K Number
K111395
Date Cleared
2011-08-01

(74 days)

Product Code
Regulation Number
870.1330
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PrimeWire PRESTIGE® Plus Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

Device Description

The PrimeWire PRESTIGE® Plus pressure guide wire is a steerable guide wire with a pressure transducer mounted 3 cm proximal to the tip. The PrimeWire PRESTIGE® Plus guide wire measures pressure when used with the SmartMap®, s5™ Family, and ComboMap® systems. The PrimeWire PRESTIGE® Plus guide wire has a diameter of 0.014" (0.36 mm) and is available in lengths of 185 cm or 300 cm, and also in straight or pre-shaped tips. The PrimeWire PRESTIGE® Plus guide wire is packaged preconnected to the connector with a torque device to facilitate navigation through the vasculature.

AI/ML Overview

The document K111395 describes the Volcano PrimeWire PRESTIGE® Plus Pressure Guide Wire, a steerable guide wire with a pressure transducer, which is substantially equivalent to the predicate device, PrimeWire PRESTIGE® Pressure Guide Wire (K100930).

1. Table of Acceptance Criteria and Reported Device Performance:

The document lists "Design Input Description" as the acceptance criteria and the "Required Testing, Leveraged, or Justification" as the reported device performance.

Acceptance Criteria (Design Input Description)Reported Device Performance (Required Testing, Leveraged, or Justification)
Working Length - product shall come in 185cm and 300 cm lengthsTesting
Flexible Distal Length - length of flexible section shall be ~ 30 cm in lengthTesting
Distal Radiopaque Length - length of distal tip shall be 3.0 cmTesting
Maximum outer diameter of .0145"Justification – 100% verified in manufacturing
The distal 1.0 cm of the wire shall be equivalent stiffness to the predicate device tip.Testing
Better torqueability than predicate device in distal or tortuous anatomy.Testing
Maintain lubricity within the body throughout procedure (approx. 60 minutes).Testing
Particulate generation in simulated use shall meet USP 788: Maximum of 3000 particles $\ge$ 10 microns; maximum of 300 particles $\ge$ 25 microns.Testing
Sensor Housing to Core Tensile shall be $\ge$ 1.0 pounds.Testing
Tip Tensile $\ge$ 1.0 pounds.Testing
Turns to Failure $\ge$ 10 turnsTesting
Wire connector must withstand 10 insertions in connector.Testing
Sensor shall have same accuracy as predicate deviceLeveraged from previously cleared predicate device
Sensor drift shall not exceed 5 mmHG in 10 minutesLeveraged from previously cleared predicate device and 100% verified in manufacturing
Subjecting PLUS to clinically relevant tortuosity shall not cause erratic output(No explicit testing method mentioned; implied by "No change from predicate device" for this input)
3 year shelf lifeTesting
BiocompatibilityTesting
PackagingLeveraged from previously cleared predicate device
SterilizationLeveraged from previously cleared predicate device

Biocompatibility Acceptance Criteria and Performance:

Test DescriptionResults
Cytotoxicity Study using the Colony Assay Extraction MethodPass
ISO Intracutaneous Study (Irritation Test), Extract 0.9% NaClPass
ISO Intracutaneous Study (Irritation Test), Extract Sesame OilPass
ISO Systemic Toxicity Study, Sesame OilPass
ISO Systemic Toxicity Study, 0.9% NaClPass
ISO Maximum Sensitization Study, Extract Sesame OilPass
ISO Maximum Sensitization Study, Extract 0.9% NaClPass
Pyrogen Study, Material Mediated, 0.9% NaClPass
ASTM Partial Thromboplastin Time - Plasma ExtractPass
ASTM Hemolysis CMF PBS ExtractPass
C3a Complement Activation (Hemocompatibility), Normal Human Serum ExtractPass
SC5b-9 Complement Activation (Hemocompatibility), Human Serum ExtractPass
In Vivo Thromboresistance Study, Peripheral Vessel / Jugular VeinPass
Preliminary Extraction, Japanese MHLWPass
Exaggerated Extraction Method 1 or 2, Japanese MHLWPass
Maximum Sensitization Study, Method 1 or 2, MHLWPass

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: The document does not specify the exact sample sizes for each of the performance tests (e.g., how many wires were tested for tensile strength or torqueability). It generally refers to "Testing" being performed.
  • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. Given the nature of a 510(k) submission for a medical device modification, the testing is typically conducted by the manufacturer (Volcano Corporation, US-based) in a controlled laboratory setting.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The tests performed are physical and chemical characteristic evaluations of the device, rather than diagnostic interpretations requiring human expert consensus.

4. Adjudication method for the test set:

Not applicable. The tests are objective physical and chemical measurements rather than subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a medical instrument (pressure guide wire), not an AI-assisted diagnostic tool or software.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used:

The ground truth for most performance criteria is defined by established engineering and material science standards, ISO standards (e.g., ISO 10993-1 for biocompatibility), USP standards (e.g., USP 788 for particulate generation), and the performance characteristics of the legally marketed predicate device (K100930). For example, "Sensor shall have same accuracy as predicate device" means the predicate device's accuracy serves as the ground truth.

8. The sample size for the training set:

Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

9. How the ground truth for the training set was established:

Not applicable. This device does not involve a "training set" or AI.

{0}------------------------------------------------

K111395

Volcano Corporation July 25, 2011

PrimeWire PRESTIGE®Plus Pressure Guide Wire Special 510(k)

510 (K) Summary

AUG - 1 2011

PrimeWire PRESTIGE® Plus Pressure Guide Wire

Date Prepared:July 7, 2011
Submitted by:Volcano Corporation3661 Valley Centre Dr.Suite 200San Diego, CA 92130
Contact person:Marilyn PourazarSenior Director of Regulatory Affairs
Phone number:(858) 720-4116
Facsimile number:(858) 720-0612
Device Name:PrimeWire PRESTIGE® Plus Pressure Guide Wire
Classification name:> 870.1330 Catheter guide wireClassII
> 870.2870 Catheter tip pressure transducerII

Predicate Device:

The Volcano PrimeWire PRESTIGE® Plus Pressure Guide Wire is substantially equivalent to the following:

510(k) NumberProduct NameClearance Date
K100930PrimeWire PRESTIGE® Pressure Guide WireApril 30, 2010

Device Description:

The PrimeWire PRESTIGE® Plus pressure guide wire is a steerable guide wire with a pressure transducer mounted 3 cm proximal to the tip. The PrimeWire PRESTIGE® Plus guide wire measures pressure when used with the SmartMap®, s5™ Family, and ComboMap® systems. The PrimeWire PRESTIGE® Plus guide wire has a diameter of 0.014" (0.36 mm) and is available in lengths of 185 cm or 300 cm, and also in straight or pre-shaped tips. The PrimeWire PRESTIGE® Plus guide wire is packaged preconnected to the connector with a torque device to facilitate navigation through the vasculature.

l

{1}------------------------------------------------

Intended Use:

The PrimeWire PRESTIGE® Plus Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

Device Technological Characteristics and Comparison to Predicate Device:

The Volcano Corporation PrimeWire PRESTIGE® Plus Pressure Guide Wire Device is substantially equivalent to the predicate device, PrimeWire PRESTIGE® Pressure Guide Wire.

The PrimeWire PRESTIGE® Plus Pressure Guide Wire Device uses the same fundamental scientific technology and has the same intended use as that of the predicate device.

Performance Data:

Applicable testing was performed in accordance with the Design Verification Plan including a Risk Analysis addressing the impact of enhancements to the device and components. The test results indicate the PrimeWire PRESTIGE® Plus Pressure Guide Wire is comparable to the predicate device.

Design Input DescriptionDesign ChangeRequired Testing,Leveraged orJustification
Working Length - product shall comein 185cm and 300 cm lengthsNo change from predicate device.Testing
Flexible Distal Length - length offlexible section shall be ~ 30 cm inlengthNo change from predicate device.Testing
Distal Radiopaque Length - lengthof distal tip shall be 3.0 cmNo change from predicate device.Testing
Design Input DescriptionDesign ChangeRequired Testing,Leveraged orJustification
Maximum outer diameter of .0145"No change from predicate device.Justification –100% verified inmanufacturing
The distal 1.0 cm of the wire shall beequivalent stiffness to the predicatedevice tip.No change from predicate device.Testing
Better torqueability than predicatedevice in distal or tortuous anatomy.Replaced silicon-based coating withhydrophilic coatingTesting
Maintain lubricity within the bodythroughout procedure (approx. 60minutes).Replaced silicon-based coating withhydrophilic coatingTesting
Particulate generation in simulated useshall meet USP 788: Maximum of 3000particles $\ge$ 10 microns; maximum of300 particles $\ge$ 25 microns.Replaced silicon-based coating withhydrophilic coatingTesting
Sensor Housing to Core Tensile shallbe $\ge$ 1.0 pounds.No change from predicate device.Testing
Tip Tensile $\ge$ 1.0 pounds.No change from predicate device.Testing
Turns to Failure $\ge$ 10 turnsNo change from predicate device.Testing
Wire connector must withstand 10insertions in connector.No change from predicate device.Testing
Sensor shall have same accuracy aspredicate deviceNo change from predicate device.Leveraged frompreviously clearedpredicate device
Sensor drift shall not exceed 5 mmHGin 10 minutesNo change from predicate device.Leveraged frompreviously clearedpredicate device and100% verified inmanufacturing
Subjecting PLUS to clinically relevanttortuosity shall not cause erratic outputNo change from predicate device.
3 year shelf lifeReplaced silicon-based coating withhydrophilic coatingTesting
BiocompatibilityReplaced silicon-based coating withhydrophilic coatingTesting
PackagingNo change from predicate device.Leveraged frompreviously clearedpredicate device
SterilizationNo change from predicate device.Leveraged frompreviously clearedpredicate device

Summary of Performance Characteristics

K111395 2

{2}------------------------------------------------

{3}------------------------------------------------

Biocompatibility Studies:

The biocompatibility of the PrimeWire PRESTIGE® Plus was verified in accordance with ISO 10993-1, Biological Evaluation of Medical Devices and FDA Memorandum # G95-1. Test results confirmed biocompatibility of the pressure wire was tested as an external communicating, circulating blood contact, limited exposure (<24 hrs) device.

The following table summarizes biocompatibility test results on the following PrimeWire PRESTIGE® Plus materials:

Test DescriptionResults
Cytotoxicity Study using the Colony Assay Extraction MethodPass
ISO Intracutaneous Study (Irritation Test), Extract 0.9% NaClPass
ISO Intracutaneous Study (Irritation Test), Extract Sesame OilPass
ISO Systemic Toxicity Study, Sesame OilPass
ISO Systemic Toxicity Study, 0.9% NaClPass
ISO Maximum Sensitization Study, Extract Sesame OilPass
ISO Maximum Sensitization Study, Extract 0.9% NaClPass
Pyrogen Study, Material Mediated, 0.9% NaClPass
ASTM Partial Thromboplastin Time - Plasma ExtractPass
ASTM Hemolysis CMF PBS ExtractPass
C3a Complement Activation (Hemocompatibility), NormalHuman Serum ExtractPass
SC5b-9 Complement Activation (Hemocompatibility), HumanSerum ExtractPass
In Vivo Thromboresistance Study, Peripheral Vessel / JugularVein (vessel to be determined based on device evaluation)(Note: the control device will be identified in the NAMSAprotocol before test execution)Pass
Preliminary Extraction, Japanese MHLWPass
Exaggerated Extraction Method 1 or 2, Japanese MHLW (testtype depends on preliminary extraction results)Pass
Maximum Sensitization Study, Method 1 or 2, MHLW (testtype depends on preliminary extraction results)Pass

Image /page/3/Picture/6 description: The image contains a sequence of digits and a letter. The sequence starts with the letter 'K', followed by four instances of the digit '1', and then the digits '395'. There is also a small superscript '4' to the right of the number.

{4}------------------------------------------------

Volcano Corporation July 25, 2011

Conclusion:

The Volcano PrimeWire PRESTIGE® Plus Pressure Guide Wire Device has the same performance specifications, fundamental scientific technology and intended use as that of the predicate device, PrimeWire PRESTIGE® Pressure Guide Wire. The design enhancements to the device do not raise any new questions regarding safety or efficacy. The performance data and a declaration of conformity with design controls support a determination of continuing substantial equivalence of the modified device to the predicate device.

$\qquad K11139\overline{5}^{5}$

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three wavy lines representing the eagle's body. The eagle is facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Volcano Corporation c/o Ms. Marilyn Pourazar Senior Director, Regulatory Affairs 3661 Valley Centre Drive Suite 200 San Diego, CA 92130

AUG - 1 2011

Re: K111395 Trade/Device Name: PrimeWire PRESTIGE® Plus Pressure Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter, Percutaneous Regulatory Class: Class II Product Code: DQX Dated: July 12, 2011 Received: July 13, 2011

Dear Ms. Pourazar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr re and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractly and i warranties. We remind you, however, that device labeling must be truthful and not much you

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements if the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{6}------------------------------------------------

Page 2 - Ms. Marilyn Pourazar

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersQffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

K

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

510(k) Number (if known): K11395

Device Name: PrimeWire PRESTIGE® Plus Pressure Guide Wire Device

Indications for Use:

The PrimeWire PRESTIGE® Plus Pressure Guide Wire Device is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

Prescription Use x (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
----------------------------------------------------------

(Division Sign-Off)

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Numbe

Page 1 of

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.