K Number
K121856

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-07-11

(15 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters Line Extension are recommended for use in Percutaneous Transluminal Angioplasty of the renal, tibial, popiteal, femoral and peroneal arteries. These catheters are not for use in coronary arteries.

Device Description

The subject device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters Line Extension, are small vessel balloon catheters consisting of an over the wire catheter with a balloon fixed at the distal tip. The semi-compliant, low profile balloon has two radiopaque marker bands to delineate the working length of the balloon and aid in balloon placement. The coaxial, over the wire catheter is compatible with 0.014" and 0.018" guidewires, as labeled.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, specifically a line extension of the Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than conducting extensive clinical studies to determine specific performance metrics against acceptance criteria in the same way a diagnostic AI device would.

Therefore, many of the requested categories for AI/diagnostic device studies (like sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this SDO (Software as a Medical Device) type of submission.

However, I can extract the relevant information regarding acceptance criteria and the studies performed to demonstrate equivalence.


Acceptance Criteria and Device Performance Study for Ultraverse® 014 and 018 PTA Balloon Dilatation Catheter Line Extension

This submission is a 510(k) for a "line extension" of an existing device (Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters). The core of a 510(k) is to demonstrate substantial equivalence to a previously cleared predicate device, rather than proving performance against novel clinical acceptance criteria. The modifications were to "add balloon and catheter lengths."

The "acceptance criteria" in this context refer to engineering and performance specifications evaluated through in vitro tests to ensure the new lengths perform comparably to the predicate and meet established safety and functional requirements.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test PerformedReported Device Performance Statement
Mechanical/PhysicalDimensional TestingComparable to the predicate device
Distensibility TestingComparable to the predicate device
Fatigue TestingComparable to the predicate device
Balloon Burst StrengthComparable to the predicate device
Balloon Burst ModeComparable to the predicate device
Balloon to Shaft Tensile (Strength)Comparable to the predicate device
ReinsertionComparable to the predicate device
Catheter Shaft LeaksComparable to the predicate device
Sheath CompatibilityComparable to the predicate device
TrackabilityComparable to the predicate device
FunctionalInflation TimeComparable to the predicate device
Deflation TimeComparable to the predicate device

Summary of Performance: The document states: "The results from these tests demonstrate that the technological characteristics and performance criteria of the subject Ultraverse® 014 & 018 PTA Balloon Dilatation Catheter Line Extension is comparable to the predicate device and that it can perform in a manner equivalent to devices currently on the market for the same intended use."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified in the provided text for each individual test. These are typically in vitro bench tests, and sample sizes are determined by engineering and statistical requirements for device testing rather than patient case numbers.
  • Data Provenance: Not applicable in the context of geographical origin or retrospective/prospective patient data. The data is from in vitro laboratory testing conducted by the manufacturer, Bard Peripheral Vascular, Inc.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not Applicable. This is an in vitro device performance study, not a clinical diagnostic performance study requiring expert ground truth for patient data.

4. Adjudication Method for the Test Set

  • Not Applicable. As this is an in vitro device performance study, there is no "adjudication method" in the sense of reconciling disagreements among human readers or experts on patient data. Test results are typically objective measurements against engineering specifications.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and effect size

  • Not Applicable. This type of study is for evaluating diagnostic performance of an AI system, not for a physical medical device's line extension.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is not an AI/algorithm-based device. "Standalone performance" here would refer to the device's functional and mechanical performance, which was evaluated through the in vitro tests listed.

7. The type of ground truth used

  • Engineering Specifications / Predicate Device Performance: The "ground truth" or reference for comparison in this 510(k) submission is the performance of the legally marketed predicate device and established engineering specifications and standards for balloon dilatation catheters. The tests are designed to confirm the new device lines perform within acceptable ranges relative to these benchmarks.

8. The sample size for the training set

  • Not Applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

  • Not Applicable. No training set exists for this type of device.

{0}------------------------------------------------

K12 1856

Ultraverse® 014 and 018 PTA Balloon Dilatation Catheter Line Extension

510(k) Summary 21 CFR 807.92

JUL 1 1 2012

Page 11 of

As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a summary of the information upon which substantial equivalence determination is based as follows:

1. Submitter Information:

Applicant: Bard Peripheral Vascular, Inc.

1625 West Third Street

Tempe, Arizona 85281

Phone: 480-379-2841

Fax: 480-449-2546

Contact: Erin Fox, Regulatory Affairs Specialist II

Date: June 25, 2012

2. Subject Device:

Device Trade Name:Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters
Common or Usual Name:Catheter, Angioplasty, Peripheral, Transluminal/
Catheter, Percutaneous
Classification:Class II
Classification Panel:Cardiovascular

3. Predicate Device:

Device Trade Name:Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters
Cleared 510(K) Number:K093965
Date of Cleared 510(K):03/17/10

4. Summary of Change:

The modifications from the predicate device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters, to the subject device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheter Line Extension, were to add balloon and catheter lengths.

{1}------------------------------------------------

  1. Device Description:

The subject device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters Line Extension, are small vessel balloon catheters consisting of an over the wire catheter with a balloon fixed at the distal tip. The semi-compliant, low profile balloon has two radiopaque marker bands to delineate the working length of the balloon and aid in balloon placement. The coaxial, over the wire catheter is compatible with 0.014" and 0.018" guidewires, as labeled.

6. Indications for Use of Device:

The Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters, are recommended for use in Percutaneous Transluminal Angioplasty of the renal, tibial, popliteal, femoral and peroneal arteries. These catheters are not for use in coronary arteries.

7. Technological Comparison to Predicate Device:

The technological characteristics of the subject device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheter Line Extension, are substantially equivalent to those of the predicate device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters, in terms of intended use, indications for use, materials, fundamental scientific technology, target population, operating principle, packaging, sterility assurance level and method of sterilization.

8. Performance Data:

To demonstrate the substantial equivalence of the subject device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheter Line Extension, to the predicate device, the technological characteristics and performance criterion were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following additional in vitro tests were performed:

  • Dimensional Testing .
  • Distensibility Testing �
  • . Fatigue Testing
  • Inflation Time �
  • Deflation Time .

{2}------------------------------------------------

  • Balloon Burst Strength .
  • Balloon Burst Mode .
  • . Balloon to Shaft Tensile
  • Reinsertion .
  • . Catheter Shaft Leaks
  • Sheath Compatibility .
  • . Trackability

The results from these tests demonstrate that the technological characteristics and performance criteria of the subject Ultraverse® 014 & 018 PTA Balloon Dilatation Catheter Line Extension is comparable to the predicate device and that it can perform in a manner equivalent to devices currently on the market for the same intended use.

Kla 1856 ... Page 13 of 3

9. Conclusions:

The Ultraverse® 014 & 018 PTA Balloon Dilatation Catheter Line Extension met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test.protocols and/or customer inputs. The Ultraverse 014 & 018 PTA Balloon Dilatation Catheter Line Exentsion are substantially equivalent to the legally marketed predicate device, the Ultraverse® 014 & 018 PTA Balloon Dilatation Catheters.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Bard Peripheral Vascular c/o Ms. Erin Fox 1625 West 3rd St. Tempe, AZ 85281-1740

JUL 1 1 2012

Re: K121856

Trade/Device Name: Ultraverse PTA catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II (two) Product Code: LIT Dated: June 25, 2012 Received: June 26, 2012

Dear Mr. Fox:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 -- Ms. Eric Fox

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

M.S. Gillham

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Indications for Use:

K121856

Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters Line Extension

The Ultraverse® 014 and 018 PTA Balloon Dilatation Catheters Line Extension are recommended for use in Percutaneous Transluminal Angioplasty of the renal, tibial, popiteal, femoral and peroneal arteries. These catheters are not for use in coronary arteries.

Prescription Use X CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (Part21 (21CFR 801 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

CMAZAHLLLL

ardiovascular Devices 510(k) Number

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).