(131 days)
The PedCAT is intended to be used for 3-D imaging of the foot & ankle region, to visualize and assess the osseous and certain soft tissue structures, including joint spaces, bone angles and fractures. This modality is anticipated to be applicable to pediatric* cases as well as adults*, when appropriate diagnosis of a given foot condition is considered necessary.
The PedCAT is a dedicated X-Ray imaging device that acquires a 360 degree rotational X-ray sequence, reconstructs a three-dimensional matrix of the examined volume and produces two dimensional views of this volume. The PedCAT can measure distances and thickness on two dimensional images. Images produced by the PedCAT can be printed or exported on magnetic and optical media. The PedCAT gantry is comprised of an X-ray source, image detector, and motorized gantry. The gantry facilitates the acquisition of a full Xray sequence by the software. The software receives the two dimensional images acquired by the detector, transforms them into three dimensional images and displays them on the workstation monitor for viewing.
This document describes the CurveBeam PedCAT, a computed tomography x-ray system intended for 3-D imaging of the foot & ankle region. The submission is a 510(k) summary, which generally focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed clinical study results against specific acceptance criteria for performance metrics.
The provided text focuses on establishing substantial equivalence to a predicate device (Imaging Sciences DVT Scanner, K051980) based on similar intended use and technological characteristics. The study that proves the device meets "acceptance criteria" in this context is primarily a comparison to the predicate device and phantom testing to show similar imaging performance.
Here's a breakdown of the requested information based on the provided document:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" with numerical targets for clinical performance metrics (e.g., sensitivity, specificity for fracture detection). Instead, the performance evaluation aims to demonstrate that the PedCAT's imaging characteristics are "authentic and accurate representations of the imaged object, within the expected tolerance for a CBCT device, and very similar to the predicate device."
| Acceptance Criteria (Implied by Substantial Equivalence Claim) | Reported Device Performance (Summary) |
|---|---|
| Imaging Quality - Spatial Resolution | Evaluated using an ACR phantom. Result: "visible line pairs" and "authentic and accurate representations of the imaged object, within the expected tolerance for a CBCT device, and very similar to the predicate device." |
| Imaging Quality - Uniformity of Hounsfield Units (HU's) | Evaluated in water equivalent material phantom. Result: "authentic and accurate representations of the imaged object, within the expected tolerance for a CBCT device, and very similar to the predicate device." |
| Imaging Quality - HU values of various density materials | Evaluated in an ACR phantom. Result: "authentic and accurate representations of the imaged object, within the expected tolerance for a CBCT device, and very similar to the predicate device." |
| Imaging Quality - Osseous details and joint spaces (Anatomic) | Evaluated using an anatomic foot phantom (foot skeleton enclosed in soft tissue equivalent material). Result: "showed osseous details and joint spaces very similar to the predicate device, albeit on a different anatomy but on similar tissue densities." Further confirmed by clinical scans: "substantially similar to the predicate device, in terms of osseous details and joint spaces." |
| Scatter Radiation | Measured in the 2-scan orbit mode. Results: "values were found to be in a very low range and comparable to the predicate device, although slightly higher than single-orbit scans in the PedCAT." |
| Software Functionality (Frame capture & Reconstruction) | Frame capture tools are "identical" (supplied by Varian) and both PedCAT and predicate use FDK back-projection algorithm for CBCT reconstruction. Implied performance is comparable due to shared core components and algorithms. |
| Compliance with Standards | Explicit Criterion: The PedCAT complies with applicable FDA and international standards pertaining to electrical, mechanical, software, EMC, and radiation safety of medical and/or laser devices. Reported Performance: "The PedCAT Computed tomography x-ray system complies with applicable FDA and international standards pertaining to electrical, mechanical, software, EMC, and radiation safety of medical and / or laser devices." |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size (Clinical): "A large collection of patients have been scanned at 2 clinical sites." The exact number of patients is not specified.
- Test Set Sample Size (Phantom):
- ACR phantom
- Water equivalent material phantom
- Anatomic foot phantom (foot skeleton enclosed in soft tissue equivalent material)
- Data Provenance:
- Clinical: Retrospective or Prospective is not explicitly stated, but it was performed "under IRB framework," suggesting a structured approach to clinical data collection. Country of origin not specified.
- Phantom: Bench testing. Specific country not specified, but the submission is to the FDA in the USA.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not specify the number of experts or their qualifications used to establish ground truth for the clinical data. The statement "the results have been found to be substantially similar to the predicate device, in terms of osseous details and joint spaces" implies a comparison by qualified personnel, but details are not provided. For phantom studies, the "ground truth" is the known characteristics of the phantoms themselves.
4. Adjudication method for the test set
The document does not describe a formal adjudication method for the clinical test set (e.g., 2+1, 3+1). The evaluation appears to be a qualitative comparison of image characteristics ("substantially similar... in terms of osseous details and joint spaces") rather than an adjudication of specific findings against a consensus ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC comparative effectiveness study was not explicitly mentioned or described.
- AI Assistance: The PedCAT is described as an "X-Ray imaging device" and the software focuses on image reconstruction and display. There is no mention of Artificial Intelligence (AI) or AI-assisted reading features, so the concept of human reader improvement with/without AI assistance is not applicable to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance assessment was done, primarily through phantom studies. The characteristics of the resultant volume on the ACR phantom and the water equivalent phantom were evaluated for spatial resolution, uniformity of Hounsfield Units, and HU values of various density materials. The anatomic foot phantom also assessed osseous details and joint spaces. These evaluations are based on the output of the device's imaging and reconstruction algorithms without human interpretation in the loop. The clinical evaluation implicitly assesses the algorithm's output as well, but in a physician-reviewed context.
7. The type of ground truth used
- Phantom Studies: The ground truth was the known physical properties and structures of the phantoms (ACR phantom, water equivalent material phantom, anatomic foot phantom).
- Clinical Studies: The implicit ground truth for comparison was the image quality and diagnostic information provided by the predicate device, particularly regarding "osseous details and joint spaces." There is no mention of pathology, surgery, or long-term outcomes data used as ground truth. This is typical for 510(k) submissions focusing on substantial equivalence in imaging.
8. The sample size for the training set
The document does not mention a training set, as the PedCAT is an imaging device and not described as using machine learning or AI models that require a separate training set. The "software domain" section describes image capture tools and reconstruction algorithms (FDK back-projection), which are not typically "trained" in the machine learning sense. Statistical models that might be 'trained' or parameterized for image reconstruction are established algorithms, not typically requiring a distinct "training set" in a regulatory submission like this.
9. How the ground truth for the training set was established
Not applicable, as no training set for machine learning or AI was described.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "Curve Beam" in a stylized font. Above the text is a curved, dark shape that resembles an arc or a beam. The text is the main focus of the image, with the curved shape acting as a visual element that complements the text.
APR 1 0 2012
KI1354 175 Titus Ave, Suite 300 Warrington, PA 18976 Toll Free: 866-400-0035 Phone: 267-483-8081 Fax: 267-483-8086
ನ್ನ
510(k) Summary
| 510 (k) Submitter/Owner | CurveBeam, LLC175 Titus Ave, Suite 300Warrington, PA 18976Phone: 267-483-8081Fax: 267-483-8086 |
|---|---|
| Contact Person | David W. CowanVice President of Operations267-483-8087Email: dave.cowan@curvebeam.com |
| Date Prepared | November 30, 2011 |
| Trade Name | PedCAT |
| Common Name | Computed tomography x-ray system |
| Classification Name | Computed tomography x-ray system |
Predicate Devices:
| Company | Device name | Product Code | 510(k) |
|---|---|---|---|
| Imaging Sciences | DVT Scanner | JAK | K051980 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Curve Beam. The logo consists of the words "Curve Beam" in a simple, sans-serif font. Above the text is a curved line that arches over the words, adding a visual element to the logo. The logo is simple and modern.
175 Titus Ave, Suite 300 Warrington, PA 18976 Toll Free: 866-400-0035 Phone: 267-483-8081 Fax: 267-483-8086
Indications for Use:
The PedCAT is intended to be used for 3-D imaging of the foot & ankle region, to visualize and assess the osseous and certain soft tissue structures, including joint spaces, bone angles and fractures. This modality is anticipated to be applicable to pediatric* cases as well as adults*, when appropriate diagnosis of a given foot condition is considered necessary.
- Patient parameters: 50 lbs to 400 lbs Groin area at least 22" above the floor
Device Description:
The PedCAT is a dedicated X-Ray imaging device that acquires a 360 degree rotational X-ray sequence, reconstructs a three-dimensional matrix of the examined volume and produces two dimensional views of this volume. The PedCAT can measure distances and thickness on two dimensional images. Images produced by the PedCAT can be printed or exported on magnetic and optical media. The PedCAT gantry is comprised of an X-ray source, image detector, and motorized gantry. The gantry facilitates the acquisition of a full Xray sequence by the software. The software receives the two dimensional images acquired by the detector, transforms them into three dimensional images and displays them on the workstation monitor for viewing
Substantial Equivalence Summary:
The intended use of the PedCAT Computed Tomography x-ray system and the predicate device is substantially similar, with certain inconsequential differences described henceforth. The PedCAT Computed Tomography x-ray system has the same technological characteristics as the predicate device in terms of the deployed Cone Beam CT technology for 3-D imaging of anatomical structures of similar tissue density ranges. Although the intended use differs in the target anatomy, the intended diagnostic data (osseous tissue details) and applicable tissue densities are very similar. The PedCAT also differs in its mechanical layout and the patient support structure, since it is intended for 3-D imaging of Foot & Ankle in a weight-bearing as well as non-weight bearing position.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "Curve Beam" with a stylized graphic above it. The graphic is a curved, dark shape that arches over the text. The text is in a simple, sans-serif font, with the letters evenly spaced. The overall design is clean and modern.
175 Titus Ave, Suite 300 Warrington, PA 18976 Toll Free: 866-400-0035 Phone: 267-483-8081 Fax: 267-483-8086
The patient support structure is a circular rigid platform, where the patient stands. or places one or both feet if he/she sits on the adjacent seat. In the predicate device, the patient sits on a chair, while the targeted anatomy (maxillofacial region) is placed within a head support system consisting of a chin support and a headrest.
The rotating gantry in the PedCAT, on which the x-ray source and the detector panel are mounted, is enclosed in a low-density doughnut shaped plastic cover. permitting the patient's feet to be in the "doughnut-hole" opening. The predicate device has no such cover.
The above mechanical differences have no impact on how the projection data is captured and reconstructed, as compared to any CBCT device, including the predicate device. To establish this, test scans were performed on an ACR phantom, water equivalent material phantom, and an anatomic foot phantom (foot skeleton enclosed in soft tissue equivalent material). The characteristics of the resultant volume on the ACR phantom and the water equivalent phantom were authentic and accurate representations of the imaged object, within the expected tolerance for a CBCT device, and very similar to the predicate device. The specific parameters evaluated included spatial resolution (visible line pairs). uniformity of Hounsfield Units (HU's) in water equivalent material, and HU values of various density materials in the ACR phantom. The foot phantom scan showed osseous details and joint spaces very similar to the predicate device, albeit on a different anatomy but on similar tissue densities.
Further testing was performed on the clinical side. A large collection of patients have been scanned at 2 clinical sites under IRB framework, and the results have been found to be substantially similar to the predicate device, in terms of osseous details and joint spaces.
The position of the rectangular x-ray beam with respect to the target anatomy is very similar to the predicate device. In both cases, the beam is centered slightly below the center of the Field of View.
Another difference in details of a specific image acquisition mode in the PedCAT is that it performs 2 scan orbits to extend the diameter of the Field of View, with the x-ray beam and the detector offset to the left and right halves of the extended Field of View respectively. This is equivalent to having a "wider detector", but the projection geometry and other characteristics of the image acquisition process remain unchanged. The equivalence was established via test scans described above, both clinical and non-clinical (bench), which exhibited virtually identical results as compared to other scan modes. The scatter radiation was measured in
Revised 04/07/2012
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CurveBeam. The logo consists of the word "CurveBeam" in a sans-serif font, with the first letter "C" stylized with a curved line above it. The word is in black and the background is white. The logo is simple and modern.
175 Titus Ave, Suite 300 Warrington, PA 18976 Toll Free: 866-400-0035 Phone: 267-483-8081 Fax: 267-483-8086
a comprehensive manner in this mode (2 scan orbits) as well, and the values were found to be in a very low range and comparable to the predicate device, although slightly higher than single-orbit scans in the PedCAT.
In the software domain, although there are expected user interface differences with the predicate device, as well as differences in certain implementation details, the frame capture tools are identical (provided by Varian, the supplier of the detector panel) and both use FDK back-projection algorithm for CBCT reconstruction.
The above tests and analysis establishes that despite certain implementation differences, the PedCAT is substantially equivalent to the predicate device.
Safety and Effectiveness Information:
The PedCAT Computed tomography x-ray system is a Class II medical device and a Class 1M laser product.
The PedCAT Computed tomography x-ray system complies with applicable FDA and international standards pertaining to electrical, mechanical, software, EMC, and radiation safety of medical and / or laser devices.
Conclusion:
CurveBeam, LLC has demonstrated through its comparison of characteristics with the predicate device and comparison of performance testing with the predicate device that the PedCAT Computed tomography x-ray system is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the HHS emblem, which consists of a stylized caduceus with three snakes intertwined around a staff. The emblem is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration ' 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
APR 1 0 2012
Mr. David W. Cowan Vice President of Operations CurveBeam, LLC 175 Titus Avenue, Suite 300 WARRINGTON PA 18976
Re: K113548
Trade/Device Name: PedCAT Computed tomography x-ray system Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK and OAS Dated: March 6, 2012 Received: March 7, 2012
Dear Mr. Cowan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Form
510 (k) Number (if known) (
Device Name: PedCAT Computed tomography x-ray system
Indications for Use:
The PedCAT is intended to be used for 3-D imaging of the foot & ankle region, to visualize and assess the osseous and certain soft tissue structures, including joint spaces, bone angles and fractures. This modality is anticipated to be applicable to pediatric* cases as well as adults*, when appropriate diagnosis of a given foot condition is considered necessary.
- Patient parameters: 50 lbs to 400 Ibs Groin area at least 22" above the floor
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Device
Evaluation and Safety
510(k) K113548
Page 1 of
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.