K Number
K040943
Date Cleared
2004-06-15

(64 days)

Product Code
Regulation Number
882.5550
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Strata® Valve is a shunt component designed to provide continued Cerebrospinal Fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design allows the physician to non-invasively adjust valve pressure/flow performance level pre- and post-implantation by using an external magnetic Adjustment Tool without the need for radiographic confirmation.

Device Description

After surgical implantation, the Strata Valve provides a non-invasive method to address changing patient needs in the management of hydrocephalus. The unique valve design allows the physician to adjust the pressure/flow performance level by using a magnetic Adjustment Tool included with the Strata® Valve adjustment kit without the need for radiographic confirmation.

The Strata® Handtools, include the Locator Tool, Indicator Tool and Adjustment Tool. Ail three tools are required to set or change the pressure/performance level setting of the Strata® Valve.

AI/ML Overview

Acceptance Criteria and Study Details for Medtronic PS Medical Strata® Valves

Here's a breakdown of the acceptance criteria and the study details as provided in the 510(k) summary for the Medtronic PS Medical Strata® Valves and Handtools:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document presents acceptance criteria implicitly through the reported performance rather than explicit pre-defined targets. The performance metrics focus on the accuracy of the Indicator Tool in correlating with the x-ray confirmed pressure/performance level setting.

Metric (Implied Acceptance Standard)Reported Device Performance
Correlation between Indicator Tool pressure/performance level setting and x-ray at Discharge98.6%
Correlation between Indicator Tool pressure/performance level setting and x-ray for ALL IN CORRES Adjustment Data98.8%

Note: The table in the original document has some unreadable or garbled text under "Concellential" and "DJT". Only the clear percentage values have been included.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document only mentions "Discharge Data" and "ALL IN CORRES Adjustment Data," which refer to categories of observations rather than a specific number of patients or evaluations.
  • Data Provenance: The study is referred to as "clinical study data," implying prospective data collection during patient follow-up. The country of origin is not specified, but given the 510(k) submission to the FDA, it is likely that parts or all of the clinical data were collected in the US or in compliance with US regulatory standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the submitted 510(k) summary. The ground truth for valve setting is explicitly stated as "x-ray," but details on who interpreted these x-rays or their qualifications are absent.

4. Adjudication Method for the Test Set

This information is not provided in the submitted 510(k) summary. Given the reliance on "x-ray" as the ground truth, it's possible that a single, definitive x-ray interpretation was used, or if multiple were involved, the adjudication method is not mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was it done? No. The study described focuses on the correlation between the device's Indicator Tool and x-ray confirmation, not on comparing human reader performance with and without AI assistance. The device in question is a physical medical device (valve and handtools) for hydrocephalus management, not an AI diagnostic tool.

6. Standalone (Algorithm Only) Performance Study

  • Was it done? No. The device is a mechanical shunt system with external adjustment tools. The "performance" being evaluated is the accuracy of the adjustment tool's indication against an x-ray confirmation of the physical valve's setting. There is no AI algorithm involved for standalone performance evaluation.

7. Type of Ground Truth Used

  • Ground Truth: X-ray confirmation of the valve's pressure/performance level setting. The summary explicitly states: "Comparison of the Indicator Tool pressure/performance level setting and corresponding x-ray at time of Discharge and during follow-up."

8. Sample Size for the Training Set

This information is not provided as the device is not an AI/ML algorithm that typically requires a distinct "training set." The clinical data described pertains to the performance evaluation rather than the training of an algorithm.

9. How the Ground Truth for the Training Set Was Established

This is not applicable as there is no mention of a training set for an AI/ML algorithm in the context of this device. The development and verification of the physical valve and tools would likely involve engineering tests, bench testing, and potentially pre-clinical studies, but these are distinct from "training an algorithm." The clinical study data mentioned refers to the human-clinical performance validation.

{0}------------------------------------------------

JUN 1 5 2004

510(k) Summary

This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

Establishment Registration Number:2021898
Address of Manufacturer:Medtronic Neurosurgery125 Cremona DriveGoleta CA, 93117Telephone: (805) 968-1546 ext. 1770Fax: (805) 968-9336
Contact Person:Jeffrey Henderson
Date:April 8, 2004
Trade or Proprietary Name:Medtronic PS Medical Strata ® Valves andHandtools
Common usual or Classification Name:Central Nervous System Fluid Shunts andComponents. 21 CFR 882.5550

Description:

After surgical implantation, the Strata Valve provides a non-invasive method to address changing patient needs in the management of hydrocephalus. The unique valve design allows the physician to adjust the pressure/flow performance level by using a magnetic Adjustment Tool included with the Strata® Valve adjustment kit without the need for radiographic confirmation.

The Strata® Handtools, include the Locator Tool, Indicator Tool and Adjustment Tool. Ail three tools are required to set or change the pressure/performance level setting of the Strata® Valve.

Intended Use:

The Strata® Valve is a shunt component designed to provide continued Cerebrospinal Fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design allows the physician to non-invasively adjust valve pressure/flow performance level pre- and post-implantation by using an external magnetic Adjustment Tool without the need for radiographic confirmation.

Predicate Device Identification:

The Strata® Valve is substantially equivalent to the Medtronic PS Medical Strata® Valve based on the same intended use, materials, design and dimensions.

K040943 1/2

{1}------------------------------------------------

K040943 2/2

510(k) Summary

Device Testing:

The Strata® Valve is substantially equivalent to the Medical Strata® Valve based The Othata "Varo is oubota.histing, biocompatibility studies, and clinical study data.

Comparison of the Indicator Tool pressure/performance level setting and corresponding x-ray at time of Discharge and during follow-up, yielded the following results:

Discharge Data98.6%
ALL IN CORRESAdjustment Data98.8%
Concellentialית DJTAnd An A C C C C C C . C . C . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 . 0 .SERE ENGLE EN LEE MERICAN

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the words "Public Health Service" in bold, black font. The words are aligned horizontally and are of similar size. The background is plain white.

Image /page/2/Picture/2 description: The image is a black and white emblem for the U.S. Department of Health & Human Services. The emblem features a stylized eagle with three curved lines representing its wings. The eagle is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. The text is arranged around the upper half of the circle.

JUN 1 5 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Jeffrey Henderson Vice President, Quality & Regulatory Affairs Medtronic Neurosurgery 125 Cremona Drive Goleta, California 93117-5500

Re: K040943

Trade/Device Name: PS Medical Strata® Valve Regulation Number: 21 CFR 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: II Product Code: JXG Dated: April 9, 2004 Received: April 20, 2004

Dear Mr. Henderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Jeffrey Henderson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

2. Modified Indications for Use Statement

Indications for Use

510(k) Number (if known): K040943

Device Name: PS Medical Strata® Valve

Indications For Use:

"The Strata® Valve is a shunt component designed to provide continued Cerebrospinal Fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design allows the physician to non-invasively adjust valve pressure/flow ourfly - The don't and post-implantation by using an external magnetic Adjustment Tool without the need for radiographic confirmation."

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of

iriam C. Provost

(Division Sign-Off) Division of General, Restorative. and Neurological Devices

510(k) Number_ K 040943

004

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).