K Number
K112098
Device Name
TBD
Manufacturer
Date Cleared
2011-10-19

(89 days)

Product Code
Regulation Number
886.4390
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LensAR-fs Laser System is indicated for anterior capsulotomy and laser phaco fragmentation during cataract surgery. The anterior capsulotomy and laser phaco fragmentation procedures may be performed either individually or consecutively during the same surgery.

Device Description

The predicate LensAR Laser System is an ophthalmic surgical laser that has been cleared for use in anterior capsulotomy in cataract surgery (K090633) and anterior capsulotomy and laser phaco fragmentation performed individually or consecutively during the same surgery (K102727). In new laser device with modification to the pulse width of the laser (LensAR-fs Laser System) is intended for use in the same indication as cleared in K102727. The system employs a mode-locked Yb:YAG laser which generates a high frequency series of ultrashort (1500 femtosecond), low energy pulses at a wavelength of 1030, my The system is technologically the same as defined for the predicate device, i.e., designed to cut the lens and lens capsular tissue, with minimal collateral damage, by the mechanisms of plasma mediated ablation and photodisruption of targeted itssue at the beam focus. The precision capsulotomy and lens fragmentation is generated by computercontrolled scanning of the position of the laser beam focus in three dimensions at the target location. The laser energy is delivered to the eye through a disposable, patient interface device that consists of two separately packaged parts: a commercially wailable Suction Ring with Spring Loaded Syringe to fixate the eye and an Index Matching Eye Docking (IMED) device designed to match the refractive index of the comea to optimize beam targeting accuracy.

AI/ML Overview

The LensAR-fs Laser System is an ophthalmic surgical laser indicated for anterior capsulotomy and laser phaco fragmentation during cataract surgery. The acceptance criteria and the study proving the device meets these criteria are detailed below:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
Accuracy and Reproducibility of Capsulotomy Incisions (ex vivo)The device produces anterior capsulotomy incision patterns that are accurate and predictable in size and shape.
Equivalence of Geographical Cutting Patterns (human donor ocular globes for phaco fragmentation)Equivalence of the geographical cutting patterns to those established for the predicate LensAR Laser was established. The fs laser causes the lens to be divided, employing linear cuts, into a number of radial sections, similar in shape to those created by the conventional manual "phaco chop" surgical technique, and then further divided by applying a set of concentric cylindrical cuts through the radial sections.
Ocular Safety EquivalenceAn optical radiation hazard analysis by Dr. Sliney confirms the continuing ocular safety equivalence to the predicate device detailed in 510(k) K102727.
Technological Characteristics Comparability (with predicate devices)The pulse width, minor software and hardware changes are incorporated while remaining consistent with the technological characteristics of the predicate device LensAR Laser. The technical characteristics of the LensAR-fs Laser fall between those of the LensAR Laser and the LenSx Laser... and raise no new issues with respect to the safety and effectiveness of the device. The technology and mechanism of laser-tissue interaction are similar to other ultrashort pulse lasers identified in the table above. The differences are insignificant and do not affect safety or effectiveness.
Indication for Use Statement Consistency (with predicate devices)The indication for use statement for anterior capsulotomy and laser phaco fragmentation for the LLS-fs is the same as that of the predicate devices.
Absence of New Issues of Safety or Effectiveness (compared to predicate devices)The activities and information provided do not identify any new issues of safety or effectiveness.

The study aimed to demonstrate substantial equivalence to predicate devices (LensAR Laser System K090633 and K102727, and LenSx 550 Laser System K094052 and K082947) by showing similar performance characteristics and no new safety or effectiveness concerns despite changes in pulse width and minor hardware/software modifications.

2. Sample Size and Data Provenance for the Test Set

  • Ex vivo porcine eyes: The exact number is not specified, but the testing was done on "ex vivo porcine eyes."
  • Human donor ocular globes: The exact number is not specified, but evaluations were done in "human donor ocular globes."
  • Data Provenance: The studies were preclinical and conducted on ex vivo biological material (porcine eyes, human donor globes). The country of origin is not specified, nor is whether the data was retrospective or prospective, though preclinical studies of this nature are typically prospective in design.

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the number or qualifications of experts used to establish a "ground truth" for the test set in the traditional sense of clinical studies with expert image interpretation. The evaluation was primarily based on objective measurements and observations of the laser's physical effects.

  • For Ocular Safety: An optical radiation hazard analysis was performed by Dr. Sliney. His specific qualifications are not detailed beyond "Dry Sliney confirms."

4. Adjudication Method for the Test Set

No explicit adjudication method (e.g., 2+1, 3+1) is mentioned, as the studies were preclinical and focused on physical and technical performance rather than subjective interpretation by multiple human readers.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done for this device. The submission explicitly states that "thus no new clinical evaluations were required," relying instead on pre-clinical data and comparison to predicate devices. Therefore, no effect size for human readers improving with AI vs. without AI assistance is provided or relevant to this submission.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

The device is a laser system, not an AI algorithm. Its performance is inherent to its physical operation, and it is designed for use by a human surgeon. Therefore, a "standalone algorithm only" performance study is not applicable. The device's performance was evaluated based on the physical outcomes of its laser application.

7. Type of Ground Truth Used

The ground truth was established through:

  • Objective Measurements/Observations: For capsulotomy incisions (e.g., accuracy, predictability in size and shape) and phaco fragmentation patterns (e.g., division into radial and concentric sections).
  • Technical Comparability: Comparison of the device's technical characteristics (e.g., pulse width, laser-tissue interaction) with established predicate devices.
  • Optical Radiation Hazard Analysis: Expert assessment of ocular safety.

8. Sample Size for the Training Set

This information is not applicable. The device is a physical laser system, not an artificial intelligence or machine learning model that requires a "training set" in the computational sense.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

OCT 1 9 2011

KILOGGE

510(K) PREMARKET NOTIFICATION SUMMARY OF SAFETY AND EFFECTIVENESS (SS&E)

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).

SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED

  • a. Applicant: LensAR. Inc. 2800 Discovery Drive, Suite 100 Orlando, FL 32826 Tel: (407) 641-4889 Fax: (407) 386-7228
  • b. Contact Person: S.K. McGarvey Regulatory Consultant skmcgg@aol.com Cell: (702) 524-3701 Fax: (407) 386-7228

DATE SUMMARY PREPARED: July 20, 2011

NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME

a. Trade/Proprietary Name:LensAR-fs Laser System
b. Common/Usual Name:LensAR-fs Laser System
c. Classification Name:Ophthalmic Laser, Phacofragmentation System
d. Classification Code(s):21 CFR 886.4390 OOE; 21 CFR 886.4670 HQC

PREDICATE DEVICES

510(K) #TRADE NAMEMANUFACTURER
K090633 andK102727LensAR Laser SystemLensAR, Inc.
K094052 andK082947LenSx 550 Laser System Model 550LenSx Lasers, Inc.

DEVICE DESCRIPTION

The predicate LensAR Laser System is an ophthalmic surgical laser that has been cleared for use in anterior capsulotomy in cataract surgery (K090633) and anterior capsulotomy and laser phaco fragmentation performed individually or consecutively during the same surgery (K102727). In new laser device with modification to the pulse width of the laser (LensAR-fs Laser System) is intended for use in the same indication as cleared in K102727.

{1}------------------------------------------------

The system employs a mode-locked Yb:YAG laser which generates a high frequency series of ultrashort (1500 femtosecond), low energy pulses at a wavelength of 1030, my The system is technologically the same as defined for the predicate device, i.e., designed to cut the lens and lens capsular tissue, with minimal collateral damage, by the mechanisms of plasma mediated ablation and photodisruption of targeted itssue at the beam focus. The precision capsulotomy and lens fragmentation is generated by computercontrolled scanning of the position of the laser beam focus in three dimensions at the target location. The laser energy is delivered to the eye through a disposable, patient interface device that consists of two separately packaged parts: a commercially wailable Suction Ring with Spring Loaded Syringe to fixate the eye and an Index Matching Eye Docking (IMED) device designed to match the refractive index of the comea to optimize beam targeting accuracy.

STATEMENT OF INTENDED USE

The LensAR-fs Laser System is indicated for anterior capsulotomy and laser phaco fragmentation during cataract surgery. The anterior capsulotomy and laser phaco fragmentation procedures may be performed either individually or consecutively diring the same surgery.

TECHNOLOGICAL CHARACTERISTICS COMPARISON

The LensAR-fs Laser System is consistent with the technological characteristics of the predicate device LensAR Laser for use in Anterior Capsulotomy (K090633) and in Laser Phaco Fragmentation (K102727). This LensAR-fs Laser surgical device incorporates a change in pulse width and minor software and hardware changes for use in the same Indications for Use: Anterior Capsulotomy and Laser Phaco Fragmentation during cataract surgery.

The LensAR-fs Laser System is of comparable type and is substantially equivalent to the following predicate devices:

510(k) NumberClearance DateDevice Description
K102727Anterior Capsulotomy &Laser Phaco Fragmentation03/16/2011LensAR Laser System - Laser Technologyand Indications for Use predicate device
K090633Anterior Capsulotomy05/13/2010LensAR Laser System - Laser Technologyand Indications for Use predicate device
K094052Anterior Capsulotomy &Lens Fragmentation04/23/2010LenSx 550 Laser System - Laser Technologyand Indications for Use predicate device
K082947Anterior Capsulotomy08/14/2009LenSx 550 Laser System - Laser Technologyand Indications for Use predicate device

{2}------------------------------------------------

  • The activities used to evaluate the LensAR-fs Laser System (LLS-fs) and the . information and reports provided in this 510(k) submission do not identify any new issues of safety or effectiveness. The optical radiation hazard analysis done by Dry Sliney confirms the continuing ocular safety equivalence to the predicate device detailed in 510(k) K102727 for the predicate device.
  • The LensAR-fs Laser technology and mechanism of laser-tissue interaction are . similar to other ultrashort pulse lasers identified in the table above.
  • The indication for use statement for anterior capsulotomy and laser phaco . fragmentation for the LLS-fs is the same as that of the predicate devices detailed in the table above.
  • The differences between the LLS-fs and the predicate devices are insignificant and do . not affect the safety or effectiveness of the device.

BRIEF SUMMARY OF PRECLINICAL AND CLINICAL PERFORMANCE TEST RESULTS

The performance data supporting substantial equivalence of the LensAR-fs Laser System to the predicate LensAR device are summarized as follows:

Summary of Pre-Clinical Testing

Testing and analyses performed included accuracy and reproducibility of capsulotomy incisions in ex vivo porcine eyes. The data demonstrated that the Lers on explanot produces anterior capsulotomy incision patterns that are accurate and predictable in size and shape, which confirm equivalence to those achieved with the predicate LensAR Laser device (K090633 and K102727).

Additional evaluations were done in human donor ocular globes to establish the equivalence of the geographical cutting patterns to those established for the predicate LensAR Laser in 510(k) K102727. In the same manner, as demonstrated by the LensAR predicate device use, the fs laser causes the lens to be divided, employing linear cuts, into a number of radial sections, similar in shape to those created by the conventional manual "phaco chop" surgical technique, and then further divided by applying a set of concentric cylindrical cuts through the radial sections.

The technical characteristics of the LensAR-fs Laser fall between those of the LensAR Laser and the LenSx Laser as seen in the comparative analysis supplied in the supporting documentation and raise no new issues with respect to the safety and effectiveness of the device in its intended use. Thus no new clinical evaluations were required.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

LensAR, Inc. c/o Ms. Shirley McGarvey Regulatory Consultant 2800 Discovery Drive, Suite 100 Orlando, FL 32826

Re: K112098

Trade/Device Name: LensAR-fs Laser System for Anterior Capsulotomy and Laser Phaco Fragmentation Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: OOE, HOC Dated: September 27, 2011 Received: September 29, 2011

Dear Ms. McGarvey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Shirley McGarvey

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,
Kesia Alexander

Image /page/4/Picture/7 description: The image contains a handwritten word "for" in cursive script. The letter "f" has a long, curved descender extending below the baseline, and the "o" and "r" are connected with smooth strokes. The writing appears to be done with a pen or marker, and the lines are clear and legible.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):K112098
Device Name:LensAR-fs Laser System for Anterior Capsulotomy and Laser Phaco Fragmentation
Indications for Use:The LensAR-fs Laser System is indicated for anterior capsulotomy and laser phaco fragmentation during cataract surgery. The anterior capsulotomy and laser phaco fragmentation procedures may be performed either individually or consecutively during the same surgery.

Prescription Use: X (Part 21 CFR 801 Subpart D) And/Or

Over-the-Counter Use: (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Ors. C. Callaway for DLK
Dr. Vivian S. Cox

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K112098

LensAR, Inc. - 510(k) for LensAR-fs Laser System for Anterior Capsulotomy and Laser Phaco Fragmentation

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.