K Number
K123541
Date Cleared
2013-04-02

(134 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The syngo.CT Neuro Perfusion software package is designed to evaluate areas of brain perfusion. The software processes images or volumes that were reconstructed from continuously acquired CT data after the injection of contrast media.

It generates the following result volumes:

  • Cerebral blood flow (CBF)
  • Cerebral blood volume (CBV)
  • Local bolus timing (time to start (TTS), time to peak (TTP), time to drain (TTD))
  • Mean transit time (MTT)
  • Transit time to the center of the IRF (TMax)
  • Flow extraction product (permeability)
  • Temporal MIP
  • Temporal Average
  • Baseline Volume
  • Modified dynamic input data

The software also allows the calculation of mirrored regions or volumes of interest and the visual inspection of time attenuation curves. One clinical application is to visualize the apparent blood perfusion and the parameter mismatch in brain tissue affected by acute stroke.

Areas of decreased perfusion appear as areas of changed signal intensity:

  • Lower signal intensity for CBF and CBV
  • Higher signal intensity for TTP, TTD, MTT, and TMax

A second application is to visualize blood brain barrier disturbances by modeling extra-vascular leakage of blood into the interstitial space. This additional capability may improve the differential diagnosis of brain tumors and be helpful in therapy monitoring.

Device Description

The syngo. CT Neuro Perfusion software allows the quantitative evaluation of dynamic CT data of the brain acquired during the injection of a compact bolus of iodinated contrast material. It mainly aids in the early differential diagnosis of acute ischemic stroke. Blood-brain-barrier (BBB) imaging also supports the diagnostic assessment of brain tumors.

By providing images of e.g. cerebral blood flow (CBF), cerebral blood volume (CBV), time to peak (TTP), and Mean Transit Time (MTT) from one set of dynamic CT images or volumes, syngo.CT Neuro Perfusion allows a quick and reliable assessment of the type and extent of cerebral perfusion disturbances. The underlying approaches have been validated in extensive clinical studies and have been in routine clinical use for more than 10 vears.

The current syngo.CT Neuro Perfusion implementation allows simultaneous multi-slice processing and supports the workflow requirements in a stroke workflow. The availability of flow extraction product imaging extends the option to the diagnosis of brain tumors.

AI/ML Overview

The provided text does not contain detailed information about the acceptance criteria or a specific study that proves the device meets those criteria. It mainly focuses on the device's substantial equivalence to a predicate device and its indications for use.

However, based on the information available, I can infer some aspects related to non-clinical testing and general acceptance.

Here’s an attempt to structure the answer based on the provided text, highlighting what is present and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria & Standards (Inferred from text)Reported Device Performance (Inferred from text)
Conformity to IEC 60601-1-6 (Usability)Non clinical tests were conducted during product development to fulfill these requirements.
Conformity to IEC 62304 (Software Lifecycle)Non clinical tests were conducted during product development to fulfill these requirements. The testing results support that all software specifications have met the acceptance criteria.
Conformity to ISO 14971 (Risk Management)Risk analysis was completed and risk control implemented to mitigate identified hazards.
Conformity to DICOM Standard (2008)DICOM conformity is fully covered by syngo.via implementations.
Mitigation of identified hazardsRisk analysis completed and risk control implemented.
Software specifications performanceAll software specifications have met the acceptance criteria, as supported by testing results.
Verification and Validation for Substantial EquivalenceTesting for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
Safe and effective use based on labelingDevice labeling contains instructions for use and necessary cautions/warnings for safe and effective use.
No new potential safety risk compared to predicateSiemens' opinion is that the device does not introduce any new potential safety risk and performs as well as the predicate device.

2. Sample size used for the test set and the data provenance

The document does not specify a sample size for a test set or provide details on data provenance (e.g., country of origin, retrospective/prospective study design). The discussion of testing is general and relates to non-clinical software verification and validation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The filing describes non-clinical testing for software verification and validation rather than a clinical performance study with expert ground truth.

4. Adjudication method for the test set

This information is not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned in the provided text. The document describes a software package for post-processing CT data and does not detail studies on human reader performance with or without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document discusses "non clinical tests" for software verification and validation, and states that "all the software specifications have met the acceptance criteria." This implies a form of standalone performance assessment against predefined specifications, but the specifics of how "standalone" this was (e.g., if it involved simulated data or real patient data processed without human intervention for evaluation) are not detailed. It's not a clinical standalone study in the sense of diagnostic accuracy against a ground truth.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document refers to "non clinical tests" and "software specifications" rather than clinical ground truth like pathology or expert consensus from a clinical study. The "ground truth" for these non-clinical tests would likely be the expected output or behavior according to the software's functional requirements and design specifications.

8. The sample size for the training set

The document does not specify a sample size for a training set. The software likely relies on pre-established algorithms for generating perfusion maps, which would have been developed and "trained" (or validated) on various datasets over many years, as indicated by: "The underlying approaches have been validated in extensive clinical studies and have been in routine clinical use for more than 10 years." However, specifics about this device's training set are absent.

9. How the ground truth for the training set was established

The document mentions that "The underlying approaches have been validated in extensive clinical studies and have been in routine clinical use for more than 10 years." This suggests that the ground truth for the "training" (or more accurately, the development and historical validation of the underlying algorithms) would have been established through clinical studies, but the specific methods (e.g., expert consensus, correlation with other imaging modalities, or patient outcomes) are not detailed in this 510(k) summary for this particular device.

{0}------------------------------------------------

APR 0 2 2013

510(K) Summary FOR SYNGO.CT NEURO PERFUSION

Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355

Date Prepared: November 14, 2012,

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

  1. Contact Person: Mrs. Kimberly Mangum Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway D02 Malvern, PA 19355-1406 Phone: (610) 448-1772 Fax: (610) 448-1778 Email: kimberly.manqum@siemens.com
    1. Device Name and Classification syngo.CT Neuro Perfusion Product Name: Propriety Trade Name: syngo.CT Neuro Perfusion Classification Name: Computed Tomography X-ray System Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Device Class: Class II Product Code: 90JAK

{1}------------------------------------------------

3. Substantial Equivalence:

Siemens syngo. CT Neuro Perfusion post processing software package is substantially equivalent to the following medical devices in commercial distribution:

Predicate Device NameFDA ClearanceNumberFDA Clearance Date
syngo® Volume Perfusion CT-NeuroK07323801/03/2008

4. Device Description:

The syngo. CT Neuro Perfusion software allows the quantitative evaluation of dynamic CT data of the brain acquired during the injection of a compact bolus of iodinated contrast material. It mainly aids in the early differential diagnosis of acute ischemic stroke. Blood-brain-barrier (BBB) imaging also supports the diagnostic assessment of brain tumors.

By providing images of e.g. cerebral blood flow (CBF), cerebral blood volume (CBV), time to peak (TTP), and Mean Transit Time (MTT) from one set of dynamic CT images or volumes, syngo.CT Neuro Perfusion allows a quick and reliable assessment of the type and extent of cerebral perfusion disturbances. The underlying approaches have been validated in extensive clinical studies and have been in routine clinical use for more than 10 vears.

The current syngo.CT Neuro Perfusion implementation allows simultaneous multi-slice processing and supports the workflow requirements in a stroke workflow. The availability of flow extraction product imaging extends the option to the diagnosis of brain tumors.

5. Summary of Technical Characteristics of the Subject Device as Compared with the Predicate Device:

syngo.CT Neuro Perfusion software package is designed to evaluate areas of brain perfusion and visualize blood brain barrier disturbances. Syngo. CT Neuro Perfusion has similar intended use as the predicate syngo® Volume Perfusion-CT Neuro (K073238, clearance date 01/03/2008). syngo.CT Neuro Perfusion is designed to be operated on syngo.via platform in a single or multi user environment. New or modified features provided with syngo.CT Neuro Perfusion are provided in the device description.

6. Nonclinical Testing:

syngo.CT Neuro Perfusion is designed to fulfill the requirements of following standards.

  • IEC 60601-1-6 : 2006; Medical electrical equipment Part 1-6: General . requirements for basic safety and essential performance - Collateral Standard: Usability

{2}------------------------------------------------

  • IEC 62304 Ed. 1.0, "Medical Device Software Software Lifecycle . Processes"
  • . ISO 14971:2007; Medical devices - Application of risk management to medical devices
  • DICOM (Digital Imaging and Communications in Medicine) Standard: . 2008DICOM conformity is fully covered by syngo.via implementations.

Non clinical tests were conducted for syngo. CT Neuro Perfusion software package during product development. The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria.

Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.

7. Indications for Use:

The syngo. CT Neuro Perfusion software package is designed to evaluate areas of brain perfusion. The software processes images or volumes that were reconstructed from continuously acquired CT data after the injection of contrast media.

It generates the following result volumes:

  • Cerebral blood flow (CBF) .
  • . Cerebral blood volume (CBV)
  • . Local bolus timing (time to start (TTS), time to peak (TTP), time to drain (TTD))
  • . Mean transit time (MTT)
  • Transit time to the center of the IRF (TMax) .
  • Flow extraction product (permeability) .
  • . Temporal MIP
  • . Temporal Average
  • . Baseline Volume
  • Modified dynamic input data .

The software also allows the calculation of mirrored regions or volumes of interest and the visual inspection of time attenuation curves. One clinical application is to visualize the apparent blood perfusion and the parameter mismatch in brain tissue affected by acute stroke.

Areas of decreased perfusion appear as areas of changed signal intensity:

  • Lower signal intensity for CBF and CBV .
  • Higher signal intensity for TTP, TTD, MTT, and TMax .

A second application is to visualize blood brain barrier disturbances by modeling extra-vascular leakage of blood into the interstitial space. This additional capability may improve the differential diagnosis of brain tumors and be helpful in therapy monitoring.

{3}------------------------------------------------

General Safety and Effectiveness Concerns: 8.

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk manaqement is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

Conclusion as to substantial Equivalence: ல்

In summary, Siemens is of the opinion that the syngo.CT Neuro Perfusion software package does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate device.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human services, with three curved lines forming a wing-like shape.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 2, 2013

Siemens Medical Systems, Inc. % Ms. Kimberly Mangum Technical Specialist, Regulatory Submissions 51 Valley Stream Parkway, G-01 MALVERN PA 19355

Re: K123541

Trade/Device Name: syngo.CT Neuro Perfusion Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: March 8, 2013 Received: March 13, 2013

Dear Ms. Mangum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{5}------------------------------------------------

Page 2 - Ms. Kimberly Mangum

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Image /page/5/Picture/6 description: The image shows a stylized signature or logo. The signature appears to read "Ramongh. Merde" in a cursive style. The signature is surrounded by a geometric pattern that includes a series of nested squares and triangles. The overall design is intricate and appears to be hand-drawn.

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K123541

Device Name:

synqo.CT Neuro Perfusion

Indications for Use:

The syngo.CT Neuro Perfusion software package is designed to evaluate areas of brain perfusion. The software processes images or volumes that were reconstructed from continuously acquired CT data after the injection of contrast media.

It generates the following result volumes:

  • . Cerebral blood flow (CBF)
  • . Cerebral blood volume (CBV)
  • Local bolus timing (time to start (TTS), time to peak (TTP), time to drain (TTD)) .
  • Mean transit time (MTT) ●
  • Transit time to the center of the IRF (TMax) ●
  • Flow extraction product (permeability) .
  • Temporal MIP ●
  • Temporal Average ●
  • . Baseline Volume
  • Modified dynamic input data ●

The software also allows the calculation of mirrored regions or volumes of interest and the visual inspection of time attenuation curves. One clinical application is to visualize the apparent blood perfusion and the parameter mismatch in brain tissue affected by acute stroke.

Areas of decreased perfusion appear as areas of changed signal intensity:

  • · Lower signal intensity for CBF and CBV
  • Higher signal intensity for TTP. TTD. MTT. and TMax .

A second application is to visualize blood brain barrier disturbances by modeling extravascular leakage of blood into the interstitial space. This additional capability may improve the differential diagnosis of brain tumors and be helpful in therapy monitoring.

Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Janine M. Morris -S
2013.04.02
09:49:55 -04'00'

(Division Sign Off)
Division of Radiological Health
Office of In Vitro Diagnostics and Radiological Health

510(k)K123541
-----------------

Page 1 of 1

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.