(292 days)
The portable Medela® Invia Motion negative pressure wound therapy (NPWT) system is indicated to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudates and infectious material. It is intended for the use in hospitals, clinics, Long Term Care (LTC) and Home Care (HC) settings on adult patients with chronic, acute, subacute, traumatic, dehisced wounds, partial-thickness burns, ulcers (such as diabetic, neuropathic, pressure or venous insufficiency), flaps and grafts.
The Medela Invia Motion Negative Pressure Wound Therapy (NPWT) System is comprised of the Invia Motion NPWT Pump, canister/tubing set, power supply, carrying case, patient and user instructions, and Invia NPWT kits. The INVIA Motion is also compatible with Avance NPWT kits manufactured by Mölnlyke Healthcare.
Invia Motion NPWT pump is a suction pump for Negative Pressure Wound Therapy with an optical and acoustic status display. Invia Motion NPWT pump is a single patient use pump for continuous or intermittent operation and has a lifetime of 60 days.
Invia Motion NPWT pump is portable and can be operated independent of the electrical Invia Motion power supply due to a rechargeable battery. Acoustic and optical signals are triggered for variances from the set values as well as for faults.
Invia Motion NPWT system is intended for use in a home or other health care facility by medical personnel or trained lay users adhering to the instructions for use. The user may not be hard of hearing or deaf and must have normal visual acuity.
The provided text describes the Medela AG Invia Motion Negative Pressure Wound Therapy System and its 510(k) summary. However, it does not contain specific acceptance criteria, performance metrics, or details of a study designed to prove the device meets such criteria in a quantifiable manner as commonly expected for medical device evaluations that utilize AI or advanced diagnostics.
Instead, the non-clinical tests described focus on regulatory compliance, usability, and bench testing to demonstrate performance equivalence to predicate devices, rather than establishing direct clinical efficacy or performance against predefined, quantitative acceptance criteria via a clinical study with a specified sample size and ground truth.
Therefore, many sections of your requested output cannot be populated from the provided text. I will fill in what information is available and note when information is not present.
1. Table of Acceptance Criteria and Reported Device Performance
| Criterion Description | Acceptance Criterion | Reported Device Performance |
|---|---|---|
| Non-Clinical Tests | ||
| Electrical Safety | Compliance with IEC 60601-1 | Complies with IEC 60601-1 |
| Electromagnetic Compatibility | Compliance with IEC 60601-1-2 | Complies with IEC 60601-1-2 |
| Home Healthcare Environment Use | Compliance with IEC 60601-1-11 | Complies with IEC 60601-1-11 |
| Biocompatibility | Compliance with ISO 10993-1 | Complies with ISO 10993-1 |
| Sterilization Validation | (Not specified beyond "Sterilization validation") | (Validation completed) |
| Usability - Healthcare Professionals (HCPs) | Successful completion of tasks and resolution of close-calls/use-errors | All close-calls and use-errors observed during HCP use-scenarios and sub-tasks were resolved during the evaluation and did not affect patient safety. |
| Usability - Patient Lay Users (LUs) | 100% successful completion of target population use tasks | All 15 LU end-users (100%) successfully completed their target population use tasks by performing routine maintenance of the device and identifying and troubleshooting device acoustic and optical display signals to maintain safe and effective performance of the device. |
| Bench Testing (Performance and Reliability) | Design specifications met, equivalent performance to predicate device in maintaining set pressures and removing fluids in simulated wounds. | Bench testing demonstrated that the design specifications were met, including device performance and reliability. Comparison of performance with predicate device concluded equivalence in maintaining set pressures across the specification range and in removing fluids in simulated wounds. |
| Clinical Tests | (No clinical acceptance criteria or studies described) | (No clinical studies were conducted). |
2. Sample size used for the test set and the data provenance
-
Usability Study (a form of test set):
- Sample Size: 15 Healthcare Professionals (HCPs) and 15 patient Lay Users (LUs).
- Data Provenance: Conducted in the United States. Prospective (evaluation of users interacting with the device).
-
Bench Testing:
- Sample Size: Not specified.
- Data Provenance: Not specified (likely in-house lab testing).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Usability Study: The "ground truth" for the usability study was the successful completion of tasks and the identification/resolution of use errors. This was observed and assessed by the study administrators/evaluators, whose qualifications are not specified.
- Bench Testing: The "ground truth" for bench testing would be the engineering specifications and performance characteristics of the predicate device for comparison. No experts in the sense of clinical decision-makers were involved in establishing this ground truth from the provided text.
4. Adjudication method for the test set
- Usability Study: The text implies direct observation and assessment of user performance against predefined tasks and safety criteria. There is no mention of independent adjudication or a consensus method (like 2+1 or 3+1) among multiple observers.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study was done. The device is a Negative Pressure Wound Therapy System, not an AI-assisted diagnostic or imaging interpretation device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This is not applicable as the device is a therapeutic system, not an algorithm, and is operated by a human user (HCP or lay user).
7. The type of ground truth used
- Usability Study: Expert observation/assessment of user task completion and safety during device operation.
- Bench Testing: Engineering specifications and performance data of the device itself, and comparative performance data from predicate devices in simulated wound environments.
- Clinical: No clinical ground truth was established as no clinical studies were performed.
8. The sample size for the training set
- Not applicable. The device is a therapeutic system, not a machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
- Not applicable. (See point 8).
{0}------------------------------------------------
K113678 Page 1/3
Medela AG Invia Motion Negative Pressure Wound Therapy System
510(k) Summary
1
20,
OCT
In accordance with 21 CFR 807.92 the following summary of information is provided:
DATE: August 30, 2012
SUBMITTER:
Medela AG Lättichstrasse 4b 6341 Baar / Switzerland Phone +41 (0)41 769 52 47
PRIMARY CONTACT PERSON:
Adrienne Lenz, RAC Member Pathway Regulatory Consulting, LLC T 262-290-0023
SECONDARY CONTACT PERSON:
Markus Bütler Vice President Quality Management and Regulatory Affairs Medela AG
DEVICE:
TRADE NAME: Invia Motion
COMMON/USUAL NAME: Negative Pressure Wound Therapy System
CLASSIFICATION NAMES: 878.4780 Powered Suction Pump
PRODUCT CODE: OMP
PREDICATE DEVICE(S):
K080357 Medela Invia Wound Therapy
K083375 Smith and Nephew Renasys Go
K093526 KCI USA, Inc. V.A.C.Via Negative Pressure Wound Therapy System
{1}------------------------------------------------
K113678 Page 2/3
Medela AG Invia Motion Negative Pressure Wound Therapy System
DEVICE DESCRIPTION:
The Medela Invia Motion Negative Pressure Wound Therapy (NPWT) System is comprised of the Invia Motion NPWT Pump, canister/tubing set, power supply, carrying case, patient and user instructions, and Invia NPWT kits. The INVIA Motion is also compatible with Avance NPWT kits manufactured by Mölnlyke Healthcare.
Invia Motion NPWT pump is a suction pump for Negative Pressure Wound Therapy with an optical and acoustic status display. Invia Motion NPWT pump is a single patient use pump for continuous or intermittent operation and has a lifetime of 60 days.
Invia Motion NPWT pump is portable and can be operated independent of the electrical Invia Motion power supply due to a rechargeable battery. Acoustic and optical signals are triggered for variances from the set values as well as for faults.
Invia Motion NPWT system is intended for use in a home or other health care facility by medical personnel or trained lay users adhering to the instructions for use. The user may not be hard of hearing or deaf and must have normal visual acuity.
INTENDED USE:
The portable Medela® Invia Motion negative pressure wound therapy (NPWT) system is indicated to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudates and infectious material. It is intended for the use in hospitals, clinics, Long Term Care (LTC) and Home Care (HC) settings on adult patients with chronic, acute, subacute, traumatic, dehisced wounds, partial-thickness burns, ulcers (such as diabetic, neuropathic, pressure or venous insufficiency), flaps and grafts.
DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
The Invia Motion Negative Pressure Wound Therapy System uses the same fundamental technology as the Invia Wound Therapy system for most features. The indications for use of Invia Motion are the same as KCI's V.A.C Via NPWT System. Invia Motion is similar to the predicate devices in its indications for use, contra-indications and patient population. The user interface is also similar to the predicated devices. The main difference is in the volume of the canister, which is reduced but still sufficient for the expected daily volumes. The smaller size improves mobility. The pump is also intended for a single patient and has a useful life of 60 days, which is sufficient for single patient use.
SUMMARY OF NON-CLINICAL TESTS:
The Invia Motion Negative Pressure Wound Therapy System complies with voluntary standards for electrical safety, electromagnetic compatibility, biocompatibility, sterilization and use in the
{2}------------------------------------------------
K 113678 Page 3/3
Medela AG Invia Motion Negative Pressure Wound Therapy System
home healthcare environment. The following quality assurance measures were applied to the development of the system:
- Risk Analvsis .
- Software Validation .
- Electrical safety and electromagnetic compatibility testing per IEC 60601-1 and IEC . 60601-1-2 standards, respectively
- . Testing for use in the home healthcare environment per IEC 60601-1-11
- Biocompatibility testing per ISO 10993-1 standards .
- Sterilization validation .
- . Usability testing was conducted and demonstrated that the production versions of the labeling and device design provided first-time Invia® Motion™ NPWT System end-users with clear, concise instructions and the ability to avoid patterns of preventable use errors or difficulties when operating the device that could result in harms to the product end-users. Fifteen (15) Health Care Professionals (HCP) and 15 patient lay users (LU) were enrolled to participate in the study, which was conducted in the United States with English speakers. Representative product training for their user-population was provided and a training decay time was included prior to the evaluation. All 15 LU endusers (100%) successfully completed their target population use tasks by performing routine maintenance of the device and identifying and troubleshooting device acoustic and optical display signals to maintain safe and effective performance of the device. All close-calls and use-errors observed during the HCP use-scenarios and sub-tasks were resolved during the evaluation and did not affect patient safety.
- Bench Testing demonstrated that the design specifications were met, including device . performance and reliability. Comparison of performance of the Invia Motion with its compatible NPWT kits to the predicate device concluded that the Invia Motion is equivalent in maintaining set pressures across the specification range and in removing fluids in simulated wounds.
SUMMARY OF CLINICAL TESTS:
The Invia Motion Negative Pressure Wound Therapy System has not been the subject of clinical testing. A clinical evaluation of published literature has been conducted for Invia Motion.
CONCLUSION:
Medela AG considers the Invia Motion Negative Pressure Wound Therapy System to be as safe, as effective, and substantially equivalent to the predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol, which is a staff entwined with a serpent, representing medicine and health. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medela AG % Pathway Regulatory Consulting, LLC Ms. Adrienne Lenz, RAC W324S3649 County Road E Dousman, Wisconsin 53118
OCT
1 2012
Re: K113678
Trade/Device Name: Invia Motion Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 · Regulation Name: Powered suction pump Regulatory Class: II Product Code: OMP Dated: August 30, 2012 Received: September 04, 2012
Dear Ms. Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Adrienne Lenz, RAC
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Millerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K113678
Invia Motion Negative Pressure Wound Therapy System Device Name:
Indications for Use:
Invia Motion NPWT System:
The portable Medela® Invia Motion negative pressure wound therapy (NPWT) system is indicated to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudates and infectious material. It is intended for the use in hospitals, clinics, Long Term Care (LTC) and Home Care (HC) settings on adult patients with chronic, acute, subacute, traumatic, dehisced wounds, partial-thickness burns, ulcers (such as diabetic, neuropathic, pressure or venous insufficiency), flaps and grafts.
Prescription Use X
AND/OR
Over-The-Counter Use__
(Part 21 CFR 801 Subpart D)
(Part 21 CFR 801 Subpart C) .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krone for MXM
(Division Sign Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113678
§ 878.4780 Powered suction pump.
(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.