizes disposables that are already legally marketed:

  • ACCU-CHEK FlexLink Plus .
  • ACCU-CHEK Ultraflex ●
  • ACCU-CHEK Tender .
  • ACCU-CHEK Rapid-D .
  • ACCU-CHEK 3.15 ml Plastic Cartridge .

The ACCU-CHEK Aviva Combo utilizes disposables that are already legally marketed:

  • ACCU-CHEK Aviva Plus test strips .
  • · ACCU-CHEK Aviva control solutions

The ACCU-CHEK Combo System is meant only for single-patient use; the device may not be used for screening or diagnosis of diabetes or for multiplepatient use.

AI/ML Overview

Here's an analysis of the provided text to extract information about acceptance criteria and the supporting study for the ACCU-CHEK Combo System:

1. Table of Acceptance Criteria and Reported Device Performance:

The document provides specific performance criteria for the ACCU-CHEK Spirit Combo Insulin Pump. While it generally states that the ACCU-CHEK® Combo System performed equivalently to predicate devices, specific numerical acceptance criteria for the ACCU-CHEK Aviva Combo blood glucose monitor itself are not explicitly detailed in the provided text beyond general statements of accuracy.

System/ComponentAcceptance Criteria (Target)Reported Device Performance (Result)
ACCU-CHEK Spirit Combo Insulin Pump
Delivery Rates: 1.0 I.U./hAccuracy ± 5%Accuracy ± 5%
Delivery Rates: 0.05 I.U./hAccuracy ± 30%Accuracy ± 30%
Occlusion bolus volume: U100 (at 1.0 IU/h feed rate)≤ 1.5 IU≤ 1.5 IU
Occlusion bolus delay time: U100 (at 1.0 IU/h)≤ 2 h≤ 2 h
Occlusion bolus delay time: U100 (at 0.05 IU/h)≤ 24 h≤ 24 h
ACCU-CHEK Aviva Combo blood glucose monitorAccuracy (general)Accurate blood glucose test results
Blood sample volume(Not specified)0.6 µL
Test result time(Not specified)5 seconds

2. Sample Size Used for the Test Set and Data Provenance:

The document states, "Laboratory and Human Factors studies demonstrate that the ACCU-CHEK® Combo System performed in an equivalent manner to the predicate devices and is suitable for its intended use."

  • Sample Size for Test Set: Not explicitly mentioned in the provided text for either laboratory or human factors studies.
  • Data Provenance: Not specified whether the data was retrospective or prospective. The country of origin of the data is also not mentioned, although the submitting entity is Roche Diabetes Care, AG, based in Switzerland.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This information is not provided in the given text.

4. Adjudication Method for the Test Set:

This information is not provided in the given text.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

  • MRMC Study Done: No. The document refers to "human factors studies" but does not describe them as MRMC studies to compare human readers with and without AI assistance. The device itself is an insulin pump and blood glucose monitor, not an AI-assisted diagnostic tool for interpretation by human readers.
  • Effect Size: Not applicable, as no MRMC comparative effectiveness study of the described nature was mentioned.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

  • The ACCU-CHEK Combo System's components (insulin pump and blood glucose meter) inherently perform their functions in an "algorithm only" or standalone manner once activated, providing direct outputs (insulin delivery, glucose readings). The document implies that the "Laboratory studies" would form part of this standalone performance assessment for the pump's accuracy and the meter's accuracy, although specific details are limited. The "human factors studies" likely address human interaction with the device for usability and safety, rather than an "AI without human-in-the-loop" performance in a diagnostic sense.

7. The Type of Ground Truth Used:

For the ACCU-CHEK Spirit Combo Insulin Pump, the ground truth for accuracy of delivery rates, occlusion volume, and delay times would be established through laboratory measurements using precision instruments and calibrated test setups.

For the ACCU-CHEK Aviva Combo blood glucose monitor, "plasmalike results" are mentioned, suggesting a comparison of the device's glucose readings against a reference laboratory method (e.g., a highly accurate glucose analyzer).

8. The Sample Size for the Training Set:

This information is not provided in the given text. (Note: For medical devices like this, the concept of a "training set" in the machine learning sense isn't typically used. Instead, development and verification often involve extensive engineering testing and clinical validation.)

9. How the Ground Truth for the Training Set Was Established:

This information is not provided in the given text. (As mentioned above, the concept of a "training set" doesn't directly apply here in the AI context.)

{0}------------------------------------------------

K111353

JUL 1 7 2012 510(k) Summary of Safety and Effectiveness

IntroductionAccording to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
Submitted ByRoche Diabetes Care, AGKirchbergstrass 190, PostfachCH-3401 Burgdorf, Switzerland
USA Contact Person:Scott ThielAnson Group9001 Wesleyan Road, Suite 200Indianapolis, IN 46268
Device NameProprietary name: ACCU-CHEK Combo SystemCommon name: Insulin infusion pump, glucose test system, and accessoriesClassification name:Pump, infusion, insulin (21 CFR 880.5725); Class II; Product Code: LZG,Glucose test system (21 CFR 862.1345); Class II; Product Code: NBW,Drug dosing calculator (21 CFR 868.1890); Class II; Product Code: NDC
Predicate DevicesACCU-CHEK Spirit Insulin Pump - K060876ACCU-CHEK Aviva Plus blood glucose monitoring system - K101299ACCU-CHEK Pocket Compass Diabetes Management Software - K082595

Continued on next page

. *

:

ון י

{1}------------------------------------------------

Device Description The ACCU-CHEK Combo System includes:

  • An ACCU-CHEK Spirit Combo insulin infusion pump and optional . accessory software (ACCU-CHEK 360° Insulin Pump Configuration software), and
  • . An ACCU-CHEK Aviva Combo blood glucose monitor. Both devices utilize disposables that are already legally marketed.

The ACCU-CHEK Spirit Combo utilizes disposables that are already legally marketed:

  • ACCU-CHEK FlexLink Plus .
  • ACCU-CHEK Ultraflex ●
  • ACCU-CHEK Tender .
  • ACCU-CHEK Rapid-D .
  • ACCU-CHEK 3.15 ml Plastic Cartridge .

The ACCU-CHEK Aviva Combo utilizes disposables that are already legally marketed:

  • ACCU-CHEK Aviva Plus test strips .
  • · ACCU-CHEK Aviva control solutions

The ACCU-CHEK Combo System is meant only for single-patient use; the device may not be used for screening or diagnosis of diabetes or for multiplepatient use.

Continued on next page

{2}------------------------------------------------

The ACCU-CHEK Combo System is indicated for the treatment of insulin-Intended Use requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.

The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician. The ACCU-CHEK 360° Insulin Pump Configuration software facilitates the monitoring and programming of the pump settings.

The ACCU-CHEK Aviva Combo blood glucose system is intended for selftesting outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitoring the effectiveness of diabetes control. The ACCU-CHEK Aviva Combo blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the

ACCU-CHEK Aviva Combo blood glucose monitoring system. The ACCU-CHEK Aviva Plus test strips are for use with the ACCU-CHEK Aviva Combo blood glucose meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The ACCU-CHEK Aviva Combo meter can also be used to interface with, and remotely control, the

ACCU-CHEK Spirit Combo insulin infusion pumps via radio frequency communication. The ACCUCHEK Aviva Combo meter is also indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data.

For in vitro diagnostic use

The ACCU-CHEK Aviva Combo system is intended for single patient use and should not be shared.

Continued on next page

{3}------------------------------------------------

Technical Characteristics

The ACCU-CHEK Spirit Combo Insulin Pump is an external, portable insulin pump designed for continuous delivery of insulin. The design allows the delivery of 0.05 to 25.0 units of U100 insulin per hour in basal rates and 0.1 up to 25.0 units of U100 insulin per meal or meal bolus. Factory settings default to 0.05 to 25.0 units of U100 insulin per hour in basal rates and 0.1 up to 25.0 units of U100 insulin per meal or meal bolus. ACCU-CHEK 360° Insulin Pump Configuration Software is needed to adjust the hourly basal rates in the basal rate profiles that are set to zero by the factory.

The physical method used to accomplish the pumping mechanism action is a state-of-the-art syringe mechanism.

This mechanism consists of:

  • Brushless DC motor to drive the piston rod .
  • Telescoping piston rod connected with a thread to the cartridge plunger . (plunger retention) to prevent free flow and siphoning
  • Force sensor to detect occlusions in the fluid path .

The ACCU-CHEK Spirit Combo Insulin pump utilizes a single AA alkaline or lithium battery, or rechargeable NiMH accumulator.

The ACCU-CHEK Spirit Combo Insulin Pump accuracy has been verified for the following delivery rates and operational conditions.

AreaDescription
Delivery Rates:1.0 I.U./h Accuracy ± 5%
0.05 I.U./h Accuracy ± 30%
Occlusion bolus volume: U100Maximum Occlusion Volume with afeed rate of 1.0 IU/h”• 1.5 IU
Occlusion bolus delay time: U100• 2 h with 1.0 IU/h• 24 h with 0.05 IU/h

The ACCU-CHEK Aviva Combo meter was developed to utilize the ACCUCHEK Aviva Plus test system's technology and performance characteristics. These test strips are calibrated to deliver plasmalike results. Blood from your fingertip reacts with the chemicals in the test strip to create a harmless electrical current in the test strip. The meter reads the current and gives you the blood glucose result. The ACCU-CHEK Aviva Combo meter and ACCU-CHEK Aviva Plus test strip are designed to be simple and easy to use. It provides accurate blood glucose test results in 5 second while requiring a very small (0.6 µ) sample.

Continued on next page

{4}------------------------------------------------

Laboratory and Human Factors studies demonstrate that the ACCU-CHEK® Performance Summary Combo System performed in an equivalent manner to the predicate devices and is suitable for its intended use.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Roche Diagnostics C/O Mr. Scott Thiel Senior Regulatory Consultant Anson Group 9001 Wesleyan Road, Suite 200 Indianapolis. Indiana 46268

JUL 1 7 .2012

Re: K111353

Trade/Device Name: ACCU-CHEK® Combo System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG, NBW & NDC Dated: July 9, 2012 Received: July 10, 2012

Dear Mr. Thiel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Thiel

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements modi that I Dr mas made statutes and regulations administered by other Federal agencies. or the Act of any I ederal the Act's requirements, including, but not limited to: registration 1 od intest compty with the mart 807); labeling (21 CFR Part 801); medical device reporting and houng (21 er readical device-related adverse events) (21 CFR 803); good manufacturing (reporting of measur as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), nlease go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Touriner of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Fw. D. Stetson

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): _

Device Name: ACCU-CHEK® Combo System

Indications for Use:

The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.

The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician. The ACCU-CHEK 360° Insulin Pump Configuration software facilitates the monitoring and programming of the pump settings.

The ACCU-CHEK Aviva Combo blood glucose system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitoring the effectiveness of diabetes control. The ACCU-CHEK Aviva Combo blood glucose monitoring system should not be used for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing should NOT be used with the ACCU-CHEK Aviva Combo blood glucose monitoring system. The ACCU-CHEK Aviva Plus test strips are for use with the ACCU-CHEK Aviva Combo blood glucose meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The ACCU-CHEK Aviva Combo meter can also be used to interface with, and remotely control, the ACCU-CHEK Spirit Combo insulin infusion pumps via radio frequency communication. The ACCU-CHEK Aviva Combo meter is also indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data.

For in vitro diagnostic use

The ACCU-CHEK Aviva Combo system is intended for single patient use and should not be shared.

Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Lola for RSE Hielm

Page 1 of 1

Division Sign-Uff) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K11130

VOL 007 page000003

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).