(120 days)
Not Found
No
The description focuses on mechanical features and materials, with no mention of AI or ML.
No
The device is described as an intravascular catheter used for sampling blood, monitoring blood pressure, or administering fluids and blood. While it facilitates the administration of therapies, the device itself is not a therapeutic agent; it is a tool for diagnostic and administrative purposes.
No
Explanation: The device is a catheter used for sampling blood, monitoring blood pressure, or administering fluids and blood, which are therapeutic and monitoring functions, not diagnostic. While sampling blood can be used for diagnostic purposes, the device itself is not performing the diagnostic analysis.
No
The device description clearly outlines physical components such as a catheter, septum, stabilization platform, and needle-shielding mechanism, indicating it is a hardware device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is for "inserted into a patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure or administer fluids and blood intravascularly." This describes a device used in vivo (within the body) for direct patient care and intervention.
- Device Description: The description details a physical catheter and needle system designed for insertion into a blood vessel.
- Lack of IVD Characteristics: An IVD is a medical device used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health. The description of the Introcan Safety® 3 does not mention any testing or analysis of samples. While it can be used to sample blood, the device itself is not performing the diagnostic test on that sample.
Therefore, the Introcan Safety® 3 Closed Intravascular Catheter is a device used for accessing the vascular system for various medical procedures, not for performing in vitro diagnostic tests.
N/A
Intended Use / Indications for Use
The Introcan Safety® 3 Closed Intravascular Catheter is inserted into a patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure or administer fluids and blood intravascularly. The 18-22 gauge catheters may be used with power injectors for which the maximum pressure setting is 300 psi with a luer lock connection only.
Product codes (comma separated list FDA assigned to the subject device)
FOZ
Device Description
The Introcan Safety® 3 Closed IV Catheter consists of an over-the-needle, peripheral intravascular catheter made of radiopaque polyurethane, an integrated bidirectional septum, a stabilization platform, and a passive safety needle-shielding mechanism.
The Introcan Safety® 3 will be offered in the following gauge sizes and lengths:
18ga x 1-3/4" (45mm)
18ga x 1-1/4" (32 mm)
20ga x 1-1/4" (32mm)
20ga x 1" (25mm)
22ga x 1" (25mm)
24ga x ¾" (19mm)
Introcan Safety® 3 design is described as a closed IV catheter since it protects clinicians and patients from blood exposure. Since the needle is withdrawn through a septum that seals after the needle has been removed, blood is thus contained within the Introcan Safety® 3 device. The pressure exerted on the needle as it passes through the septum wipes blood from the needle further reducing potential blood exposure.
The Introcan Safety® 3 catheter has an integrated stabilization platform is designed to improve catheter stability while minimizing catheter movement within the vessel.
The passive safety needle-shielding mechanism of the Introcan Safety® 3 is located inside the catheter hub. Upon withdrawal of the needle, the safety shield engages as the needle passes through the catheter hub and deploys automatically to shield the needle tip. The safety shield protects during disposal, aiding in the prevention of needlestick injuries. Once the safety shield engages and shields the needle tip, the user is unable to re-insert the needle which aids in the prevention of catheter shearing.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Vascular system
Indicated Patient Age Range
Not Found. May be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Testing was conducted to demonstrate the performance of the Introcan Safety® 3 Closed IV Catheter and substantial equivalence with the predicate devices: B. Braun Medical Inc.'s Introcan Safety® and BD's Nexiva™ Closed IV Catheter System Closed IV Catheter System. The functional performance testing conducted with the proposed device demonstrates that the Introcan Safety® 3 Closed IV Catheter is as safe and effective as the predicate devices.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 880.5200 Intravascular catheter.
(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).
0
AUG 3 0 2011
l of 3
page
B. Braun Medical Inc. |
---|
510(k) Premarket Notification |
Introcan Safety® 3 Closed IV Catheter |
5. 510(k) SUMMARY
DATE: | August 19, 2011 |
---|---|
SUBMITTER: | B. Braun Medical Inc. |
901 Marcon Boulevard | |
Allentown, PA 18109-9341 | |
610-266-0500, ext. 2966 | |
Contact: Angela J. Caravella, Regulatory Affairs Specialist | |
Phone: (610) 596-2966 | |
Fax: (610) 266-4962 | |
E-mail: angela.caravella@bbraun.com | |
DEVICE NAME: | Introcan Safety® 3 Closed IV Catheter |
COMMON OR | |
USUAL NAME: | Safety Intravascular Catheter |
DEVICE | |
CLASSIFICATION: | Class II, per 21 CFR §880.5200: Intravascular Catheter |
PREDICATE DEVICES: | Introcan Safety® IV TM Catheter, B. Braun Medical, Inc., |
K021094 and K020785, Class II, DYB and FOZ, 880.5200: | |
Intravascular Catheter | |
Nexiva TM Closed IV Catheter System, Becton Dickinson, | |
K032843, Class II, FOZ, 880.5200: Intravascular Catheter | |
DESCRIPTION: | The Introcan Safety® 3 Closed IV Catheter consists of an |
over-the-needle, peripheral intravascular catheter made of | |
radiopaque polyurethane, an integrated bidirectional septum, | |
a stabilization platform, and a passive safety needle- | |
shielding mechanism. | |
The Introcan Safety® 3 will be offered in the following | |
gauge sizes and lengths: | |
18ga x 1-3/4" (45mm) | |
18ga x 1-1/4" (32 mm) | |
20ga x 1-1/4" (32mm) | |
20ga x 1" (25mm) | |
22ga x 1" (25mm) | |
24ga x ¾" (19mm) |
- '
1
K111234 page 2 of 3
.
| | Introcan Safety® 3 design is described as a closed IV
catheter since it protects clinicians and patients from blood
exposure. Since the needle is withdrawn through a septum
that seals after the needle has been removed, blood is thus
contained within the Introcan Safety® 3 device. The
pressure exerted on the needle as it passes through the
septum wipes blood from the needle further reducing
potential blood exposure. |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Introcan Safety® 3 catheter has an integrated
stabilization platform is designed to improve catheter
stability while minimizing catheter movement within the
vessel. |
| | The passive safety needle-shielding mechanism of the
Introcan Safety® 3 is located inside the catheter hub. Upon
withdrawal of the needle, the safety shield engages as the
needle passes through the catheter hub and deploys
automatically to shield the needle tip. The safety shield
protects during disposal, aiding in the prevention of
needlestick injuries. Once the safety shield engages and
shields the needle tip, the user is unable to re-insert the
needle which aids in the prevention of catheter shearing.
This device may be used for any patient population with |
| | consideration given to adequacy of vascular anatomy and
appropriateness for the solution being infused and duration
of therapy |
| INTENDED USE: | The Introcan Safety® 3 Closed Intravascular Catheter is
inserted into a patient's vascular system for short term use
(less than 30 days) to sample blood, monitor blood pressure
or administer fluids and blood intravascularly. The 18-22
gauge catheters may be used with power injectors for which
the maximum pressure setting is 300 psi with a luer lock
connection only. |
| SUBSTANTIAL
EQUIVALENCE: | Three predicate devices are utilized for substantial
equivalence, B. Braun Medical Inc's Introcan Safety®
(K021094 and K020785) and Becton Dickinson's Nexiva™
Closed IV Catheter System (K032843). Introcan Safety® 3
Closed IV Catheter has similar intended uses to B. Braun
Inc.'s Introcan Safety®, in that they both act as an
intravascular catheter that is inserted into a patient's vascular
system for short term use (less than 30 days) to sample |
2
K111236 page 3 of 3
blood, monitor blood pressure or administer fluids and blood intravascularly, the same passive safety needle-shielding feature is used to aid in the prevention of needlestick injuries, and some gauge sizes can be used with power injectors up to 300 psi.
Introcan Safety® 3 is similar to the Introcan Safety® , containing the same design elements, similar gravity flow rates, and intended uses with the only feature added is a valve that controls the flow of blood within the device until the catheter is attached to a luer compatible associated device such as an IV set, IV extension set, syringe, or filter which will engage the valve and allow for flow of fluid through the catheter.
Introcan Safety® 3 has a similar intended use to BD's Nexiva™ Closed IV Catheter System in that they both act as an intravascular catheter that is inserted into a patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure or administer fluids and blood intravascularly, they both contain a needleshielding feature to aid in the prevention of needlestick injuries, blood is contained within the device to prevent blood exposure, both designs include a stabilization platform to reduce catheter movement within the vessel, they may both be used in any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure, and similar gauge sizes are suitable for use with power injectors rated for a maximum 300 psi and at a defined flow rate.
The Introcan Safety® 3 was developed to achieve a design that prevents the exposure of blood which matches the intended use of BD's Nexiva™ Closed IV Catheter System device. Each of the other intended uses for the Nexiva® are the same as the Introcan Safety® 3.
CONCLUSION:
Testing was conducted to demonstrate the performance of the Introcan Safety® 3 Closed IV Catheter and substantial equivalence with the predicate devices: B. Braun Medical Inc.'s Introcan Safety® and BD's Nexiva™ Closed IV Catheter System Closed IV Catheter System. The functional performance testing conducted with the proposed device demonstrates that the Introcan Safety® 3 Closed IV Catheter is as safe and effective as the predicate devices.
3
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human figures, representing the department's mission to protect the health of all Americans and provide essential human services. The graphic is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
B. Braun Melsungen AG C/O Ms. Angela J. Caravella Regulatory Affairs Specialist B. Braun Medical, Incorporated 901 Marcon Bloulevard Allentown, Pennsylvania 18109
AUG 3 0 2011
Re: K111236
Trade/Device Name: Introcan Safety® 3 Closed Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: August 19, 2011 Received: August 22, 2011
Dear Ms. Caravella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
4
Page 2 - Ms. Caravella
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH1's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthony D. man
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
5
K111236
B. Braun Medical Inc. 510(k) Premarket Notification Introcan Safety® 3 Closed IV Catheter
4. INDICATIONS FOR USE STATEMENT
Page _________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known):
Device Name:
Introcan Safety® 3 Closed IV Catheter
Indications For Use:
Introcan Safety® 3 Closed Intravascular Catheter is inserted into a patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure or administer fluids and blood intravascularly.
The 18-22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi with a luer lock connection only.
Prescription Use __ > (Per 21 CFR 801.109)
OR
Over-The-Counter Use - - -
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rld C. A 8/30/4
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111236