K Number
K122237

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-01-03

(161 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Biophotas LifeLight device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

The blue spectrum light is intended to reduce the appearance of mild to moderate acne vulgaris.

Device Description

The LifeLight® is a lightweight device which uses specific wavelengths of light, produced by light emitting diodes (LEDs), to manage aesthetic conditions. LifeLight produces light in the near infrared region of the spectrum (880 nm) intended to provide topical heating to tissue. Blue light (464 nm) is intended to help reduce the appearance of mild to moderate acne. The console of the device contains the electronics of the device and an automatic shut-off safety feature. The LifeLight operates for 30 minutes per use.

AI/ML Overview

The provided text describes a 510(k) submission for the Biophotas LifeLight® device. It outlines the device's intended use and compares it to a predicate device, but it does not include a comprehensive study that establishes specific acceptance criteria or provides detailed performance results against those criteria.

However, it does mention some nonclinical testing related to the infrared LED array's ability to elevate tissue temperature.

Here's an attempt to answer your request based on the available information, with significant limitations due to the lack of a dedicated study section:


1. Table of Acceptance Criteria and Reported Device Performance

Based on the provided text, specific formal "acceptance criteria" and a detailed "reported device performance" table are not explicitly stated. The document primarily focuses on demonstrating substantial equivalence to a predicate device.

However, we can infer a non-clinical "performance" from the testing described for the infrared function:

Acceptance Criteria (Inferred)Reported Device Performance (from Nonclinical Test)
Ability to elevate topical tissue temperature for temporary reliefElevated skin to a temperature of between 40 – 45°C over the course of ten minutes of treatment.

It's crucial to note that this is an inference based on the description of a nonclinical test, not a formally defined acceptance criterion with a quantitative target. The blue light for acne indication does not have any performance data presented in this document.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a sample size for any test set for either the infrared or blue light indications. It broadly refers to "Device testing" for the infrared array.

The data provenance for this unquantified testing is implied to be from the manufacturer's internal "nonclinical" testing. There is no information regarding country of origin or whether it was retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth and Qualifications

The document does not mention the use of experts to establish ground truth for any test set. The nonclinical test for the infrared function appears to be a direct physical measurement (temperature elevation), not requiring expert adjudication of outcomes.

4. Adjudication Method

Since no experts are mentioned and the performance data is a direct physical measurement (temperature), there was no adjudication method described or likely required for the presented (limited) testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done or described. This document focuses on demonstrating substantial equivalence of a physical therapy device, not a diagnostic algorithm requiring human reader performance analysis.

6. Standalone (Algorithm Only) Performance Study

The Biophotas LifeLight® is a physical therapy device utilizing LEDs for therapeutic heating and acne treatment. It does not involve a standalone algorithm in the way an AI diagnostic device would. Therefore, a standalone (algorithm only) performance study (as understood in the AI context) was not performed or relevant to this type of device.

7. Type of Ground Truth Used

For the infrared indication, the "ground truth" for the nonclinical testing appears to be direct physical measurement (temperature elevation). For the blue light indication, no specific ground truth or performance data is presented in this document.

8. Sample Size for the Training Set

The device is not an AI/ML algorithm. Therefore, there is no training set in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for an AI/ML algorithm, this question is not applicable.


Summary of Limitations:

The provided document is a 510(k) summary for a physical therapy device. It aims to demonstrate substantial equivalence to a predicate, not to report a comprehensive clinical or non-clinical study with detailed acceptance criteria and performance metrics typically found for AI/ML devices. The "performance" data is limited to a single non-clinical measurement for the infrared function, and no such data is present for the blue light acne indication.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "biophotas" with a stylized graphic to the left of the word. The graphic is a series of dots arranged in a radiating pattern. The word "bio" is solid black, while the word "photas" is outlined in black.

JAN 0 3 2013

1122287

Traditional 510(k) Summary As required in 21CFR 807.92, we hereby submit this 510(k) Summary:

510(k) owners name, address, phone, fax, contact person & prep date

The 510(k) owner is Biophotas, Inc., 250 El Camino Real, #114, Tustin, CA 92780

Phone: (714) 838 - 1956

Fax :

The contact person is Mr. Shepard G. Bentley, RAC

510(k) preparation date: 20 July, 2012

Name of the device, trade name, proprietary name, and classification name

Trade name: Biophotas LifeLight®

Common name: Infrared Lamp

Classification name: Lamp Infrared, Therapeutic Heating (ILY)

Predicate Device:

The legally marketed predicate for the Biophotas LifeLight® is:

K060792, illumiMed LED Phototherapy device

Device Description:

The LifeLight® is a lightweight device which uses specific wavelengths of light, produced by light emitting diodes (LEDs), to manage aesthetic conditions. LifeLight produces light in the near infrared region of the spectrum (880 nm) intended to provide topical heating to tissue. Blue light (464 nm) is intended to help reduce the appearance of mild to moderate acne.

Indication for Use/Intended Use:

The Biophotas LifeLight™ device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. The blue spectrum is intended to reduce mild to moderate inflammatory acne vulgaris.

The Intended Use information from the predicate product is equivalent and is reproduced below:

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "biophotas" in a stylized font. To the left of the word is a design made up of many small circles. The circles are arranged in a pattern that resembles a starburst or a network of connections. The word "biophotas" is written in lowercase letters, and the letters are a dark gray color.

K060792 illumiMed LED Phototherapy device

The illumiMed™ device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. The blue spectrum is intended to reduce mild to moderate acne vulgaris. Use of the combination of blue, red and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions, specifically indicated to treat mild to moderate inflammatory acne vulgaris.

Technological characteristics:

The Biophotas LifeLight® is a lightweight, portable light-emitting diode (LED) device which emits light energy. A blue LED array in the 464 nm spectrum is used in treatment for mild to moderate inflammatory acne. The near infrared LED array provides mild topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle pain. The console of the device contains the electronics of the device and an automatic shut-off safety feature. The LifeLight operates for 30 minutes per use. The LifeLight shares substantially equivalent wavelengths to its predicate devices for the related indications, and substantially equivalent energy source, energy delivery, target populations and locations for use.

echnicaracteristic
Wavelength464nm, 640nm, 880nm430nm, 660nm, 940nm
Electricalpower110 - 120 volts110 - 120 volts
TreatmentTwice weekly, 30 minutes per treatmentEvery three days, 9 minutes per treatment
EnergyDelivery(standard dose)465 nm - 19.6 Joules/cm2430 nm - 18.9 Joules/cm2+
Energy Output465 nm - 10.9 mW/cm2430 nm - 35 mW/cm2
Indicated UseAcne, PainAcne, Pain
UsePrescriptionPrescription
  • = information taken from company user's manual †= information from predicate performance calculation

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "biophotas" with a stylized graphic to the left. The graphic is made up of many small circles arranged in a radial pattern. The word "biophotas" is written in a sans-serif font, with each letter outlined in black.

Nonclinical and clinical test data:

The Biophotas LifeLight® and the predicate devices have satisfied product safety testing to the IEC 60601-1 standard, and the electromagnetic safety testing to the IEC 60601-1-2 standard. Device testing to demonstrate the ability of the infrared LED array to elevate topical tissue temperature confirmed that the LifeLight elevated skin to a temperature of between 40 – 45 C measured over the course of ten minutes of treatment, during which mid-treatment readings the device was deactivated.

In summary, the BioPhotas LifeLight product is substantially equivalent to the predicate, utilizes mature technology and materials and provides a safe and effective addition to the substantial number of products cleared for these indications.

End of 510{k) Summary Section

Substantial Equivalence

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an eagle with its wings spread, symbolizing the federal government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Biophotas, Incorporated % Bentley Biomedical Consulting, LLC Mr. Shepard G. Bentley, RAC Director, Regulatory Affairs 250 El Camino Real #110 Tustin, California 92780

Re: K12237

Trade/Device Name: Biophotas LifeLight® Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: ILY, GEX Dated: December 14, 2012 Received: December 17, 2012

Dear Mr. Bentley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

January 3, 2013

{4}------------------------------------------------

Page 2 - Mr. Shepard G. Bentley, RAC

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Peter D. Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K122237

Device Name: Biophotas Lifelight

Indications For Use: The Biophotas LifeLight device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

The blue spectrum light is intended to reduce the appearance of mild to moderate acne vulgaris.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Neil R Ogden 2013.01.02 16:11:27 -05'00'

(Division Sign-Off) for MXM
Division of Surgical Devices
510(k) Number K122237

Page 1 of __ 1

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.