K Number
K112258

Validate with FDA (Live)

Device Name
NEUROTECH PLUS
Date Cleared
2012-01-09

(157 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Neurotech Plus device is indicated for the following:

In combination Neuromuscular Electrical Stimulation (NMES) and Transcutaneous Electrical Nerve Stimulation (TENS) modes, models 431, 432, 433, 434, 436, 439 & 440:

NMES Indications:
Maintain or increase the range of motion
Prevention or retardation of disuse atrophy
Re-educate muscles
Relax muscle spasms
Increase local blood circulation
Prevention of venous thrombosis of the calf muscles immediately after surgery

TENS Indications:
Symptomatic relief and management of chronic, intractable pain
Relief of pain associated with arthritis
Adjunctive treatment in the management of post-surgical and post-trauma pain
Adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee (models 431, 432 and 433 only)

In Neuromuscular Electrical Stimulation (NMES) only mode, models 437 & 441:

NMES Indications:
Maintain or increase the range of motion
Prevention or retardation of disuse atrophy
Re-educate muscles
Relax muscle spasms
Increase local blood circulation
Prevention of venous thrombosis of the calf muscles immediately after surgery

In Transcutaneous Electrical Nerve Stimulation (TENS) only mode, model 438:

TENS Indications:
Symptomatic relief and management of chronic, intractable pain
Relief of pain associated with arthritis
Adjunctive treatment in the management of post-surgical and post-trauma pain

Device Description

The Neurotech Plus is a portable, two-channel; battery operated system which can provide both Neuromuscular Electrical Stimulation (NMES) and Transcutaneous Electrical Nerve Stimulation (TENS). The device is intended be available by prescription only. Included with the Neurotech Plus control unit, are a lead wire assembly, electrodes and instructions for use.

The Neurotech Plus contains ten program sets which have been each allocated an individual type number. Seven of these types offer a combination of NMES and TENS programs. There are two types which offer only NMES programs and one which has only TENS programs.

The lead-wire assembly contains the ID chip (EEPROM) that identifies the required program set. Each of the program set options are pre-programmed during manufacturing and no access to the configuration is available to either the end user or clinician. When each lead wire is connected to the unit and the outputs (A, B, C or D) to the electrodes, the two channels may be operated independently through the unit. Detailed diagrams for correct device usage and placement of the electrodes are available as part of the instructions for use.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (Neurotech Plus, Type 413) and, as such, it does not describe an acceptance criteria study in the way one might for an AI/ML device or software performance. This document focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and intended use, rather than presenting a performance study with acceptance criteria and results.

However, I can extract information related to the device's characteristics and the comparison made to establish substantial equivalence, which is a form of meeting "acceptance criteria" in the context of a 510(k):

1. Table of Acceptance Criteria and Reported Device Performance

In a 510(k) submission, "acceptance criteria" is not explicitly stated as performance metrics like sensitivity/specificity for a diagnostic AI. Instead, substantial equivalence is established by demonstrating that the new device has the same intended use and similar technological characteristics to a legally marketed predicate device, or that any differences do not raise new questions of safety or effectiveness. The "reported device performance" is essentially the detailed technical specifications of the Neurotech Plus and how they compare to the predicate devices.

The tables (7.1 through 7.10) in the document provide a detailed comparison of the Neurotech Plus models with their respective predicate devices across various electrical stimulation parameters. These comparisons serve as the basis for demonstrating that the new device performs "as well as or better than" the listed predicate.

Below is an example of an abstracted table, as the original tables (7.1-7.10) are extensive and detailed for each model. I will use the information from Table 7.1 for Model 431 as an example comparison, highlighting how the new device's parameters compare to the predicate devices.

Abstracted Comparison Table (Example for Neurotech Plus Model 431 vs. Predicates)

CharacteristicAcceptance Criteria (Predicate Device K061516: Staodyn Max / K971437: Bionicare / K082011: MediTens)Neurotech Plus Model 431 Reported PerformanceComparison Outcome (for this submission's purpose)
WaveformPulsed, symmetric, biphasic (Staodyn Max, MediTens); Pulsed, Monophasic (Bionicare)Pulsed, symmetric, biphasicSimilar to Staodyn Max & MediTens
ShapeRectangular (Staodyn Max); Rectangular, with interphase interval (MediTens); Exponential, Spike (Bionicare)Rectangular, with interphase intervalSimilar to MediTens
Max Output Voltage (500 Ω)30.0V (Staodyn Max); 12V (Bionicare); 37.5V (MediTens)35.0VWithin range of predicates (or comparable)
Max Output Current (500 Ω)60.0mA (Staodyn Max); 24mA (Bionicare); 75.0mA (MediTens)70.0mAWithin range of predicates (or comparable)
RMS Output Voltage (500 Ω)8.9V (Staodyn Max); 4.3V (Bionicare); 6.5V (MediTens)7.8VWithin range of predicates (or comparable)
RMS Output Current (500 Ω)17.8mA (Staodyn Max); 8.6mA (Bionicare); 13mA (MediTens)15.7mAWithin range of predicates (or comparable)
Primary Phase Duration60-350uS (Staodyn Max); 640 μδ (Bionicare); 150uS (MediTens)250uSWithin range of predicates (or comparable)
Pulse Duration700uS (Staodyn Max); 640 μS (Bionicare); 400μς (MediTens)400-600uS (uS both phases + 100µS interphase delay)Within range of predicates (or comparable)
Frequency80 to 125 Hz (Staodyn Max); 100 Hz (Bionicare); 4 to 99 Hz (MediTens)4 to 125HzWithin range of predicates (or comparable)
Net Charge (500 Ω)0 μC (Staodyn Max, MediTens); 21 μC (Bionicare) *Method for 0μC specified for Staodyn & MediTens0 μC @ 500 Ω (Symmetric, biphasic and leading polarity alternates for each successive pulse)Similar to Staodyn Max & MediTens
Max Phase Charge (500 Ω)9 to 21 μC (Staodyn Max); 21 μC (Bionicare); 7.5 to 11.25 μC (MediTens)10.5 to 17.5 μCWithin range of predicates (or comparable)
Max Current Density0.71mA/cm² (Staodyn Max); 0.08 mA/cm² (Bionicare); 0.672mA/cm² (MediTens)0.80mA/cm²Within range of predicates (or comparable)
Max Average Current5.3mA (Staodyn Max); 2.1mA (Bionicare); 2.25mA (MediTens)3.5mAWithin range of predicates (or comparable)
Max Average Power Density6.3 mW/cm² (Staodyn Max); 0.34 mW/cm² (Bionicare); 4.4mW/cm² (MediTens)6.2mWWithin range of predicates (or comparable)

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states: "No new clinical tests have been submitted as part of this premarket notification." Therefore, there is no "test set" in the context of patient data, nor is there information on sample size or data provenance from a prospective or retrospective study. The primary "test" for this submission is a comparison of engineering specifications against predicate devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Since no new clinical tests were conducted and no patient-specific "test set" was used, there were no experts involved in establishing a "ground truth" for such a test set. The "ground truth" for a 510(k) is implicitly the established safety and effectiveness of the predicate devices.

4. Adjudication Method for the Test Set

Not applicable, as no new clinical test set requiring adjudication was used.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

Not applicable. This is a 510(k) for a TENS/NMES device, not an AI/ML diagnostic or treatment assistance software. No MRMC study was performed.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Not applicable. This is a hardware medical device (Neurotech Plus, Type 413), not a standalone algorithm.

7. The Type of Ground Truth Used

In the context of a 510(k), the "ground truth" for demonstrating safety and effectiveness relies on the previously established regulatory clearance of the predicate devices. The technological characteristics and intended use of the Neurotech Plus were compared to these already approved devices. The ground truth is effectively the regulatory precedent set by the predicate devices.

8. The Sample Size for the Training Set

Not applicable, as this is not an AI/ML device and therefore does not involve a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for bio-medical research ltd. The logo consists of the letters "bmr" in a bold, sans-serif font. Below the letters is the text "bio-medical research ltd." in a smaller, sans-serif font. The letters "bmr" are connected by a thin, curved line.

K112258

This 510k Summary is being submitted in accordance with the requirements of 21 CFR 807.92.

1. Contact Details

Name:Anne-Marie Keenan
Address:Bio-Medical Research Ltd.,Parkmore Business Park, WestGalway, Ireland
Telephone:+353 91 774300
Fax:+353 91 774302
E-Mail:akeenan@bmr.ie
Prepared:November ,2, 2011

2. Device Name and Regulation

Trade Name of Device:Neurotech Plus, Type 413
Regulation Number:21 CFR 882.5890
Regulation NameTranscutaneous electrical nerve stimulator for pain relief
Product Code:GZJ, NYN, IPF
Device Class:II

3. Identification of Equivalent Legally Marketed Device

510(k) Number:K082011
Manufacturer:Bio-Medical Research Ltd.
Trade Name:MediStim XP, Type 281

{1}------------------------------------------------

510(k) Number:K082011
Manufacturer:Bio-Medical Research Ltd.
Trade Name:MediTens XP, Type 458
510(k) Number:K061516
Manufacturer:Compex Technologies
Trade Name:Staodyn® Max, Model 4470
510(k) Number:K971437
Manufacturer:Murray Electronics
Trade NameBionicare Stimulator System Model Bio-1000

4. Description of Device

The Neurotech Plus is a portable, two-channel; battery operated system which can provide both Neuromuscular Electrical Stimulation (NMES) and Transcutaneous Electrical Nerve Stimulation (TENS). The device is intended be available by prescription only. Included with the Neurotech Plus control unit, are a lead wire assembly, electrodes and instructions for use.

The Neurotech Plus contains ten program sets which have been each allocated an individual type number. Seven of these types offer a combination of NMES and TENS programs. There are two types which offer only NMES programs and one which has only TENS programs.

The lead-wire assembly contains the ID chip (EEPROM) that identifies the required program set. Each of the program set options are pre-programmed during manufacturing and no access to the configuration is available to either the end user or clinician. When each lead wire is connected to the unit and the outputs (A, B, C or D) to the electrodes, the two channels may be operated independently through the unit. Detailed diagrams for correct device usage and placement of the electrodes are available as part of the instructions for use.

5. Statement of Intended Use/Indications for Use

Intended Use: Dependent on the program chosen, the Neurotech Plus, Type 413 can deliver Neuromuscular Electrical Stimulation (NMES) for the activation of muscle for rehabilitation

('

{2}------------------------------------------------

and/or Transcutaneous Electrical Nerve Stimulation (TENS) for the activation of nerves for neuromodulation and management of pain.

Indications for Use:

i

Models 431, 432, 433, 434, 436, 439, & 440 offer a combination of NMES or TENS programs.

NMES Indications for Use

Maintain or increase the range of motion Prevention or retardation of disuse atrophy Re-educate muscles Relax muscle spasms Increase local blood circulation Prevention of venous thrombosis of the calf muscles immediately after surgery

TENS Indications for Use

Symptomatic relief and management of chronic, intractable pain Relief of pain associated with arthritis Adjunctive treatment in the management of post-surgical and post-trauma pain Adjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee (models 431, 432, 433 only)

Models 437, 441 (NMES only indications for use)

Maintain or increase the range of motion Prevention or retardation of disuse atrophy Re-educate muscles Relax muscle spasms Increase local blood circulation Prevention of venous thrombosis of the calf muscles immediately after surgery

Model 438 (TENS only indications for use)

Symptomatic relief and management of chronic, intractable pain Relief of pain associated with arthritis Adjunctive treatment in the management of post-surgical and post-trauma pain

Sale of the device has been restricted to sale under a prescription order from a licensed practitioner.

{3}------------------------------------------------

6. Summary of Technological Characteristics

There are no new technological characteristics that could affect safety or effectiveness of the Neurotech Plus device. No new clinical tests have been submitted as part of this premarket notification.

The Neurotech Plus device complies with the following international standards:

  • IEC 60601-1 (1998) + A1 (1991) + A2 (1995) Medical Electrical Equipment Part 1: . General Requirements for Safety.
  • . IEC 60601-2-10 (1987) + A1 (2001) Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators.
  • IEC 60601-1-2:2001 Medical electrical equipment Part 1-2: General requirements . for safety - Collateral standard: Electromagnetic compatibility - Requirements and Tests :
  • I.S. EN ISO 14971 2007 .

6.1 Program Sets

Tables 6.1.1 to 6.1.10 show the program parameters for each model of the Neurotech Plus.

{4}------------------------------------------------

Table 6.1+.1Model 431ScreenTextIntensityRate(Hz or pps)Pulse Width(usec)Cont(sec.)Relax(sec.)RampUp(sec.)RampDown(sec.)Additional FunctionsTreatmentTime(minutes)
1P1Adj125 to 4;modulates(decreases) between125 to 4 in 12second cycle -remains at 125 for 3seconds and 4 for 3seconds of cycletime(see Note 1)150 to 250;modulates(increases) between150 to 250 in 12second cycle -remains at 150 for 3seconds and 250 for3 seconds of cycletime(see Note 1)Open
2P2Adj99200Burst Toggle - 4 Hz, 250 usec Activation andoutput remains constant after pressing the burstbutton. Press Burst again to cancel. No rampup/down will be implementedOpen
3P3AdjCh 1: 4Ch 2: 125200Criss-cross Function, Burst - Ch1 125 Hz, 200 usec& Ch2 125 Hz, 200 usec. Activation and outputremains constant for 10 minutes after burst buttonpressed. No ramp up/down will be implemented(See note 2)Open
4P4Adj4250Burst - 125 Hz, 175 usec. Activation and outputremains constant while pressing burst button. Noramp up/down will be implementedOpen
5P5Adj503005151.030
6P6Adj503005101.0Trigger30
7P7Adj50300551.0Trigger30
8P8Adj353005101.0Trigger30
9P9Adj8300551.0Trigger20
10P10Adj100701002005151.09010
Adj100701002005151.09020
Adj10010090
Table 6.1.2Model 432ScreenTextIntensityRate(Hz or pps)Pulse Width(usec)Contract(sec.)Relaxation(sec.)Ramp Up(Sec.)RampDown(Sec.)Additional FunctionsTreatment Time(minutes)
1P1Adj125 to 4;modulates (decreases)between 125 to 4 in 12second cycle - remains at125 for 3 seconds and 4for 3 seconds of cycletime150 to 250;modulates (increases)between 150 to 250 in12 second cycle -remains at 150 for 3seconds and 250 for 3seconds of cycle timeBurst-4Hz, 250 usec.Activation and outputremains constantwhile pressing burstbutton. No rampup/down will beimplementedOpen
2P2Adj99200Open
3P3AdjCh 1: 4Ch 2: 125200Criss-cross Function,Burst - Ch1 125 Hz,200 usec & Ch2 125Hz, 200 usec.Activation and outputremains constant for10 minutes after burstbutton pressed. Noramp up/down will beimplemented.(See note 2)Open
4P4Adj4250Burst-125Hz, 175usec. Activation andoutput remainsconstant whilepressing burst button.No ramp up/down willbe implementedOpen
5P5Adj503005151.00.5Trigger30
6P6Adj503005101.00.5Trigger30
7P7Adj50300551.00.5Trigger30
8P8Adj353005101.00.5Trigger20
9P9Adj8300551.0.5Trigger20
10P10Adj701001002005151.00.570
701001002005151.00.570

.

Page 5 of 33

{5}------------------------------------------------

:

Page 6 of 33

:

:

{6}------------------------------------------------

Table 6.1.3Model 433ScreenTextIntensityRate(Hz or pps)Width(usec)Contraction(sec.)Relaxation(sec.)Ramp Up(Sec.)RampDown(Sec.)AdditionalFunctionsTreatmentTime(minutes)
1P1Adjustable125 to 4;modulates(decreases)between 125 to4 in 12 secondcycle - remainsat 125 for 3seconds and 4for 3 seconds ofcycle time150 to 250;modulates(increases)between 150 to250 in 12second cycle -remains at 150for 3 secondsand 250 for 3seconds ofcycle timeOpen
2P2Adjustable125175Open
3P3Adjustable4250Burst -Frequency of 4Hz, pulse widthof 250 usec.Activation andoutput remainsconstant whilepressing burstbutton. Willramp down over 3sec. and up over 3secOpen
4P4Adjustable503005151.00.5Trigger30
5P5Adjustable503005101.00.5Trigger30
6P6Adjustable50300551.00.5Trigger30
7P7Adjustable353005101.00.5Trigger30
8P8Adjustable8300551.00.5Trigger20
9P9Adjustable100100151.00.5Burst -Frequency of 125Hz. pulse widthof 175 usec.Activation andoutput remainsconstant whilepressing burstbutton. No rampupodwn will beimplemented.70
10P10Adjustable702005151.00.5Trigger20
Table 6.1.4Model 434ScreenTextIntensityRate(Hz or pps)Width(usec)Contraction(sec.)Relaxation(sec.)RampUp(Sec.)RampDown(Sec.)AdditionalFunctionsTreatmentTime(minutes)
1P1Adjustable80120ContinuousStimulation2Open
2P2Adjustable2180ContinuousStimulation2Open
3P3Adjustable8070-120ContinuousStimulation2Open
4P4Adjustable80/2180ContinuousStimulation2Open
5P5Adjustable9/2300ContinuousStimulation220
6P6Adjustable8300551.00.5Trigger20
7P7Adjustable353005101.00.5Trigger30
8P8Adjustable503005151.00.5Trigger30
9P9Adjustable503005101.00.5Trigger30
10P10Adjustable50300551.00.5Trigger30
Table 6.1.5Model 436ScreenTextIntensityRate(Hz or pps)Width(usec)Contraction(sec.)Relaxation(sec.)Ramp Up(Sec.)Ramp Down(Sec.)Additional FunctionsTreatmentTime(minutes)
1P1Adjustable125 to 4:modulates(decreases)between 125 to 4in 12 secondcycle - remains at125 for 3 secondsand 4 for 3seconds of cycletime150 to 250;modulates(increases)between 150 to250 in 12 secondcycle - remains at150 for 3 secondsand 250 for 3seconds of cycletimeNANANANANoneOpen
2P2Adjustable125 to 4;modulates(decreases)between 125 to 4in 12 secondcycle - remains at125 for 3 secondsand 4 for 3seconds of cycletime50 to 100;modulates(increases)between 50 to 100in 12 second cycle- remains at 50 for3 seconds and 100for 3 seconds ofcycle timeNANANANANoneOpen
3P3Adjustable125150NANANANABurst - Frequency is80hz, pulse width 300usec with 8 pulses perburst;2 bursts per second. 12second cycle time - 2seconds modulationfrom normal to burst, 8seconds at burst, 2seconds modulatingfrom burst to normal.Frequency willincrement/decrementduring the transitionto/from burst.Open
4P4Adjustable803001025(RemoveBeep)1.51.5Trigger30
5P5Adjustable503001020 (RemoveBeep)1.51.5Trigger30
6P6Adjustable35300551.51.5Trigger30

·

Page 7 of 33

t

{7}------------------------------------------------

Page 8 of 33

{8}------------------------------------------------

.

t

{9}------------------------------------------------

Relax time5 seconds5 seconds5 secondsseconds
Ramp Down Time2 seconds2 seconds2 seconds2 seconds
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Contract Time------------------------------------------------------------------------------------------------------------------------------------------------------------------------------seconds3 secondsANNUAL ALL PROPERTY LA CONSULT LE LEASE LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEsecondsseconds-----------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-Ramp-Up time2seconds2seconds-And And Commend Commend Comments of Children2seconds2seconds
A . A . APulse Width2000 JsCarles Market225 Us2500s2700s
Carreland ContractFrequency0Hz55Hz------------------------------------------------------------------------------------------------------------------------------------------------------------------------------55Hz------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A50Hz------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
DurationSTERE COLORIO CONSULTION20 minutesA Land State State of States of Active Andrewi25 minutes30 minutes40 minutes
Table 6.1.6Model 437ProgramParticle Property of Children Comments of ChildrenProgram211-2017 ComProgram 3Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children States of thProgram
Table 6.1.7Model 438ScreenTextIntensityRate(Hz or pps)Width(usec)PulseTreatment Time(minutes)
1P1Adjustable125 to 4;modulates (decreases) between 125 to 4 in 12 second cycle – remains at 125 for 3 seconds and 4 for 3 seconds of cycle time150 to 250;modulates (increases) between150 to 250 in 12 second cycle -remains at 150 for 3 seconds and250 for 3 seconds of cycle timeNA45
2P2Adjustable125 to 4;modulates (decreases) between 125 to 4 in 12 second cycle – remains at 125 for 3 seconds and 4 for 3 seconds of cycle time50 to 100;modulates (increases) between 50to 100 in 12 second cycle -remains at 50 for 3 seconds and100 for 3 seconds of cycle timeNA45
3P3Adjustable125175NA45
4P4Adjustable4250NA45
5P5AdjustableChannel 1 - 4Channel 2 - 125175NA45
6P6Adjustable801758 pulses per burst;2 bursts per second45
7P7Adjustable125 to 50;modulates (decreases) between 125 and 50 in 8 second cycle (4/4)175NA45
8P8Adjustable125250 to 150;modulates (decreases) between250 and 150 in 8 second cycle(4/4)NA45
9P9Adjustable125 to 50;modulates (decreases) between 125 to 50 in 6 second cycle (3/3)50 to 100;modulates (increases) between 50to 100 in 6 second cycle (3/3)NA45
10P10Adjustable125 to 50;modulates (decreases) hetween 125 and 50 in 10 second cycle (5/5)150 to 250;modulates (increases) between150 and 250 in 10 second cycle(5/5)NA45

Page 10 of 33

-�

{10}------------------------------------------------

Table 6.1.8Model 439ScreenTextIntensityRate(Hz or pps)Width(usec)Contraction(sec.)Relaxation(sec.)Ramp Up(Sec.)Ramp Down(Sec.)AdditionalFunctionsTreatmentTime(minutes)
1P1Adjustable50300551.00.5Trigger30
2P2Adjustable5030010101.01.0Trigger30
3P3Adjustable5030010201.51.0Trigger30
4P4Adjustable5030010301.51.0Trigger30
5P5Adjustable352508251.51.0Trigger30
6P6Adjustable35300551.01.0Trigger30
7P7Adjustable1025010101.51.0Trigger30
8P8Adjustable4250Burst-125 Hz, 175 usec.Activation and output remains constant while pressing burst button.Will ramp up over 3 sec. and down over 3 sec45
9P9Adjustable125175No-trigger45
10P10Adjustable125 to 4; modulates (decreases) between 125 to 4 in 12 second cycle - remains at 125 for 3 sec. and 4 for 3 sec. of cycle time150 to 250; modulates (increases) between 150 to 250 in 12 sec. cycle - remains at 150 for 3 sec. and 250 for 3 sec. of cycle timeNo-trigger45

Page 11 of 33

·

・・

{11}------------------------------------------------

Table 6.1.9Model 440ScreenTextIntensityRate(Hz or pps)Width(usec)Contraction(sec.)Relaxation(sec.)Ramp Up(Sec.)RampDown(Sec.)Additional FunctionsTreatmentTime(minutes)
1P1Adjustable125 to 4;modulates(decreases)between 125 to 4in 12 second cycle- remains at 125for 3 seconds and4 for 3 seconds ofcycle time150 to 250;modulates(increases)between 150 to250 in 12 secondcycle - remains at150 for 3 secondsand 250 for 3seconds of cycletimeOpen
2P2Adjustable99 to 50;modulates(decreases)between 99 and 50in 8 second cycle(4/4)150 to 250;modulates(increases)between 150 to250 in 8 secondcycle (4/4)Open
3P3Adjustable125175Criss-cross Function(See note 3)Open
4P4Adjustable42501.0
5P5Adjustable503005101.00.5Trigger30
6P6Adjustable5030010101.00.5Trigger30
7P7Adjustable5030010201.00.5Trigger30
8P8Adjustable5030010301.00.5Trigger30
9P8Adjustable35300551.00.5Trigger20
10P10Adjustable1030010101.00.5Trigger20
ScreenTextIntensityRate(Hz or pps)Width(usec)Contraction(sec.)Relaxation(sec.)RampUp(Sec.)RampDown(Sec.)AdditionalFunctionsTreatmentTime(minutes)
5.1.10441P1Adjustable503005101.00.5Trigger20
P2Adjustable5030010101.00.5Trigger20
P3Adjustable5030010201.00.5Trigger20
P4Adjustable5030010301.00.5Trigger20
P5Adjustable35300551.00.5Trigger20
P6Adjustable4300Continuous StimulationNo-Trigger20
P7Adjustable103002051.50.5Trigger20
P8Adjustable7030010501.00.5Trigger20
P9Adjustable993005301.00.5trigger30

.

.

Page 12 of 33

ς

-� .

{12}------------------------------------------------

Substantial Equivalence Comparison Tables 1

analysis carried out between the proposed Neurotech Plus and the proposed device is as safe, as effective and Each model of the Neurotech Plus, Type 413 device has been compared to it's predicate device at the worst case program. Based on comparati performs as well or better than the listed predicate

{13}------------------------------------------------

t

·

:

Table 7.1Neurotech PlusModel 431Predicate DeviceStaodyn MaxK061516PredicateBionicareK971437PredicateMediTensK082011
Mode or Program Name(For calculation of maximum rms current, and power density)Plan 3 (Highest Output)Program E (Highest output)Only has 1 modeProgram 399Hz, 150μS
Waveform (e.g pulsed monophasic, biphasic)Pulsed, symmetric, biphasicPulsed, symmetric, biphasicRectangularPulsed, MonophasicExponential, SpikePulsed, symmetric, biphasic
Shape (e.g rectangular, spike, rectified sinusoidal)Rectangular, with interphase intervalRectangular, with interphase interval
Maximum Output Voltage (volts)(+/- 10%)35.0V @ 500 Ω75.0V @ 2 kΩ70.3V @ 10 kΩ30.0V @ 500 ΩNot availableNot available12V 500ΩNot availableNot available37.5V @ 500 Ω70.0V@ 2 kΩ71.8V @ 10 kΩ
Maximum Output Current (specify units)(+/- 10%)70.0mA @ 500 Ω37.5mA @ 2 kΩ7.0mA @ 10 kΩ60.0 mA @ 500 ΩNot availableNot available24mA @ 500ΩNot availableNot available75.0mA @ 500 Ω35.0mA @ 500 Ω7.0mA @ 10 kΩ
RMS Output Voltage (volts)(+/- 10%)7.8V @ 500 Ω8.9V @ 500 Ω4.3VEstimated assuming a square pulse shape 0.64ms6.5V@500Ω
16.8V @ 2 kΩ7.9V @ 10 kΩNot availableNot availableNot availableNot available14.2V@2kΩ4.95V@10kΩ
RMS Output Current (specify units)(+/- 10%)15.7mA @ 500 Ω17.8mA @ 500 Ω8.6mA13mA@500Ω
8.4mA @ 2 kΩ0.8mA @ 10 kΩNot availableNot availableNot availableNot available7.1mA@2kΩ0.5mA@10kΩ
Duration of primary (depolarizing) phase (usec)250uS60-350uS640 μδ150uS
Pulse Duration (usec)400-600uS (uS both phases +100µS interphase delay)700uS (350 uS both phases)640 με400μς
Frequency (Hz)[ or Rate (pps) ]4 to 125Hz80 to 125 Hz100 Hz4 to 99 Hz
For multiphasic waveforms only:Symmetrical phases?YesYesN/AYes
Phase duration (include100 to 250μ S60-350utN/A100 to 150uS
Table 7.1Neurotech PlusModel 431Predicate DeviceStaodyn MaxK061516PredicateBionicareK971437PredicateMediTensK082011
(units) state range, ifapplicable)(both phases, ifasymmetrical)
Net Charge(microcoulombs (uC) per pulse) (if zero,state method of achieving zero net charge.)0 μC @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for eachsuccessive pulse.0 μC @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for eachsuccessive pulse.21 μC @ 500 Ω0 μC @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for eachsuccessive pulse.
Maximum Phase Charge (uC)10.5 to 17.5 μC @ 500 Ω9 to 21 μC @ 500 Ω21 μC @ 500 Ω7.5 to 11.25 μC @ 500 Ω
Maximum Current Density (mA/cm², r.m.s){@ 500 Ω Where T is the duration of averaging forworst case (highest) output}$0.80mA/cm^2(T=1sec)$Program EUsing 2" square electrode=17.8 mA/25 cm²= 0.71mA/cm²$0.08 mA/cm^2$i.e. 8.6mA/108 cm²Using 12 x 9 cm electrodes$0.672mA/cm^2$5cm round electrode)
Maximum Average Current (average absolute value),mA { @ 500 Ω]3.5mA5.3mA2.1mA2.25mA
Maximum Average Power Density, (W/ cm²), (usingsmallest electrode conductive surface area){ @ 500 Ω Where T is the output duration}6.2mW (T=1sec)50mm round electrode$6.3 mW/cm^2$Program EUsing 2" square electrode$0.34 mW/cm^2$$4.4mW/cm^2$50mm round electrode
Burst Mode (i.e., pulse (a) Pulses per bursttrains):N/ABurst mode is simply an alternatefrequency-pulse witdhcombination. There are actuallyno intermittent bursts8N/AN/ABurst mode is simply an alternatefrequency-pulse witdhcombination. There are actuallyno intermittent bursts
(b) Bursts per secondN/A2N/AN/A
(c) Burst duration(seconds)N/A0.09N/AN/A
(d) Duty Cycle ( Line (b) xLine (c) )N/A.18N/AN/A
ON Time (seconds)N/AN/AN/AN/A
OFF Time (seconds)N/AN/AN/AN/A

.

"}

{14}------------------------------------------------

·

.

Page 15 of 33

ન્દ્ર

יי

{15}------------------------------------------------

Table 7.2Mode or Program NameNeurotech PlusModel 432Predicate DeviceStaodyn MaxK061516PredicateBionicareK971437PredicateMediTensK082011Program 3 (99Hz, 150μS)
Waveform (e.g pulsed monophasic, biphasic)Plan 3Program E (Highest output)Only has 1 mode
Shape (e.g rectangular, spike, rectified sinusoidal)Pulsed, symmetric, biphasicRectangular, with interphase intervalPulsed, symmetric, biphasicRectangularPulsed, MonophasicExponential, SpikePulsed, symmetric, biphasicRectangular, with interphase interval
Maximum Output Voltage (volts)(+/-10%)35.0V @ 500 Ω30.0V @ 500 Ω12V 500Ω37.5V @ 500 Ω
75.0V @ 2 kΩNot availableNot available70.0V @ 2 kΩ
70.3V @ 10 kΩNot availableNot available71.8V @ 10 kΩ
Maximum Output Current (specify units)(+/-10%)70.0mA @ 500 Ω60.0 mA @ 500 Ω24mA @ 500Ω75.0mA @ 500 Ω
37.5mA @ 2 kΩNot availableNot available35.0mA @ 500 Ω
7.0mA @ 10 kΩNot availableNot available7.0mA @ 10 kΩ
RMS Output Voltage (volts)(+/-10%)7.8V @ 500 Ω8.9V @ 500 Ω4.3VEstimated assuming a square pulseshape 0.64ms6.5V@500Ω
16.8V @ 2 kΩNot availableNot available14.2V@2kΩ
7.9V @ 10 kΩNot availableNot available4.95V@10kΩ
RMS Output Current (specify units)(+/- 10%)15.7mA @ 500 Ω17.8mA @ 500 Ω8.6mA13mA@500Ω
8.4mA @ 2 kΩNot availableNot available7.1mA@2kΩ
0.8mA @ 10 kΩNot availableNot available0.5mA@10kΩ
Duration of primary (depolarizing) phase (usec)150-250uS60-350uS640 μS150uS
Pulse Duration (usec)400-600uS (uS both phases + 100μs interphase delay)700uS (350 uS both phases)640 μS400μS
Both phases
Frequency (Hz)[ or Rate (pps) ]4-125Hz80 to 125 Hz100 Hz4 to 99 Hz
For multiphasic waveforms only:Symmetrical phases?N/ANoYes
Phase duration (include units) state range, if60-350uSN/A100 to 150uS
Table 7.2
Neurotech PlusModel 432Predicate DeviceStaodyn MaxK061516Predicate BionicareK971437Predicate MediTensK082011
applicable)(both phases, asymmetrical) ifNet Charge(microcoulombs (uC) per pulse) (if zero, state method of achieving zero net charge.)0 μC @ 500 ΩSymmetric, biphasic and leading polarity alternates for each successive pulse.0 μC @ 500 ΩSymmetric, biphasic and leading polarity alternates for each successive pulse.21 μC @ 500Ω0 μC @ 500 ΩSymmetric, biphasic and leading polarity alternates for each successive pulse.
Maximum Phase Charge (uC)14.0 μC @ 500 Ω9 to 21 μC @ 500 Ω21 μC @ 500Ω7.5 to 11.25 μC @ 500 Ω
Maximum Current Density (mA/cm², r.m.s){@ 500 Ω Where T is the output duration}0.80mA/cm²(T=1sec)Program EUsing 2" square electrode=17.8 mA/25 cm²= 0.71mA/cm²0.08 mA/cm²i.e. 8.6mA/108 cm²Using 12 x 9 cm electrodes0.672mA/cm²5cm round electrode)
Maximum Average Current (average absolute value), mA {@ 500 Ω}3.5mA5.3mA2.1mA2.25mA
Maximum Average Power Density, (W/ cm²),(using smallest electrode conductive surface area){@ 500 Ω Where T is the output duration}6.2mW (T=1sec)6.3 mW/cm²Program EUsing 2" square electrode0.34 mW/cm²4.4mW/ cm²50mm round electrode
Burst Mode (i.e., pulse trains):Burst mode is simply an alternate frequency-pulse witdh combination.There are actually no intermittent bursts8N/ABurst mode is simply an alternate frequency-pulse witdh combination. There are actually no intermittent bursts
(a) Pulses per burstN/A
(b) Bursts per secondN/A2N/AN/A
(c) Burst duration (seconds)N/A0.09N/AN/A
(d) Duty Cycle [Line (b) x Line (c)]N/A.18N/AN/A
ON Time (seconds)N/AN/AN/AN/A
OFF Time (seconds)N/AN/AN/AN/A
Additional Features (specify, if applicable)N/AN/AN/AN/A

.

4

{16}------------------------------------------------

r

.

.

.

,

יי

{17}------------------------------------------------

Table 7.3Mode or Program NameWaveform (e.g pulsed monophasic, biphasic)Shape (e.g rectangular, spike, rectified sinusoidal)Maximum Output Voltage (volts)(+/- 10%)Maximum Output Current (specify units)(+/- 10%)RMS Output Voltage (volts)(+/- 10%)RMS Output Current (specify units)(+/- 10%)Duration of primary (depolarizing) phase (usec)Pulse Duration (usec)Frequency (Hz)[ or Rate (pps) ]Symmetrical phases?For multiphasic waveforms only:Phase duration (include units) state range, if
Neurotech PlusModel 433Plan 2Pulsed, symmetric, biphasicRectangular, with interphase interval35.0V @ 500 Ω75.0V @ 2 kΩ70.3V @ 10 kΩ70.0mA @ 500 Ω37.5mA @ 2 kΩ7.0mA @ 10 kΩ7.3V @ 500 Ω14.6mA @ 500 Ω7.8mA @ 2 kΩ0.8mA @ 10 kΩ150-250uS400-600uS (uS both phases +100µS interphase delay)4-125HzYes60-350uS
Predicate DeviceStaodyn MaxK061516Program E (Highest output)Pulsed, symmetric, biphasicRectangular30.0V @ 500 ΩNot availableNot available60.0 mA @ 500 ΩNot availableNot available8.9V @ 500 Ω17.8mA @ 500 ΩNot availableNot availableNot available700uS (350 uS both phases)80 to 125 HzYesN/A
PredicateBionicareK971437Only has 1 modePulsed, MonophasicExponential, Spike12V 500ΩNot availableNot available24mA @ 500ΩNot availableNot available4.3VEstimated assuming a square pulse shape 0.64ms8.6mANot availableNot availableNot available640 µS100 Hz
PredicateMediTensK082011Program 399Hz, 150μSPulsed, symmetric, biphasicRectangular, with interphase interval37.5V @ 500 Ω70.0V@ 2 kΩ71.8V @ 10 kΩ75.0mA @ 500 Ω35.0mA @ 500 Ω7.0mA @ 10 kΩ6.5V@500ΩNot available14.2V@2kΩ4.95V@10kΩNot available150uS400μ SBoth phases4 to 99 HzYes100 to 150uS

Page 18 of 33

:

.

{18}------------------------------------------------

Table 7.3Neurotech PlusModel 433Predicate DeviceStaodyn MaxK061516PredicateBionicareK971437PredicateMediTensK082011
applicable)(both phases,if asymmetrical)
Net Charge(microcoulombs (uC) per pulse) (if zero,state method of achieving zero net charge.)0 μC @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for eachsuccessive pulse.0 μC @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for eachsuccessive pulse.21 μC @ 500Ω0 μC @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for eachsuccessive pulse.
Maximum Phase Charge (uC)12.3 µC @ 500 Ω9 to 21 μC @ 500 Ω21 µC @ 500Ω7.5 to 11.25 µC @ 500 Ω
Maximum Current Density (mA/cm², r.m.s){@ 500 Ω Where T is the output duration)0.75mA/cm²(T=1sec)Program EUsing 2" square electrode=17.8 mA/25 cm²= 0.71mA/cm²0.08 mA/cm²i.e. 8.6mA/108 cm²Using 12 x 9 cm electrodes0.672mA/cm²5cm round electrode)
Maximum Average Current (average absolute value),mA {@ 500 Ω)3.06mA5.3mA2.1mA2.25mA
Maximum Average Power Density, (W/ cm²), (usingsmallest electrode conductive surface area)1 @ 500 Ω Where T is the output duration}5.5mW (T=1sec)6.3 mW/cm²Program EUsing 2" square electrode0.34 mW/cm²4.4mW/cm²50mm round electrode
Burst Mode (i.e., pulse (a) Pulses per bursttrains):N/ABurst mode is simply an alternatefrequency-pulse witdh combination.There are actually no intermittentbursts8N/AN/ABurst mode is simply an alternatefrequency-pulse witdhcombination. There are actuallyno intermittent bursts
(b) Bursts per secondN/A2N/AN/A
(c) Burst duration(seconds)N/A0.09N/AN/A
(d) Duty Cycle [ Line(b) x Line (c)]N/A.18N/AN/A
ON Time (seconds)N/AN/AN/AN/A
OFF Time (seconds)N/AN/AN/AN/A
Additional Features (specify, if applicable)N/AN/AN/AN/A

Page 19 of 33

.

ệ :

·

{19}------------------------------------------------

Table 7.4Mode or Program NameWaveform (e.g pulsed monophasic, biphasic)Shape (e.g rectangular, spike, rectified sinusoidal)Maximum Output Voltage (volts)(+/- 10%)Maximum Output Current (specify units)(+/- 10%)RMS Output Voltage (volts)(+/- 10%)RMS Output Current (specify units)(+/- 10%)Duration of primary (depolarizing) phase (usec)Pulse Duration (usec)Frequency (Hz) or Rate (pps) ]For multiphasic waveforms only:Symmetrical phases?Phase duration (include units) state range, if applicable)(both phases, if asymmetrical)Net Charge(microcoulombs (uC) per pulse) (if zero, state method of achieving zero net charge.)
Neurotech PlusModel 434Plan 4Pulsed, symmetric, biphasicRectangular, with interphase interval35.0V @ 500 Ω75.0V @ 2 ΚΩ70.3V @ 10 ΚΩ70.0mA @ 500 Ω37.5mA @ 2 ΚΩ7.0mA @ 10 ΚΩ5.4V @ 500 Ω11.6V @ 2 ΚΩ6.3V @ 10 ΚΩ10.8mA @ 500 Ω5.8mA @ 2 ΚΩ0.6mA @ 10 ΚΩ120-180uS340-460uS (uS both phases + 100μS interphase delay)2-80HzYes100 to 250μ S0 μC @ 500 ΩSymmetric, biphasic and leading polarity
Predicate DeviceStaodyn MaxK061516Program E (Highest output)Pulsed, symmetric, biphasicRectangular30.0V @ 500 ΩNot availableNot available60.0 mA @ 500 ΩNot availableNot available8.9V @ 500 ΩNot availableNot available17.8mA @ 500 ΩNot availableNot available60-350uS700uS (350 uS both phases)80 to 125 HzYes60-350uS0 μC @ 500 ΩSymmetric, biphasic and
Predicate DeviceMediTensK082011Program 3 (99Hz, 150µS)Pulsed, symmetric, biphasicRectangular, with interphase interval37.5V @ 500 Ω70.0V @ 2 ΚΩ71.8V @ 10 ΚΩ75.0mA @ 500 Ω35.0mA @ 500 Ω7.0mA @ 10 ΚΩ6.5V@500Ω14.2V@2kΩ4.95V@10kΩ13mA@500Ω7.1mA@2kΩ0.5mA@10ΚΩ150uS400μ S4 to 99 HzYes100 to 150uSΟ μC @ 500 ΩSymmetric, biphasic and leading

!

י

  • 8

{20}------------------------------------------------

alternates for each successive pulse.leading polarity alternates for each successive pulse.polarity alternates for each successive pulse.
Maximum Phase Charge (µC)10.5µC @ 500 Ω9 to 21 µC @ 500 Ω7.5 to 11.25 µC @ 500 Ω
Maximum Current Density (mA/cm², r.m.s){@ 500 Ω Where T is the output duration}0.55mA/cm²(T=1sec)Program EUsing 2" square electrode=17.8 mA/25 cm²= 0.71mA/cm²0.672mA/cm²5cm round electrode)
Maximum Average Current (average absolute value), mA{ @ 500 Ω}1.68mA5.3mA2.25mA
Maximum Average Power Density, (W/ cm²), (using smallest electrode conductive surface area){@ 500 Ω Where T is the output duration}3.0mW (T=1sec)6.3 mW/cm²Program EUsing 2" square electrode4.4mW/ cm²50mm round electrode
Burst Mode (i.e., pulse trains):(a) Pulses per burstN/ABurst mode is simply an alternatefrequency-pulse witdh combination. Thereare actually no intermittent bursts8N/ABurst mode is simply an alternatefrequency-pulse witdh combination.There are actually no intermittentbursts
(b) Bursts per second2
(c) Burst duration(seconds)0.09
(d) Duty Cycle [ Line (b) x Line (c) ].18
ON Time (seconds)N/AN/AN/A
OFF Time (seconds)N/AN/AN/A
Additional Features (specify, if applicable)N/AN/AN/A

:

Page 21 of 33

. . . .

{21}------------------------------------------------

Table 7.5Neurotech PlusModel 436Predicate DeviceStaodyn MaxK061516Neurotech PlusModel 436Predicate DeviceMediStim XPK082011
Mode or Program NamePlan 3 (TENS)Plan 4 (NMES)
Waveform (e.g pulsed monophasic, biphasic)Pulsed, symmetric, biphasicPulsed, symmetric, biphasicPulsed, symmetric, biphasicPulsed, symmetric, biphasic
Shape (e.g rectangular, spike, rectified sinusoidal)Rectangular, with interphase intervalRectangularRectangular, with interphase intervalRectangular, with interphase interval
Maximum Output Voltage (volts)(+/- 10%)35.0V @ 500 Ω75.0V @ 2 kΩ70.3V @ 10 kΩ30.0V @ 500 ΩNot availableNot available35.0V @ 500 Ω75.0V @ 2 kΩ70.3V @ 10 kΩ37.5V @ 500 Ω70.0V@ 2 kΩ71.8V @ 10 kΩ
Maximum Output Current (specify units)(+/- 10%)70.0mA @ 500 Ω37.5mA @ 2 kΩ7.0mA @ 10 kΩ60.0 mA @ 500 ΩNot availableNot available70.0mA @ 500 Ω37.5mA @ 2 kΩ7.0mA @ 10 kΩ75.0mA @ 500 Ω35.0mA @ 500 Ω7.0mA @ 10 kΩ
RMS Output Voltage (volts)(+/- 10%)6.8V @ 500 Ω14.5V @ 2 kΩ7.8V @ 10 kΩNot availableNot availableNot available7.7V @ 500 Ω16.4V @ 2 kΩ6.3V @ 10 kΩ6.5 V @ 500 Ω12.1V @ 2 kΩ7.1V @ 10 kΩ
RMS Output Current (specify units)(+/- 10%)13.6mA @ 500 Ω7.3mA @ 2 kΩ0.8mA @ 10 kΩNot availableNot availableNot available15.3mA @ 500 Ω8.2mA @ 2 kΩ0.6mA @ 10 kΩ13.0mA @ 500 Ω6.0mA @2 kΩ0.7mA @ 10 kΩ
Duration of primary (depolarizing) phase (usec)50-250uS60-350uS300uS150-300uS
Pulse Duration (usec)200-600uS (uS bothphases + 100μςinterphase delay)700uS (350 uS bothphases)600 uS (100 uS bothphases + 100μςinterphase delay)400-700us (100 uS bothphases + 100μ Sinterphase delay)
Frequency (Hz)[ or Rate (pps) ]4-125HzNot available35-80Hz2-100Hz
For multiphasic waveforms only:Symmetrical phases?Phase duration (include units) staterange, if applicable)(both phases, ifasymmetrical)N/AN/AN/AN/A
Net Charge(microcoulombs (uC) per pulse) (if zero, state method of achievingzero net charge.)0 μC @ 500 ΩSymmetric, biphasic andleading polarity alternates0 μC @ 500 ΩSymmetric, biphasic andleading polarity alternates0 μC @ 500 ΩSymmetric, biphasic andleading polarity alternates0 μC @ 500 ΩSymmetric, biphasic andleading polarity alternates
Table 7.5Neurotech PlusModel 436Predicate DeviceStaodyn MaxK061516Neurotech PlusModel 436Predicate DeviceMediStim XPK082011
for each successive pulse.for each successive pulse.for each successive pulse.for each successive pulse.
Maximum Phase Charge (uC)$10.5µC @ 500 Ω$Not available$21.0µC @ 500 Ω$$21.0µC @ 500 Ω$
Maximum Current Density (mA/cm², r.m.s){ @ 500 Ω Where T is the output duration}$0.69mA/cm²(T=1sec)$Not available$0.78mA/cm²(T=1sec)$$mA/cm²(T=1sec)$
Maximum Average Current (average absolute value), mA { @ 500 Ω}2.63mANot available3.36mANot available
Maximum Average Power Density, (W/ cm²), (using smallest electrodeconductive surface area){ @ 500 Ω Where T is the output duration}$4.7mW (T=1sec)$Not available$6.0mW (T=1sec)$Not available
Burst Mode (i.e., pulse trains):(a) Pulses per burstN/AN/AN/AN/A
(b) Bursts per secondN/AN/AN/AN/A
(c) Burst duration (seconds)N/AN/AN/AN/A
(d) Duty Cycle [Line (b) x Line (c) ]N/AN/AN/AN/A
ON Time (seconds)N/AN/AN/AN/A
OFF Time (seconds)N/AN/AN/AN/A

Page 22 of 33

{22}------------------------------------------------

ー・

:

.

{23}------------------------------------------------

Table 7.6Neurotech PlusModel 437Predicate DeviceMediStim XPK082011
Mode or Program NamePlan 3 (NMES)
Waveform (e.g pulsed monophasic, biphasic)Pulsed, symmetric, biphasicPulsed, symmetric, biphasic
Shape (e.g rectangular, spike, rectified sinusoidal)-Rectangular, with interphase intervalRectangular, with interphase interval
Maximum Output Voltage (volts)(+/- 10%)35.0V @ 500 Ω37.5V @ 500 Ω
75.0V @ 2 ΚΩ70.0V @ 2 ΚΩ
70.3V @ 10 ΚΩ71.8V @ 10 ΚΩ
Maximum Output Current (specify units)(+/- 10%)70.0mA @ 500 Ω75.0mA @ 500 Ω
37.5mA @ 2 ΚΩ35.0mA @ 500 Ω
7.0 mA @ 10 ΚΩ7.0mA @ 10 ΚΩ
RMS Output Voltage (volts)(+/-10%)5.8V @ 500 Ω6.5 V @ 500 Ω
12.4V @ 2 ΚΩ12.1V @ 2 ΚΩ
5.2V @ 10 ΚΩ7.1V @ 10 ΚΩ
RMS Output Current (specify units)(+/- 10%)11.6mA @ 500 Ω13.0mA @ 500 Ω
6.2mA @ 2 ΚΩ6.0mA @ 2 ΚΩ
0.5mA @ 10 ΚΩ0.7mA @ 10 ΚΩ
Duration of primary (depolarizing) phase (usec)200-270uS150-300uS
Pulse Duration (usec)500-640 uS (100 uS both phases + 100µSinterphase delay)400-700us (100 uS both phases + 100μSinterphase delay)
Frequency (Hz)[ or Rate (pps) ]50-60Hz2-100Hz
For multiphasic waveforms only:Symmetrical phases?N/AN/A
Phase duration (include units) staterange, if applicable)(both phases, ifasymmetrical)N/AN/A
Net Charge(microcoulombs (uC) per pulse) (if zero, state method of achievingzero net charge.)0 μC @ 500 ΩSymmetric, biphasic and leading polarity alternatesfor each successive pulse.0 μC @ 500 ΩSymmetric, biphasic and leading polarityalternates for each successive pulse.
Maximum Phase Charge (uC)17.5μC @ 500 Ω21.0µC @ 500 Ω
Maximum Current Density (mA/cm², r.m.s){@ 500 Ω Where T is the output duration)0.59mA/cm²(T=1sec)
Table 7.6Neurotech PlusModel 437Predicate DeviceMediStim XPK082011
Maximum Average Current (average absolute value), mA {@ 500 Ω}1.93mANot available
Maximum Average Power Density, (W/ cm2), (using smallest electrodeconductive surface area) { @ 500 Ω Where T is the output duration}3.4mW (T=1sec)Not available
Burst Mode (i.e., pulse trains):(a) Pulses per burstN/AN/A
(b) Bursts per secondN/AN/A
(c) Burst duration (seconds)N/AN/A
(d) Duty Cycle [ Line (b) x Line (c) ]N/AN/A
ON Time (seconds)N/AN/A
OFF Time (seconds)N/AN/A
Table 7.7Neurotech PlusModel 438Predicate DeviceStaodyn MaxK061516
Code or Program NamePlan 8
Waveform (e.g pulsed monophasic, biphasic)Pulsed, symmetric, biphasicPulsed, symmetric, biphasic
Shape (e.g rectangular, spike, rectified sinusoidal)Rectangular, with interphase intervalRectangular
Maximum Output Voltage (volts)(+/- 10%)35.0V @ 500 Ω75.0V @ 2 kΩ70.3V @ 10 kΩ30.0V @ 500 ΩNot availableNot available
Maximum Output Current (specify units)(+/- 10%)70.0mA @ 500 Ω37.5mA @ 2 kΩ7.0mA @ 10 kΩ60.0 mA @ 500 ΩNot availableNot available
RMS Output Voltage (volts)(+/- 10%)8.7V @ 500 Ω18.7V @ 2 kΩ7.8V @ 10 kΩNot availableNot availableNot available
RMS Output Current (specify units)(+/- 10%)17.5mA @ 500 Ω9.4mA @ 2 kΩ0.8mA @ 10 kΩNot availableNot availableNot available
Duration of primary (depolarizing) phase (usec)50-250uS60-350uS
Pulse Duration (usec)200-600uS (uS both phases + 100μS interphase delay)700uS (350 uS both phases)
Frequency (Hz)[ or Rate (pps) ]for multiphasic waveforms only:4-125HzN/A
Symmetrical phases?N/AN/A
Phase duration (include units) state range, if applicable)(both phases, if asymmetrical)N/AN/A
Net Charge(microcoulombs (uC) per pulse) (if zero, state method of achieving zero net charge.)0 μC @ 500 ΩSymmetric, biphasic and leading polarity alternates for each successive pulse.0 μC @ 500 ΩSymmetric, biphasic and leading polarity alternates for each successive pulse.
Maximum Phase Charge (uC)17.5µC @ 500 Ω
Maximum Current Density (mA/cm².r.m.s)0.89mA/cm²(T=1sec)
Ω Where T is the output duration}
Average Current (average absolute value), mA { @ 500 Ω}4.38mAN/A
Average Power Density, (W/ cm²), (using smallest electrode surface area)7.8mW (T=1sec)N/A
Ω Where T is the output duration}
(i.e., pulse trains):
(a) Pulses per burstN/A
(b) Bursts per secondN/A
(c) Burst duration (seconds)N/A
(d) Duty Cycle [ Line (b) x Line (c) ]N/A
e (seconds)N/A
ne (seconds)N/A

{24}------------------------------------------------

.

Page 25 of 33

.

and the comments of the comments of the comments of

.

. . . . . .

{25}------------------------------------------------

Page 26 of 33

{26}------------------------------------------------

.

.

{27}------------------------------------------------

Table 7.8Neurotech PlusModel 439Predicate DeviceStaodyn MaxK061516PredicateMediTensK082011
Mode or Program NamePlan 9 (TENS)Program E (Highest output)Program 3
99Hz, 150µS
Waveform (e.g pulsed monophasic, biphasic)Pulsed, symmetric, biphasicPulsed, symmetric, biphasicPulsed, symmetric, biphasic
Shape (e.g rectangular, spike, rectified sinusoidal)Rectangular, with interphaseintervalRectangularRectangular, with interphaseinterval
Maximum Output Voltage (volts)(+/- 10%)35.0V @ 500 Ω75.0V @ 2 kΩ70.3V @ 10 kΩ30.0V @ 500 ΩNot availableNot available37.5V @ 500 Ω70.0V @ 2 kΩ71.8V @ 10 kΩ
Maximum Output Current (specify units)(+/- 10%)70.0mA @ 500 Ω37.5mA @ 2 kΩ7.0mA @ 10 kΩ60.0 mA @ 500 ΩNot availableNot available75.0mA @ 500 Ω35.0mA @ 500 Ω7.0mA @ 10 kΩ
RMS Output Voltage (volts)(+/- 10%)6.8V @ 500 Ω14.6V @ 2 kΩ7.8V @ 10 kΩ8.9V @ 500 ΩNot availableNot available6.5V @500Ω14.2V @2kΩ4.95V@10kΩ
RMS Output Current (specify units)(+/- 10%)13.6mA @ 500 Ω7.3mA @ 2 kΩ0.8mA @ 10 kΩ17.8mA @ 500 ΩNot availableNot available13mA@500Ω7.1mA@2kΩ0.5mA@10kΩ
Duration of primary (depolarizing) phase (usec)150-250uS60-350uS150uS
Pulse Duration(usec)400-600uS (uS both phases +100µS interphase delay)700uS (350 uS both phases)400µSBoth phases
Frequency (Hz) [ or Rate (pps) ]For multiphasic waveformsonly:4-125Hz80 to 125 Hz4 to 99 Hz
Symmetrical Phases?Phase DurationYes100 to 250µSYes60-350uSYes100 to 150uS
Net Charge(microcoulombs (uC) per pulse) (if zero, state method ofachieving zero net charge.)0 µC @ 500 ΩSymmetric, biphasic andleading polarity alternates foreach successive pulse.0 µC @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for eachsuccessive pulse.0 µC @ 500 ΩSymmetric, biphasic andleading polarity alternates foreach successive pulse.
Maximum Phase Charge (uC)10.6μC @ 500 Ω9 to 21 µC @ 500 Ω7.5 to 11.25 μC @ 500 Ω
Table 7.8Neurotech PlusModel 439Predicate DeviceStaodyn MaxK061516PredicateMediTensK082011
Maximum Current Density (mA/cm², r.m.s){@ 500 Ω Where T is the output duration)0.70mA/cm²(T=1sec)Program EUsing 2" square electrode=17.8 mA/25 cm²= 0.71mA/cm²0.672mA/cm²5cm round electrode)
Maximum Average Current (average absolute value),mA (@ 500 Ω)2.66mA5.3mA2.25mA
Maximum Average Power Density, (W/ cm²), (using smallest electrodeconductive surface area) {@ 500 Ω Where T is the output duration}4.7mW (T=1sec)6.3 mW/cm²Program EUsing 2" square electrode4.4mW/cm²50mm round electrode
Burst Mode (i.e., pulse trains):(a) Pulses per burstN/ABurst mode is simply analternate frequency-pulse widthcombination. There are actuallyno intermittent bursts8N/ABurst mode is simply analternate frequency-pulse widthcombination. There are actuallyno intermittent bursts
(b) Bursts per second2
(c) Burst duration (seconds)0.09
(d) Duty Cycle [ Line (b) x Line (c) ].18
ON Time (seconds)N/AN/AN/A
OFF Time (seconds)N/AN/AN/A
Additional Features (specify, if applicable)N/AN/AN/A
Table 7.9
Neurotech PlusModel 440Predicate DeviceStaodyn MaxK061516PredicateMediTensK082011
Mode or Program NamePlan 9 (TENS)Program E (Highest output)Program 3 (99Hz, 150μS)
Waveform (e.g pulsed monophasic, biphasic)Pulsed, symmetric, biphasicPulsed, symmetric, biphasicPulsed, symmetric, biphasic
Shape (e.g rectangular, spike, rectified sinusoidal)Rectangular, with interphase intervalRectangularRectangular, with interphase interval
Maximum Output Voltage (volts)(+/-10%)35.0V @ 500 Ω30.0V @ 500 Ω37.5V @ 500 Ω
75.0V @ 2 ΚΩNot available70.0V @ 2 ΚΩ
70.3V @ 10 ΚΩNot available71.8V @ 10 ΚΩ
Maximum Output Current (specify units) (+/- 10%)70.0mA @ 500 Ω60.0 mA @ 500 Ω75.0mA @ 500 Ω
37.5mA @ 2 ΚΩNot available35.0mA @ 500 Ω
7.0mA @ 10 ΚΩNot available7.0mA @ 10 ΚΩ
RMS Output Voltage (volts)(+/-10%)6.8V @ 500 Ω8.9V @ 500 Ω6.5V@500Ω
14.6V @ 2 ΚΩNot available14.2V@2ΚΩ
7.8V @ 10 ΚΩNot available4.95V@10ΚΩ
RMS Output Current (specify units)(+/- 10%)13.6mA @ 500 Ω17.8mA @ 500 Ω13mA@500Ω
7.3mA @ 2 ΚΩNot available7.1mA@2kΩ
0.8mA @ 10 ΚΩNot available0.5mA@10ΚΩ
Duration of primary (depolarizing) phase (usec)150-250uS60-350uS150uS
Pulse Duration (usec)400-600uS (uS both phases + 100μδinterphase delay)700uS (350 uS both phases)400μS
Frequency (Hz)[ or Rate (pps) ]4-125Hz80 to 125 Hz4 to 99 Hz
For multiphasic waveformsonly:N/AYesYes
Net Charge(microcoulombs (uC) per pulse)(if zero, state method of achieving zero net charge.)100 to 250μς0 μC @ 500 ΩSymmetric, biphasic and leading polarityalternates for each successive pulse.60-350uS0 μC @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for each successivepulse100 to 150uS0 μ€ @ 500 ΩSymmetric, biphasic and leadingpolarity alternates for each successivepulse.
Maximum Phase Charge (uC)10.6μC @ 500 Ω9 to 21 μC @ 500 Ω7.5 to 11.25 µε @ 500 Ω
Maximum Current Density (mA/cm², r.m.s){@ 500 Ω Where T is the output duration}0.70mA/cm²(T=1sec)Program EUsing 2" square electrode=17.8 mA/25 cm²0.672mA/cm²5cm round electrode)
Table 7.9Neurotech PlusModel 440Predicate DeviceStaodyn MaxK061516PredicateMediTensK082011
Maximum Average Current (average absolute value),mA { @ 500 Ω}2.66mA= 0.71mA/cm²5.3mA2.25mA
Maximum Average Power Density. (W/ cm²), (using smallestelectrode conductive surface area){ @ 500 Ω Where T is the output duration)4.7mW (T=1sec)6.3 mW/cm²Program EUsing 2" square electrode4.4mW/cm²50mm round electrode
Burst Mode (i.e.. pulse trains):(a) Pulses per burstN/A8N/A
Burst mode is simply an alternatefrequency-pulse witdh combination.There are actually no intermittent burstsBurst mode is simply an alternatefrequency-pulse witdh combination.There are actually no intermittent bursts
(b) Bursts per second2
(c) Burst duration (seconds)0.09
(d) Duty Cycle[ Line (b) x Line (c)].18
ON Time (seconds)N/AN/AN/A
OFF Time (seconds)N/AN/AN/A
Additional Features (specify, if applicable)N/AN/AN/A
Table 7.10
Neurotech PlusModel 441Predicate DeviceMediStim XPK082011
Mode or Program NamePlan 9 (NMES)
Waveform (e.g pulsed monophasic, biphasic)Pulsed, symmetric, biphasicPulsed, symmetric, biphasic
Shape (e.g rectangular, spike, rectified sinusoidal)Rectangular, with interphase intervalRectangular, with interphase interval
Maximum Output Voltage (volts)(+/-10%)35.0V @ 500 Ω75.0V @ 2 kΩ70.3V @ 10 kΩ37.5V @ 500 Ω70.0V@ 2 kΩ71.8V @ 10 kΩ
Maximum Output Current (specify units)(+/-10%)70.0mA @ 500 Ω37.5mA @ 2 kΩ7.0 mA @ 10 kΩ75.0mA @ 500 Ω35.0mA @ 500 Ω7.0mA @ 10 kΩ
RMS Output Voltage (volts)(+/-10%)8.5V @ 500 Ω18.3V @ 2 kΩ7.0V @ 10 kΩ6.5 V @ 500 Ω12.1V @ 2 kΩ7.1V @ 10 kΩ
RMS Output Current (specify units)(+/- 10%)17.1mA @ 500 Ω9.1mA @ 2 kΩ0.7mA @ 10 kΩ13.0mA @ 500 Ω6.0mA @ 2 kΩ0.7mA @ 10 kΩ
Duration of primary (depolarizing) phase (usec)300uS150-300uS
Pulse Duration (usec)400uS (100 uS both phases + 100μSinterphase delay)400-700us (100 us both phases + 100µSinterphase delay)
Frequency (Hz) [ or Rate (pps) ]10-99Hz2-100Hz
For multiphasic waveforms only:Symmetrical phases?N/AN/A
Phase duration (include units) staterange, if applicable)(both phases, ifasymmetrical)N/AN/A
Net Charge(microcoulombs (uC) per pulse) (if zero, state method of achievingzero net charge.)0 μC @ 500 ΩSymmetric, biphasic and leading polarityalternates for each successive pulse.0 μC @ 500 ΩSymmetric, biphasic and leading polarityalternates for each successive pulse.
Maximum Phase Charge (uC)21.0μC @ 500 Ω21.0µC @ 500 Ω
Maximum Current Density (mA/cm², r.m.s){@ 500 Ω Where T is the output duration)Maximum Average Current (average absolute value), mA {@ 500 Ω)Maximum Average Power Density, (W/ cm² ), (using smallest electrode conductive surface area)@ 500 Ω Where T is the output duration)
0.87mA/cm²(T=1sec)4.16mA7.4mW (T=1sec)Not available
Burst Mode (i.e., pulse trains):(a) Pulses per burstN/AN/A
(b) Bursts per secondN/AN/A
(c) Burst duration (seconds)N/AN/A
(d) Duty Cycle [Line (b) x Line (c)]N/AN/A
ON Time (seconds)N/AN/A
OFF Time (seconds)N/AN/A

{28}------------------------------------------------

·

:

Page 29 of 33

{29}------------------------------------------------

Page 30 of 33

.

{30}------------------------------------------------

.

{31}------------------------------------------------

:

.

{32}------------------------------------------------

{33}------------------------------------------------

Image /page/33/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, rendered in a stylized, minimalist design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Biomedical Research, LTD c/o Ms. Anne-Marie Keenan Quality & Regulatory Engineer Parkmore Business Park West Galway, Ireland

Re: K112258

Trade/Device Name: Neurotech Plus, Type 413 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: GZJ, NYN, IPF Dated: November 30, 2011 Received: December 5, 2011

Dear Ms. Keenan:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bection of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to tegans manative possible in the Medical Device Artic commerce prior to May 28, 1776, the encoment and of the Federal Food. Drug, devices that have been recuire approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require approvine the general controls provisions of the Act. The You may, merciole, market the device, basjon to the most for annual registration, listing of general controls provisions of the rict interest and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Flease note: CDICT does not overaliate and more be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see acove) interior major regulations affecting your device can be may be subject to additional controls: "Existing migral" - " " " " a ddition, FDA may found in the Code of I caccai resgments concerning your device in the Federal Register.

JAN - 9 2012

{34}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours,

Ja

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{35}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Neurotech Plus, Type 413

Indications for Use:

The Neurotech Plus device is indicated for the following:

In combination Neuromuscular Electrical Stimulation (NMES) and Transcutaneous Electrical Nerve Stimulation (TENS) modes, models 431, 432, 433, 434, 436, 439 & 440:

NMES IndicationsTENS Indications
Maintain or increase the range of motionSymptomatic relief and management of chronic, intractable pain
Prevention or retardation of disuse atrophyRelief of pain associated with arthritis
Re-educate musclesAdjunctive treatment in the management of post-surgical and post-trauma pain
Relax muscle spasmsAdjunctive therapy in reducing the level of pain and symptoms associated with osteoarthritis of the knee (models 431, 432 and 433 only)
Increase local blood circulation
Prevention of venous thrombosis of the calf muscles immediately after surgery

In Neuromuscular Electrical Stimulation (NMES) only mode, models 437 & 441:

NMES Indications
Maintain or increase the range of motion
Prevention or retardation of disuse atrophy
Re-educate muscles
Relax muscle spasms
Increase local blood circulation
Prevention of venous thrombosis of the calf muscles immediately after surgery

In Transcutaneous Electrical Nerve Stimulation (TENS) only mode, model 438:

TENS Indications
Symptomatic relief and management of chronic, intractable pain
Relief of pain associated with arthritis
Adjunctive treatment in the management of post-surgical and post-trauma pain
Prescription UseX
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K112258

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).