K Number
K111149

Validate with FDA (Live)

Date Cleared
2011-12-30

(249 days)

Product Code
Regulation Number
876.5010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HANAROSTENT® BILIARY (NNN) is indicated for the palliation of malignant strictures in the biliary tree.

Device Description

The HANAROSTENT® Biliary (NNN) is a self-expanding uncoated tubular prosthesis designed to maintain patency of bile duct strictures caused by malignant tumors. It consists of a self-expanding metal stent and delivery system. The Biliary stent is made of Nickel Titanium alloy (Nitinol) wire, which expands at body temperature. The stent is deployed with supplied introducers for percutaneous and endoscopic use.

The Stent is available in two diameters (8 and 10mm) and eight lengths (40mm, 50mm, 60mm, 70mm, 80mm, 90mm, 100mm, and 120mm). User preference, individual patient condition and/or anatomy determine the appropriate type and size for use. The HANAROSTENT® Biliary (NNN) is intended for single use only.

AI/ML Overview

This document describes a 510(k) premarket notification for the HANAROSTENT® Biliary (NNN) device, a self-expanding metal stent indicated for the palliation of malignant strictures in the biliary tree. The submission focuses on demonstrating substantial equivalence to a predicate device (Niti-S Biliary Stent & Introducer) rather than providing detailed acceptance criteria and a standalone human factors study.

Here's an analysis of the provided information, addressing your specific points:

1. A table of acceptance criteria and the reported device performance

The provided document does not explicitly state specific acceptance criteria (e.g., numerical thresholds for performance) for general device functionality, nor does it provide a direct table of reported device performance in this format.

Instead, the submission relies on demonstrating substantial equivalence to an existing predicate device (Niti-S Biliary Stent & Introducer) through comparison of design, materials, and intended use. The performance summary refers to "bench testing" that was performed to "confirm the safety and effectiveness" as compared to the predicate device.

Acceptance Criteria (Implied by Substantial Equivalence Review)Reported Device Performance (Summary)
Mechanical Performance:Confirmed through bench testing (compression force, expansion force, deployment force and accuracy, withdrawal force). Not explicitly quantified against specific criteria, but deemed comparable to predicate.
Biocompatibility:Assessed in accordance with ISO10993-1:2003.
Sterility:EO Sterilization (same as predicate).
Functionality (Deployment/Placement):Capable of percutaneous and endoscopic placement, release by pulling outer sheath (same as predicate).
Material Composition:Nitinol stent, Gold markers, Introducer: Teflon, PEEK, ABS, SUS 304, PC (comparable to predicate with minor differences).
Intended Use:Palliation of malignant strictures in the biliary tree (identical to predicate).

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not mention any clinical study involving a "test set" of patient data. The performance claims are based on bench testing and biocompatibility assessment, not on clinical outcomes with a specific sample size. Therefore, there's no information on data provenance (country, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as there was no clinical "test set" and thus no ground truth established by experts in a clinical context for this submission. The evaluation was based on engineering bench testing and regulatory assessment of substantial equivalence.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This question is not applicable as there was no clinical "test set" requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. The device is a medical stent, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is irrelevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The device is a medical stent, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

As this submission relies primarily on bench testing and comparison to a predicate device, the "ground truth" for the device's technical performance would be based on engineering standards and measurements, and for biocompatibility, ISO 10993-1:2003 standards. There is no mention of clinical ground truth like pathology or outcomes data.

8. The sample size for the training set

This question is not applicable. The device is a physical medical device, not an AI model, so there is no "training set."

9. How the ground truth for the training set was established

This question is not applicable as there is no training set for this device.

In summary:

This 510(k) submission for the HANAROSTENT® Biliary (NNN) is a predication-based submission. The "study that proves the device meets the acceptance criteria" is fundamentally a comparison to a legally marketed predicate device (Niti-S Biliary Stent & Introducer) supported by bench testing and biocompatibility assessments. The submission's core argument is that the new device has "the same device characteristics, composition, function, and intended use" as the predicate, with minor differences not raising new questions of safety or efficacy. The FDA's letter concurs with the substantial equivalence claim, allowing the device to proceed to market with specific labeling limitations regarding vascular use.

{0}------------------------------------------------

K111149
page 1 of 3

Pre-Market Notification 510(k) Summary

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

1. Sponsor Information:

Company Name & Address:M.I.Tech Co., Ltd.241-3 Habuk-ri, Jinwi-myeon,Pyeongtaek-si, Gyeonggi-do451-864 Republic of Korea
Contact Person:Danny Lee
Contact Title:Manager
Contact Phone Number:+82 31 662-5645
Contact Fax Number:+82 31 662-5648

Date of Summary:

2. Device Name and Classification:

Common and Usual Name:Biliary Catheter
Proprietary Name:HANAROSTENT® Biliary (NNN)
Classification Name:Biliary Catheter
Classification Regulation:21 CFR 876.5010
Regulatory Class:Class 2
Product Code:FGE
Performance Standards:No applicable performance standards have beenissued under section 514 or under section 513(b) ofthe Food, Drug and Cosmetic Act.

October 21, 2011

3. Predicate Device(s):

Niti-S Biliary Stent & Introducer, K073667

4. Description of Device:

The HANAROSTENT® Biliary (NNN) is a self-expanding uncoated tubular prosthesis designed to maintain patency of bile duct strictures caused by malignant tumors. It consists of a self-expanding metal stent and delivery system. The Biliary stent is made of Nickel Titanium alloy (Nitinol) wire, which expands at body temperature. The stent is deployed with supplied introducers for percutaneous and endoscopic use.

The Stent is available in two diameters (8 and 10mm) and eight lengths (40mm, 50mm, 60mm, 70mm, 80mm, 90mm, 100mm, and 120mm). User preference, individual patient condition and/or anatomy determine the appropriate type and size for use. The HANAROSTENT® Biliary (NNN) is intended for single use only.

క. Indications for Use:

The HANAROSTENT® Biliary (NNN) is indicated for the palliation of malignant structures in the biliary tree.

{1}------------------------------------------------

K111149
page 2 of 3

Proprietary NameHANAROSTENT® Biliary (NNN)Niti-S Biliary Stent & Introducer
Indications for useThe HANAROSTENT® Biliary (NNN)is indicated for the palliation ofmalignant strictures in the biliarytree.The Niti-S Biliary Stent is indicatedfor the palliation of malignantstrictures in the biliary tree.
Design(Stent)Image: HANAROSTENT Biliary StentImage: Niti-S Biliary Stent
Nitinol wireNitinol wire
Cylinder shape, Diamond pattern,Cylinder shape, Diamond pattern
Flare shapeNon Flare shape
Length; 40, 50, 60, 70, 80, 90, 100,120mmLength; 40, 50, 60, 70, 80, 90, 100,120 mm
Diameter; 8, 10mmDiameter; 8, 10mm
Marker; 12 gold markersMarkers; 8 Pt/Ir2 STS 316L
Design(Introducer)Image: HANAROSTENT IntroducerImage: Niti-S Introducer
Braided tube typeLength; 60, 180cmDiameter; 2.36mmCo-axial tube typeUsable Length; 50,60,180,190cmDiameter; 2.7, 2.8mm
Single UseYesYes
SterileEO SterilizationEO Sterilization
Method ofPlacementPercutaneous,EndoscopicPercutaneous,Endoscopic
Method ofDeploymentRelease by pulling Outer sheathRelease by pulling Outer sheath
MaterialsStent - NitinolStent - Nitinol
Marker - GoldMarker - Pt/Ir, STS316L
Introducer - Teflon, PEEK, ABS, SUS304, PCIntroducer - Teflon, PE, ABS

6. Comparison with Predicate Device(s):

The HANAROSTENT® Biliary (NNN) has the same device characteristics, composition, function, and intended use as the predicate device, such as the Niti-S Biliary Stent & Introducer.

The minor differences between the devices are introducer type, marker materials, and dimensional differences. These minor differences do not raise new questions of safety or efficacy of the device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "K111149 page 3 of 3". The text appears to be handwritten. The first line contains a series of numbers and letters, while the second line indicates that it is page 3 of 3.

Based on the comparison of intended use and technical features, the HANAROSTENT® Biliary (NNN) is substantially equivalent to the identified predicate device.

7. Performance Summary:

Appropriate bench testing including compression force, expansion force, deployment force and accuracy, withdrawal force and corrosion tests were performed to confirm the safety and effectiveness of the HANAROSTENT® Biliary (NNN) stent as compared to the identified predicate device. Biocompatibility safety was assessed in accordance with ISO10993-1:2003.

Based on the performance comparisons, the HANAROSTENT® Biliary (NNN) is substantially equivalent to the identified predicate device.

8. Conclusions:

In accordance with the Federal Food, Drug and Cosmetic Act. 21 CFR Part 807, and based on the information provided in this premarket notification M.I.Tech Co., Ltd. concludes that the HANAROSTENT® Biliary (NNN) stent and introducer system is safe and effective, performs at least as well, and is substantially equivalent to the predicate devices as described herein and similar devices legally commercially available.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

M.I.Tech Co., Ltd. % Mr. Paul Sumner Vice President Arkin Consulting Group 1733 Canton Lane MARIETTA GA 30062

DEC 30 2011

Re: K11149

Trade/Device Name: HANAROSTENT® Biliary (NNN) Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: December 19, 2011 Received: December 22, 2011

Dear Mr. Sumner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{4}------------------------------------------------

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This resure will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be may of cables to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse ievents) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Mark A. Milliken

fo

Christy Foreman Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K111149

Device Name: HANAROSTENT® BILIARY (NNN)

Indications For Use: The HANAROSTENT® BILIARY (NNN) is indicated for the palliation of malignant strictures in the biliary tree.

Prescription Use __ > (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

:

Colin M. Pollard

Gastro-Rena

Page 1 of 1

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.