Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200104
    Device Name
    Oxylight
    Date Cleared
    2020-06-04

    (139 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Oxylight is intended for dermatological use by physicians and healthcare professionals for the following:

    LED Technology is intended for:

    -Blue LED = 465nm - to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

    -Red LED 625nm- for treatment of superficial, benign vascular and pigmented lesions.

    • Yellow LED 590nm - treatment of periorbital wrinkles and rhytides.

    Microdermabrasion is intended for exfoliation of the skin.

    Oxygen spray is intended to refresh the skin.

    Device Description

    The Oxylight system is a skin therapy system that uses three major modalities delivered by different attachments powered from a single energy source and operated from a universal control unit. The treatments include, LED Light Panel Therapy (Class 2), Microdermabrasion (Class 1, 510k exempt) and Oxygen Spray. The Oxygen Spray is delivered through the same panel used for LED light therapy and separate handpieces.

    AI/ML Overview

    The provided document is a 510(k) Pre-market Notification, which focuses on demonstrating substantial equivalence to existing legally marketed devices, rather than providing extensive clinical study data to prove novel efficacy or meet specific acceptance criteria for a new clinical claim. Therefore, much of the requested information regarding acceptance criteria, specific device performance against those criteria, and detailed clinical study methodologies is not available in this document.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not present a table of specific clinical acceptance criteria or human-centered performance metrics. Instead, it compares the technical characteristics of the Oxylight device (the "subject device") with predicate devices to argue for substantial equivalence in safety and efficacy.

    The relevant "performance data" section focuses on safety testing against recognized electrical and electromagnetic compatibility standards.

    StandardTest TitleTest ArticleAcceptance Criteria (Implicit)Reported Performance
    IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012Medical electrical equipment Part 1: General requirements for basic safety and essential performanceOxyLightCompliance with standard IEC 60601-1Pass
    EN 60601-1-2:2015, EN 61000-3-2:2014, EN 61000-2-3:2013Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and testsSapphire 3 OxyLightCompliance with standard EN 60601-1-2 et al.Pass
    ISO 14971:2012Medical devices – Application of risk management to medical devicesNot directly tested (risk management applied)Application of risk management principlesApplied

    2. Sample Size Used for the Test Set and the Data Provenance

    The document explicitly states: "Clinical testing was not performed with this device."

    Therefore, there is no test set in the sense of a patient cohort, nor data provenance from such a set. The "test article" for safety testing was the device itself (OxyLight / Sapphire 3 OxyLight).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Since no clinical testing was performed, no ground truth was established by experts for a test set in a clinical context.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done, as indicated by the statement "Clinical testing was not performed with this device."
    The document does not mention any human reader performance or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The Oxylight is a physical device (LED panel, microdermabrasion, oxygen spray) for dermatological use, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    For the purpose of substantial equivalence, the "ground truth" implicitly relies on the established safety and efficacy of the predicate devices with similar technical characteristics and indications for use. For the safety tests conducted, the ground truth was compliance with the specified international safety and EMC standards.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/algorithm-based device requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1