(147 days)
A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.
The Powder-Free Nitrile Examination Gloves, Orange is substantial equivalent to Powder-Free Nitrile Examination Gloves, Black Color with 510k number K090194 that is currently marketed. The device modification is a change in color from Black to Orange. Another device modification is the addition of adding a citric acid coating to the interior surface of the gloves to achieve a pH of 5.0 - 5.5 for the inner surface of the glove.
The provided document is a 510(k) Summary for Powder-Free Nitrile Examination Gloves, Orange Color. It describes the device, its intended use, and a summary of technological characteristics compared to standards. The document does not describe a study in the context of an AI/ML medical device with the information requested (e.g., sample sizes for test/training, number of experts, adjudication methods, MRMC studies, or specific ground truth types like pathology or outcomes data).
Instead, this document is a submission to the FDA demonstrating substantial equivalence for a medical device (gloves) by confirming its compliance with established performance standards.
Therefore, many of the requested fields cannot be filled from the provided text as they are not applicable to this type of device submission.
Here's an attempt to answer based on the provided document, using "N/A" where information is not present or not relevant to the type of device:
1. Table of acceptance criteria and the reported device performance
| Characteristics | Standards (Acceptance Criteria) | Device Performance |
|---|---|---|
| Dimensions | ASTM D6319-10 | Meets |
| Physical Properties | ASTM D6319-10 | Meets |
| Freedom from pinholes | ASTM D6319-10 / FDA 21 CFR 800.20 | Meets, AQL 2.5 |
| Powder-Free | ASTM D 6124-06 | Meets < 2mg/glove |
| Biocompatibility | Primary Skin Irritation in Rabbits | Passes (Not a primary skin irritant) |
| Dermal Sensitization | Passes (Not a contact sensitizer) |
2. Sample size used for the test set and the data provenance
- Sample Size: Not explicitly stated. The tests are described as conforming to ASTM standards and FDA regulations, which imply specific sample sizes for various tests (e.g., AQL 2.5 for pinholes), but these specific numbers are not detailed in the summary.
- Data Provenance: Not specified, but generally, such tests are performed by the manufacturer or a contracted lab to confirm compliance with regulatory standards.
- Retrospective/Prospective: Not applicable in the context of this device type. These are physical/chemical performance tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of experts: N/A. The "ground truth" for these tests are the established ASTM standards and FDA regulations.
- Qualifications of those experts: N/A.
4. Adjudication method for the test set
- N/A. Compliance is determined by meeting pre-defined physical and chemical specifications, not by expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- N/A. This is not an AI/ML device, and no MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- N/A. This is not an AI/ML device.
7. The type of ground truth used
- The "ground truth" for the device's performance is compliance with established physical, chemical, and biological standards (ASTM D6319-10, ASTM D 6124-06, FDA 21 CFR 800.20, and biocompatibility tests like Primary Skin Irritation and Dermal Sensitization). It is based on pre-defined engineering and regulatory specifications.
8. The sample size for the training set
- N/A. No "training set" in the context of an AI/ML algorithm is applicable to this device.
9. How the ground truth for the training set was established
- N/A. No "training set" is applicable.
{0}------------------------------------------------
120890
Image /page/0/Picture/1 description: The image shows a logo for ShenWei USA Inc. The logo features a stylized black hand above the text "ShenWei usa inc." The text is in a bold, sans-serif font, with "ShenWei" being larger than "usa inc."
Shen Wei USA Inc. 33278 Central Ave. Suite 102 Union City, CA.94587
510(k) Summary
AUG 1 7 2012
1.0 Submitter:
Name: Shen Wei USA Inc. Address: 33278 Central Ave. Suite 102 Union City, CA. 94587
| Phone No: | 510-429-8692 |
|---|---|
| Fax No: | 510-487-5347 |
Manufacturer:
Name: ZHANGJIAGANG JIAMEI RUBBER PRODUCTS CO., LTD Address: FENGHUANG TOWN, ZHANGJIAGANG CITY, CHINA 215614 Phone No: 520-845-0023 Fax No: 220-842-03 I 1
Date 510k Summary was prepared: July 8th, 2012
2.0 Contact Person:
| Name: | Albert Li |
|---|---|
| Phone No: | 510-429-8692 |
| Fax No: | 510-487-5347 |
3.0 Name of the device:
| Trade Name: | Powder-Free Nitrile Examination Gloves, Orange Color |
|---|---|
| Device Name: | Powder-Free Nitrile Examination Gloves, Orange Color |
| Common Name: | Patient Examination Gloves |
| Classification Name: | Patient examination glove (21 CFR 880.6250, Product Code FMC) |
| Product Code: | Nitrile-80LZA |
4.0 Predicate Device Information:
The predicate device is K090194, Powder-Free Nitrile Examination Gloves, Black Color, 07/07/2009. Class I Powder Free Nitrile Examination gloves, 80LZA, that meets all requirements of ASTM D 6319-10 and FDA 21 CFR 800.20.
{1}------------------------------------------------
0
5.0 Description of the Device Modification:
The Powder-Free Nitrile Examination Gloves, Orange is substantial equivalent to Powder-Free Nitrile Examination Gloves, Black Color with 510k number K090194 that is currently marketed. The device modification is a change in color from Black to Orange. Another device modification is the addition of adding a citric acid coating to the interior surface of the gloves to achieve a pH of 5.0 - 5.5 for the inner surface of the glove.
6.0 Labeling and Intended Use of the Device:
Draft labels for Powder-Free Nitrile Examination Gloves, Orange can be found in Attachment 1. The only change to labeling is before it was labeled black glove and now the present labels being labeled orange glove and added "(@) pH 5.0-5.5".
The Powder-Free Nitrile Examination Gloves, Orange Color is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only. Indication for Use Statement can be found in Attachment 2.
7.0 Summary of the Technological Characteristics of the Device
The Powder-Free Nitrile Examination Gloves, Orange Color are summarized with the following technological characteristics compared to ASTM or equivalent standards.
| Characteristics | Standards | Device Performance |
|---|---|---|
| Dimensions | ASTM D6319-10 | Meets |
| Physical Properties | ASTM D6319-10 | Meets |
| Freedom from pinholes | ASTM D6319-10 | Meets, AQL 2.5 |
| FDA 21 CFR 800.20 | ||
| Powder-Free | ASTM D 6124-06 | Meets < 2mg/glove |
| Biocompatibility | Primary Skin Irritation in | Passes |
| Rabbits | (Not a primary skin irritant) | |
| Dermal Sensitization | Passes (Not a contact | |
| sensitizer) |
The test methods used are the same as those submitted in the original submission. A declaration of conformity with design controls is included in Attachment 3.
8.0 Conclusion: .
The Powder-Free Nitrile Examination Gloves, Orange Color will perform according the glove performance standards referenced in section 7.0 above and meet ASTM standards, and FDA requirements for water leak test on pinhole
{2}------------------------------------------------
:
:
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
:
3
AQL. Also, this device is substantial equivalent in safety and performance to currently marketed devices.
1
.
.
י
. . . . .
:
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be an abstract caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 1 7 2012
Shen Wei (USA), Incorporated Mr. Albert Li Director of Compliance 33278 Central Avenue, Suite 102 Union City, California 94587
Re: K120890
Trade/Device Name: Powder-Free Nitrile Examination Gloves, Orange Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: July 13, 2012 Received: July 17, 2012
Dear Mr. Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Mr. Li
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), } please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Avato
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Attachment Two
INDICATION FOR USE
Applicant: Shen Wei (USA) Inc.
Device Name: Powder-Free Nitrile Examination Gloves, Orange Color
Indication For Use:
A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Prescription Use Per 21CFR 801.109 OR
Over-The Counter X (Optional Format 1-2-96)
Alani Airti
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K120890
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.