K Number
K961000

Validate with FDA (Live)

Date Cleared
1996-10-02

(204 days)

Product Code
Regulation Number
870.4350
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CAPIOX® SX18 Hollow Fiber Oxygenator is used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integral heat exchanger is used to warm or cool the blood or perfusion fluid flowing through the device.

The CAPIOX SX18 Hardshell Reservoir (detachable) is used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains filters to remove particulate matter and defoamers to facilitate air bubble removal.

Device Description

CAPIOX® SX18 Hollow Fiber Oxygenator contains an integrated heat exchanger and a detachable Hardshell Reservoir. This design permits an integrated system for ease of use as well as independent use of the oxygenator and of the Hardshell reservoir to accommodate a variety of circuit configurations.

The SX18 oxygenator is a membrane oxygenator consisting of microporous hollow fibers with an integrated heat exchanger, consisting of stainless steel tubes.

The detachable Hardshell reservoir has a rotatable connection with the oxygenator which permits optimum positioning for connections in the circuit. The venous blood inlet port is also rotatable to permit minimizing tubinq lengths which could result in lower circuit priming volumes.

The Hardshell reservoir contains a defoamer and a screen filter in the venous blood inlet section. The total capacity of the reservoir is 4,000 mL.

The cardiotomy section of the hardshell reservoir contains a defoamer and a cardiotomy filter to facilitate qas bubble removal and the removal of particulates/emboli from suctioned blood entering the reservoir.

A detachable sampling system is positioned at the top of the hardshell reservoir which contains 3 three-way stopcocks. These stopcocks can be used for sampling. The sampling system contains a one-way valve permitting withdrawal of liquid samples but prohibiting entry of air into the blood exiting the oxygenator.

AI/ML Overview

The provided text describes the CAPIOX® SX18 Hollow Fiber Oxygenator with detachable Hardshell Reservoir and its substantial equivalence to the Cobe CML Excel. However, the document does not contain a detailed study proving the device meets specific acceptance criteria with reported device performance statistics, sample sizes, ground truth establishment, or expert-based reviews.

The information provided largely focuses on demonstrating "substantial equivalence" to a predicate device (Cobe CML Excel) rather than detailing a specific study with quantitative acceptance criteria and results.

Here's a breakdown of what can be extracted and what is missing:

1. Table of Acceptance Criteria and Reported Device Performance

This specific table is not provided in the document. The text mentions "performance tests" for "Gas Transfer" and "Pressure Drop of Heat Exchanger and Oxygenator" but does not give specific numerical acceptance criteria or the measured performance values for the SX18. It only states, "some differences in performance were observed... however, these differences do not present clinically significant differences."

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided. There is no mention of a specific "test set" in the context of a clinical or ex-vivo study with a defined sample size.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided. The document makes no mention of expert review or ground truth establishment for a test set.

4. Adjudication Method for the Test Set

This information is not provided. There is no mention of an adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

This information is not provided. The document describes a comparison between the SX18 and CML Excel devices, but not a study involving human readers' performance with and without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This information is not applicable as the device is a medical oxygenator, not an algorithm or AI system. The "performance" discussed relates to the physical device's function.

7. Type of Ground Truth Used

This information is not explicitly stated in terms of a formal "ground truth" as typically understood in AI/algorithm studies. The performance comparison against the predicate device (Cobe CML Excel) serves as the basis for "substantial equivalence." The implicit ground truth is the established performance and safety profile of the Cobe CML Excel and generally accepted physiological requirements for oxygenation during cardiopulmonary bypass.

8. Sample Size for the Training Set

This information is not applicable as the device is not an AI/algorithm-based system requiring a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as above.


Summary of Available Information Related to Performance and Equivalence:

The document focuses on demonstrating Substantial Equivalence through comparison with a predicate device, the Cobe CML Excel, based on:

  • Intended Use: Same
  • Design and Materials: Similar, with some differences (e.g., integrated heat exchanger location, detachable reservoir, type of tubing, specific membrane technology - hollow fibers vs. plate membrane). The document asserts that these differences "do not present significant differences in the function and intended uses of the devices."
  • Technology and Principles of Operation: Substantially equivalent (e.g., membrane technology, gravity/vacuum collection, defoamers, filters).
  • Specifications: Some differences (e.g., priming volume, membrane surface area, maximum blood flow range). Claims these "do not affect the substantial equivalence... since both provide adequate gas exchange for clinical use."
  • Performance: "Comparison... was conducted relative to the extended acceptable blood flow rate specification to include 0.5 - 2 LPM."
    • Tests performed: Gas Transfer, Pressure Drop of Heat Exchanger and Oxygenator.
    • Result: "some differences in performance were observed... however, these differences do not present clinically significant differences."

Additional Safety Information Provided:

  • Gas Transfer relative to time
  • Heat Exchanger Performance Factor
  • Sterilization validation (SAL of 10-5)
  • Ethylene oxide residual limits
  • Manufacturing control tests (100% performance and leak testing)
  • Blood contacting materials tested per FDA GPM #G95-1 (ISO-10993)

In conclusion, while the document asserts the device's performance and safety through comparison with a predicate, it does not provide the specific-study details, quantitative acceptance criteria, or performance data in the format requested. The primary method of "proof" is the demonstration of substantial equivalence rather than a detailed report of a specific study meeting predefined, quantitative acceptance criteria.

{0}------------------------------------------------

K961000

Image /page/0/Picture/1 description: The image shows a date stamp. The stamp reads "OCT - 2 1996". The text is in black ink and is slightly rotated.

SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE

Proprietary Device Name:CAPIOX® SX18 Hollow Fiber Oxygenatorwith detachable Hardshell Reservoir
Classification Name:Cardiopulmonary bypass oxygenator, heatexchanger, reservoir, defoamer, bloodfilter and manifold.

Reason for Submission:

To extend the minimum flow rate specification.

Intended Use:

The CAPIOX® SX18 Hollow Fiber Oxygenator is used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integral heat exchanger is used to warm or cool the blood or perfusion fluid flowing through the device.

The CAPIOX SX18 Hardshell Reservoir (detachable) is used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains filters to remove particulate matter and defoamers to facilitate air bubble removal.

Description

CAPIOX® SX18 Hollow Fiber Oxygenator contains an integrated heat exchanger and a detachable Hardshell Reservoir. This design permits an integrated system for ease of use as well as independent use of the oxygenator and of the Hardshell reservoir to accommodate a variety of circuit configurations.

The SX18 oxygenator is a membrane oxygenator consisting of microporous hollow fibers with an integrated heat exchanger, consisting of stainless steel tubes.

{1}------------------------------------------------

  • II Summary and Certification Summary of Safety and Effectiveness
    The detachable Hardshell reservoir has a rotatable connection with the oxygenator which permits optimum positioning for connections in the circuit. The venous blood inlet port is also rotatable to permit minimizing tubinq lengths which could result in lower circuit priming volumes.

The Hardshell reservoir contains a defoamer and a screen filter in the venous blood inlet section. The total capacity of the reservoir is 4,000 mL.

The cardiotomy section of the hardshell reservoir contains a defoamer and a cardiotomy filter to facilitate qas bubble removal and the removal of particulates/emboli from suctioned blood entering the reservoir.

A detachable sampling system is positioned at the top of the hardshell reservoir which contains 3 three-way stopcocks. These stopcocks can be used for sampling. The sampling system contains a one-way valve permitting withdrawal of liquid samples but prohibiting entry of air into the blood exiting the oxygenator.

Substantial Equivalence

The CAPIOX® SX18 Oxygenator and Hardshell Reservoir are substantially equivalent to the Cobe CML Excel as follows:

Intended use: same

Design and Materials:

Both the CAPIOX SX18 and the Cobe CML Excel have integrated heat exchangers. The SX18 heat exchanger is coupled with the oxygenator, while the CML Excel heat exchanger is coupled with the reservoir. This design causes blood in the SX18 to flow from the reservoir to the pump and then to the heat exchanger/oxygenator, while the blood flows from the reservoir/heat exchanger to the pump and into the oxygenator in the CML Excel.

{2}------------------------------------------------

  • II Summary and Certification Summary of Safety and Effectiveness
    The hardshell reservoir for the SX18 is detachable while it is not detachable for the CML Excel.

The heat exchanger of the SX18 uses straight stainless steel tubes as blood conduits while the CML Excel uses coiled stainless steel tubes.

Gas exchange is accomplished through hollow polypropylene fibers in the SX18 and via a polypropylene plate membrane in the CML Excel.

Both devices incorporate hardshell reservoirs (open systems) which collect blood from the venous and cardiotomy returns. They both incorporate filters and defoamers/air separators to facilitate air and particulate removal.

Although some design dissimilarities exist, the performance testing results demonstrate that these differences do not present significant differences in the function and intended uses of the devices.

Technology and Principles of Operation

Both devices use membrane technology. The SX18 uses hollow fibers while the CML Excel uses a membrane plate. Both devices utilize gravity and/or external vacuum (cardiotomy) for blood collection into the reservoir. Air removal is facilitated by defoamers and the tendency of air to rise through liquid. Particulate removal is facilitated by the blood flow pathway through filters contained in the reservoirs. Some form of pumping mechanism is utilized to transfer blood from the reservoir component to the oxygenator component.

The technology and principles of operation for the SX18 and the CML Excel are substantially equivalent.

{3}------------------------------------------------

  • II Summary and Certification Summary of Safety and Effectiveness

Specifications

The priming volume for the SX18 is lower than that for the CML Excel (470 versus 850 mL, respectively). The membrane surface area for the SX18 is 1.8 m² and 3.0 m² for the Cobe CML Excel.

The maximum blood flow range for the SX18 is 7 LPM and for the Excel is 8 LPM (minimum is 0.5 LPM).

Theses differences do not affect the substantial equivalence of the devices since both provide adequate gas exchange for clinical use.

Performance

Comparison of the SX18 and CML Excel performance was conducted relative to the extended acceptable blood flow rate specification to include 0.5 - 2 LPM. The following tests were performed:

•Gas Transfer •Pressure Drop of Heat Exchanger and Oxygenator

In summary, some differences in performance were observed between the SX18 and the CML Excel, however, these differences do not present clinically significant differences.

The SX18 and the CML Excel are substantially equivalent in intended use, design and materials, technology/principles of operation, specifications and performance. Differences as described above do not raise new issues of safety or effectiveness.

Terumo's statement that this device is substantially equivalent to any other device is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action.

{4}------------------------------------------------

  • II Summary and Certification Summary of Safety and Effectiveness

Additional Safety Information

·Gas Transfer relative to time ·Heat Exchanger Performance Factor

•Sterilization conditions have been validated to provide a Sterility Assurance Level (SAL) of 10-5.

·Ethylene oxide residuals will not exceed the maximum residue limits proposed for Part 821 of Title 21 in the Federal Register of June 23, 1978 (or as finalized or amended).

•Manufacturing control tests include 100% performance and leak testing.

·Blood contacting materials were tested in accordance with the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, " Biological Evaluation of Medical Devices Part 1: Evaluation and Testing (External communicating devices/Circulating Blood/Limited contact duration) .

Date Prepared March 5, 1996

  • Prepared by: Sandi Hartka, M.A.S., R.A.C. Submissions Supervisor Regulatory Affairs
    • for: Terumo Medical Corporation 2100 Cottontail Lane Somerset, NJ 08873

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”