K Number
K113181
Manufacturer
Date Cleared
2011-12-12

(45 days)

Product Code
Regulation Number
866.6020
Panel
PA
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immunomagnetically selected and aligned. This product is for in vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

Device Description

The CELLTRACKS ANALYZER II® is a semi-automated fluorescence microscope, consisting of the analyzer, a dedicated computer with CELLTRACKS® software, monitor, keyboard, mouse and uninterruptible power supply (UPS). Use of this product requires training and should be used under the supervision of laboratory management. The CELLTRACKS ANALYZER II® is for analysis of rare cells that are isolated from biological fluids including whole blood. It is used in conjunction with the CELLTRACKS® AUTOPREP® System, which automates and standardizes the sample preparation with specific reagent kits.

AI/ML Overview

Here's an analysis of the provided text regarding the CellTracks Analyzer II® device's acceptance criteria and studies:

Based on the provided 510(k) summary, the device under review (CellTracks Analyzer II®) is an upgrade to an existing predicate device (CellTracks Analyzer II® K060110). This submission primarily focuses on software verification and validation, rather than a new standalone clinical study with specific acceptance criteria in terms of diagnostic performance metrics like accuracy, sensitivity, or specificity against a clinical ground truth.

The document does not explicitly state specific acceptance criteria in terms of diagnostic performance metrics for the device. Instead, the acceptance criterion appears to be demonstrating substantial equivalence to the predicate device through software verification and validation.

The "study" that proves the device meets the acceptance criteria is the software verification and validation activities. The summary states: "Equivalence was demonstrated through software verification and validation." This implies that the software changes (upgrade from version 2.4 to 2.5.0, operating system update, anomaly solutions, new features) were thoroughly tested to ensure they did not negatively impact the device's function and continued to perform as expected, consistent with the predicate device.

Here's a breakdown of the requested information based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied by the 510(k) Summary):

  • Maintain substantial equivalence to the predicate device (CellTracks Analyzer II® K060110).
  • Ensure the software upgrade (from version 2.4 to 2.5.0) does not introduce new safety or effectiveness issues.
  • The software performs its intended functions (image acquisition, analysis, cell selection, review, reporting, archiving, logic and interface to PC) consistently with the predicate.

Reported Device Performance (from the 510(k) Summary):

  • The performance of the CellTracks Analyzer II® (modified) is demonstrated to be substantially equivalent to the predicate device.
  • No new issues of safety or effectiveness have been raised.
  • Changes made include an upgraded operating system, solutions to software anomalies, new features for customer satisfaction, and software text string translations.
Acceptance Criteria (Implied)Reported Device Performance
Maintain substantial equivalence to predicate device.Device demonstrated substantial equivalence to the predicate device.
No new issues of safety or effectiveness.No new issues of safety or effectiveness have been raised.
Software functions (image acquisition, analysis, etc.) remain constant.All algorithms and logic for image acquisition, analysis, cell selection, review, reporting and archiving, and the interface remain the same.

2. Sample Size Used for the Test Set and Data Provenance

The 510(k) summary does not explicitly mention a "test set" in the context of clinical cases or patient samples for diagnostic performance evaluation or sample sizes for such tests. Given that the submission is for a software upgrade to an already cleared device and demonstrates equivalence through software verification and validation, it's highly probable that testing involved internal software validation datasets and possibly samples used in regression testing, but no specific number of patient samples or studies on human data are described as part of this submission for establishing substantial equivalence. The data provenance is not specified.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the summary. Since the submission focuses on software changes and substantial equivalence through software verification/validation rather than a new clinical performance study, there's no mention of experts establishing ground truth for a test set of patient data.


4. Adjudication Method for the Test Set

Since no clinical "test set" with expert assessment is mentioned, no adjudication method is described.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done according to this 510(k) summary. The submission focuses on software verification and validation to demonstrate equivalence to a predicate device, not on assessing human reader improvement with or without AI assistance. The device itself is a "semi-automated fluorescence microscope" for enumeration, implying technical functionality rather than direct AI assistance to human diagnostic interpretation in the typical MRMC study context.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

The 510(k) summary does not explicitly describe a standalone performance study of the algorithms. However, the device description states it is a "semi-automated fluorescence microscope" and explicitly mentions that "All of the algorithms associated with image acquisition, analysis, cell selection, review, reporting and archiving and the logic and interface to the PC remain the same" as the predicate. This suggests the algorithms have a standalone function within the semi-automated enumeration process, but formal standalone performance metrics (e.g., accuracy, precision of enumeration against a 'true' count) for the upgraded software version are not presented as part of this summary to demonstrate acceptance criteria. The verification and validation activities would have confirmed that these algorithms continue to perform their intended function.


7. The Type of Ground Truth Used

Given the nature of the device (immunomagnetic cell selection and enumeration) and the focus on software updates for an existing device, the "ground truth" during software verification and validation would likely involve:

  • Reference data/simulated data: To test algorithm changes and ensure correct processing.
  • Known sample counts/calibrated controls: For enumeration accuracy testing.
  • Comparison to the predicate device's output: To ensure consistent results post-update.

However, the specific type of ground truth (e.g., pathology, outcomes data, expert consensus on images) is not detailed in this summary. It's more likely related to internal technical validation against established benchmarks for cell enumeration and image processing.


8. The Sample Size for the Training Set

The 510(k) summary does not mention a training set in the context of machine learning. The device described appears to use established algorithms for image analysis and cell enumeration rather than a machine learning model that requires a "training set" in the modern sense. Therefore, this question is not applicable based on the provided document.


9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned (see point 8), this information is not applicable.

§ 866.6020 Immunomagnetic circulating cancer cell selection and enumeration system.

(a)
Identification. An immunomagnetic circulating cancer cell selection and enumeration system is a device that consists of biological probes, fluorochromes, and other reagents; preservation and preparation devices; and a semiautomated analytical instrument to select and count circulating cancer cells in a prepared sample of whole blood. This device is intended for adjunctive use in monitoring or predicting cancer disease progression, response to therapy, and for the detection of recurrent disease.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Immunomagnetic Circulating Cancer Cell Selection and Enumeration System.” See § 866.1(e) for availability of this guidance document.