K Number
K111623
Device Name
I-ENDO DUAL
Date Cleared
2012-01-10

(214 days)

Product Code
Regulation Number
872.4200
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

I-Endo Dual is indicated for use by dentists in standard endodontic procedures using endodontic files and endodontic drills.

Device Description

The I ENDO DUAL is an electric motor-driven handpiece intended for root canal preparation procedures in the endodontic industry. It works in both reciprocating mode and continuous rotation.

The motor allows the dentist to program 10 settings of their choice for speed and torque control in rotating mode and to program 10 setting of their choice for speed and angles in reciprocating mode.

The device consists of a control unit with an LCD screen for selection of settings. A foot pedal is connected to the control unit so the dentist can selectively activate and deactivate the motor.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device meets the acceptance criteria for the "I-ENDO DUAL" dental motor.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Standards Met)Reported Device Performance (Compliance)
IEC 60601-1 (General Safety Norms)Complies
IEC 60601-1-2 (EMC)Complies
IEC 55011 (EMC Conducted RF emissions)Complies
IEC 60529 (Degrees of protection by enclosures - IP Code)Complies
IEC 61000-4-2 (EMC Electrostatic Discharge Immunity)Complies
IEC 61000-4-3 (Radiated RF immunity)Complies
IEC 61000-4-4 (Fast Transient Immunity)Complies
IEC 61000-4-5 (Pulse immunity)Complies
IEC 61000-4-6 (Conducted RF immunity)Complies
IEC 61000-4-11 (Supply Voltage Hole Immunity)Complies
ISO 11498:1997 (Dental handpieces, Dental low-voltage electrical motors)Complies

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state a sample size in terms of the number of devices tested for the performance evaluations. It generally refers to "the I-ENDO DUAL" being tested.

The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective. The testing was performed by a "Notified Laboratory."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not provided. The study focuses on compliance with technical standards through laboratory testing rather than expert-derived ground truth.

4. Adjudication Method for the Test Set:

This information is not applicable and is not provided. The testing involves standardized laboratory measurements against defined technical standards, not expert adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

No, an MRMC comparative effectiveness study was not done. The document describes technical performance testing against standards and a comparison to a predicate device based on technical characteristics and intended use, not on human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

No, a standalone (algorithm only) performance study was not done, as the I-ENDO DUAL is a physical dental motor, not an AI algorithm. The performance data relates to the device's adherence to safety and electromagnetic compatibility standards.

7. The Type of Ground Truth Used:

The "ground truth" used in this context refers to the specified technical requirements and limits defined within the cited international and national standards (e.g., IEC 60601-1, IEC 60601-1-2, ISO 11498:1997). The device performance was compared against these predefined physical and electrical safety benchmarks.

8. The Sample Size for the Training Set:

This information is not applicable and is not provided. The I-ENDO DUAL is a physical medical device, not an AI system that requires a "training set."

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable and is not provided, as the device is not an AI system requiring a training set.

{0}------------------------------------------------

Summary of Safety and Effectiveness – "510 (k) Summary"

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMD 1990 and CFR 807.92.

SUBMITTER INFORMATION

a.Company NameSATELEC
b.Company Address17 Avenue Gustave EiffelBP 3021633708 Merignac CedexFrance
c.Company phone+01133556340607
Company fax01133556349292
d.Contact PersonRick RosatiSatelecc/o ACTEON, Inc.124 Gaither Drive, Suite 140MT Laurel, NJ 08054Tel : 800 289-6367 ext 39Fax : 856 222-4726Email :Rick.Rosati@us.acteongroup.com
e.Date Summary PreparedNovember, 11, 2011

{1}------------------------------------------------

DEVICE IDENTIFICATION

a.Trade/Proprietary Names:I-ENDO DUAL
b.Common nameDental Motor
c.Classification NameDental Handpiece andAccessories
d.CFR Number872.4200
e.Device ClassI
f.Product CodeEBW

IDENTIFICATION OF PREDICATE DEVICE

CompanyDevice510(k) No.Date Cleared
Dentsply Int.E3 Torque ControlMotorK10365326 May. 2011

DEVICE DESCRIPTION

The I ENDO DUAL is an electric motor-driven handpiece intended for root canal preparation procedures in the endodontic industry. It works in both reciprocating mode and continuous rotation.

The motor allows the dentist to program 10 settings of their choice for speed and torque control in rotating mode and to program 10 setting of their choice for speed and angles in reciprocating mode.

The device consists of a control unit with an LCD screen for selection of settings. A foot pedal is connected to the control unit so the dentist can selectively activate and deactivate the motor.

{2}------------------------------------------------

INDICATIONS FOR USE

The I ENDO DUAL is a medical device designed for use by dentists for use with dental root canal instruments in continuous rotation with torque control or in reciprocating movement.

SUBSTANTIAL EQUIVALENCE

The I-ENDO DUAL is substantially equivalent to the E3 Torque Control Motor in commercial distribution by Dentsply International.

The fundamental technical characteristics of the I-ENDO DUAL are similar to those of the predicate device and are listed on the comparison chart provided in this 510(k) submission.

TECHNICAL CHARACTERISTICS

I-ENDO DUAL was designed and developed to provide a microprocessor controlled endodontic system with similar performances compared to predicate device. I-ENDO DUAL :

  • is equivalent in functions to the predicate device, ●
  • has adjustable speeds and torque values (related to the reduction rate of the handpiece . selected) that are fully customisable by the end user,
  • . has automatic motor shutdown/autoreverse system which provides to switch off/reverse the motor whenever set torque is reached,
  • has reciprocating movements fully customizable. .

PERFORMANCE DATA

I-ENDO DUAL was tested in accordance with the technical requirements of IEC 60601-1 and IEC 60601-1-2.

All evaluation of the I-ENDO DUAL were performed by Notified Laboratory, and all the results comply to standard listed below.

The conclusions drawn from the performance test are that I-ENDO DUAL comply with:

  • EMC Conducted RF emissions IEC 55011 .
  • Degrees of protection provided by enclosures (IP Code) IEC 60529 .
  • Medical Electrical Equipment Part 2,"Genaral Safety Norms" IEC 60601-1 .
  • IEC 60601-1-2 Medical Electrical Equipment Part 1 EMC .
  • IEC 61000-4-2 EMC Electrostatic Discharge Immunity ●
  • IEC 61000-4-3 Radiated RF immunity .

{3}------------------------------------------------

K111623 4

  • IEC 61000-4-4 Fast Transient Immunity .
  • IEC 61000-4-5 Pulse immunity ●
  • IEC 61000-4-6 Conducted RF immunity ●
  • IEC 61000-4-11 Supply Voltage Hole Immunity .
  • ISO 11498:1997 Dental handpieces Dental low-voltage electrical motors .

510(k) CHECKLIST

This notification contains all information required by 21 CFR 807.87.

A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission.

CONCLUSION AS TO SUBSTANTIAL EQUIVALENCE

I-ENDO DUAL is substantially equivalent to the E3 Torque Control Motor (K103653) based on equivalence of the intended use, the non-clinical performance data and technological characteristics. Performance testing data of the I-ENDO DUAL was compared to the substantially equivalent of the E3 Torque Control Motor (K103653). This included electrical safety, electromagnetic compatibility, and non-clinical performance testing of both hardware and software functions. The I-ENDO DUAL does not raise any new issues of substantially equivalent or performance of the product when compared to the E3 Torque Control Motor (K103653. These test results support a determination of substantial equivalence.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SATELEC-ACTEON Group C/O Mr. Rick Rosati Quality Manager ACETON, Incorporated 124 Gaither Drive, Suite 140 Mt Laurel, New Jersey 08054

JAN 1 0 2012

Re: K111623

Trade/Device Name: I-ENDO DUAL Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EBW Dated: January 2, 2012 Received: January 4, 2012

Dear Mr. Rosati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Rosati

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing ( vopering on set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Indication for use Statement

510(k) Number:

111623 To be assigned by FDA

Device Name:

Endodontic Micromotor

Model:

I-ENDO DUAL

Indications for Use:

I-Endo Dual is indicated for use by dentists in standard endodontic procedures using endodontic files and endodontic drills.


(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use______________________________________________________________________________________________________________________________________________________________

OR

Over-The -Counter Use

(Per 21 CFR 801.109)

Susan Ranne

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K111623

SATELEC / By Dental - Acteon Group I-ENDO DUAL

§ 872.4200 Dental handpiece and accessories.

(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.