K Number
K102003

Validate with FDA (Live)

Date Cleared
2010-10-22

(99 days)

Product Code
Regulation Number
874.4420
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To access and treat the frontal recesses, sphenoid sinus ostia and maxillary ostia/ethmoid infundibula in adults using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.

Device Description

The XprESS Multi-Sinus Dilation Tool is intended to remodel or recreate the sinus outflow tract via trans-nasal balloon dilation. The XprESS device combines features of a curved suction tip and a frontal ostium seeker with the tissue expansion effect of balloon dilation. The familiar features of this device enable a physician to track the device to the sinus ostium using endoscopic visualization. Since the distal end of the device is re-shapeable, one balloon can be modified to work on multiple sinuses within the same patient.

The XprESS Multi-Sinus Dilation Tool has been tested to withstand multiple inflations and device tip manipulations (up to 25) in a surgical case wherein all 6 sinus ostia are being dilated.

The XprESS device curved suction tip has a 2 mm atraumatic ball tip with a 1 mm inside diameter. A suction tube may be connected to the proximal barbed fitting to provide active suction.

The XprESS Multi-Sinus Dilation Tool is provided sterile and for single use only.

The items packaged with the Xpress Multi-Sinus Dilation Tool include the Inflation Device and the Infusion Line.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the XprESS Multi-Sinus Dilation Tool (K102003):

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state quantitative acceptance criteria in a pass/fail format. However, it describes performance testing conducted to ensure design specifications were met and the device performed as intended.

Acceptance Criterion (Implicit)Reported Device Performance
Withstand multiple inflationsTested to withstand multiple inflations.
Withstand device tip manipulationsTested to withstand device tip manipulations (up to 25).
Support utility in maxillary sinusCadaver study conducted to support utility in the maxillary sinus.
Meet design specificationsPerformance testing showed that the device meets design specifications.
Perform as intendedPerformance testing showed that the device performed as intended.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The text mentions a "cadaver study" but does not specify the number of cadavers used.
  • Data Provenance: The cadaver study is presumably a prospective study conducted for the purpose of this submission. The country of origin is not explicitly stated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not provide information on the number or qualifications of experts involved in establishing ground truth for the cadaver study. It's likely that a surgeon or medical professional familiar with sinus anatomy and procedures would have performed or overseen the procedures in the cadaver study.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the cadaver study.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study described is a performance study on a cadaver model, not a comparative study involving human readers or clinical outcomes.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable. The XprESS Multi-Sinus Dilation Tool is a physical medical device, not an algorithm or AI system. Therefore, the concept of "standalone algorithm performance" doesn't apply. The performance study was for the device itself.

7. The Type of Ground Truth Used

For the cadaver study, the "ground truth" would have been the direct anatomical observation of the sinus outflow tracts after dilation using the device. This would involve visual inspection and potentially physical measurements or dissections within the cadaver to confirm that the remodeling or recreation of the sinus outflow tract occurred as intended.

8. The Sample Size for the Training Set

This question is not applicable. This device is not an AI/ML algorithm that requires a "training set." The term "design verification testing" covered various engineering tests for the device's physical properties and function.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as there is no "training set" for this type of medical device. The "ground truth" for the device's design specifications would have been established through engineering design requirements, material specifications, and functional targets determined by the manufacturer.

{0}------------------------------------------------

K102003

510(k) Summary

Date Prepared:September 20, 2010
Submitter Information:Entellus Medical, Inc.6705 Wedgwood Court, NorthMaple Grove, MN 55311
Establishment Registration:3006345872
Contact Information:Karen E. PetersonVice President Clinical, Regulatory and Quality(763) 463-7066kpeterson@entellusmedical.com
Device Information:
Trade Name:XprESS Multi-Sinus Dilation Tool
Common Name:Sinus Balloon Dilation System
Classification Name:ENT Manual Surgical Instrument
Product Code:LRC
Regulation Number:Class I, 21 CFR 874.4420
Predicate Devices:

XprESS Multi-Sinus Dilation Tool [K093007 (Entellus Medical Balloon Device)] FinESS Sinus Treatment System [K081542]

Device Description:

The XprESS Multi-Sinus Dilation Tool is intended to remodel or recreate the sinus outflow tract via trans-nasal balloon dilation. The XprESS device combines features of a curved suction tip and a frontal ostium seeker with the tissue expansion effect of balloon dilation. The familiar features of this device enable a physician to track the device to the sinus ostium using endoscopic visualization. Since the distal end of the device is re-shapeable, one balloon can be modified to work on multiple sinuses within the same patient.

The XprESS Multi-Sinus Dilation Tool has been tested to withstand multiple inflations and device tip manipulations (up to 25) in a surgical case wherein all 6 sinus ostia are being dilated.

The XprESS device curved suction tip has a 2 mm atraumatic ball tip with a 1 mm inside diameter. A suction tube may be connected to the proximal barbed fitting to provide active suction.

Image /page/0/Figure/8 description: The image shows a medical device with several labeled parts. The device has a curved suction tip on one end and a proximal barbed fitting on the other end. There is also a balloon slide mechanism and a balloon inflation luer, which are located on the main body of the device. The image provides a clear view of the device's components and their arrangement.

XprESS Multi-Sinus Dilation Tool

{1}------------------------------------------------

The XprESS Multi-Sinus Dilation Tool is provided sterile and for single use only.

The items packaged with the Xpress Multi-Sinus Dilation Tool include the Inflation Device and the Infusion Line.

Indication for Use:

To access and treat the frontal recesses, sphenoid sinus ostia and maxillary ostia/ethmoid infundibula in adults using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.

Contraindications:

  • Do not use this XprESS device in patients who are allergic to nickel or barium sulfate. .
  • Do not attach the XprESS device directly to the CT Image Guidance systems. This may . result in inaccurate device positioning.

Technological Characteristics:

The device has identical technological characteristics (i.e., design, materials, chemical composition, function, mode of operation, packaging and sterilization) as the predicate device [K093007]. Both devices are sterilized using Ethylene Oxide (EtO), are validated per ISO 11135-1, and have a Sterility Assurance Level (SAL) of 10°. Both devices are for single use only. Items packaged with both devices include an inflation device and an infusion line. Both devices are biocompatible per ISO 10993-1.

Substantial Equivalence:

No design modifications were necessary to support the expanded indications for use: therefore this device is identical to the currently marketed XprESS device [K093007] in terms of technological characteristics including: the materials, design, function, packaging and sterilization. The intended use of the device is the same as the intended use of the predicate . devices. The expanded indication for use in the maxillary sinus is similar to the FinESS Sinus Treatment System [K081542].

Performance Data:

Performance testing of the XprESS device consisted of design verification testing and a cadaver study. Design verification testing included testing for additional balloon inflations and device tip manipulations. A cadaver study was conducted to support the utility of this device in the maxillary sinus. Biocompatibility, sterilization, packaging, distribution, shelf life testing, animal and clinical data were not submitted. Performance testing showed that the device meets design specifications and performed as intended.

Conclusion

In conclusion, the device is substantially equivalent based on a comparison of intended use, indications for use, and technological characteristics.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles a bird.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Entellus Medical,Inc. c/o Ms. Karen Peterson Vice President Clinical, Regulatory and Quality 6705 Wedgwood Court North Maple Grove, MN 55311

Re: K102003

Trade/Device Name: XprESS Multi-Sinus Dilation Tool (JD-100) Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose & Throat Manual surgical instrument Regulatory Class: Class I Product Code: LRC Dated: August 27, 2010 Received: August 30, 2010

Dear Ms. Peterson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

OCT 2 2 2010

..

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kesia Alexander

for

MR
DR
DR

Malvina B. Eydelman. M.D. Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K10.2003

Indications for Use Statement

K102003 510(k) Number:

XprESS Multi-Sinus Dilation Tool Device Name:

Indications for Use:

OCT 2 2 2010

To access and treat frontal recesses, sphenoid sinus ostia and maxillary ostialethmoid infundibula in adults using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.

Prescription Use × (21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Rudy CRNP
(Division Sign-Off)

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K102003

§ 874.4420 Ear, nose, and throat manual surgical instrument.

(a)
Identification. An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.(b)
Classification Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 874.9.