(28 days)
The JETSTREAM System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.
The Jetstream Navitus L System is an atherectomy catheter system designed with an expandable cutting tip intended for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to an external collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectomy procedure.
The Jetstream Navitus L System consists of two primary components: a Catheter with Control Pod and a Console, which are packaged separately. Each of these system components is described generally as follows:
- Jetstream Navitus L Catheter with Control Pod: A sterile, single-use unit . consisting of an electrically-driven Catheter with attached Control Pod. As with the predicate device, the Jetstream Navitus L Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities, and the Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double pouched tray.
- PV Console: A reusable compact PV Console, with two (2) peristaltic pumps for . aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The PV Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.
This 510(k) is for modifications to the Jetstream Navitus System. The primary modifications of this 510(k) are to increase the cutting tip and expandable blade diameters (i.e., from 2.1 to 2.4 mm for the cutting tip, and from 3.0 to 3.4 mm for the expandable blades), and decrease the overall catheter length (from 135 to 120 cm).
The provided document does not describe a study in the context of an AI/ML medical device. Instead, it is a 510(k) summary for a medical device called the "JETSTREAM Navitus L System," which is a peripheral atherectomy catheter system. The "acceptance criteria" and "device performance" in this document refer to the results of in vitro engineering and performance tests conducted to demonstrate substantial equivalence to a predicate device, not the performance of an AI/ML algorithm.
Therefore, many of the requested categories (like sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set size, and how ground truth for training was established) are not applicable to this type of device submission.
Here's an attempt to extract the relevant information based on the prompt's request, acknowledging the fundamental difference in the nature of the device and its evaluation:
1. Table of acceptance criteria and the reported device performance:
The document lists various tests performed to demonstrate substantial equivalence. It doesn't explicitly state quantitative acceptance criteria for each test in a table format, but it implies that the results were deemed comparable to the predicate device. The "reported device performance" is the conclusion that the modified device's technological and performance characteristics are comparable to the predicate.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Dimensional verification meets design specifications and is comparable to predicate. | "Demonstrate that the technological and performance characteristics of the modified Jetstream Navitus System are comparable to the predicate Jetstream Navitus System." |
| Heat generation within acceptable limits and comparable to predicate. | (Same as above) |
| System reliability/life test results comparable to predicate. | (Same as above) |
| Aspiration efficiency & crossing time comparable to predicate. | (Same as above) |
| Speed drop flexibility comparable to predicate. | (Same as above) |
| Material liberation (Teflon & Polyimide) comparable to predicate. | (Same as above) |
| Rotational speed comparable to predicate. | (Same as above) |
| Accessory compatibility maintained. | (Same as above) |
| Infusion & aspiration flow rates comparable to predicate. | (Same as above) |
| Catheter pull force comparable to predicate. | (Same as above) |
| Catheter trackability and pushability comparable to predicate. | (Same as above) |
| Torque to failure comparable to predicate. | (Same as above) |
| Contrast injection performance comparable to predicate. | (Same as above) |
| Aorta strip test results comparable to predicate. | (Same as above) |
| Layered dissection model performance comparable to predicate. | (Same as above) |
| Overall safety and effectiveness of modifications supported. | "Support the safety and effectiveness of the modifications that are the subject of this 510(k)..." |
| Modified device performs equivalently to predicate with identical intended use. | "...and ensure the modified device can perform in a manner equivalent to the predicate Jetstream Navitus System with the identical intended use." |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document does not specify the sample sizes (e.g., number of catheters, number of tests) for each in vitro test. It mentions "in vitro testing."
- Data Provenance: The data provenance is "in vitro testing," meaning laboratory tests, not human subject data. Country of origin is not specified but implicitly from the manufacturer's testing. Retrospective or prospective is not applicable to in vitro engineering tests; these are designed experiments.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable as the evaluation involved in vitro device performance characteristics and engineering measurements, not clinical interpretation requiring expert ground truth establishment.
4. Adjudication method for the test set:
This is not applicable as there was no expert review or human interpretation of cases. The tests were objective engineering measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size:
This is not applicable as this is an atherectomy catheter system, not an imaging or diagnostic device that would typically undergo an MRMC study. There is no AI component.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is not applicable as there is no algorithm or AI component in this device. It is a mechanical catheter system with electronic controls.
7. The type of ground truth used:
The "ground truth" here refers to objective engineering and performance standards or measurements established by the manufacturer's test protocols and comparison to the predicate device's known characteristics. It's not expert consensus, pathology, or outcomes data in the sense of clinical studies.
8. The sample size for the training set:
This is not applicable as there is no AI/ML algorithm requiring a training set.
9. How the ground truth for the training set was established:
This is not applicable as there is no AI/ML algorithm requiring a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Pathway Medical Technologies. The logo consists of a circular graphic on the left, with a stylized mountain and path inside. To the right of the graphic is the word "PATHWAY" in large, sans-serif font, with the words "MEDICAL TECHNOLOGIES" in a smaller font underneath.
FEB 2 4 2012
510(k) SUMMARY
General Information:
January 26, 2012 Date of Summary Preparation: Pathway Medical Technologies, Inc. Name and Address of Manufacturer: 10801 120th Ave NE Kirkland, Washington 98033 Contact Person: Brit Baird Regulatory Affairs Manager Phone: 425-636-4137 Fax: 425-636-4001 JETSTREAM Navitus™ L System Device Trade Names: Peripheral Atherectomy Catheter Common Name: 21 CFR 870.4875 Regulation Number: Intraluminal Artery Stripper Regulation Name: Class II Regulatory Class: Cardiovascular Classification Panel: MCW Product Code:
Pathway Medical Technologies, Inc.
{1}------------------------------------------------
Performance Standards: Performance Standards do not currently exist for these devices. None are established under Section 514.
Device Description: The Jetstream Navitus L System is an atherectomy catheter system designed with an expandable cutting tip intended for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to an external collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectomy procedure.
The Jetstream Navitus L System consists of two primary components: a Catheter with Control Pod and a Console, which are packaged separately. Each of these system components is described generally as follows:
- Jetstream Navitus L Catheter with Control Pod: A sterile, single-use unit . consisting of an electrically-driven Catheter with attached Control Pod. As with the predicate device, the Jetstream Navitus L Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities, and the Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double pouched tray.
- PV Console: A reusable compact PV Console, with two (2) peristaltic pumps for . aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The PV Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.
This 510(k) is for modifications to the Jetstream Navitus System. The primary modifications of this 510(k) are to increase the cutting tip and expandable blade diameters (i.e., from 2.1 to 2.4 mm for the cutting tip, and from 3.0 to 3.4 mm for the expandable blades), and decrease the overall catheter length (from 135 to 120 cm).
Indications for Use: The JETSTREAM System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.
{2}------------------------------------------------
Substantially Equivalent Devices: Pathway Medical cites the Jetstream Navitus™ System (K110626) as the primary predicate device for the aforementioned modifications and substantial equivalence basis. However, the design rationale for and device testing of the modified device also includes references to the additional predicate Pathway Medical devices listed in the table below:
| Predicate Devices | Pathway MedicalPredicate 510(k) |
|---|---|
| Jetstream G3® SF 1.6 System | K111229 |
| Jetstream G3® SF System | K101334 |
| K101221 | |
| Jetstream G3® System | K093456 |
| K092332 | |
| Jetstream G3® L System | K100462 |
| K093918 | |
| Jetstream G2® NXT System | K091509 |
| Jetstream Pathway PVTM Atherectomy System | K082186 |
| Pathway PVTM Atherectomy System | K081328 |
Testing Summary: To demonstrate substantial equivalence of the modified Jetstream Navitus System to the predicate Jetstream Navitus System, the technological and performance characteristics were evaluated using in vitro testing for the primary (and supporting) modifications, as outlined below:
- Dimensional Verification .
- Heat Generation .
- . System Reliability/Life Test
- . Aspiration Efficiency & Crossing Time
- Speed Drop Flexibility .
- . Material Liberation (Teflon & Polyimide)
- . Rotational Speed
- Accessory Compatibility .
- . Infusion & Aspiration Flow Rates
- . Catheter Pull
- . Catheter Trackability and Pushability
- Torque to Failure
- . Contrast Injection
- Aorta Strip Test .
- Lavered Dissection Model .
The results from these tests:
- demonstrate that the technological and performance characteristics of the . modified Jetstream Navitus System are comparable to the predicate Jetstream Navitus System,
{3}------------------------------------------------
- support the safety and effectiveness of the modifications that are the subject of . this 510(k), and
- ensure the modified device can perform in a manner equivalent to the predicate . . Jetstream Navitus System with the identical intended use.
Conclusion (Statement of Equivalence): The data and information presented within this submission (including in vitro testing) and the similarities between the modified and predicate devices support a determination of substantial equivalence, and therefore market clearance of the modified Jetstream Navitus System through this 510(k) Premarket Notification.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB 2 4 2012
Pathway Medical Technologies, Inc. c/o Mr. Brit Baird Regulatory Affairs Manager 10801 120th Ave NE Kirkland, WA 98033
Re: K120242
Trade/Device Name: Jetstream Navitus L Systems Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II (two) Product Code: MCW Dated: January 26, 2012 Received: January 27, 2012
Dear Mr. Baird:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Brit Baird
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 Jou active possible . cov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Pour may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
M. A. Killebrew
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
INDICATIONS FOR USE
K12024Z 510(k) Number (if known):
Device Name:
JETSTREAM Navitus™ L System
Indications for Use: The JETSTREAM System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D)
. :
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---|---|
| -------------------------------------------------------- | -- |
. H. Wilhelm
(Division Sign-Off) (Bivision of Cardiovascular Devices
16120242 510(k) Number_
CONFIDENTIAL
57
§ 870.4875 Intraluminal artery stripper.
(a)
Identification. An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)(b)
Classification. Class II (performance standards).