K Number
K111254
Manufacturer
Date Cleared
2011-08-22

(110 days)

Product Code
Regulation Number
874.4420
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Relieva Solo Elite Sinus Balloon Catheter is an instrument intended to dilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

For children aged 17 and under, the Relieva Solo Elite Sinus Balloon Catheter is intended to dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

Device Description

The Relieva Solo Elite Sinus Balloon Catheter is a flexible catheter that is intended to dilate sinus ostia. The shaft allows for inflation of the sinus balloon and permits the passage of a sinus guidewire or sinus illumination system to facilitate access to the target sinus ostia. A hypotube is incorporated on the proximal end to provide rigidity during insertion through a sinus guide catheter. The subject device also has the capability to irrigate the sinus through the distal tip and three side holes on the distal end.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (Relieva Solo Elite Sinus Balloon Catheter), which typically demonstrates substantial equivalence to predicate devices rather than proving clinical efficacy or diagnostic accuracy with extensive human studies. Therefore, many of the requested elements for AI/device performance studies (like MRMC, expert consensus, training set details) are not applicable to this type of submission.

However, I can extract the relevant performance data and acceptance criteria based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

The submission states that the Relieva Solo Elite Sinus Balloon Catheter met various performance acceptance criteria. While the specific numerical acceptance criteria values are not given, the categories of criteria and the confirmation that they were met are provided.

Acceptance Criteria CategoryReported Device Performance
Dimensional SpecificationsMet all criteria
Balloon Burst PressureMet all criteria
Joint Separation ForceMet all criteria
Deflation TimeMet all criteria
Irrigation Flow RateMet all criteria
Balloon Cycle FatigueMet all criteria
Sterility (SAL 10^-6)Met criteria (validated per AAMI/ANSI/ISO 11135-1: 2007)
Ethylene Oxide ResidualsMet ISO 10993-7:2008 requirements
Shelf LifeEstablished per ASTM F1980-07, ASTM F88/F88M-09, ISTA 2A-11, and ASTM F2096-04 requirements

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify sample sizes for each individual test or the data provenance (e.g., country of origin, retrospective/prospective). These are typically detailed in the test reports themselves, which are referenced but not included in this summary document. The tests conducted are primarily engineering performance tests rather than clinical studies with patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device submission focuses on engineering performance and substantial equivalence, not diagnostic accuracy requiring expert panel review for ground truth.

4. Adjudication Method for the Test Set

Not applicable. This device submission focuses on engineering performance tests, not clinical performance requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This device is a manual surgical instrument, not an AI-powered diagnostic device, and no MRMC study is mentioned or relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is not an algorithm or AI system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the performance tests would be the established engineering standards and material specifications. For example:

  • Dimensional Specifications: Engineering drawings and measurement standards.
  • Balloon Burst Pressure: Material failure limits and design specifications.
  • Sterility: Microbiological validation guidelines (e.g., AAMI/ANSI/ISO 11135-1).
  • Ethylene Oxide Residuals: Toxicology standards (e.g., ISO 10993-7).

8. The Sample Size for the Training Set

Not applicable. This is not an AI/machine learning device that uses a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. This is not an AI/machine learning device.

{0}------------------------------------------------

111254

Image /page/0/Picture/1 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top right of the "t" to above the "A", creating a semi-circular shape over the word. The text is black against a white background.

:

AUG 2 2 2011

510(k) SUMMARY

Sponsor/Submitter:Acclarent, Inc.1525-B O'Brien DriveMenlo Park, California 94025
Contact Person:Kim KySr. Regulatory Affairs SpecialistPhone: (650) 687-4473Fax: (650) 687-4449
Date of Submission:August 11, 2011
Device Trade Name:Relieva Solo Elite Sinus Balloon Catheter
Common Name:Sinus Balloon Catheter
Device Classification:Class I
Regulation Number:21 CFR 874.4420
Classification Name:Ear, nose, and throat manual surgical instrument
Product Code:LRC
Predicate Devices:Relieva Sinus Balloon Catheter (K073041)Xpress Multi-Sinus Dilation Tool (K102003)
Device Description:The Relieva Solo Elite Sinus Balloon Catheter is a flexiblecatheter that is intended to dilate sinus ostia. The shaft allowsfor inflation of the sinus balloon and permits the passage of asinus guidewire or sinus illumination system to facilitateaccess to the target sinus ostia. A hypotube is incorporated onthe proximal end to provide rigidity during insertion through asinus guide catheter. The subject device also has the capabilityto irrigate the sinus through the distal tip and three side holeson the distal end.
Indications for Use:The Relieva Solo Elite Sinus Balloon Catheter is aninstrument intended to dilate sinus ostia and spaces within theparanasal sinus cavities for diagnostic and therapeuticprocedures. It is also intended to irrigate from within a targetsinus for therapeutic procedure and to facilitate diagnosticprocedures.For children aged 17 and under, the Relieva Solo Elite SinusBalloon Catheter is intended to dilate sinus ostia and spacesassociated with the maxillary sinus for diagnostic andtherapeutic procedures. It is also intended to irrigate fromwithin a target sinus for therapeutic procedures and to

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "t", creating a visual element above the word. The overall impression is that of a company logo or brand name.

์ไ

Technological Characteristics:

Performance Data:

Summary of Substantial Equivalence:

facilitate diagnostic procedures.

The Relieva Solo Elite Sinus Balloon Catheter is a device that allows for dilation of sinus ostia with the added capability to irrigate. Dilation is achieved via a non-compliant balloon located on the distal end of the device.

The Relieva Solo Elite Sinus Balloon Catheter met all performance acceptance criteria including dimensional specifications; balloon burst pressure, joint separation force, deflation time, irrigation flow rate, and balloon cycle fatigue.

The sterilization process is validated per AAMI/ANSI/ISO 11135-1: 2007 and demonstrated a sterility assurance level of 10 . The method used for sterilization validation is the overkill (half-cycle approach) in a fixed chamber. Ethylene oxide residuals were tested and meet ISO 10993-7:2008 requirements. The subject device is not tested or labeled as "non-pyrogenic".

Shelf life was established per ASTM F1980-07 ASTM F88/F88M-09, ISTA 2A-11, and ASTM F2096-04 requirements.

The Relieva Solo Elite Sinus Balloon Catheter is substantially equivalent to predicate device as confirmed through relevant tests.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MID 20993-0002

Acclarent, Inc. Ms. Kim Ky Manager, Regulatory & Clinical 1525-B O'Brien Dr Menlo Park, CA 94025

AUG 2 2 2011

Re: K111254

Trade/Device Name: Relieva Solo Elite Sinus Balloon Catheter Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, nose, and throat manual surgical instrument Regulatory Class: I Product Code: LRC Dated: July 18, 2011 Received: July 19, 2011

Dear Ms. Ky,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Evelin Mz. Jr.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

III 25̄4

Image /page/4/Picture/1 description: The image shows the word "Acclarent" in a stylized font. The word is in black and appears to be a company logo. There is a curved line above the "t" with a dot above the line.

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):K111254
Trade Name:Relieva Solo Elite Sinus Balloon Catheter
Common Name:Sinus Balloon Catheter
Indications For Use:The Relieva Solo Elite Sinus Balloon Catheter is an instrumentintended to dilate sinus ostia and spaces within the paranasalsinus cavities for diagnostic and therapeutic procedures. It isalso intended to irrigate from within a target sinus for therapeuticprocedures and to facilitate diagnostic procedures.For children aged 17 and under, the Relieva Solo Elite SinusBalloon Catheter is intended to dilate sinus ostia and spacesassociated with the maxillary sinus for diagnostic and therapeuticprocedures. It is also intended to irrigate from within a targetsinus for therapeutic procedures and to facilitate diagnosticprocedures.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anatolur

(Division Sign-Off) (Division Sign-Orr)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K111254.

Page __ of

(Posted November 13, 2003)

§ 874.4420 Ear, nose, and throat manual surgical instrument.

(a)
Identification. An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.(b)
Classification Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 874.9.