K Number
K121139
Manufacturer
Date Cleared
2012-05-02

(16 days)

Product Code
Regulation Number
878.3300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AccuMesh Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during the laparoscopic repair of soft tissue defects (e.g. hernia repair).

Device Description

The AccuMesh Positioning System is a manual laparoscopic device intended to facilitate hernia mesh delivery and placement in laparoscopic ventral hernia repair. The AccuMesh Positioning System comprises two main sections: operation handle and deployment section which are connected via a tube. Included in the packaging configuration is a furling sleeve intended to assist the surgeons to fur! the mesh tightly and rapidly over the AccuMesh positioning system.

AI/ML Overview

The provided text describes a 510(k) submission for the Covidien AccuMesh Positioning System, specifically regarding the addition of a furling sleeve. It claims substantial equivalence to previously cleared devices (K101218 and K103269). However, the document does not contain the detailed performance data, acceptance criteria, or study specifics required to answer many of your questions.

Here's an analysis of what can and cannot be answered based on the provided text:

1. Table of acceptance criteria and the reported device performance:

  • Acceptance Criteria: Not explicitly stated in the document.
  • Reported Device Performance: Not detailed in the document. The text only says: "Performance testing was conducted using sterilized samples of the proposed design of the AccuMesh Positioning System with the furling sleeve. The furling sleeve interface and compatibility with laparoscopic ports and components of the AccuMesh Positioning System were evaluated with various commercially available mesh products." This is a high-level statement about the type of testing, not the results or specific metrics.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified. The document only mentions "sterilized samples" without a number.
  • Data Provenance: Not specified. It does not mention country of origin or whether it was retrospective or prospective. Given the nature of a 510(k) for a device like this, it would typically involve prospective lab/bench testing or potentially animal studies, but the document doesn't specify.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This device is a manual surgical instrument, not an AI or imaging diagnostic tool that typically relies on expert ground truth for performance evaluation in the way you're asking. Its performance evaluation likely focuses on mechanical and functional aspects, compatibility, and usability by surgeons.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI-assisted diagnostic or imaging device. It's a manual surgical instrument.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is a manual surgical instrument, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not Applicable/Not Specified. For a manual surgical tool, "ground truth" would relate to its ability to perform its stated function (e.g., correctly deliver and position mesh without failure, demonstrating compatibility). The document does not detail how this "truth" was established beyond the general statement about performance testing.

8. The sample size for the training set:

  • Not Applicable. This is a physical device, not an AI model requiring a training set.

9. How the ground truth for the training set was established:

  • Not Applicable. See point 8.

Summary of what is missing from the provided text to answer your questions:

The provided text is a 510(k) summary focused on the administrative aspects and the claim of substantial equivalence for a manual surgical instrument. It explicitly states that "Performance testing was conducted using sterilized samples..." but does not provide any specific data, metrics, acceptance criteria, or detailed methodology from those tests. Therefore, most of your questions, which are geared towards performance evaluation of AI/diagnostic tools, cannot be answered from this document.

{0}------------------------------------------------

K121139

page 1/2

Covidien

AccuiMesh Positioning System

4.0 510(K) SUMMARY

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(K) summary have been provided in conformance with 21 CFR §807.98.

Submitter Information

Name: Address:

Covidien 15 Crosby Drive Bedford, MA 01730

Establishment Registration: 9615742

Name of contact person:

James McMahon Senior Manager, Regulatory Affairs Covidien 15 Crosby drive Bedford, MA 01730 USA Phone: (781) 839 1787

April 12, 2012 Date prepared:

AccuMesh™ Positioning System Trade or proprietary name:

Mesh Deployer Common or usual name: Mesh, Surgical, Deployer Classification name: General and Plastic Surgery (79) Classification panel:

21 CFR 878.3300 Requiation:

ORQ Product Code:

Legally marketed devices to which equivalence is claimed:

Accumesh Positioning System (510k cleared as the PatchAssist device - K101218 Accumesh Positioning System PatchAssist Large (510k cleared as the PatchAssist device - K103269

Reason for 510(k) submission:

To obtain market clearance on the Accumesh Positioning System with furling sleeve.

Device description:

The AccuMesh Positioning System is a manual laparoscopic The nobalities. Intended to facilitate hernia mesh delivery and placement in laparoscopic ventral hernia repair. The

CONFIDENTIAL

8

{1}------------------------------------------------

K12.1139

page 4/2

Covidien

AccuMesh Positioning System

AccuMesh Positioning System comprises two main sections: operation handle and deployment section which are connected via a tube. Included in the packaging configuration is a furling sleeve intended to assist the surgeons to fur! the mesh tightly and rapidly over the AccuMesh positioning system.

Intended use of the device: Intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic repair of soft tissue defects (e.g. hernia repair).

Indications for use:

Indicated to facilitate the delivery of soft tissue prosthetics during laparoscopic repair of soft tissue defects (e.g. hernia repair).

Summary comparing the technological characteristics of the subject and predicate devices:

The proposed addition of the AccuMesh Furling Sleeve maintains the same fundamental design as the previously cleared AccuMesh (PatchAssist) device (K101218 and K103269). Both designs are intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic repair of soft tissue defects (e.g. hernia repair).

Performance data:

Performance testing was conducted using sterilized samples of the proposed design of the AccuMesh Positioning System with the furling sleeve. The furling sleeve interface and compatibility with laparoscopic ports and components of the AccuMesh Positioning System were evaluated with various commercially available mesh products.

CONFIDENTIAL

9

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Covidien % Mr. James McMahon Senior Manager, Regulatory Affairs 15 Crosby Drive Bedford, Massachusetts 01730

Re: K121139

Trade/Device Name: AccuMesh™ Positioning System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: ORQ Dated: April 13, 2012 Received: April 16, 2012

Dear Mr. McMahon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

MAY - 2 2012

{3}------------------------------------------------

Page 2 - Mr. James McMahon

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
for Peter
Mark N. Molkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K 12.1139

Device Name: AccuMesh™ Positioning System

Indications For Use: The AccuMesh Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during the laparoscopic repair of soft tissue defects (e.g. hernia repair).

AND/OR Prescription Use × (Part 21 CFR 801 Subpart D)

510(k) Number

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices

Page 1 of __

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.