K Number
K100288

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-08-03

(183 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NeoMed Enteral Only Extension Set is intended for use as an extension set for nasogastric/oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors which help mitigate the risk of accidental misconnection with an I.V. system to the enteral system or the enteral system to an I.V. system.

Device Description

The NeoMed Enteral Only Extension Set is a sterile disposable for single patient use only device designed to help minimize the potential for inadvertent delivery of enteral feedings through the intravenous route. The device consists of flexible PVC tubing designed to connect existing feeding tubes (nasogastric, oralgastric, etc) to various delivery systems in including pumps and syringes. The set consists of tubing with an enteral connector (catheter tip) and an oral syringe connector not compatible with IV tubing or stopcocks. The NeoMed Enteral Only Extension Set provides an orange stripe for easy quick identification of enteral feeding lines as well as an "Enteral Only" tag and slide clamp to provide the additional safety assurance for connection errors.

AI/ML Overview

The provided text describes the NeoMed Enteral Only Extension Set, a Class II medical device. However, it does not contain the information needed to fill out a table of acceptance criteria, detailed device performance, or a comprehensive study description as requested.

The document is a 510(k) summary, which focuses on establishing substantial equivalence to a predicate device rather than presenting detailed clinical study results or specific performance criteria with numerical targets.

Here's a breakdown of what can be extracted and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (Target)Reported Device Performance (Result)
Mechanical TestingNot specified quantitativelyPassed
TensileNot specified quantitativelyPassed
StrainNot specified quantitativelyPassed
ElongationNot specified quantitativelyPassed
Physical PropertiesNot specified quantitativelyPassed
Dimensional VerificationNot specified quantitativelyPassed
Flow RateNot specified quantitativelyPassed
Hub CompatibilityNot specified quantitativelyPassed
Biocompatibility TestingPassed per ISO 10993-1Passed
CytotoxicityNot specified quantitativelyPassed
SensitizationNot specified quantitativelyPassed
IrritationNot specified quantitativelyPassed
Packaging ChallengesNot specified quantitativelyPassed
Package IntegrityNot specified quantitativelyPassed
Product StabilityNot specified quantitativelyPassed

Explanation of Table: The document states that "Design verification functional test results demonstrate that the NeoMed Enteral Only Extension Set performs its intended use." It then lists the types of tests conducted: "mechanical: tensile, strain, elongation and physical/properties testing comprised of dimensional verification, flow rate, and hub compatibility. Packaging challenges for package integrity and product stability were conducted with aging studies. All materials have been evaluated in accordance with ISO 10993-1: Biological Evaluation of Medical Devices – part 1: Evaluation and Testing and the three tests conducted - cytotoxicity, sensitization, and irritation pass."

*Crucially, the specific quantitative acceptance criteria (e.g., "tensile strength > X N") and the numerical results (e.g., "tensile strength = Y N") for these tests are NOT provided in this document. The document only states that the tests were conducted and "pass" or "demonstrate that the device performs its intended use."


2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified. The document refers to "functional tests conducted" but does not give the number of units tested for each type of performance test.
  • Data Provenance: Not specified (e.g., country of origin of data). The testing appears to be internal validation by the manufacturer, NeoMed, Inc., or its contracted labs.
  • Retrospective or Prospective: N/A for this type of bench/functional testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This document describes bench and biocompatibility testing for a medical device (an extension set), not a diagnostic or prognostic device that requires expert-established ground truth from patient data. The "ground truth" for these tests would be the established engineering standards or regulatory requirements.

4. Adjudication method for the test set

  • Not Applicable. As above, this is laboratory and bench testing, not a clinical study involving human readers or subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No such study was done. This device is an enteral extension set, a physical medical device, not an AI-powered diagnostic or decision support system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Engineering Standards and Regulatory Guidelines. For the functional and biocompatibility tests, the "ground truth" is adherence to established industry standards (e.g., ISO 10993-1 for biocompatibility) and the device's design specifications.

8. The sample size for the training set

  • Not Applicable. There is no "training set" as this device does not involve machine learning or AI.

9. How the ground truth for the training set was established

  • Not Applicable. As above, no training set.

{0}------------------------------------------------

K100288 page 1 of 2

NEOMED
incorporated

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

DATEAugust 2, 2010
APPLICANTNeoMed100 Londonderry CtSuite 112Woodstock, GA 30188Tony Lair, PresidentTel: 888-876-2225e-mail: lair1@concentric.net
OFFICIALCORRESPONDENTPenny Northcutt, RAC, CQARegulatory Consultant for NeoMed, Inc.REGSolutions, LLCTel: 678-428-6978Fax: 678-513-0937e-mail: pennynorthcutt@theregsolutions.com
TRADE NAMENeoMed Enteral Only Extension Set
CLASSIFICATIONNAMETubes, Gastrointestinal (and accessories)
DEVICECLASSIFICATIONAND PRODUCTCODEClass II per 21 CFR §876.5980Gastroenterology/UrologyProduct Code: KNT
PREDICATEDEVICE NAMERespironics Enteral Only Extension Set, K082654

SUBSTANTIAL EQUIVALENCE:

The NeoMed Enteral Only Extension Set is substantially equivalent to the Respironics Enteral Only Extension Set cleared under K082654.

Both devices have similar indications for use, materials, product design, and method of operation. Bench testing has demonstrated that the NeoMed Enteral Only Extension Set is functionally equivalent to the proposed predicate extension set. Any minor differences have no impact affect safety or effectiveness of the NeoMed Enteral Only Extension Set.

{1}------------------------------------------------

K1002 88

Page 2 of 2

NEÖMED
incorporated

DESCRIPTION OF THE DEVICE:

The NeoMed Enteral Only Extension Set is a sterile disposable for single patient use only device designed to help minimize the potential for inadvertent delivery of enteral feedings through the intravenous route. The device consists of flexible PVC tubing designed to connect existing feeding tubes (nasogastric, oralgastric, etc) to various delivery systems in including pumps and syringes. The set consists of tubing with an enteral connector (catheter tip) and an oral syringe connector not compatible with IV tubing or stopcocks. The NeoMed Enteral Only Extension Set provides an orange stripe for easy quick identification of enteral feeding lines as well as an "Enteral Only" tag and slide clamp to provide the additional safety assurance for connection errors.

INTENDED USE/INDICATIONS FOR USE:

The NeoMed Enteral Only Extension Set is intended for use as an extension set for nasogastric/oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors which help mitigate the risk of accidental connection of an I.V. system to the enteral system or the enteral system to an I.V. system.

This device is indicated for use in neonatal and pediatric patients in connection with an enteral feeding tube to provide nutrition via nasal or oral gastric placements.

PERFORMANCE DATA:

The NeoMed Enteral Only Extension Set materials have a long history of use in extension set and feeding tube manufacturing and are biocompatible. Design verification functional test results demonstrate that the NeoMed Enteral Only Extension Set performs its intended use.

The functional tests conducted were mechanical: tensile, strain, elongation and physical/properties testing comprised of dimensional verification, flow rate, and hub compatibility. Packaging challenges for package integrity and product stability were conducted with aging studies. All materials have been evaluated in accordance with ISO 10993-1: Biological Evaluation of Medical Devices – part 1: Evaluation and Testing and the three tests conducted - cytotoxicity, sensitization, and irritation pass,

CONCLUSION:

Based on the performance testing, it can be concluded that the NeoMed Enteral Only Extension Set is equivalent to the Respironics Enteral Only Extension Set predicate with respect to intended use, materials, design, and technological characteristics,

Traditional 510(k) NeoMed, Inc.

NeoMed Enteral Only Extension Set

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo features a stylized depiction of an eagle with three lines representing its wings. The words "DEPARTMENT OF HEALTH & HU." are arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G60 Silver Spring, MD 20993-0002

AUG 3 -

2010

Neomed, Inc. c/o Ms. Penny Northcutt Executive Director REGSolutions, LLC 717 Lakeglen Drive SUWANEE GA 30024

· Re: K100288

Trade/Device Name: Neomed enteral only extension set Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: KNT Dated: July 28, 2010 Received: July 29, 2010

Dear Ms. Northcutt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in intestode commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drive and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listion of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{3}------------------------------------------------

Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Hubert Lehman MD

erbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K100288

Indications for Use

AUG 3- 2010

K100288

Device Name: Neomed Enteral Only Extension Set

Indications For Use: The NeoMed Enteral Only Extension Set is intended for use as an extension set for nasogastric/oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors which help mitigate the risk of accidental misconnection with an I.V. system to the enteral system or the enteral system to an I.V. system.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .

Concurrence of CDRH, Office of Device Evaluation (ODE)

Heled Remu

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K100288

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.