K Number
K112156
Date Cleared
2011-10-17

(82 days)

Product Code
Regulation Number
882.5550
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CODMAN Certas Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. The CODMAN Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting.

Device Description

The Certas Programmable Valve System is designed for shunting cerebrospinal fluid (CSF) for the treatment of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular pressure and drainage of CSF. The Therapy Management System allows the user to non-invasively adjust the opening pressure setting before and after implant. The Certas Programmable Valve will be available with and without the legally marketed SIPHONGUARD CSF Control Device and the Codman BACTISEAL Catheters. The Certas Programmable Valve includes the same legally marketed accessories that are available with the Hakim Programmable Valve. The Certas Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. The Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting. The SIPHONGUARD device is designed to reduce the potential hazards of excessive lowering of intraventricular pressure (with respect to atmospheric pressure) when a patient is in an upright position. The BACTISEAL catheters are intended for use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid is indicated.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and study information for the Codman Certas Programmable Valve System:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text describes pre-clinical testing for the Codman Certas Programmable Valve System. While it doesn't explicitly state quantitative "acceptance criteria" values in a table, it lists the types of tests performed and the general outcome.

Acceptance Criterion (Test Performed)Reported Device Performance
Leak TestingPerforms as intended, safe and effective.
Pressure-Flow TestingPerforms as intended, safe and effective.
Overpressure TestingPerforms as intended, safe and effective.
Bursting Pressure TestingPerforms as intended, safe and effective.
Dynamic Break Strength TestingPerforms as intended, safe and effective.
Long-Term Stability TestingPerforms as intended, safe and effective.
Puncture TestingPerforms as intended, safe and effective.
MRI Translational Attraction (ASTM F2052)MR Conditional in a 3 Tesla MRI system.
MRI Torque (ASTM F2213)MR Conditional in a 3 Tesla MRI system.
MRI Artifact (ASTM F2119)MR Conditional in a 3 Tesla MRI system.
MRI Heating (ASTM F2182)MR Conditional in a 3 Tesla MRI system.
Biocompatibility (ISO 10993-1, FDA G95-1)All materials meet requirements.
Sterilization (ISO 17665 Part 1 & 2, 11737)Sterility assurance level of 10^-6^ achieved.
Overall Substantial EquivalenceThe Certas Valve and Certas Therapy Management System are substantially equivalent to predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide specific sample sizes for the pre-clinical tests or details on the provenance of data (e.g., country of origin, retrospective/prospective). The testing appears to be primarily bench or laboratory-based, as is common for pre-clinical performance evaluations of medical devices.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not mention the use of experts to establish a "ground truth" for the test set. The tests described are primarily objective, quantitative engineering and materials tests (e.g., pressure-flow, strength, MRI compatibility, biocompatibility, sterility) rather than clinical evaluations requiring expert interpretation of outcomes.

4. Adjudication Method for the Test Set

Not applicable. As noted above, the tests are objective pre-clinical evaluations, not interpretive clinical assessments that would require an adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document describes pre-clinical (bench) testing to demonstrate substantial equivalence based on performance characteristics, materials, and principles of operation. It does not mention any clinical studies, MRMC studies, or human reader performance evaluations.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. The device is a physical medical implant and associated management system, not an AI algorithm. The performance evaluation is for the physical device itself.

7. Type of Ground Truth Used

The "ground truth" for the pre-clinical tests would be the established scientific and engineering standards and methodologies defined by the cited ISO and ASTM standards. For example, for sterility, the ground truth is a sterility assurance level of 10^-6^. For MRI compatibility, the ground truth is the criteria defined by ASTM F2503 for "MR Conditional."

8. Sample Size for the Training Set

Not applicable. The device is a physical medical device, not an AI/ML algorithm that requires a training set. The "training" of the device itself refers to its design and manufacturing processes based on engineering principles and regulatory standards.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As above, this is a physical medical device, not an AI/ML algorithm. The "ground truth" for its development would be based on established engineering principles, material science, and prior knowledge from predicate devices and relevant standards.

{0}------------------------------------------------

K112156

OCT 1 7 2011

510(k) Summary

Medos International Sarl Company Name: Chemin-Blanc 38 Company Address: CH 2400 LeLocle, Switzerland Jocelyn Raposo Contact Person: Senior Regulatory Affairs Specialist (508) 828-3421 Phone: JRaposo@its.jnj.com Email: July 26, 2011 Date: Name of the Device: Codman Certas Programmable Valve System Propriety / Trade Name: Codman Certas Therapy Management System Hydrocephalus Shunt System Common Name: Class II (JXG) Classification: Central Nervous System Fluid Shunt and Components Legally Marketed Predicate Devices: Codman Hakim Programmable Valve K974739 -Codman Hakim Shunt System K091346 -Codman SIPHONGUAFID CSF Control Device K992173 -Codman Bactiseal Catheters K003322 -Medtronic PS Medical® Strata® Type® Valve K060681 -Medtronic PS Strata® Valves and Handtools K040943 -Sophysa Polaris K090342 -The Certas Programmable Valve System is designed for shunting Device Description: cerebrospinal fluid (CSF) for the treatment of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular pressure and drainage of CSF. The Therapy Management System allows the user to non-invasively adjust the opening pressure setting before and after implant.

p.1/3

{1}------------------------------------------------

K112156

The Certas Programmable Valve will be available with and without the legally marketed SIPHONGUARD CSF Control Device and the Codman BACTISEAL Catheters.

The Certas Programmable Valve includes the same legally marketed accessories that are available with the Hakim Programmable Valve.

The Certas Programmable Valve is an implantable device that Intended Use: provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

The Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting.

The SIPHONGUARD device is designed to reduce the potential hazards of excessive lowering of intraventricular pressure (with respect to atmospheric pressure) when a patient is in an upright position.

The BACTISEAL catheters are intended for use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid is indicated.

Technological Comparison:

The Certas Programmable Valve is substantially equivalent to the predicate Hakim Programmable Valve, and the Medtronic Strata and Polaris Sophysa Valves and Adjustment Tool Kits. Substantial equivalence is based on intended use, performance characteristics, materials, and principles of operation.

Preclinical testing was performed to demonstrate that the Certas Performance Data: Programmable Valve and Therapy Management System performs as intended and is safe and effective. Testing was conducted in accordance with ISO 7197 and included Leak, Pressure-Flow, Overpressure, Bursting Pressure, Dynamic Break Strength, Long-Term Stability and Puncture testing. Magnetic Resonance Imaging (MRI) testing was performed according to ASTM F2052, F2213, F2119, and F2182 for Translational Attraction, Torque, Artifact, and Heating, respectively. The MRI testing demonstrated that the

p.2/3

{2}------------------------------------------------

K112156

Certas valve is MR Conditional in a 3 Tesla MRI system according to ASTM F2503. Biocompatibility was assessed and all of the materials meet the requirements of ISO 10993-1 and FDA Blue Book Memorandum G95-1. Sterilization testing was performed according to ISO 17665 Part 1 and 2 and ISO 11737 demonstrating that a sterility assurance level of 10° was achieved. The test data establish that the Certas Valve and the Certas Therapy Management System are substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN" around the left side. To the right of the seal is a stylized symbol that resembles an eagle or bird with three curved lines representing its wings or feathers.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OGT 1 7 2011

Codman & Shurtleff, Inc. c/o Jocelyn Raposo Senior Regulatory Affairs Specialist 325 Paramount Drive Raynham MA 02767-0350

Re: K112156

Trade/Device Name: Codman Certas Programmable Valve System, Codman Certas Therapy Management System

Regulation Number: 21 CFR 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: Class II Product Code: JXG Dated: July 26, 2011 Received: July 27, 2011

Dear Mr. Raposo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalYroblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

E. K. Etem
Malvina B. Eydelman, M.D.

Malvina B. Eydelman, M Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

K112156

Device Name:

CODMAN Certas Programmable Valve CODMAN Certas Therapy Management System

Indications For Use:

The CODMAN Certas Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

The CODMAN Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

JOE HUTTER

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

Page 1 of

510(k) Number K112156

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).