K Number
K112995
Device Name
GEM PREMIER 4000
Date Cleared
2012-05-03

(209 days)

Regulation Number
862.1600
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2 pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb) parameters. Total bilirubin can also be quantitated from heparinized plasma samples when analyzed in the tBili/CO-Ox mode. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity. Total bilirubin measurements are used in the diagnosis and management of biliary tract obstructions, liver disease and various hemolytic diseases and disorders involving the metabolism of bilirubin. In neonates, the level of total bilirubin is used to aid in assessing the risk of kernicterus.

Device Description

The GEM Premier 4000 is a portable critical care system. The potassium (K+) sensor membrane on the GEM Premier 4000 is being modified to lower the valinomycin concentration, along with a proportional decrease in the amount of counterion.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: GEM® Premier 4000 with modified K+ sensor membrane

1. Table of Acceptance Criteria and Reported Device Performance (Summary derived from the document):

Test TypeAcceptance Criteria (Implied/Expected)Reported Device Performance (with modified K+ sensor)
PrecisionK+ results for whole blood precision must be "within specification" (details of specification not explicitly stated, but implied by conclusion).Syringe Mode:- Level 1 (2.90 mmol/L): Total Imprecision %CV = 1.51%- Level 2 (3.88 mmol/L): Total Imprecision %CV = 1.32%- Level 3 (7.41 mmol/L): Total Imprecision %CV = 0.86%Full Capillary Mode:- Level 1 (3.06 mmol/L): Total Imprecision %CV = 3.55%- Level 2 (4.06 mmol/L): Total Imprecision %CV = 2.71%- Level 3 (7.44 mmol/L): Total Imprecision %CV = 2.16%Micro Capillary Mode:- Level 1 (3.34 mmol/L): Total Imprecision %CV = 3.31%- Level 2 (4.25 mmol/L): Total Imprecision %CV = 2.66%- Level 3 (7.83 mmol/L): Total Imprecision %CV = 1.27%Conclusion: All K+ results for whole blood precision were within specification.
LinearitySupport the current claimed reportable range of 0.2 to 19.0 mmol/L.Syringe Mode: y = 1.0144x - 0.0612 (R² = 0.9999), y = 1.0085x - 0.0038 (R² = 0.9999), y = 1.0438x - 0.165 (R² = 1)Full Capillary Mode: y=1.0115x + 0.0251 (R² = 0.9997); y = 1.0302x - 0.1058 (R² = 0.9998); y = 1.0324x - 0.0435 (R² = 0.9996)Micro Capillary Mode: y = 1.068x - 0.136 (R² = 0.9999), y = 1.0791x - 0.2189 (R² = 0.9998), y = 1.1017x - 0.2881 (R² = 0.9994)Conclusion: Linearity results support the claimed reportable range of 0.2 to 19.0 mmol/L for all sampling modes.
InterferencesNo clinically significant interference from common substances, or a specific limitation added if observed.Only citrate at concentrations ≥ 7.3 mmol/L showed a clinically significant interference effect. A limitation was added to labeling: "Blood collection tubes containing sodium citrate as an additive will produce a clinically significant change in potassium and should be avoided."
Method Comparison (Internal)K+ performance comparable to the GEM Premier 3000.Syringe: Slope = 0.978, R² = 0.998Full Capillary: Slope = 0.980, R² = 0.997Micro Capillary: Slope = 0.987, R² = 0.997Conclusion: K+ performance is comparable to the GEM Premier 3000 for all sample modes.
Field Site Testing (External)K+ performance comparable to the GEM Premier 3000/3500 in a clinical setting.Syringe: Slope = 1.050, R² = 0.989 (Range 2.0 to 7.5 mmol/L)Full Capillary: Slope = 1.018, R² = 0.957 (Range 2.3 to 5.6 mmol/L)Micro Capillary: Slope = 0.996, R² = 0.963 (Range 2.3 to 5.6 mmol/L)Conclusion: K+ performance in the clinical setting is comparable to the GEM Premier 3000/3500 for all sample modes.

2. Sample Size Used for the Test Set and Data Provenance:

  • Precision Study:
    • Sample Size: 3 sample levels (whole blood). Each level was assayed twice per day in eight replicates for five days. This totals 2 (assays/day) * 8 (replicates) * 5 (days) = 80 measurements per level per analyzer for each mode. With 2 analyzers, this is 160 measurements per level per mode.
    • Data Provenance: Not explicitly stated, but implied to be laboratory-based (internal testing) using whole blood samples. It's retrospective in the sense that the data was collected for the purpose of this submission, but not from an ongoing patient cohort.
  • Linearity Study:
    • Sample Size: Whole blood samples at 7 different K+ concentrations. Each concentration was tested in duplicate on the reference Flame Photometer and in triplicate on 3 different GEM Premier 4000 analyzers for each of the 3 sample modes. This implies 7 (concentrations) * 3 (replicates) * 3 (analyzers) = 63 measurements per sample mode.
    • Data Provenance: Not explicitly stated, but implied to be laboratory-based (internal testing) using manipulated whole blood samples.
  • Interference Study:
    • Sample Size: Substances were tested, but the number of samples or replicates is not specified.
    • Data Provenance: Not explicitly stated, but implied to be laboratory-based (internal testing).
  • Method Comparison (Internal):
    • Sample Size: 220 samples for each sample mode (syringe, full capillary, micro capillary).
    • Data Provenance: Internal study, likely laboratory-based.
  • Field Site Testing (External):
    • Sample Size: 454 samples for syringe mode, 304 samples for full capillary, and 304 samples for micro capillary.
    • Data Provenance: Clinical setting at three field sites. This implies prospective collection or anonymized retrospective patient samples from those sites.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • This is not an AI/imaging device study, so the concept of experts establishing ground truth for a test set in the traditional sense is not directly applicable.
  • For the Linearity study, "Flame atomic emission photometry (flame photometry)" was used as the reference method (ground truth). This is an established analytical technique, not a human expert.
  • For Method Comparison and Field Site Testing, the predicate devices (GEM Premier 3000 and 3500) served as the reference standard for comparison, rather than human expert opinion.

4. Adjudication Method for the Test Set:

  • Not applicable as this is a medical device performance study, not a study involving human interpretation needing adjudication. The "ground truth" was established by objective reference methods or comparison to predicate devices.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, this was not an MRMC comparative effectiveness study. This is a performance study for an in vitro diagnostic (IVD) device (blood analyzer) measuring specific analytes, not a study evaluating human reader performance with or without AI assistance for tasks like image interpretation.

6. Standalone (Algorithm Only) Performance:

  • Yes, the studies are primarily standalone performance assessments of the device itself. The device, the GEM Premier 4000 with the modified K+ sensor, is an automated analytical instrument. The studies evaluated its analytical performance (precision, linearity, interference, method comparison) without direct human intervention in the K+ measurement process beyond sample collection and instrument operation.

7. Type of Ground Truth Used:

  • Objective Reference Methods and Predicate Device Comparison:
    • Precision: Internal specifications (implied criterion).
    • Linearity: Flame atomic emission photometry (a recognized gold standard analytical method for K+).
    • Interferences: Clinical significance based on established medical understanding of K+ variations.
    • Method Comparison: GEM Premier 3000 (the previous version of the device).
    • Field Site Testing: GEM Premier 3000/3500 (predicate devices in a clinical setting).

8. Sample Size for the Training Set:

  • Not Applicable. This submission describes modifications to an existing device's sensor and subsequent validation studies. There is no mention of a "training set" in the context of machine learning or AI models. The device's underlying technology (potentiometric measurement) is based on electrochemical principles, not trained algorithms.

9. How the Ground Truth for the Training Set Was Established:

  • Not Applicable. As there is no training set for an AI model, this question is not relevant to the described device and studies. The device's performance is governed by its sensor design and calibration, which are validated through the studies outlined.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos for Werfen Group and Instrumentation Laboratory. The Werfen Group logo is a globe-like image with horizontal lines and the text "Werfen Group" below it. The Instrumentation Laboratory logo is a geometric shape resembling a cube with the text "Instrumentation Laboratory" to the right of it.

K112995

MAY - 3 2012

510(k) Summary

This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. .

Applicant Contact Information:

Applicant:Instrumentation Laboratory Co.
Address:180 Hartwell RoadBedford, MA 01730
Contact Person:Carol Marble, Regulatory Affairs Director
Phone Number:781-861-4467
Fax Number:781-861-4207
E-mail:cmarble@ilww.com
Preparation Date:April 25, 2012

Proprietary Name:

GEM® Premier 4000 with iQM® (Intelligent Quality Management) GEM® CVP 1 and 2 (Calibration Valuation Product) with CO-Ox GEM® CVP 3 and 4 (Calibration Valuation Product) Hematocrit GEM® CVP 5 (Calibration Valuation Product) tBili

Regulatory Information:

  • GEM Premier 4000 with iQM (Intelligent Quality Management) .
DescriptionCFR SectionDevice ClassProduct Code
Blood gases and blood pH862.1120Class IICHL
Sodium test system862.1665Class IIJGS
Potassium test system862.1600Class IICEM
Calcium test system862.1145Class IIJFP
Chloride test system862.1170Class IICGZ
Glucose test system862.1345Class IICGA
Lactic acid test system862.1450Class IKHP
Automated hematocrit instrument864.5600Class IIGKF
Carboxyhemoglobin assay864.7425Class IIGHS
Automated hemoglobin system864.5620Class IIGKR
Whole blood hemoglobin assays864.7500Class IIGLY
Bilirubin (Total or Direct)Test System862.1110Class IICIG
(Total and Unbound) in theNeonate Test System862.1113Class I,ReservedMQM

{1}------------------------------------------------

Regulatory Information (Cont.):

  • GEM CVP 1 and 2 (Calibration Valuation Product) with CO-Ox .
  • GEM CVP 5 (Calibration Valuation Product) tBili .
DescriptionCFR SectionDevice ClassProduct Code
Quality Control Material862.1660Class IJJY

GEM CVP 3 and 4 (Calibration Valuation Product) Hematocrit

DescriptionCFR SectionDevice ClassProduct Code
Hematocrit Control864.8625Class IIGLK

Predicate Device:

GEM Premier 4000 with iOM

Original K061974; Update K093623

Description of Device Modification:

The potassium (K+) sensor membrane on the GEM Premier 4000 is being modified to lower the valinomycin concentration, along with a proportional decrease in the amount of counterion. This modification is being made to bring the K+ performance on the GEM Premier 4000 back to the original labeled claims as cleared under K061974 (no change to K+ claims under K093623).

  • There are no changes to the following with this submission: .
    • No expansion or change in indications for use /intended use .
    • . No change to performance claims, except to introduce sodium citrate as an interferent
    • No change to software ●
    • No change to instrument or cartridge hardware beyond the K+ sensor modifications .
    • No change to the fluidic design ●
    • · No change to sensor fundamental technology, including for the K+ sensor
    • No change to cartridge bag solutions or other measuring sensors ●
    • No change to labeling of the cartridge ●

{2}------------------------------------------------

Indications for Use (No Change from K093623):

The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2 pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb) parameters. Total bilirubin can also be quantitated from heparinized plasma samples when analyzed in the tBili/CO-Ox mode. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity. Total bilirubin measurements are used in the diagnosis and management of biliary tract obstructions, liver disease and various hemolytic diseases and disorders involving the metabolism of bilirubin. In neonates, the level of total bilirubin is used to aid in assessing the risk of kernicterus.

Intelligent Quality Management (iQM) is used as the quality control and assessment system for the GEM Premier 4000 system. iQM is an active quality process control program designed to provide continuous monitoring of the analytical process with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls. Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed.

As part of this program, GEM CVP (Calibration Valuation Product) with CO-Ox, GEM CVP tBili and GEM CVP Hematocrit are external solutions intended to complete the calibration process and final accuracy assessment of the iQM cartridge calibration following warm-up. The reported values for GEM CVP (two levels for pH, blood gases, electrolytes, metabolites, total bilirubin, CO-Oximetry and hematocrit) must meet IL's specifications before the iQM cartridge can be used for patient sample measurements. Once the cartridge calibration is verified, the internal iQM program monitors the status of the system during the cartridge use life.

Comparison to Predicate Device:

Following are the similarities and differences between the current and modified K+ sensor membrane on the GEM Premier 4000:

  • Similarities: .
    • Same Indications for Use / Intended use .
    • Same Test Environment .
    • Same Operating Principle (Potentiometric Measurement) ●
    • Same Membrane Geometry .
    • Same Membrane Components .
    • Same Performance Characteristics, except to introduce sodium citrate as an interferent .
  • Differences limited to membrane formulation: .
    • Lowered valinomvcin concentration .
    • Proportional decrease in the amount of counterion .

Standards Referenced (if applicable):

  • CLSI EP5-A2, Evaluation of Precision Performance of Clinical Chemistry Devices ●
  • CLSI EP6-A. Evaluation of the Linearity of Quantitative Measurement Procedures: A . Statistical Approach
  • CLSI EP7-A2, Interference Testing in Clinical Chemistry ●
  • CLSI EP9-A2. Method Comparison and Bias Estimation Using Patient samples ●
  • CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantitation ●
  • CLSI EP-25-A. Evaluation of Stability of In Vitro Diagnostic Reagents .
  • CLSI C29-A2, Standardization of Sodium and Potassium Ion Selective Electrode Systems to ● the Flame Photometric Reference Method

{3}------------------------------------------------

Performance Characteristics Summary:

A series of studies were conducted that evaluated the performance of the GEM Premier 4000 system with the modified K+ sensor, including:

  • Precision .
  • . Linearity
  • Interferences .
  • Method Comparison ●
  • Field Site Testing ●

Whole Blood Precision

A precision study was conducted to evaluate the performance of the GEM Premier 4000 with the modified K+ sensor using whole blood. Each whole blood sample level was assayed twice per day in eight replicates for five days on two GEM Premier 4000 analyzers using the three sample modes: syringe, full capillary and micro capillary.

The results are presented in Tables 1-3 below.

Table 1: Whole Blood Precision Summary - Syringe Mode
---------------------------------------------------------------
LevelK+ Mean(mmol/L)NWithin RunSDWithin Run%CVTotalImprecisionSDTotalImprecision%CV
Level 12.901600.041.51%0.041.51%
Level 23.881600.051.32%0.051.32%
Level 37.411600.060.86%0.060.86%
LevelK+ Mean(mmol/L)NWithin RunSDWithin Run%CVTotalImprecisionSDTotalImprecision%CV
Level 13.061600.113.55%0.113.55%
Level 24.061600.112.71%0.112.71%
Level 37.441600.121.58%0.162.16%
Table 3: Whole Blood Precision Summary - Micro Capillary Mode
-------------------------------------------------------------------
LevelK+ Mean(mmol/L)NWithin RunSDWithin Run%CVTotalImprecisionSDTotalImprecision%CV
Level 13.341600.113.31%0.113.31%
Level 24.25158*0.112.66%0.112.66%
Level 37.831600.101.27%0.101.27%

· One sample duplicate pair was discarded due to a pre-analytical issue.

Conclusion: All K+ results for whole blood precision were within specification.

{4}------------------------------------------------

Performance Characteristics Summarv (Cont.):

Linearity

A linearity study was performed to evaluate the performance of the GEM Premier 4000 with the modified K+ sensor. Whole blood samples were manipulated through spiking and diluting to 7 different K+ concentrations spanning the claimed reportable range of 0.2 to 19 mmol/L. Flame atomic emission photometry (flame photometry), the preferred method to standardize direct ionselective electrode analyzers for determination of K+, acted as the reference method in testing. Each concentration was tested in duplicate on the reference Flame Photometer and in triplicate on 3 different GEM Premier 4000 analyzers using the 3 sample modes: syringe, full capillary and micro capillary.

Graphs of the observed values (for each of the three GEM Premier 4000 instruments) versus the expected values (flame photometry) are provided below for the three sample modes: syringe (Figure 1), full capillary (Figure 2) and micro capillary (Figure 3). .

Image /page/4/Figure/4 description: This image is a graph titled "GEM Premier 4000 K+ Linearity Syringe Mode (3 GEM 4000 Instruments)". The graph plots "K+ from GEM 4000, mmol/L" on the y-axis and "K+ Target by Flame Photometry, mmol/L" on the x-axis. There are three trendlines plotted on the graph, with equations y = 1.0144x - 0.0612 (R^2 = 0.9999), y = 1.0085x - 0.0038 (R^2 = 0.9999), and y = 1.0438x - 0.165 (R^2 = 1).

Figure 1: GEM Premier 4000 in Syringe Mode (3 Instruments)

{5}------------------------------------------------

Image /page/5/Figure/0 description: The image is a graph titled "GEM Premier 4000 K+ Linearity Capillary Mode (3 GEM 4000 Instruments)". The x-axis is labeled "K+ Target by Flame Photometry, mmol/L" and ranges from 0.0 to 30.0. The y-axis is labeled "K+ from GEM 4000, mmol/L" and ranges from 0.0 to 30.0. There are three trendlines plotted on the graph, with equations and R^2 values displayed: y=1.0115x + 0.0251, R^2 = 0.9997; y = 1.0302x - 0.1058, R^2 = 0.9998; y = 1.0324x - 0.0435, R^2 = 0.9996.

GEM Premier 4000 in Micro Capillary Mode (3 Instruments) Figure 3:

Image /page/5/Figure/2 description: The image is a graph titled "GEM Premier 4000 K+ Linearity Micro Capillary Mode (3 GEM 4000 Instruments)". The graph shows the relationship between K+ from GEM 4000, mmol/L and K+ Target by Flame Photometry, mmol/L. There are three trendlines on the graph, with equations y = 1.068x - 0.136, y = 1.0791x - 0.2189, and y = 1.1017x - 0.2881, and R^2 values of 0.9999, 0.9998, and 0.9994 respectively.

Conclusion: The linearity results support the current claimed reportable range of 0.2 to 19.0 mmol/L for the modified K+ sensor on the GEM Premier 4000, using all sampling modes (syringe, full capillary and micro capillary).

{6}------------------------------------------------

Interferences

Interference testing was performed on the substances listed below using the GEM Premier 4000 with the modified K+ sensor. These substances represent endogenous substances found among common abnormalities (e.g., lipemia), common prescription and over-the-counter drugs, sample additives such as anticoagulants and preservatives, abnormal biochemical metabolites, and medications often prescribed for critically ill patients.

  • Conclusion: Only citrate at concentrations at ≥ 7.3 mmol/L showed a clinically significant interference effect on the modified K+ sensor. Based on the results, the following limitation is being added to the labeling: "Blood collection tubes containing sodium citrate as an additive will produce a clinically significant change in potassium and should be avoided."

Method Comparison

An internal method comparison study was performed to compare the GEM Premier 4000 with the modified K+ sensor vs. GEM Premier 3000. All three sample modes were tested.

The summary results for each sample mode are presented in Table 5 below.

r0able
œ
NRange of SamplesSlope(95% CI)Intercept(95% CI)R2
Syringe2200.2 to 20.1 mmol/L0.978(0.972 to 0.984)0.143(0.112 to 0.174)0.998
Full Capillary2200.2 to 19.5 mmol/L0.980(0.973 to 0.987)0.169(0.135 to 0.204)0.997
Micro Capillary2200.2 to 20.6 mmol/L0.987(0.980 to 0.995)0.276(0.236 to 0.317)0.997

Conclusion: The above results support that across the reportable range, the K+ performance on the GEM Premier 4000 with the modified K+ sensor is comparable to the GEM Premier 3000 for all sample modes (syringe, full capillary and micro capillary).

Field Site Testing

External field testing was performed to compare the GEM Premier 3000/3500 vs. GEM Premier 4000 with the modified K+ sensor in the clinical setting. All three sample modes were tested at three field sites.

The pooled summary results for each sample mode are presented in Table 6 below.

{7}------------------------------------------------

Table 6:

NRange of SamplesSlope(95% CI)Intercept(95% CI)R2
Syringe4542.0 to 7.5 mmol/L1.050(1.040 to 1.061)-0.106(-0.147 to -0.064)0.989
Full Capillary3042.3 to 5.6 mmol/L1.018(0.993 to 1.043)-0.001(-0.100 to 0.099)0.957
Micro Capillary3042.3 to 5.6 mmol/L0.996(0.974 to 1.018)0.218(0.130 to 0.307)0.963

Conclusion: The above results support that in the clinical setting, the K+ performance on the GEM Premier 4000 with the modified K+ sensor is comparable to the GEM Premier 3000/3500 for all sample modes (syringe, full capillary and micro capillary).

Final Conclusion:

Analysis of the validation and verification test results support that the GEM Premier 4000 with the modified K+ sensor membrane is safe, effective, and substantially equivalent to the predicate device.

{8}------------------------------------------------

10903 New Hampshire Avenue Silver Spring, MD 20993

Instrumentation Laboratory Co. c/o Carol Marble 180 Hartwell Rd Bedford, MA 01730

MAY - 3 2012

Re: K112995

Trade Name: GEM Premier 4000 with iQM (Intelligent Quality Management), GEM CVP 1 and 2 (Calibration Valuation Product) with CO-Ox, GEM CVP 3 and 4 (Calibration Valuation Product) Hematocrit, GEM CVP 5 (Calibration Valuation Product) tBili Regulation Number: 21 CFR §862.1600 Regulation Name: Potassium test system Regulatory Class: Class II Product Codes: CEM, CHL, CGZ, CGA, CIG, GKR, GKF, GHS, GLK, GLY, JGS, JFP, JJY, KHP, MQM Dated: March 29, 2012 Received: March 30, 2012

Dear Ms. Marble:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{9}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you active of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 096-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance...

You may obtain other general information on your responsibilities under the Act from the Tou may other back Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

Signature

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use Statement

K112995

510(k) Number (if known):

GEM® Premier 4000 with iQM® (Intelligent Quality Management) Device Names: GEM® CVP 1 and 2 (Calibration Valuation Product) with CO-Ox GEM® CVP 3 and 4 (Calibration Valuation Product) Hematocrit GEM® CVP 5 (Calibration Valuation Product) tBili

Indications for Use:

The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb) parameters. Total bilirubin can also be quantitated from heparinized plasma samples when analyzed in the tBili/CO-Ox mode. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity. Total bilirubin measurements are used in the diagnosis and management of biliary tract obstructions, liver disease and various hemolytic diseases and disorders involving the metabolism of bilirubin. In neonates, the level of total bilirubin is used to aid in assessing the risk of kernicterus.

Intelligent Quality Management (iQM) is used as the quality control and assessment system for the GEM Premier 4000 system. iQM is an active quality process control program designed to provide continuous monitoring of the analytical process with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls. Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed.

As part of this program, GEM CVP (Calibration Valuation Product) with CO-Ox, GEM CVP tBili and GEM CVP Hematocrit are external solutions intended to complete the calibration process and final accuracy assessment of the iQM cartridge calibration following warm-up. The reported values for GEM CVP (two levels for pH, blood gases, electrolytes, metal bilirubin, CO-Oximetry and hematocrit) must meet IL's specifications before the iQM cartridge can be used for patient sample measurements. Once the cartridge calibration is verified, the internal iQM program monitors the status of the system during the cartridge use life.

Prescription Use V (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

AND/OR

Ruse Clue

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) 七 11 2 9 9 5

§ 862.1600 Potassium test system.

(a)
Identification. A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.(b)
Classification. Class II.