K Number
K130359

Validate with FDA (Live)

Date Cleared
2013-03-13

(28 days)

Product Code
Regulation Number
870.4350
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The (X-Coated) Capiox® SX25 and SX18 Hollow Fiber Oxygenators with/without Detachable Hardshell Reservoirs are used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery for periods up to 6 hours. The integral heat exchanger is used to warm or cool blood or perfusion fluid as it flows through the device. The (detachable) hardshell reservoir is used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The Hardshell Reservoir is also used for post-operative chest drainage and autotransfusion procedures to aseptically return the blood to the patient for blood volume replacement. The Hardshell Reservoir is also used with the vacuum-assisted venous return technique during cardiopulmonary bypass. (The X-Coating™ is a polymer coating that is applied to blood contacting surfaces of the oxygenator to reduce the adhesion of platelets to the surfaces of the device.)

The Capiox® Hardshell Reservoir is a hardshell reservoir used to store blood during extracorporeal circulation from both the venous line. And the cardiotomy line. The reservoir contains filters to remove particulate matter and defoamers to facilitate air bubble removal. The Hardshell Reservoir is also used for post-operative chest drainage and autotransfusion procedures to aseptically return blood to the patient for blood volume replacement. The Hardshell Reservoir is also used with the vacuum-assisted venous return technique during cardiopulmonary bypass. The Hardshell Reservoir contains X-Coating, which is intended to reduce platelet adhesion on the surfaces of the device. The device may be used for procedures lasting up to 6 hours.

Device Description

The modified and predicate Capiox SX Oxygenator utilizes porous fiber technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of a patient during cardiopulmonary bypass. A fiber bundle offers the porous membrane surface to sufficiently permit the movement of gases through the walls of the hollow fibers via diffusion. The modified and predicate Capiox SX Oxygenator has an integrated heat exchanger that is comprised of stainless steel encased in a polycarbonate housing. The stainless steel acts as heat transfer material that permits heat that is generated from a temperature controlled external water bath to transverse across the walls of the stainless steel to effect the necessary temperature change upon circulating blood. With respect to the filtration of blood, the modified and predicate Capiox Hardshell Reservoir relies upon mechanical entrapment of particulates and emboli within the filter media as a means to remove those particulates from the blood. The design of the modified Capiox SX oxygenator device is unaffected by the changes being incorporated at this time. The subject of this Special 510(k) is a modification being made to the Hardshell Reservoir. The design of the Hardshell Reservoir component remains identical to the design of the original reservoir that was cleared by FDA with (K961000, K962667, K993772, and K013526) except that a positive pressure relief valve will be included in the lid of the reservoir. The intent of the relief valve is to eliminate excessive pressure that could accumulate in a reservoir during bypass procedures. The materials that are used in the construction of the CAPIOX® SX Oxygenator/Hardshell Reservoir may include, but are not limited to, nylon, polycarbonate, stainless steel, polyvinyl chloride, polyurethane, polyester, polypropylene, polyethylene, and X-Coating™.

AI/ML Overview

The provided text describes a 510(k) summary for a modified medical device, the Capiox® SX Oxygenator and Hardshell Reservoir. The document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device's efficacy through extensive clinical studies against specific performance criteria.

Therefore, many of the requested categories for acceptance criteria and studies (like sample sizes for test/training sets, expert qualifications, ground truth establishment, MRMC studies, or standalone algorithm performance) are not applicable or not provided in this type of regulatory submission. This document describes in-vitro performance evaluations for substantial equivalence, not a study to establish clinical efficacy or AI performance.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not provide a table with explicit acceptance criteria and corresponding reported performance metrics in the format of a typical efficacy study. Instead, it lists performance evaluations conducted to demonstrate substantial equivalence to predicate devices. The "reported device performance" is implied to meet the expectations established by the predicate devices.

Performance Evaluation CategoryImplied Acceptance Basis
Assessment of Reservoir Pressure during Simulated Bypass ProceduresDevice pressure performance during simulated bypass procedures should be equivalent to or within acceptable limits compared to the predicate device, especially with the new pressure relief valve.
Positive and Negative Pressure Testing of the ReservoirThe modified reservoir, including the new valve, must withstand and perform as expected under both positive and negative pressure conditions, comparable to the predicate.
Pressure Relief Valve performance following application of vacuum to the reservoirThe pressure relief valve must function correctly to relieve pressure when vacuum is applied, as designed.
Pressure Relief Valve-to-Reservoir Interface TestingThe interface between the new valve and the reservoir must be secure and functional.
Usability TestingThe device, with the modification, should be usable for its intended purpose without introducing new usability issues.
Sterilization AssessmentThe device must remain sterilizable after modification.
Shock Drop and Vibration TestingThe device must withstand typical transportation and handling stresses; performance should be comparable to the predicate.
Artificial Conditioning to Shelf-Life of the productThe modification should not adversely affect the product's shelf-life under simulated aging conditions.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified. The evaluations mentioned are primarily in-vitro tests and assessments, not typically described with "test set" sample sizes in the context of clinical data or AI model evaluation.
  • Data Provenance: In-vitro performance evaluations. No information on country of origin or whether it's retrospective/prospective in a clinical sense. These are laboratory-based engineering tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Not Applicable. This is a 510(k) submission for a modified medical device, not a study evaluating an AI algorithm's diagnostic performance requiring expert-established ground truth. The "ground truth" here is the expected physical and functional performance of the device based on engineering principles and comparison to predicate devices.

4. Adjudication Method for the Test Set

  • Not Applicable. No human adjudication method described, as this is not a study requiring expert consensus on clinical findings or AI outputs.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. An MRMC study is relevant for evaluating the impact of AI assistance on human reader performance for diagnostic tasks. This document describes the substantial equivalence of a physical medical device (oxygenator/reservoir), not a diagnostic AI system.

6. Standalone (Algorithm Only) Performance

  • No. This is not an AI algorithm. It's a physical medical device.

7. Type of Ground Truth Used

  • Engineering Specifications and Predicate Device Performance. The "ground truth" for these performance evaluations is the established engineering design specifications for the modified device and the known, acceptable performance characteristics of the predicate devices. The goal is to show the modified device performs equivalently or within acceptable parameters.

8. Sample Size for the Training Set

  • Not Applicable. There is no "training set" in the context of AI for this type of device modification. The device design and materials are based on existing engineering knowledge and the predicate device's design.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As there is no training set mentioned in an AI context, there's no ground truth established for it. The design of the modified device is based on internal engineering processes, regulatory requirements, and the characteristics of the predicate devices.

{0}------------------------------------------------

SECTION 5 - 510(k) SummaryModified Capiox® SX Oxygenator and Hardshell ReservoirMAR 1 3 2013
Submitter Information24
Device Names24
Identification of Predicate Device25
Intended Use25
Principles of Operation and Technology27
Design and Materials27
Performance Evaluations28
Substantial Equivalence Comparison28
Substantial Equivalence Statement28
Conclusion for 510(k) Summary28

.

.

:

.

{1}------------------------------------------------

Submitter Information:

Contact: This submission was prepared in February 2013 by: Suzanne Grenier, RAC Sr. Regulatory Affairs Specialist Terumo Cardiovascular Systems Corp. 125 Blue Ball Road Elkton, MD 21921 Telephone: 1-800-262-3304, Ext. 7688

This submission was prepared for: Terumo Cardiovascular Systems Corporation 125 Blue Ball Road Elkton, MD 21921 Registration #1124841

Device Names/Classifications:

Proprietary NameClassification NameCommon Name
Capiox® SX18 Hollow FiberCardiopulmonary BypassBlood Gas Oxygenator
Oxygenator with detachableOxygenator (Code: DTZ)
Hardshell ReservoirCardiopulmonary BypassBlood Reservoir
Blood Reservoir (Code: DTN)
Capiox® SX25 Hollow FiberCardiopulmonary BypassBlood Gas Oxygenator
Oxygenator with detachableOxygenator (Code: DTZ)
Hardshell ReservoirCardiopulmonary BypassBlood Reservoir
Blood Reservoir (Code: DTN)
X-Coated Capiox® SX18Cardiopulmonary BypassBlood Gas Oxygenator
Hollow Fiber OxygenatorOxygenator (Code: DTZ)
with/without detachableCardiopulmonary BypassBlood Reservoir
Hardshell ReservoirBlood Reservoir (Code: DTN)
X-Coated Capiox® SX25Cardiopulmonary BypassBlood Gas Oxygenator
Hollow Fiber OxygenatorOxygenator (Code: DTZ)
with/without detachableCardiopulmonary BypassBlood Reservoir
Hardshell ReservoirBlood Reservoir (Code: DTN)
Capiox® Hardshell ReservoirCardiopulmonary BypassBlood Reservoir
Blood Reservoir (Code: DTN)

{2}------------------------------------------------

Predicate Devices:

The device submitted in this 510(k) maintains characteristics that are substantially equivalent in intended use, design, technology/principles of operation, materials and specifications to the following devices:

CAPIOX® SX18 Hollow Fiber Oxygenator with Detachable Hardshell Reservoir (K961000)

CAPIOX® SX25 Hollow Fiber Oxygenator with Detachable Hardshell Reservoir (K962667)

X-Coated CAPIOX® SX18 and SX25 Hollow Fiber Oxygenators with/without Detachable Hardshell Reservoirs (K993772)

Capiox® RX Hardshell Reservoir (K013526)

{3}------------------------------------------------

Intended Use:

The intended use remains the same as the intended use in the cleared submissions (K961000, K962667, K993772, and K013526). There have been no changes to the indications or intended use of the modified devices as a result of the addition of the positive pressure relief valve.

Intended Use for Oxygenator/Reservoir Product:

The (X-Coated) Capiox® SX25 and SX18 Hollow Fiber Oxygenators with/without Detachable Hardshell Reservoirs are used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery for priods up to 6 hours.

The integral heat exchanger is used to warm or cool blood or perfusion fluid as it flows through the device.

The (detachable) hardshell reservoir is used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains a venous sectiventhat is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate asir bubble removal.

The Hardshell Reservoir is also used for post-operative chest drainage and autotransfusion procedures to aseptically return the blood to the patient for blood volume replacement.

The Hardshell Reservoir is also used with the vacuum-assisted venous return technique during cardiopulmonary bypass.

(The X-Coating™ is a polymer coating that is applied to blood contacting surfaces of the oxygenator to reduce the adhesion of platelets to the surfaces of the device. )

Intended Use for Reservoir Only Product:

The Capiox® Hardshell Reservoir is a hardshell reservoir used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains filters to remove particulate matter and defoamers to facilitate air bubble removel.

The Hardshell Reservoir is also used for post-operative chest drainage and autotransfusion procedures to aseptically return blood to the patient for blood volume replacement.

The Hardshell Reservoir is also used with the vacuum-assisted venous return technique during cardiopulmonary bypass.

The Hardshell Reservoir contains X-Coating, which is intended to reduce platelet adhesion on the surfaces of the device.

The device may be used for procedures lasting up to 6 hours.

{4}------------------------------------------------

Principles of Operation and Technology:

The technological characteristics and principles of operation remain the same as that of the predicate devices previously cleared (K961000, K962667, K993772, and K013620, for all devices in this submission. There have been no changes to the fundaments lof ivi an technology of the modified devices.

The modified and predicate Capiox SX Oxygenator utilizes porous fiber technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of a patient during cardiopulmonary by the A fiber bundle offers the porous membrane surface to sufficiently permit the movement of gases through the walls of the hollow fibers via diffusion.

The modified and predicate Capiox SX Oxygenator has an integrated heat exchanger that is comprised of stainless steel encased in a polycarbonate housing. The stainless steel acts as heat transfer material that permits heat that is generated from a temperature controlled external i water bath to transverse across the walls of the stainless steel to effect the necessary temperature change upon circulating blood.

With respect to the filtration of blood, the modified and predicate Capiox Hardshell Reservoir relies upon mechanical entrapment of particulates and emboli within the filter messive its a means to remove those particulates from the blood.

Design and Materials:

With respect to the design of the oxygenator, the design of the modified Capiox SX oxygenator device is unaffected by the changes being incorporated at this time. The subject of this Special 510(k) is a modification being made to the Hardshell Reservoir.

With respect to the design of the Hardshell Reservoir component remains identical to the design of the original reservoir that was cleared by FDA with (K961000, K962667, K993772, and K013526) ~ except that a positive pressure relief valve will be included in the lid of the reservoir. The intent of the relief valve is to eliminate excessive pressure that could accumulate in a reservoir during bypass procedures.

The materials that are used in the construction of the CAPIOX® SX Oxygenator/Hardshell Reservoir may include, but are not limited to, nylon, polycarbonate, stainless steel, rollyvinyl chloride, polyurethane, polyester, polypropylene, polyethylene, and X-Coating™

Terumo Cardiovascular Systems concludes that the differences between the modified device and the predicate device do not affect the intended use of the device nor do they affect softev and effectiveness of the device when used as labeled.

{5}------------------------------------------------

Performance Evaluations:

Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. Substantial equivalence is demonstrated with the following in-vitro performance evaluations:

  • Assessment of Reservoir Pressure during Simulated Bypass Procedures .
  • Positive and Negative Pressure Testing of the Reservoir .
  • Pressure Relief Valve performance following application of vacuum to the reservoir .
  • Pressure Relief Valve-to-Reservoir Interface Testing .
  • Usability Testing .
  • Sterilization Assessment .
  • Shock Drop and Vibration Testing .
  • Artificial Conditioning to Shelf-Life of the product .

Substantial Equivalence Comparison:

In demonstrating substantial equivalence of the modified CAPIOX® SX Oxygenator/Hardshell Reservoir to the predicate CAPIOX® SX Oxygenator/Hardshell Reservoir, a comparative study and/or assessment was performed in each of the following areas:

  • Intended use .
  • Target Population .
  • Duration of use .
  • Product labeling .
  • Product design .
  • Materials .
  • Principles of Operation and Technology .
  • Device Performance .

Substantial Equivalence Statement:

The modified CAPIOX® SX Oxygenator/Hardshell Reservoir and the predicate CAPIOX® SX Oxygenator/Hardshell Reservoir are substantially equivalent in intended use, principles of operation and technology, design and materials, and performance. Any noted differences between the subject device and the predicate devices do not raise new issues of safere and effectiveness.

Conclusion:

Based upon the comparative studies and analyses, Terumo Cardiovascular Systems concludes that the modified CAPIOX® SX Oxygenator/Hardshell Reservoir is substarrights equivalent to the predicate modified CAPIOX® SX Oxygenator/Hardshell Reservoir. It is further concluded that any recognized differences noted during the assessments do not raise new issues of patient/user safety or product effectiveness.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular arrangement of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in all caps. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 13, 2013

Terumo Cardiovascular Systems Corporation c/o Suzanne Grenier 125 Blue Ball Road Elkton, MD 21921

Re: K130359

Trade/Device Name: CAPIOX SX18 and SX25 Hollow Fiber Oxygenators (with/without X-Coating), with/without Detachable Hardshell Reservoirs; and CAPIOX RX Hardshell Reservoir Regulation Number: 21 CFR 870.4350 -Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: February 12, 2013 Received: February 13, 2013

Dear Ms. Grenier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indictions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleding.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Righter.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{7}------------------------------------------------

Page 2 -- Ms. Suzanne Grenier

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050,

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH) s) Office of Compliance. Also, please note the regulation entitled, "Misbanding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of telefone to premation" (21CFR Part
CFR Part 803), please go to CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the Act from the
(800) 638-2041 or (301) 706-7100 continer Assistance at its toll-free num (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Matthew G. Hillebrenner

for Bram D. Zuckerman, M.D.
Director Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

SECTION 4 - Indications for Use Modified Capiox® SX Oxygenator and Hardshell Reservoir

510(k) Number (if known): Unknown at time of Submission

Device Name:

Capiox® SX18 Hollow Fiber Oxygenator with detachable Hardshell Reservoir

Capiox® SX25 Hollow Fiber Oxygenator with detachable Hardshell Reservoir

X-Coated Capiox® SX18 Hollow Fiber Oxygenator with/without detachable Hardshell Reservoir

X-Coated Capiox® SX25 Hollow Fiber Oxygenator with/without detachable Hardshell Reservoir

Indications For Use:

The (X-Coated) Capiox® SX25 and SX18 Hollow Fiber Oxygenators with/without Detachable Hardshell Reservoirs are used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery for periods up to 6 hours.

The integral heat exchanger is used to warm or cool blood or perfusion fluid as it flows through the device.

The (detachable) hardshell reservoir is used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air buble removal.

The Hardshell Reservoir is also used for post-operative chest drainage and autotransfusion procedures to aseptically return the blood to the patient for blood volume replacement.

The Hardshell Reservoir is also used with the vacuum-assisted venous return technique during cardiopulmonary bypass.

(The X-Coating™ is a polymer coating that is applied to blood contacting surfaces of the oxygenator to reduce the adhesion of platelets to the surfaces of the device.)

Prescription Use _____________________________________________________________________________________________________________________________________________________________OR OR
(Part 21 CFR 801 Subpart D)(Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Matthew G. Hillebrenner

{9}------------------------------------------------

t TERUMO

SECTION 4 – Indications for Use Modified Capiox® SX Oxygenator and Hardshell Reservoir

510(k) Number (if known): Unknown at time of Submission

Device Name:

Capiox® Hardshell Reservoir

Indications For Use:

The Capiox® Hardshell Reservoir is a hardshell reservoir used to store blood during extracorporeal circulation from both the venous line. And the cardiotomy line. The reservoir contains filters to remove particulate matter and defoamers to facilitate air bubble removal.

The Hardshell Reservoir is also used for post-operative chest drainage and autotransfusion procedures to aseptically return blood to the patient for blood volume replacement.

The Hardshell Reservoir is also used with the vacuum-assisted venous return technique during cardiopulmonary bypass.

The Hardshell Reservoir contains X-Coating, which is intended to reduce platelet adhesion on the surfaces
of the device of the device.

The device may be used for procedures lasting up to 6 hours.

Prescription Use XX __________________________________________________________________________________________________________________________________________________________OR OR SE
(Part 21 CFR 801 Subpart D)(Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE ~ CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Matthew G. Hillebrenner

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”