K Number
K110521
Manufacturer
Date Cleared
2011-04-19

(55 days)

Product Code
Regulation Number
870.2300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EverOn Central Display Station (CDS) is intended to provide secondary display of the information as displayed on multiple individual bed side monitoring units, on a central remote screen. The EverOn CDS is not intended to replace any part of the bed-side patient monitoring procedures. The system can be used in hospitals or hospital type and clinic environment.

Device Description

The EverOn Central Display Station (CDS) is intended to communicate with multiple EverOn bed-side monitoring devices and remotely display the information as displayed on multiple individual bed side monitoring units, on a central screen. The communication can be performed either via standard wired or via wireless LAN communication.

The transmitted information from Bed-side to CDS includes alert information and physiological parameters. The CDS can also format the alert information as obtained from bed-side units into a message that can be transmitted to external devices that can communicated with the CDS via standard TCP/IP port. Data and report files (.CSV and .PDF) generated at the bed-side units can be retrieved via the CDS by the user and can be downloaded and / or sent for printing from the CDS. The CDS includes standard hardware (PC, communication and IT hardware). EarlySense develops the application software which is used on the system's PC computer (Central Display Station).

AI/ML Overview

The provided document is a 510(k) summary for the EverOn Central Display Station (CDS) with wireless communication capability. This document focuses on demonstrating substantial equivalence to a predicate device and certifying compliance with relevant standards, rather than providing detailed performance data from a clinical study with acceptance criteria in the typical sense for a diagnostic algorithm.

Therefore, many of the requested points regarding acceptance criteria, sample sizes, ground truth establishment, expert qualifications, and specific study types (MRMC, standalone) are not applicable or not explicitly detailed in this type of regulatory submission.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

This document does not present a table of acceptance criteria and reported device performance in the manner typically seen for diagnostic algorithms. Instead, the "performance data" section focuses on demonstrating the safety and effectiveness of the modification (wireless communication) to an already cleared device and showing compliance with established standards.

The acceptance criteria are implicitly those of the standards met and the successful completion of verification and validation testing, hazard analysis, and full load bench testing.

Acceptance Criteria (Implicit)Reported Device Performance (Summary)
Compliance with EN/IEC 60601-1-4 (2000) (Safety)System reviewed for safety; no new safety issues raised.
Compliance with ISO 14971-1 (2007) (Risk Management)Hazard analysis performed, including risk analysis; specific solutions implemented.
Compliance with FDA Guidance: "General Principles of Software Validation" (11/1/02)Comprehensive software verification and validation testing performed.
Compliance with IEC 62304:2006 (Software Development Life Cycle)Software developed according to a defined life cycle.
Compliance with IEC 60812 Ed.2:2006 (FMEA)FMEA procedure used for failure modes and effects analysis.
Correct communication between bed-side and CDS with wireless (Performance)Full load bench testing performed to establish correct communication and performance with wireless.
Substantial Equivalence to Predicate Device (Overall)Believed to be substantially equivalent to the cleared EverOn CDS without wireless, without raising new safety/effectiveness issues.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated. The full load bench testing implies a robust simulation or real-world test, but the number of bed-side units or data points is not
    quantified.
  • Data Provenance: Not applicable in the context of a clinical test set. The testing was conducted by EarlySense Ltd.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable. The performance evaluation was engineering-focused (software verification, bench testing, hazard analysis) rather than clinical. There's no mention of experts establishing clinical ground truth for a test set.

4. Adjudication Method for the Test Set

Not applicable. This testing did not involve human interpretation or adjudication in the context of clinical ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This type of study is not relevant for a device like the EverOn CDS, which is a display and communication system, not a diagnostic algorithm that assists human readers in interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device itself is a display system, not a standalone diagnostic algorithm. The performance testing described (software verification, bench testing) could be considered "standalone" in the sense that it evaluates the system's function independently without continuous human interaction during the test, but it is not an AI algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance evaluation was defined by:

  • Engineering specifications and requirements (for software verification and validation).
  • Expected communication protocols and display accuracy (for bench testing).
  • Industry standards for safety, risk management, and software development (EN/IEC 60601-1-4, ISO 14971-1, IEC 62304, IEC 60812).

8. The Sample Size for the Training Set

Not applicable. The EverOn CDS is a software-driven display and communication system, not an AI/Machine Learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set was used.

{0}------------------------------------------------

K110521
page 1 of 3

APR 1 9 2011

510(K) SUMMARY

EarlySense Ltd.

EverOn Central Display System (CDS)

7.1.1 Applicant's Name:

EarlySense Ltd. 12 Tzvi st. Ramat Gan 52504, Israel Tel: +972-3-7522330 Fax: +972-3-7522340

7.1.2 Contact Person:

Dalia Argaman EarlySense Ltd. 12 Tzvi Street. Ramat Gan 52504, Israel Tel: +972-3-752-2330 Fax: +972-3-752-2340 Mail: Dalia.Argaman@EarlySense.com

7.1.3 Date Prepared:

Feb. 15, 2011

7.1.4 Trade Name:

EverOn Central Display Station

7.1.5 Classification Name:

System Network and communication, physiological monitors Medical Specialty / Panel: Cardiovascular Product Code: MSX Class: II Regulation Number: 870.2300

{1}------------------------------------------------

7.1.7 Predicate Devices

The EverOn Central Display Station (CDS) is substantially equivalent to the following:

    1. Masimo Safety Net Patient (PSN) system, K071047 (Masimo Corporation), Regulation number: 21 CFR 870.2300, System Network and communication, Physiological Monitors; Product code: MSX
    1. EverOn Central Display Station (CDS) K100376 (EarlySense Ltd) Regulation number: 21 CFR 870.2300, System Network and communication, Physiological Monitors; Product code: MSX

7.1.8 Device Description:

The EverOn Central Display Station (CDS) is intended to communicate with multiple EverOn bed-side monitoring devices and remotely display the information as displayed on multiple individual bed side monitoring units, on a central screen. The communication can be performed either via standard wired or via wireless LAN communication.

The transmitted information from Bed-side to CDS includes alert information and physiological parameters. The CDS can also format the alert information as obtained from bed-side units into a message that can be transmitted to external devices that can communicated with the CDS via standard TCP/IP port. Data and report files (.CSV and .PDF) generated at the bed-side units can be retrieved via the CDS by the user and can be downloaded and / or sent for printing from the CDS. The CDS includes standard hardware (PC, communication and IT hardware). EarlySense develops the application software which is used on the system's PC computer (Central Display Station).

7.1.9 Intended Use:

The EverOn Central Display Station (CDS) is intended to provide secondary display of the information as displayed on multiple individual bed side monitoring units, on a central remote screen. The EverOn CDS is not intended to replace any part of the bed-side patient monitoring procedures. The system can be used in hospitals or hospital type and clinic environment.

7.1.10 Performance Standards:

The EverOn Central Display Station follows and complies with the following standards:

{2}------------------------------------------------

  • Programmable electrical medical system Requirements for safety. EN/IEC 60601-1-4 (2000)
  • 트 ANSI / AAMI/ ISO - Application of risk management to medical devices. ISO 14971-1 (2007)
  • General Principles of Software Validation; Final Guidance for 트 Industry and FDA Staff, FDA, CDRH, 11/1/02
  • 해 IEC 62304:2006 Software Development Life Cycle for medical devices.
  • IEC 60812 Ed.2:2006. Analysis technique for system reliability - ■ Procedure for failure modes and effects analysis (FMEA)

7.1.12 Performance Data & Substantial Equivalence

The EverOn CDS with wireless communication capability is substantially equivalent in all aspects e.g., technological characteristics, mode of operation, performance characteristics, intended use, etc., to the Masimo Patient Safety Net (PSN) (K071047) and is identical in all aspects to the EverOn Central Display Station (K100376).

The modification:

The CDS software was modified to include the capability of wireless communication. The modified System was subjected to the following:

  • 제 Hazard analysis was performed, including risk analysis, level and solutions performed for the entire system and software.
  • The modified software went through comprehensive software verification and validation testing.
  • Full load bench testing was performed to establish the correct . communication between bed-side and CDS and establish the correct performance of the system with wireless communication.

Based on the design verification and validation processes performed as a result of risk assessment and results of the testing performed, EarlySense Ltd. believes that the CDS with wireless communication capability is substantially equivalent to the cleared EverOn CDS where communication utilized only standard wired LAN, without raising new safety and/or effectiveness issues.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. The text is in all caps and is smaller than the eagle symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

EarlySense Ltd. c/o Ms. Dalia Argaman 12 Tzvi Street. Ramat Gan 52504, Israel

APR 1 9 2011

Re: K110521

Device Name: EverOn Central Display Station (CDS) with Wireless Communication Regulation Number: 21 CFR 870.2300 Regulation Name: Physiological monitors network and communication system Regulatory Class: Class II (Two) Product Codes: MSX Dated: March 22, 2011 Received: March 30, 2011

Dear Ms. Argaman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 – Ms. Dalia Argaman

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Cor Bram D. Zuckerman, M.D. C Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: EverOn Central Display Station

Indications for Use:

The EverOn Central Display Station (CDS) is intended to provide secondary display of the information as displayed on multiple individual bed side monitoring units, on a central remote screen. The EverOn CDS is not intended to replace any part of the bed-side patient monitoring procedures. The system can be used in hospitals or hospital type and clinic environment.

Prescription Use _ V (Part 21 C.F.R. 801 Subpart D)

AND/OR

Over-The-Counter Use (Part 21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Wision Sign-Off)

ivision of Čardiovascular Devices

510(k) Number k//0a52/

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).