K Number
K101221
Date Cleared
2010-06-02

(30 days)

Product Code
Regulation Number
870.4875
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Jetstream System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.

Device Description

The Jetstream G3 System is an atherectomy catheter system designed with an expandable cutting tip for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to a collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectorny procedure.

The Jetstream G3 System consists of two primary components: (1) a Catheter with Control Pod and (2) a Console, which are packaged separately. Each of these system components is described generally as follows:

  • . Jetstream G3 Catheter with Control Pod: A sterile, single-use unit consisting of an electrically driven Catheter and Control Pod. The Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities. The Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double-pouched tray.
  • PV Console: A reusable compact Console, with two (2) peristaltic pumps for . aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.

This 510(k) is for the same device most recently cleared under 510(k) K093456, but with the primary modifications intended to ensure any ingress of aspirated material into the guidewire lumen is minimized. In addition, the PATHWAY JETWIRE™ has been added as a compatible guidewire in the product labeling.

AI/ML Overview

The Pathway Medical Technologies Jetstream G3™ System (K101221) is a peripheral atherectomy catheter system. This 510(k) submission describes modifications primarily intended to minimize ingress of aspirated material into the guidewire lumen and the addition of the PATHWAY JETWIRE™ as a compatible guidewire.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly present a table of acceptance criteria with corresponding performance values. Instead, it lists the types of in vitro testing performed to demonstrate substantial equivalence to predicate devices. The "reported device performance" is summarized as the tests demonstrating that the "technological and performance characteristics of the modified Jetstream G3 System are comparable to the predicate devices."

Here's a summary of the types of tests mentioned as "acceptance criteria" through their performance evaluation:

Acceptance Criteria (Test Type)Reported Device Performance Summary
System Reliability/LifeDemonstrated comparable performance to predicate devices.
Speed Drop FlexibilityDemonstrated comparable performance to predicate devices.
Accessory CompatibilityDemonstrated comparable performance to predicate devices.
Infusion and Aspiration Flow RatesDemonstrated comparable performance to predicate devices.
Catheter Trackability and PushabilityDemonstrated comparable performance to predicate devices.
Aspiration Efficiency and Crossing TimeDemonstrated comparable performance to predicate devices, particularly focusing on minimizing ingress into the guidewire lumen with the modifications.
Thrombus Aspiration ChallengeDemonstrated comparable performance to predicate devices.
Guidewire and Device WearDemonstrated comparable performance to predicate devices.
Sterilant (EO) ResidualsDemonstrated comparable performance to predicate devices.
Control Pod Logic VerificationDemonstrated comparable performance to predicate devices.
Intended UseConfirmed device performs for its stated intended use: atherectomy of peripheral vasculature and breaking apart/removing thrombus from upper and lower extremity peripheral arteries, comparable to predicate devices.
PyrogenicityDemonstrated comparable performance to predicate devices.

The document states, "The results from these tests: demonstrate that the technological and performance characteristics of the modified Jetstream G3 System are comparable to the predicate devices, support the safety and effectiveness of the modifications... and ensure the modified device can perform in a manner equivalent to the listed predicate devices with the identical intended use."

2. Sample Size Used for the Test Set and Data Provenance

The document only states that "in vitro testing" was performed.

  • Sample Size: The specific sample sizes for each in vitro test are not provided in this summary.
  • Data Provenance: The tests were "in vitro," meaning conducted in a controlled laboratory environment, not on human or animal subjects. The country of origin for the data is implied to be the manufacturer's location or its testing facilities, which is Kirkland, Washington, USA. The data is not retrospective or prospective patient data, but rather performance testing data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the study involved in vitro performance testing of a medical device, not a diagnostic or AI-driven system requiring expert assessment for 'ground truth' in the clinical sense. The "ground truth" for these engineering and performance tests would be defined by established engineering and biological standards, measurement protocols, and comparison to the predicate devices.

4. Adjudication Method for the Test Set

This information is not applicable as the study involved in vitro performance testing. Adjudication methods like 2+1 or 3+1 typically apply to clinical studies where multiple human readers or experts assess cases to establish a definitive diagnosis or outcome.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study involves human readers (e.g., radiologists, pathologists) evaluating cases, often with and without AI assistance, to measure diagnostic performance. The Jetstream G3™ System is a medical device for atherectomy, not an AI software diagnostic tool, and therefore, an MRMC study with human readers and AI assistance is not relevant to its evaluation in this 510(k) submission.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

No, a standalone (algorithm only) performance study was not done in the context of AI without human-in-the-loop performance. The Jetstream G3™ System is a physical medical device, not a standalone AI algorithm. The performance evaluation focuses on the physical and functional characteristics of the device itself.

7. The Type of Ground Truth Used

For the in vitro performance tests, the "ground truth" was established by engineering specifications, standardized test methods, and quantitative measurements for device characteristics (e.g., flow rates, mechanical properties, wear, aspiration efficiency). The primary "ground truth" for demonstrating equivalence was the established performance of the predicate devices. The tests aimed to prove that the modified device met the same performance standards or demonstrated comparable characteristics as previously cleared versions.

8. The Sample Size for the Training Set

This information is not applicable. The Jetstream G3™ System is a physical medical device and does not involve AI or machine learning algorithms that require a "training set" in the computational sense. The device's design and manufacturing processes are based on engineering principles and validated through testing, not machine learning training.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable for the same reasons above. There is no AI training set for this device.

{0}------------------------------------------------

K101221

Image /page/0/Picture/1 description: The image shows the logo for Pathway Medical Technologies. The logo consists of a circular graphic on the left, with a shaded upper portion and a white lower portion containing a curved shape. To the right of the graphic is the word "PATHWAY" in large, sans-serif font, with the words "MEDICAL TECHNOLOGIES" in a smaller font beneath it.

JUN - 2 2010

510(k) SUMMARY

General Information:

Date of Summary Preparation:May 28, 2010
Name and Address of Manufacturer:Pathway Medical Technologies, Inc.10801 120th Ave NEKirkland, Washington 98033
Contact Person:Brian ClearyVice President of Clinical and Regulatory AffairsPhone: 425-636-4079Fax: 425-636-4001
Trade Name:Jetstream G3™ System
Common Name:Peripheral Atherectomy Catheter
Regulation Number:21 CFR 870.4875
Regulation Name:Intraluminal Artery Stripper
Regulatory Class:Class II
Classification Panel:Cardiovascular
Product Code:MCW

{1}------------------------------------------------

Performance Standards: Performance Standards do not currently exist for these devices. None are established under Section 514.

Device Description: The Jetstream G3 System is an atherectomy catheter system designed with an expandable cutting tip for use in debulking and treating vascular disease in the peripheral vasculature. Separate lumens within the Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the peripheral treatment site through a port in the Catheter tip to a collection bag located on the Console. The distal portion of the Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectorny procedure.

The Jetstream G3 System consists of two primary components: (1) a Catheter with Control Pod and (2) a Console, which are packaged separately. Each of these system components is described generally as follows:

  • . Jetstream G3 Catheter with Control Pod: A sterile, single-use unit consisting of an electrically driven Catheter and Control Pod. The Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities. The Control Pod provides a user interface with keypad controls. The unit, its electrical connectors, tubing, and aspirant collection bag are packaged in a double-pouched tray.
  • PV Console: A reusable compact Console, with two (2) peristaltic pumps for . aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.

This 510(k) is for the same device most recently cleared under 510(k) K093456, but with the primary modifications intended to ensure any ingress of aspirated material into the guidewire lumen is minimized. In addition, the PATHWAY JETWIRE™ has been added as a compatible guidewire in the product labeling.

Indications for Use: The Jetstream System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, iliac or renal vasculature.

{2}------------------------------------------------

Substantially Equivalent Devices: Pathway Medical cites the following devices as substantially equivalent predicate devices:

  • · Jetstream G3TM System (K100462)
  • · Jetstream G3TM System (K093456)
  • · Jetstream G3TM System (K092332)
  • · Jetstream G3TM L System (K093918)
  • · Jetstream G2TM System (K083837)
  • · Pathway PVTM Atherectomy System (K081328)

Testing Summary: To demonstrate substantial equivalence of the modified Jetstream G3 System to the predicate devices, the technological and performance chracteristics were evaluated using in vitro testing as outlined below:

  • System Reliability/Life .
  • Speed Drop Flexibility .
  • Accessory Compatibility .
  • Infusion and Aspiration Flow Rates .
  • Catheter Trackability and Pushability *
  • Aspiration Efficiency and Crossing Time .
  • Thrombus Aspiration Challenge .
  • Guidewire and Device Wear .
  • Sterilant (EO) Residuals �
  • Control Pod Logic Verification �
  • Intended Use .
  • Pyrogenicity .

The results from these tests:

  • demonstrate that the technological and performance characteristics of the . modified Jetstream G3 System are comparable to the predicate devices,
  • support the safety and effectiveness of the modifications that are the subject of . this 510(k), and
  • ensure the modified device can perform in a manner equivalent to the listed . predicate devices with the identical intended use.

Conclusion (Statement of Equivalence): The data and information presented within this submission (including in vitro testing) and the similarities between the modified and predicate devices support a determination of substantial equivalence, and therefore market clearance of the modified Jetstream G3 System through this 510(k) Premarket Notification.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized eagle or bird symbol, represented by three curved lines that suggest wings or feathers. The overall design is simple and recognizable, representing the department's role in public health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN - 2 2010

Pathway Medical Technologies, Inc. c/o Mr. Brian Cleary Vice President of Clinical and Regulatory Affairs 10801 12000 Ave NE Kirkland, WA 98033

Re: K101221

Trade/Device Name: Jetstream G3 System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II (two) Product Code: MCW Dated: April 30, 2010 Received: May 3, 2010

Dear Mr. Cleary:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K101221_

Device Names: Jetstream G3TM System

Indications for Use:

The Jetstream System is intended for use in atherectomy of the peripheral vasculature and to break apart and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, illiac or renal vasculature.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .

Concurrence of CDRH, Office of Device Evaluation (ODE)

W. J. Nagel

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number K061221

Page _ of _

§ 870.4875 Intraluminal artery stripper.

(a)
Identification. An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)(b)
Classification. Class II (performance standards).