K Number
K121792
Device Name
NEUVIZ 64 MULTI-SLICE CT SCANNER SYSTEM
Date Cleared
2012-11-16

(141 days)

Product Code
Regulation Number
892.1750
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The NeuViz 64 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
Device Description
The NeuViz 64 Multi-slice CT Scanner System is composed of a gantry, a patient couch, an operator console and includes image acquisition hardware and software, and associated accessories. It is designed to be a head and whole body X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is an interactive program used for X-ray scan control, image reconstruction, and image archive/evaluation.
More Information

No
The summary describes standard CT image reconstruction and processing, and there is no mention of AI, ML, or related concepts in the device description, intended use, or performance studies.

No
The device is described as an imaging system used to acquire anatomical images for diagnostic purposes, not for treating conditions.

Yes

Explanation: The device is a CT scanner that produces cross-sectional images of the body. The "Summary of Performance Studies" states that radiologists assessed "the diagnostic quality of the images," indicating that the images are used for diagnostic purposes.

No

The device description explicitly states it is composed of hardware components like a gantry, patient couch, and operator console, in addition to software.

Based on the provided information, the NeuViz 64 Multi-Slice CT Scanner System is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to provide information about a person's health.
  • Device Function: The description clearly states that the NeuViz 64 is a Computed Tomography X-ray system. It acquires X-ray transmission data from the whole body and reconstructs it into cross-sectional images. This is an in vivo imaging technique, meaning it examines the body directly, not samples taken from the body.
  • Lack of Mention of Samples: There is no mention of the device analyzing biological samples or performing tests on them.

Therefore, the NeuViz 64 Multi-Slice CT Scanner System falls under the category of medical imaging devices, not In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

The NeuViz 64 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Product codes

JAK

Device Description

The NeuViz 64 Multi-slice CT Scanner System is composed of a gantry, a patient couch, an operator console and includes image acquisition hardware and software, and associated accessories. It is designed to be a head and whole body X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is an interactive program used for X-ray scan control, image reconstruction, and image archive/evaluation.

The NeuViz 64 Multi-slice CT Scanner System is of comparable type and substantially equivalent to the SIEMENS Sensation 64 CT system (K040665) and NeuViz 16 Multi-slice CT Scanner System(K092742).

The NeuViz 64 system consists of two variants:

  • NeuViz 64i : Low configuration;

  • NeuViz 64e: High configuration.

The differences are the capacity of Tube and High Voltage Generator, and relevant specification/performance.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Computed Tomography X-ray

Anatomical Site

whole body, head

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Clinical testing: Sample clinical images were obtained from 260 subjects at 2 sites. Radiologist assessed the diagnostic quality of the images, and the conclusion is that the images are of diagnostic quality and can meet the clinical requirements.

Key Metrics

Not Found

Predicate Device(s)

K040665, K092742

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

Neusoft

510(k)

K121792

510(K) Summary

Page 1 of 3

f

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92

General Information:

| Trade Name:
Common Name: | NeuViz 64 Multi-slice CT Scanner System
CT Scanner |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | 21 CFR Part 892.1750
Computed Tomography X-ray System |
| Classification: | Class II |
| Performance Standard: | 21 CFR Subchapter J, Federal Diagnostic X-ray Equipment
Standard |
| Manufacture: | Neusoft Medical Systems Co., Ltd.
No. 16 Shiji Road, Hunnan Industrial Area, Shenyang,
Liaoning, China, 110179 |
| Distributor: | Neusoft Medical Systems Co., Ltd.
No. 16 Shiji Road, Hunnan Industrial Area, Shenyang,
Liaoning, China, 110179 |
| Submitter: | Contact : Tian Yuehui
Title : Manager of Q&R Department
Tel : 86-24-83660646
Fax : 86-24-83660563
E-Mail : tianyh@neusoft.com |

Summary prepared :Sept.20, 2012

1

Page 2 of 3

Safety and Effectiveness information

Intended Uses:

The NeuViz 64 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Device Description:

The NeuViz 64 Multi-slice CT Scanner System is composed of a gantry, a patient couch, an operator console and includes image acquisition hardware and software, and associated accessories. It is designed to be a head and whole body X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is an interactive program used for X-ray scan control, image reconstruction, and image archive/evaluation.

The NeuViz 64 Multi-slice CT Scanner System is of comparable type and substantially equivalent to the SIEMENS Sensation 64 CT system (K040665) and NeuViz 16 Multi-slice CT Scanner System(K092742).

The NeuViz 64 system consists of two variants:

  • Neu/iz 64i : Low configuration;

  • NeuViz 64e: High configuration.

The differences are the capacity of Tube and High Voltage Generator, and relevant specification/performance.

Predicated Devices:

SIEMENS Sensation 64 CT system (K040665) NeuViz 16 Multi-slice CT Scanner System(K092742)

Statement of Substantial Equivalence:

In the opinion of Neusoft Medical Systems Co., Ltd., the NeuViz 64 system is of comparable type and substantially equivalent to the SIEMENS Sensation 64 CT system (K040665) that complies with the same or equivalent standards and has the same intended uses. All of these systems use on-board high frequency High-Voltage generator to generate X-radiation from X-ray tube. The X-ray transmission data is detected by the Solid-State detector and is reconstructed by the computer which has an interactive user interface. Both of these devices two dimensional image and 3D image that can be filmed.or electronically stored for future review.

The safety and effectiveness of the NeuViz 64 system was assured by adherence to Good Manufacturing Practices(GMP) 21CFR 820 and to International Standards ISO 13485:2003. The following quality assurance measures were applied to the development of the system:

2

Page 3 of 3

Risk analysis is performed according to ISO 14971 standard and internal risk management procedure.

Software safety is assured by the company procedures that conform to accepted practices. Quality assurance procedures are adhered to, and meeting the specifications and functional requirements if demonstrated via testing.

Requirements are tracked and design reviews are implemented.

Verification and validation activities(including performance testing, safety testing and simulated use testing) are performed.

Electrical and Mechanical safety is assured by adherence to IEC 60601-1 Standards

Radiation safety is assured by compliance with 21 CFR, Subchapter J Performance standards

Dose check is assured by compliance with NEMA XR25 Dose check standard

Clinical testing: Sample clinical images were obtained from 260 subjects at 2 sites. Radiologist assessed the diagnostic quality of the images, and the conclusion is that the images are of diagnostic quality and can meet the clinical requirements.

Based on the above considerations, it is Neusoft's opinion that NeuViz 64 system is substantially equivalent in safety and effectiveness to the predicate device: SIEMENS Sensation 64 CT system (K040665).

3

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. The eagle is positioned on the right side of the logo, and the text is on the left side.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD . 20993-0002

Neusoft Medical Systems Co., LTD. % Mr. Charles Mack Correspondent IRC USA 77325 Joyce Way Echo, OREGON 97826

November 16, 2012

Re: K121792

Trade/Device Name: NeuViz 64 Multi-Slice CT Scanner System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: II Product Code: JAK Dated: October 2, 2012 Received: October 2, 2012

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

4

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Michael D. O'Hara

Janine M. Morris Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

5

Indications for Use

510(k) Number (if known): K121792

Device Name: NeuViz 64 Multi-Slice CT Scanner System

Indications for Use:

The NeuViz 64 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray r reaturing a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Yes Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Michael, D. O'Hara

(Division Sign Off) Division of Radiological Health Office of In Vitro, Diagnostic and Radiological Health

510(k) /6121792