K Number
K113127

Validate with FDA (Live)

Device Name
MIRAGE QUATTRO
Manufacturer
Date Cleared
2012-01-06

(74 days)

Product Code
Regulation Number
868.5905
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Mirage Quattro channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.

The Mirage Quattro is to be used by adult patients (>66lb />30kg) for whom positive airway pressure has been prescribed.

The Mirage Quattro is intended for single patient re-use in the home environment and/or multi-patient reuse in the hospital/institutional environment.

Device Description

The Mirage Quattro provides an interface such that airflow from a positive pressure source is directed to the patient's nose and mouth. The mask is held in place with adjustable headgear that straps the mask to the face.

The Mirage Quattro is safe when used under the conditions and purposes intended as indicated in the labeling provided with the product.

The Mirage Quattro is a prescription device supplied non-sterile.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the ResMed Mirage Quattro device:

ResMed Mirage Quattro - Acceptance Criteria and Study Analysis

This 510(k) summary focuses on a "Special 510(k): Device Modification" for the ResMed Mirage Quattro, comparing it to a previously cleared version (K063122). The primary modification appears to be an extended therapy pressure range.

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria for parameters like CO2 performance or pressure-flow characteristics. Instead, it relies on substantiating equivalence to the predicate device.

Acceptance Criteria (Implied)Reported Device Performance
Safety and Efficacy Equivalence to Predicate Device (K063122)Conclusion: "The modified Mirage Quattro is as safe and effective as the previously cleared Mirage Quattro device."
Identical Intended Use"it has the same intended use"
Identical Technological Characteristics (overall)"it has identical technological characteristics to the previously cleared device" (except for the extended pressure range, which was verified). This includes: sealing via silicone interface, various sizes for fit, integrated vents for CO2 flushing, anti-asphyxia valve (AAV), conventional air delivery hose connection (ISO 5356-1:2004), moulded plastic/silicone components and fabric/nylon headgear, all materials deemed safe (ISO10993-1).
No New Questions of Safety or Effectiveness"the modified device did not raise any new questions of safety or effectiveness"
CO2 Performance Equivalence"The CO2 performance of the modified device and the previously cleared device are substantially equivalent."
Pressure-Flow Characteristics and Flow Impedance Equivalence"The pressure-flow characteristics and flow impedance of both the modified device and the predicate device are identical."
Operability on Same Flow Generator Settings"Both the modified device and the previously cleared device are designed to operate on the same Full Face ResMed flow generator settings." (Note: The only difference mentioned is an extension of the therapy pressure range.)
Extended Therapy Pressure Range (4-20cmH2O to 4-40cmH2O) - Verified"Verification testing of the modified device did not raise any new questions of safety and efficacy."
Reusability in Home and Hospital/Institution Environment"Both the modified device and the previously cleared device can be reused in the home and hospital / institution environment."

2. Sample Size Used for the Test Set and Data Provenance

The document states that "Bench testing is sufficient to demonstrate safety and efficacy of the modified Mirage Quattro." This indicates that no clinical "test set" with human subjects was used for this modification. The testing was entirely bench-based, meaning it likely involved laboratory equipment and simulations rather than patient data.

  • Sample Size for Test Set: Not applicable, as testing was bench-based.
  • Data Provenance: Not applicable, as no human data was collected.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable, as no clinical test set requiring expert ground truth establishment was conducted. The assessment was based on engineering and performance specifications, likely evaluated by internal ResMed engineers and regulatory affairs personnel.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set requiring adjudication was conducted.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. The document explicitly states, "Bench testing is sufficient to demonstrate safety and efficacy of the modified Mirage Quattro." This rules out any clinical studies, including MRMC studies, for this particular modification.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

This question is not directly applicable in the context of a medical device like a mask, which is a physical interface and not an algorithm or AI. The testing performed would be considered "standalone" in the sense that it was bench testing of the device itself, without human interaction.

7. The Type of Ground Truth Used

The "ground truth" for this modification was primarily:

  • The performance specifications and characteristics of the predicate device (Mirage Quattro K063122), which had previously demonstrated safety and efficacy.
  • Engineering and performance test results (e.g., pressure-flow, CO2 flush, material safety) showing that the modified device met these specifications and performed equivalently or acceptably within the extended pressure range.

8. The Sample Size for the Training Set

Not applicable, as no machine learning algorithm or AI model was involved in this device modification.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no machine learning algorithm or AI model was involved.

{0}------------------------------------------------

RESMED

())

参考文献

Special 510(k): Device Mirage Quattro
Special 510(k): Device Modification

5. 510(k) SUMMARY

K 113127

[As required by 21 CFR 807.92]

Date Prepared18 October, 2011
SubmitterMs. Tracey Bullivant,Regulatory Affairs Manager
Official ContactMr. David D'Cruz,V.P., US Medical & Regulatory Affairs9001 Spectrum Center BlvdSan Diego CA 92123 USATel: (858) 836-5984
Device Trade NameMirage Quattro™
Device CommonName/ ClassificationNameVented Full Face Mask;Accessory to Noncontinuous Ventilator (IPPB)
Classification21 CFR 868.5905, 73 BZD (Class II)
Predicate DeviceMirage Quattro™ (K063122)
DescriptionThe Mirage Quattro provides an interface such that airflow from apositive pressure source is directed to the patient's nose and mouth.The mask is held in place with adjustable headgear that straps themask to the face.
The Mirage Quattro is safe when used under the conditions andpurposes intended as indicated in the labeling provided with theproduct.
The Mirage Quattro is a prescription device supplied non-sterile.
Intended UseThe Mirage Quattro channels airflow noninvasively to a patient froma positive airway pressure device such as a continuous positiveairway pressure (CPAP) or bilevel system.
The Mirage Quattro is to be used by adult patients (>66lb />30kg)for whom positive airway pressure has been prescribed.
The Mirage Quattro is intended for single patient re-use in the homeenvironment and/or multi-patient reuse in the hospital/institutionalenvironment.
Comparison ofTechnologicalCharacteristicsComparison with previously cleared Mirage QuattroThe modified device and the previously cleared mask both provide aseal via silicone interface. Both masks are offered in varioussizes to ensure adequate fit over the extended patient population.
Both masks incorporate vents to provide continuous air flow to flushout and minimize the amount of CO2 rebreathed by the patient. Thedesign of the mask components is such that the incorporation ofthese vents does not interfere with the intended performance of themasks. Both masks also contain an anti-asphyxia valve (AAV) toenable the patient to breathe fresh air in the event that airflow fromthe flow generator is impeded.
Both masks connect to a conventional air delivery hose between themask and the positive airway-pressure source via standard conicalconnectors (ref: ISO 5356-1:2004).
Both masks are constructed using moulded plastic and siliconecomponents and fabric / nylon headgear. All the components of bothmasks are fabricated using materials deemed safe (ISO10993-1).
Both the modified device and the previously cleared device aredesigned to operate on the same Full Face ResMed flow generatorsettings. The pressure-flow characteristics and flow impedance ofboth the modified device and the predicate device are identical.
Both the modified device and the previously cleared device can bereused in the home and hospital / institution environment.
Clinical DataUse of vented masks with CPAP or Bilevel therapy is proventechnology and is well accepted by the medical community. Benchtesting is sufficient to demonstrate safety and efficacy of themodified Mirage Quattro, as was the case with the previouslycleared Mirage Quattro device.
Performance DataComparison with previously cleared Mirage QuattroThe CO2 performance of the modified device and the previouslycleared device are substantially equivalent. Both the modified deviceand the previously cleared device are designed to operate on thesame flow generator settings as specified in the User Guide. Theonly difference is an extension in the therapy pressure range from4-20cmH2O to 4-40cmH2O. Verification testing of the modifieddevice did not raise any new questions of safety and efficacy.
SubstantialEquivalenceConclusionThe modified Mirage Quattro is as safe and effective as thepreviously cleared Mirage Quattro device:- it has the same intended use:- it has identical technological characteristics to thepreviously cleared device;- the modified device did not raise any new questions ofsafety or effectiveness;- it is at least as safe and effective as the previouslyCleared Mirage Quattro (K063122).

・・ ;

{1}------------------------------------------------

!~------

. · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·

·

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of a human figure embracing the world. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JAN - 6 2012

ResMed Limited C/O Mr. David D'Cruz Vice President, US Medical Regulatory Affairs ResMed Corporation 9001 Spectrum Center Boulevard San Diego, California 92123

Re: K113127

Trade/Device Name: MIRAGE QUATTRO™ FULL FACE MASK Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: December 2, 2011 Received: December 7, 2011

Dear Mr. D'Cruz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, . . . . . labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{3}------------------------------------------------

Page 2- Mr. D'Cruz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Division Director --Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): Device Name: Indication for Use

MIRAGE QUATTRO™ FULL FACE MASK

The Mirage Quattro channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.

The Mirage Quattro is to be used by adult patients (>66lb />30kg) for whom positive airway pressure has been prescribed.

The Mirage Quattro is intended for single patient re-use in the home environment and/or multi-patient reuse in the hospital/institutional environment.

Over-The-Counter Use AND/OR Prescription Use X (Part 21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) air 21 OF N 60 Four D)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of Device Evaluation (ODE)

Page 1 of_ 1_

L. Schulthis

(Division Sian-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K113127

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).