K Number
K121369
Date Cleared
2012-12-19

(226 days)

Product Code
Regulation Number
882.5890
Panel
NE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

As a TENS device:
Symptomatic relief of chronic (long term) intractable pain
Symptomatic relief of post-traumatic acute pain and post-surgical acute pain

As an Interferential and Premodulated device:
Symptomatic relief of chronic intractable pain, acute post traumatic pain or acute post surgical pain

As a Russian device:
Temporary relaxation of muscle spasms
Prevention or retardation of disuse atrophy in post-injury type conditions
Increase local blood circulation
Muscle re-education
Maintaining or increasing range of motion

As an Burst modulated alternaing current (Aussie) device:
Temporary relaxation of muscle spasms
Prevention or retardation of disuse atrophy in post-injury type conditions
Increase local blood circulation
Muscle re-education
Maintaining or increasing range of motion

As a Microcurrent device:
Symptomatic relief of chronic intractable pain
Symptomatic relief of post-traumatic acute pain and post-surgical acute pain

Device Description

Neurodyn and Neurodyn Aussie Neuromuscular Stimulators are intended for the treatment of, relief of chronic (long term) intractable pain as adjunctive treatment of post-surgical and post-traumatic acute pain. Both devices have the same intended uses and incorporate the same technologies as the predicate the Vectra Genisys K031077. The Neurodyn Muscle Stimulator is a programmable device. It comes equipped with 5 preset clinical programs along with 10 user protocols. The user programs are adjustable and can be changed according to the patient's needs, doctor's recommendations and prescription settings.

The Neurodyn Aussie Muscle Stimulator has four output channels with independent intensity controls. Thus, four different areas can be stimulated separately or together during a therapy session. It is adjustable and can be changed according to the patient's needs, doctor's recommendations and prescription settings. It generates the medium frequency alternate current (MFAC), burst modulated alternating current (Aussie)- type of sinusoidal current with a frequency carrying 1,000 Hz or 4,000 Hz and a burst duration of 4 ms or 2 ms, modulated in pulse trains (bursts) with a variable frequency from 1 to 120Hz.

AI/ML Overview

The provided text details a 510(k) summary for the Neurodyn and Neurodyn Aussie Powered Muscle Stimulators, seeking substantial equivalence to the Chattanooga Vectra Genisys K031077. The document outlines essential performance metrics and demonstrates equivalency through non-clinical testing and comparison of device characteristics.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" in the typical sense of quantitative performance thresholds. Instead, it demonstrates substantial equivalence to a predicate device (Chattanooga Vectra Genisys K031077) by showing that the Neurodyn and Neurodyn Aussie devices share identical or very similar technological characteristics, intended uses, indications for use, safety, and effectiveness.

Therefore, the "acceptance criteria" are implicitly met by demonstrating this substantial equivalence through a detailed comparison, rather than by achieving specific numerical performance targets against a predefined standard.

FeatureAcceptance Criteria (Implied: Substantial Equivalence to Predicate)Reported Device Performance (Neurodyn / Neurodyn Aussie)Predicate Device Performance (Vectra Genisys K031077)
Intended UseIdentical to predicateIdentical
Indications for UseIdentical to predicateIdentical
Target PopulationIdentical to predicateIdentical
Human FactorsIdentical to predicateIdentical
ContraindicationsIdentical to predicateIdentical
Technological CharacteristicsIdentical (or very similar) to predicateIdentical (e.g., MFAC, materials, software)Identical
Device MaterialSimilar to predicateABS plastic panel, LCD displayABS plastic panel, LCD display
Width (in)Comparable to predicateNeurodyn Aussie: 10.6, Neurodyn: 14.69.75
Height (in)Comparable to predicateNeurodyn Aussie: 4.9, Neurodyn: 4.98.75
Depth (in)Comparable to predicateNeurodyn Aussie: 10.4, Neurodyn: 12.412.75
Weight (lbs)Comparable to predicateNeurodyn Aussie: 4.08, Neurodyn: 5.56
PerformanceIdentical to predicateIdenticalIdentical
BiocompatibilityUses FDA cleared electrodesFDA cleared electrodesFDA cleared electrodes
Mechanical SafetyIdentical to predicateIdenticalIdentical
Anatomical SitesIdentical to predicateIdenticalIdentical
Number of ChannelsComparable to predicateNeurodyn Aussie: 4, Neurodyn: 42/4
Current Types (Russian, Aussie, Interferential, Microcurrent, TENS, Premodulated)Similar array of current types to predicateVaries between Neurodyn and Neurodyn Aussie, but comparable to predicate's offerings.Vectra Genisys offers all listed. Specific Neurodyn models offer subsets.
Voltage InputIdentical to predicate100/240V, 50/60Hz Bivolt100/240V, 50/60Hz, 1.0A
OutputIdentical to predicate+24V 7.3A+24V 7.3A
Electrical ClassIdentical to predicateIIII
Electrical TypeIdentical to predicateBF TypeBF Type
Method of line current isolationComparable to predicateDouble IsolationFuse-Two 5.6A Time Lag
Patient leakage control (normal condition)Comparable to predicate (within acceptable limits)0.0508mA69μA
Patient leakage control (single fault condition)Comparable to predicate (within acceptable limits)0.0252mA31μA
Software MicroprocessorYesYesYes
Automatic overload tripYesYesYes
Automatic shut offYesYesYes
Temperature range (transport/storage)Comparable to predicate41°F-122°F59°F-104°F
Environment operating temperature rangeComparable to predicate41°F-113°F59°F-104°F
Locking featureYesKeyboard lock safety featureKeyboard lock safety feature
Treatment timer with auto shut offYesYesYes
Auto test and repeat treatmentYesYesYes
Safety standards requirementsISO 13485, IEC 60601-1, IEC 60601-2, IEC 60602-10, CESame standards (Vectra Genisys also UL 60602)ISO 13485, IEC 60601-1, IEC 60601-2, IEC 60602-10, UL 60602
Electromagnetic compatibilityIEC 60601-1-2001/A:2004Identical StandardIEC 60601-1-2001/A:2004

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document describes a non-clinical test design focused on demonstrating substantial equivalence to a predicate device. This typically involves performance verification against a known, cleared device, rather than a test on a patient sample. The data provenance is derived from engineering and design specifications and testing performed by Ibramed, a company based in Brazil (Amparo - Sao Paulo - Brasil). The document does not specify whether this was a retrospective or prospective comparison, but the data itself (device specifications) would inherently be current/prospective for the new devices and historical/retrospective for the predicate device's cleared specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This type of information is not applicable to this submission. The "ground truth" here is the established safety and effectiveness of the legally marketed predicate device (Chattanooga Vectra Genisys K031077). The assessment is based on comparison of technical specifications and performance against recognized standards, not on expert consensus regarding patient outcomes or diagnostic accuracy.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable. There is no "adjudication" in the sense of resolving discrepancies between expert readings or interpretations. The comparison is between documented technical specifications and results of non-clinical engineering tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic or AI-assisted interpretation devices, which is not the nature of a powered muscle stimulator. The submission focuses on demonstrating substantial equivalence of a therapeutic device based on its functional characteristics and safety.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. These are powered muscle stimulator devices, not algorithmic diagnostic tools. Their performance is inherent in their electrical output and physical characteristics, and they are operated by a human user (clinician or patient) as part of a therapeutic regimen.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" is the established safety and effectiveness of the legally marketed predicate device (Chattanooga Vectra Genisys K031077) and compliance with relevant international and national electrical safety and medical device standards (e.g., ISO 13485, IEC 60601 series). No clinical outcomes data or pathology samples were used as ground truth for this 510(k) submission.

8. The sample size for the training set

This is not applicable. As a medical device submission based on substantial equivalence to a predicate device and non-clinical testing, there is no "training set" in the context of machine learning or AI algorithms. The device design and manufacturing processes are likely informed by prior engineering knowledge and regulatory standards, but not a specific "training set" of data for an algorithm.

9. How the ground truth for the training set was established

This is not applicable, as there is no training set mentioned or implied in the context of this 510(k) submission.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).