K Number
K130794
Manufacturer
Date Cleared
2013-06-20

(90 days)

Product Code
Regulation Number
866.6020
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immunomagnetically selected and aligned. This product is for in vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

Device Description

The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope, consisting of the analyzer, a dedicated computer with CELLTRACKS® software, monitor, keyboard, mouse and uninterruptible power supply (UPS). The system also supports an optional Remote Review Workstation, which consists of a dedicated computer with CELLTRACKS® software, monitor, keyboard and mouse. Use of this product requires training and should be used under the supervision of laboratory management. The CELLTRACKS ANALYZER II® is for analysis of rare cells that are isolated from biological fluids including whole blood. It is used in conjunction with the CELLTRACKS® AUTOPREP® System, which automates and standardizes the sample preparation with specific reagent kits. The optional Remote Review Workstation provides the capability to review images and report results remotely.

AI/ML Overview

The provided text does not contain detailed acceptance criteria or a specific study that demonstrates the device meets these criteria in the format requested. The document is a 510(k) summary for the CELLTRACKS ANALYZER II® System, focusing on establishing substantial equivalence to a predicate device.

However, based on the information provided, here's a breakdown of what can be inferred or explicitly stated, and what is not present:

1. Table of Acceptance Criteria and Reported Device Performance:

This information is not provided in the document. The submission states "The information presented in the premarket notification demonstrates that the performance of the CELLTRACKS ANALYZER II® System (modified) is substantially equivalent to the predicate device." but does not list specific performance metrics or their acceptable ranges.

2. Sample Size Used for the Test Set and Data Provenance:

This information is not explicitly provided in the document. The document mentions "Non-clinical testing for the CELLTRACKS ANALYZER II® System included: Unit/Integration/System Verification, System Level Validation, Stress Testing, Regression Testing, Control Cell/Sample Performance Validation Testing." but does not specify the sample sizes, data origin (country), or whether the data was retrospective or prospective for these tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

This information is not provided in the document. The device is for "enumerating fluorescently labeled cells that are immunomagnetically selected and aligned," so human interpretation/adjudication might not be the primary ground truth establishment method. However, if any human review was involved in validation, it's not detailed.

4. Adjudication Method for the Test Set:

This information is not provided in the document.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

This information is not provided in the document. Given the description of the device as a "semi-automated fluorescence microscope used to enumerate fluorescently labeled cells," it's unlikely a traditional MRMC study as commonly understood for image-based diagnostic interpretations was performed or would be relevant in the same way. The focus seems to be on the accuracy and consistency of the enumeration function itself.

6. Standalone (Algorithm Only) Performance:

The device is described as "semi-automated" and includes "CELLTRACKS® software." The primary study mentioned is "non-clinical functional testing" to demonstrate substantial equivalence. The document doesn't explicitly refer to "algorithm only" performance as a distinct study. However, the system's core function is automated cell enumeration, suggesting that its performance is essentially standalone in terms of the enumeration task. The "semi-automated" aspect likely refers to the overall workflow involving sample preparation and then the device's automated analysis.

7. Type of Ground Truth Used:

This information is not explicitly provided in the document. For a cell enumeration system, the ground truth would typically be established by a highly accurate (manual or another validated automated) counting method, or potentially by using spiked samples with a known number of cells. The document mentions "Control Cell/Sample Performance Validation Testing," which implies the use of samples with known characteristics, but the method for establishing that "known" is not detailed.

8. Sample Size for the Training Set:

This information is not provided in the document. The document refers to software modifications (version 2.6.0) but does not detail any machine learning algorithms or training processes that would require a separate training set. The term "algorithm" is not used in the context of learned models. It appears to be a rule-based or image processing system rather than a deep learning model.

9. How the Ground Truth for the Training Set was Established:

This information is not provided and is likely not relevant as no machine learning training set is mentioned.


Summary of what is present:

  • Device Name: CELLTRACKS ANALYZER II® System
  • Predicate Device: CELLTRACKS ANALYZER II® (K113181)
  • Modifications: Software update (version 2.6.0) and addition of an optional Remote Review Workstation.
  • Basis for Equivalence: Non-clinical functional testing demonstrating performance is "substantially equivalent" to the predicate device.
  • Testing Types Mentioned: Unit/Integration/System Verification, System Level Validation, Stress Testing, Regression Testing, Control Cell/Sample Performance Validation Testing.
  • Intended Use/Indications: Enumerating fluorescently labeled cells that are immunomagnetically selected and aligned, for in vitro diagnostic use.

It's important to note that a 510(k) summary often provides a high-level overview. The detailed performance data, acceptance criteria, and study methodologies would be contained within the full 510(k) submission, which is not publicly available in this format.

{0}------------------------------------------------

K130794

Veridex, LLC. 1001 US Highway 202 Raritan, NJ 08869

Traditional 510(k)
CELLTRACKS ANALYZER II March 20, 2013

JUN 2 0 2013

.

8.0 510(K) SUMMARY

.

Submitter's NameVeridex, LLC
Address1001 US Highway 202Raritan, NJ 08869-0606
Telephone NumberFax Number908-927-4947908-526-5059
Contact PersonKimberly Prescott, Associate Director, Global Regulatory Affairs
Date:March 20, 2013
Name of the DeviceCELLTRACKS ANALYZER II® System
Common or UsualNameImmunomagnetic circulating cancer cell selection and enumeration system
Classification NameImmunomagnetic circulating cancer cell selection and enumeration systemDevice Class: IIProduct Code: NQIRegulation Number: 21 CFR 866.6020
Performance StandardsThere are no performance standards promulgated for this device.
Indications for UseThe CELLTRACKS ANALYZER II® is a semi-automated fluorescencemicroscope used to enumerate fluorescently labeled cells that areimmunomagnetically selected and aligned. This product is for in vitrodiagnostic use when used in tandem with specimen preparation equipmentand reagents that are legally marketed for in vitro diagnostic use with thisdevice.
Identification of theLegally MarketedDevice(Predicate Device)CELLTRACKS ANALYZER II® SystemName: Immunomagnetic circulating cancer cell selection and enumerationsystemDevice Class: IIProduct Code: NQIRegulation Number: 21 CFR 866.6020

. . . .

{1}------------------------------------------------

Veridex, LLC. 1001 US Highway 202 Raritan, NJ 08869

510(k) Summary, continued The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope, consisting of the analyzer, a dedicated computer with CELLTRACKS® software, monitor, keyboard, mouse and uninterruptible power supply (UPS). The system also supports an optional Remote Review Workstation, which consists of a dedicated computer with CELLTRACKS® software, monitor, keyboard and mouse. Use of this product requires training and should be used under the Device Description supervision of laboratory management. The CELLTRACKS ANALYZER II® is for analysis of rare cells that are isolated from biological fluids including whole blood. It is used in conjunction with the CELLTRACKS® AUTOPREP® System, which automates and standardizes the sample preparation with specific reagent kits. The optional Remote Review Workstation provides the capability to review images and report results remotely. The CELLTRACKS ANALYZER 118 (modified) is substantially equivalent to the CELLTRACKS ANALYZER II® (current - K113181, December 12, 2011). There has been no change to intended use, fundamental scientific technology, mode of operations, or specimen type/identification. There are 2 main modifications to the current CELLTRACKS Comparison to ANALYZER II® (K118131, cleared December 12, 2011) that are Predicate Device included in this Traditional 510(k) submission: The CELLTRACKS ANALYZER II® Software has been modified . to version 2.6.0, and is present on both the CELLTRACKS ANALYZER 110 PC and the Remote Review Workstation PC. A Remote Review Workstation PC is now available, supporting . identical image review and result reporting functionality that is currently supported locally on the CELLTRACKS ANALYZER II®. Non-clinical testing for the CELLTRACKS ANALYZER II® System included: Unit/Integration/System Verification . System Level Validation Description of Testing . . Stress Testing Regression Testing ● Control Cell/Sample Performance Validation Testing .

{2}------------------------------------------------

Veridex, LLC. 1001 US Highway 202 Raritan, NJ 08869

Traditional 510(k) CELLTRACKS ANALYZER II® March 20, 2013

The information presented in the premarket notification demonstrates that the performance of the CELLTRACKS ANALYZER II® System (modified) is substantially equivalent to the predicate device. Conclusion of Equivalence was demonstrated through non-clinical functional testing for Testing the modified device. The information presented in the premarket notification provides a reasonable assurance that the CELLTRACKS ANALYZER II® System (modified) is as safe and effective as the predicate device for the stated intended use.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 20, 2013

VERIDEX, LLC C/O MS. KIMBERLY PRESCOTT ASSOCIATE DIRECTOR, GLOBAL REGULATORY AFFAIRS 1001 US HIGHWAY 202 RARITAN NJ 08869

Re: K130794

Trade/Device Name: CELLTRACKS ANALYZER II® System Regulation Number: 21 CFR 866.6020

Regulation Name: Immunomagnetic circulating cancer cell selection and enumeration

system

Regulatory Class: II Product Code: NQI Dated: June 10, 2013 Received: June 13, 2013

Dear Ms. Prescott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely yours.

Maria M. Chan -S

Maria M. Chan Director

Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health

Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K130794

Device Name: CELLTRACKS ANALYZER II® System

Indications For Use:

The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immunomagnetically selected and aligned. This product is for in vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)

MariaMChan -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

K130794 510(k):

§ 866.6020 Immunomagnetic circulating cancer cell selection and enumeration system.

(a)
Identification. An immunomagnetic circulating cancer cell selection and enumeration system is a device that consists of biological probes, fluorochromes, and other reagents; preservation and preparation devices; and a semiautomated analytical instrument to select and count circulating cancer cells in a prepared sample of whole blood. This device is intended for adjunctive use in monitoring or predicting cancer disease progression, response to therapy, and for the detection of recurrent disease.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Immunomagnetic Circulating Cancer Cell Selection and Enumeration System.” See § 866.1(e) for availability of this guidance document.