(46 days)
The Oral Enteral Syringe is intended for the delivery of liquid medication, formula and breast milk.
The Oral Enteral Syringe is intended as a single use, sterile, device for the delivery of liquid medication, formula and breast milk. The Oral Enteral Syringe comes in multiple volume sizes ranging from 1cc to 60cc [e.g., 1cc, 3cc, 10cc, 20cc, 30cc and 60cc]. The VR Medical Oral Enteral Syringe, as well as the predicate device [Phillips Children's Medical Venture Oral/ Enteral Syringe (K100099)] has 'labeling' on the device itself, i.e., highly visible labeling on side of barrel marked ORAL/ENTERAL ONLY, and the utilization of bright orange colorant for the plunger and cap that easily distinguishes the syringe for feeding and not other uses.
The Oral Enteral Syringe incorporates safety connectors which eliminate the risk of IV administration through the feeding tube, i.e., the safety connectors will not mate with Luer Lock or Leur slip fittings.
The VR Medical Oral Enteral Syringe design incorporates various components, e.g., barrels, plungers, caps and end caps, which are comprised of commercially available materials with demonstrated acceptable biocompatibility, as listed below:
Barrel Material: Polypropylene 370Y
Piston Material: Thermoplastic Elastomer IR307
Plunger Material: Polypropylene 370Y
Plunger Colorant: Orange Pigment*
Cap Material: Polypropylene 370Y
Cap Colorant: Orange Pigment (P21024)
Barrel Gradation Ink: PPE Ink Series (on outside of barrel)
Nozzle Material: Polypropylene 370Y (molded with barrel)
The Oral Enteral Syringe does not contain natural rubber latex, DEHP, or BPA.
Here's an analysis of the provided text, focusing on the acceptance criteria and the study proving the device meets them:
Device: VR Medical Oral Enteral Syringe
Premarket Notification (510(k)) Number: K110853
Date of Submission: May 12, 2011
Date of Clearance: May 13, 2011
The document describes a 510(k) submission for the VR Medical Oral Enteral Syringe, asserting its substantial equivalence to predicate devices (Phillips Children's Medical Venture Oral/Enteral Syringe (K100099) and Wei Gao Group Medical Syringe (K072739)). The primary focus for substantial equivalence is on the device's design, materials, function, and safety, especially concerning its enteral-only design to prevent misconnections with IV lines.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list "acceptance criteria" in a separate, quantified section with matching "reported device performance" values for all tests. Instead, it details various tests conducted, with the implicit acceptance criterion being that the VR Medical Syringe performs equivalently to the predicate or meets established standards. The summary below interprets the "Justification" column as the implied acceptance goal or comparison.
| Specific Test | Implied Acceptance Criterion | Reported Device Performance |
|---|---|---|
| Syringe Tip Tensile Test | Performance substantially equivalent to predicate (K100099) | VR Medical Syringe tested alongside Phillips Syringe (K100099). (Implied: Met equivalence) |
| Syringe Liquid Leakage Test | Device functions according to its specifications (no leakage). | VR Medical Syringe tested and "successfully completed" in-vitro testing demonstrating it "functions according to its specifications." (Implied: No unacceptable leakage) |
| Volume Measurement Accuracy Testing (manual, ISO 7886-1, syringe pump, ISO 7886-2) | Performance substantially equivalent to predicate (K100099) and meets ISO 7886-1 and ISO 7886-2 requirements. | VR Medical Syringe tested alongside Phillips Syringe (K100099). "Successfully completed" in-vitro testing demonstrating it "functions according to its specifications." (Implied: Met accuracy requirements and equivalence) |
| ISO 7886-1 requirements (excluding tip requirements) | Meets ISO 7886-1 requirements. | VR Medical Syringe tested. "Successfully completed" in-vitro testing demonstrating it "functions according to its specifications." (Implied: Met ISO 7886-1 requirements for syringe body) |
| Size and Material Inspection | Dimensions and materials conform to design specifications. | VR Medical Syringe tested. "Successfully completed" in-vitro testing demonstrating it "functions according to its specifications." (Implied: Conformed to specifications) |
| Biocompatibility Analysis (Syringe materials) | Materials are identical or substantially equivalent to referenced predicate (K100099) or FDA-cleared. | Referenced to predicate (K100099). VR Medical device component materials and processes are "identical to referenced predicate (except orange colorant)." Confirmed presence of Polypropylene 370Y, Thermoplastic Elastomer IR307. (Implied: Biocompatible due to equivalence or known safe materials) |
| Biocompatibility Analysis (Orange colorant) | Colorants are cleared (or FDA listed) as food contact substances per 21 CFR 178.3297. | Referenced to 21 CFR 178.3297. "These colorants are cleared (or FDA listed) as food contact substance." (Implied: Biocompatible for intended use) |
| Sterilization Related Testing | Device is sterile and maintains sterility. | VR Medical Syringe tested. "Successfully completed" in-vitro testing demonstrating it "functions according to its specifications." (Implied: Demonstrated sterility and integrity) |
| Storage and Shelf Life Testing (accelerated-time aging) | Maintains functional integrity over projected shelf life. | VR Medical Syringe and actual proposed package/device tested. "Successfully completed" in-vitro testing demonstrating it "functions according to its specifications." (Implied: Expected shelf life confirmed) |
| Storage and Shelf Life Testing (real time aging) | Packaging is identical to referenced predicate device package/process, implying similar shelf life. | Referenced to predicate. VR Medical device packaging is "identical to referenced predicate device package/process." (Implied: Equivalent shelf life based on identical packaging) |
| Package Related Testing | Packaging maintains integrity and sterility. | VR Medical Syringe tested. "Successfully completed" in-vitro testing demonstrating it "functions according to its specifications." (Implied: Packaging integrity and sterility maintained) |
| ISO 80369-1 requirements | Syringe tip is not compliant with ISO 594 or other small-bore connectors except enteral feeding tips, and will not mate with Luer Lock or Luer slip fittings. It must fit commercially available enteral feeding tubes. | VR Medical Syringe tested. Analytical comparisons and lab testing confirmed: - VR Medical and Phillips Children's device tips are substantially equivalent.- VR Medical device tip is not substantially equivalent to the Wei Gao Group Syringe tip (this is okay as the primary predicate is Phillips).- VR Medical device tip is not compliant with ISO 594 standard's tip requirements.- VR Medical device tip is not compliant with other small-bore connectors listed in ISO 80369-1 except Enteral feeding tips.- The width of the VR Medical syringe tip (OD) is larger than the ISO 594 Female Luer Connector IV connector ID dimensions (4.27-4.31 mm), thus cannot physically fit an ISO 594 rigid female Luer connector.- The VR Medical Oral Enteral Syringe's tip dimensions fit well within the female connector of three different commercially available feeding tubes (MPS Acacia, AMERITUS, NeoMed). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample sizes for each test performed on the VR Medical Oral Enteral Syringe. It generally refers to "VR Medical Syringe tested" without providing the number of units.
The data provenance is primarily from in-vitro testing conducted by VR Medical Technology Co. (the applicant), based in China ("90 Gao Xin Road, ZhouZhuang Kunshan 215325, China"). The substantial equivalence also relies on retrospective data from predicate devices and referenced standards or regulations (e.g., K100099, K072739, ISO standards, 21 CFR sections). The testing described appears to be prospective for the VR Medical Syringe itself, conducted specifically for this 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
The document does not describe the use of medical experts to establish "ground truth" in the traditional sense for these engineering and performance tests. The "ground truth" for these tests is based on:
- Established engineering standards: ISO 7886-1, ISO 7886-2, ISO 80369-1.
- Regulatory precedents: 21 CFR sections for material safety (biocompatibility).
- Performance of predicate devices: K100099 serves as a benchmark for substantial equivalence.
- Analytical engineering evaluations: These would typically be performed by engineers with expertise in medical device design and testing. No specific number or qualifications are provided for these engineers.
4. Adjudication Method for the Test Set
No explicit adjudication method (like 2+1 or 3+1 consensus) is described, as the testing involves objective engineering measurements and comparisons against standards and predicate device performance rather than subjective interpretations by human readers.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done. This type of study is typically relevant for interpretative devices (e.g., imaging devices where human readers interpret results) to assess reader performance with and without AI assistance. The VR Medical Oral Enteral Syringe is a mechanical device, not an interpretative one, so an MRMC study is not applicable.
6. Standalone (Algorithm Only Without Human-In-The-Loop Performance) Study
No standalone performance study of an algorithm was conducted. This device is a physical product, not a software algorithm.
7. Type of Ground Truth Used
The ground truth used for these tests is:
- Technical Specifications/Standards: ISO standards for syringe performance (e.g., volume accuracy, general requirements), ISO 80369-1 for small-bore connectors.
- Predicate Device Performance: The performance characteristics of the Phillips Children's Medical Venture Oral/Enteral Syringe (K100099) served as a benchmark for many tests, establishing a baseline of "safe and effective" performance.
- Regulatory Compliance: Specifically, 21 CFR for material biocompatibility and FDA's previous conclusions on colorants.
- Compatibility with Commercially Available Products: The ability of the syringe's tip to fit into existing enteral feeding tubes on the market.
8. Sample Size for the Training Set
Not applicable. This device is a physical medical device, not an AI or machine learning algorithm, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
{0}------------------------------------------------
K2110853
Page 1 of 4
| SECTION 5 | 510(k) SUMMARY | MAY 13 2011 |
|---|---|---|
| Date of Submission: | May 12, 2011 | |
| Official Contact: | Lee Pan, Ph.D.CEO, VR Medical Technology Co.90 Gao Xin Road, ZhouZhuangKunshan 215325, ChinaT: 86 512 5720 0660C: (US) 6262275727lpan@vrmedical.net | |
| Proprietary Name: | VR Medical Enteral Oral Syringe | |
| Common/ Usual Name: | Gastrointestinal tubes and accessories | |
| Classification Reference: | 21 CFR 876.5980 | |
| Product Code: | FPD | |
| Predicate Devices: | Phillips Children's Medical Venture Oral/ Enteral Syringe(K100099) | |
| Wei Gao Group Medical Syringe (K072739) |
Indication for Use:
The Oral Enteral Syringe is intended for the delivery of liquid medication, formula and breast milk.
Patient Population/ Environment of Use:
The patient population is for neonates and infants. The environments for use are hospital and home environments by trained caregivers only. The Oral Enteral Syringes are disposable and for single patient use only.
Substantial Equivalence
The VR Medical Oral Enteral Syringe is substantially equivalent to commercially available oral enteral syringes. The VR Medical Oral Enteral Syringe's indication for use and FDA Product Code/ Classification Codes are identical to the Phillips Children's Medical Venture Oral/ Enteral Syringe (K100099) and thus is substantially equivalent.
The VR Medical Oral Enteral Syringe components' biocompatible materials are either identical to, or substantially equivalent to, the Phillips Children's Medical Venture Oral/ Enteral Syringe (K100099) and the Wei Gao Group Medical Syringe (K072739).
The VR Medical Oral Enteral Syringe's Technical Characteristics, Packaging and Ethylene Oxide Sterilization Characteristics, as well as Labeling Characteristics, are
{1}------------------------------------------------
Page 2 of 4
substantially equivalent to both the Phillips Children's Medical Venture Oral/ Enteral Syringe (K100099) and the Wei Gao Group Medical Syringe (K072739).
The VR Medical Oral Enteral Syringe and components successfully completed in-vitro testing that demonstrated that the device functions according to its specifications and is thus substantially equivalent in function to both the Phillips Children's Medical Venture Oral/ Enteral Syringe (K100099) and the Wei Gao Group Medical Syringe (K072739).
Design verification tests were performed on the Oral Enteral Syringe as a result of the risk analysis and product requirements. The following tests and analysis were conducted demonstrating the VR Medical Oral Enteral Syringe met the acceptance criteria.
| Specific Test | Test Model | Justification |
|---|---|---|
| Syringe Tip Tensile Test | VR Medical Syringeand Phillips Syringe(K100099) tested | Actual proposed device and predicate(K100099) tested |
| Syringe Liquid Leakage Test | VR Medical Syringetested | Actual proposed device tested |
| Volume Measurement AccuracyTesting (manual. ISO 7886-1 andsyringe pump, ISO 7886-2) | VR Medical Syringeand Phillips Syringe(K100099) tested | Actual proposed device and predicate(K100099) tested |
| ISO 7886-1 requirements(excluding tip requirements) | VR Medical Syringetested | Actual proposed device tested |
| Size and Material Inspection | VR Medical Syringetested | Actual proposed device tested |
| Biocompatibility Analysis(Syringe materials) | Referenced topredicate | VR Medical device component materials andprocesses identical to referenced predicate(except orange colorant) |
| Biocompatibility Analysis(Orange colorant) | Referenced to 21 CFR178.3297 | These colorants are cleared (or FDA listed) asfood contact substance |
| Sterilization Related Testing | VR Medical Syringetested | Actual proposed device tested |
| Storage and Shelf Life Testing(accelerated-time aging) | VR Medical Syringetested | Actual proposed package and device tested |
| Storage and Shelf Life Testing(real time aging) | Referenced topredicate | VR Medical device packaging identical toreferenced predicate device package/ process |
| Package Related Testing | VR Medical Syringetested | Actual proposed device tested |
| ISO 80369-1 requirements | VR Medical Syringetested | Actual proposed device tested |
Device Description
The Oral Enteral Syringe is intended as a single use, sterile, device for the delivery of liquid medication, formula and breast milk. The Oral Enteral Syringe comes in multiple volume sizes ranging from 1cc to 60cc [e.g., 1cc, 3cc, 10cc, 20cc, 30cc and 60cc]. The VR Medical Oral Enteral Syringe, as well as the predicate device [Phillips Children's Medical Venture Oral/ Enteral Syringe (K100099)] has 'labeling' on the device itself, i.e., highly visible labeling on side of barrel marked ORAL/ENTERAL ONLY, and the utilization of bright orange colorant for the plunger and cap that easily distinguishes the syringe for feeding and not other uses.
{2}------------------------------------------------
K110853 page 3 of 4
The Oral Enteral Syringe incorporates safety connectors which eliminate the risk of IV administration through the feeding tube, i.e., the safety connectors will not mate with Luer Lock or Leur slip fittings. Analytical comparisons and lab testing of the VR Medical device tip's dimensions to the others devices listed in the ISO 80369-1 allow the conclusions that:
- The VR Medical and Phillips Children's device tips are substantially equivalent. .
- The VR Medical device tip is not substantially equivalent to the Wei Gao Group ● Syringe tip
- The VR Medical device tip is not compliant with the ISO 594 standard's tip . requirements.
- The VR Medical device tip is not compliant with other small-bore connectors . listed in the ISO 80369-1 except Enteral feeding tips.
Syringe tip analytical engineering evaluation again demonstrates that the width (outside diameter, OD) of the VR Medical syringe tip is larger than the ISO 594 Female Luer Connector standard IV connector inner diameter (ID) dimensions [4.27-4.31 mm]. Thus, the VR Medical syringe tip cannot physically fit into an ISO 594 rigid female Leur connector and does not conform to the ISO 594 standard's requirements.
While VR Medical has searched for an applicable standard for the dimensions of feeding tubes' rigid female connectors, none were found. However, in order for hospitals to purchase feeding tubes from a variety of manufactures and be assured that these different devices will all be compatible with the hospitals' existing equipment, the industry utilizes the same female connector ID dimensions, e.g., a 'common use'. To ensure these 'common use' dimensions. the VR Medical syringe was evaluated against three different commercially available feeding tubes in relation to their rigid female connectors.
These analyses demonstrated that the VR Medical Oral Enteral Syringe's tip dimensions fit well within the female connector of commercially available feeding tubes, e.g., the MPS Acacia Enteral Feeding Tube's female connector (K080328), the AMERITUS Enteral Feeding Tube's female connector (K100526) and the NeoMed Enteral Only Extension Set's female connector (K100288).
The VR Medical Oral Enteral Syringe design incorporates various components, e.g., barrels, plungers, caps and end caps, which are comprised of commercially available materials with demonstrated acceptable biocompatibility, as listed below:
{3}------------------------------------------------
K//0853
| Syringe Component | Component Material |
|---|---|
| Barrel Material | Polypropylene 370Y |
| Piston Material | Thermoplastic Elastomer IR307 |
| Plunger Material | Polypropylene 370Y |
| Plunger Colorant | Orange Pigment* |
| Cap Material | Polypropylene 370Y |
| Cap Colorant | Orange Pigment (P21024) |
| Barrel Gradation Ink | PPE Ink Series(on outside of barrel) |
| Nozzle Material | Polypropylene 370Y (molded with barrel) |
| * It is important to note that the US FDA has already concluded in 21 CFR 178.3297 that these colorants are cleared (or FDA listed) as food contact substance, e.g., acceptable use as an indirect food additive for use as a component of materials used in manufacturing, packaging, transporting or holding food. |
The Oral Enteral Syringe does not contain natural rubber latex, DEHP, or BPA.
Conclusion
The conclusions drawn from the actual conducted (as well as the referenced) analytical engineering evaluations, the nonclinical tests and the commercial use of similar and predicate devices, demonstrate that the VR Medical Oral Enteral Syringe is as safe, as effective and performs at least as safely and effectively as the legally marketed (predicate) devices.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G604 Silver Spring, MD 20993-0002
VR Medical Technology Co. c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLD 1394 25" Street NW BUFFALO MN 55313
MAY 1 3 2011
Re: K110853
Trade/Device Name: VR Medical Oral Enteral Syringe Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: FPD Dated: April 26, 2011 Received: April 27, 2011
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{5}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hubert Lemmer MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a black and white photograph. The image is mostly white with some black spots. The black spots are scattered throughout the image.
INDICATION FOR USE STATEMENT SECTION 4
Page 1 of 1
510(k) Number (if known): K1/08SS
Device Name: VR Medical Oral Enteral Syringe
Indication for Use:
The Oral Enteral Syringe is intended for the delivery of liquid medication, formula and breast milk.
Patient Population/ Environment of Use:
The patient population is for neonates and infants. The environments for use are hospital and home environments by trained caregivers only. The Oral Enteral Syringes are disposable and for single patient use only.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/ OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
. Š
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jalun Lemun
al. and
§ 876.5980 Gastrointestinal tube and accessories.
(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.