K Number
K103736
Manufacturer
Date Cleared
2011-06-10

(170 days)

Product Code
Regulation Number
880.5860
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DMC Medical piston type syringes are single use syringes intended for injecting fluids into or withdrawing fluids from the body.

Device Description

DMC Medical Single-Use Polycarbonate Syringe is offered in the same configurations as the predicate. It is made from a calibrated hollow barrel, and a moveable piston with a plunger tip at the end of the piston. The piston shaft does not contact any part of the fluid path, rather the tip which attaches to the shaft and the barrel are fluid path components. Individual components are made from properly tested materials included in this submission. Additionally, there is a small amount of lubricant for moving the piston shaft smoothly along the inside surface of the barrel. The connector is a universal luer threaded style connector. The device is used in general medicine in clinical, hospital, or other settings of healthcare professionals.

AI/ML Overview

The information provided in the document focuses on demonstrating the substantial equivalence of the DMC Medical Single-Use Polycarbonate Syringe to a predicate device (ICU Medical Single-Use Syringe) rather than providing detailed acceptance criteria and a study proving the device meets those criteria in a typical clinical or AI/algorithm performance context.

This regulatory submission (510(k)) is for a medical device (syringe), not a software algorithm or AI model, so the "study" described is primarily non-clinical testing against recognized standards to demonstrate equivalence.

Here's an attempt to structure the available information into the requested format, with the recognition that some sections will be "N/A" or "Not Applicable" given the nature of the device and submission:


Acceptance Criteria and Device Performance for DMC Medical Single-Use Polycarbonate Syringe

The acceptance criteria for the DMC Medical Single-Use Polycarbonate Syringe are primarily based on conformity to the international standard ISO 7886-1 for sterile single-use hypodermic syringes and the biological safety standard ISO 10993-1, as well as demonstration of substantial equivalence to the predicate device, K070856.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Material EquivalenceSyringe Barrel: PolycarbonatePolycarbonate
Plunger Tip: Polyisoprene (predicate) / Elastomer (submitted)Elastomer (considered equivalent in function)
Lubricant: Medical Grade OilMedical Grade Oil
Functional EquivalenceCalibrated Barrel VolumeYES
Sterilization Method: EtOEtO
Operational Characteristics: Identical to predicate deviceOperations are identical to the predicate, allowing accurate fluid measurement via printed external housing in ml.
Safety & Performance StandardsConformity to ISO 7886-1 (Sterile Single-use Hypodermic Syringes)All syringes conform to ISO 7886-1
Sterility Assurance Level (SAL): 10-6SAL has been established to meet the 10-6 level
Conformity to ISO 10993-1 (Biological evaluation of medical devices)All syringes conform to ISO 10993-1, including minimizing residual gases.
Pressure Testing to FailureTested to failure; devices meet healthcare professional needs.
PackagingMinimizing residual gases; discourage re-usePackaged to ensure conformity with ISO 10993-1, minimizing residual gases, and discouraging re-use.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated as a numerical value for a specific test set. The submission mentions "All syringes conform to..." and "DMC Medical syringes are tested to...", implying testing was performed on a representative sample to ensure compliance.
  • Data Provenance: The testing was "Non-clinical Testing" performed by DMC Medical Ltd. to demonstrate conformity to ISO standards and substantial equivalence. No country of origin for a clinical data set is applicable, as this is a device submission based on bench testing and material equivalence.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Not Applicable. This submission is for a physical medical device (syringe) and relies on objective, standardized non-clinical testing (e.g., pressure testing, material analysis, sterility testing) against ISO standards, rather than expert-established ground truth from clinical interpretation.

4. Adjudication Method for the Test Set

  • Not Applicable. As there is no "ground truth" derived from expert consensus or interpretation, an adjudication method for such a ground truth is not relevant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study is relevant for imaging or diagnostic algorithms where human readers interpret cases. This filing is for a physical syringe and does not involve human readers interpreting data.

    • Effect Size of AI Improvement: Not applicable.

6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Not Applicable. This is not an algorithm or AI device. The "standalone performance" is the inherent function and safety of the syringe itself as demonstrated through non-clinical testing.

7. The Type of Ground Truth Used

  • Objective Standards and Physical Testing Results: The "ground truth" for this device is its adherence to established international standards (ISO 7886-1, ISO 10993-1), material specifications, and performance characteristics (e.g., pressure resistance, accurate volume measurement) as verified through non-clinical laboratory testing. The predicate device's established safety and effectiveness also serve as a comparative ground.

8. The Sample Size for the Training Set

  • Not Applicable. This is not an AI/ML device that requires training data. The syringe components and manufacturing process are developed and refined through engineering practices, not machine learning training sets.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. No training set or associated ground truth is relevant for this device.

{0}------------------------------------------------

K103736 Jun 10 2011

DMC MEDICAL Ltd.

Unit 2, Abbey House Shannon Town Centre Shannon, County Clare Ireland 353-61-7148900, Phone 353-61-7148944, Fax TSB / Bryan Wixted Preparation Date: April 6th 2011

Domestic Contact: Charmaine Henderson Phone: 925-299-8702 Fax: 925-299-8702

510(K) Summary for the:

Trade Name:DMC Medical Single-Use Polycarbonate Syringe
Common Name:Syringe, Hypodermic
Classification Name:Piston Syringe, 21 CFR 880.5860, Class II Device, Product Code: FMF

Legally Marketed Predicate Devices for Substantial Equivalence:

*K070856 - ICU Medical Single-Use Syringe

Rationale for SE:

The ICU Medical Single-Use Syringe is a Class II device that is offered pre-sterilized or in bulk for offer to other kit manufacturers. The ICU Medical Single-Use Polycarbonate Syringe is made from identical materials and by the same processes.

Description of Submitted Device:

DMC Medical Single-Use Polycarbonate Syringe is offered in the same configurations as the predicate. It is made from a calibrated hollow barrel, and a moveable piston with a plunger tip at the end of the piston. The piston shaft does not contact any part of the fluid path, rather the tip which attaches to the shaft and the barrel are fluid path components. Individual components are made from properly tested materials included in this submission. Additionally, there is a small amount of lubricant for moving the piston shaft smoothly along the inside surface of the barrel. The connector is a universal luer threaded style connector. The device is used in general medicine in clinical, hospital, or other settings of healthcare professionals.

Intended Uses of the DMC Medical Syringe:

DMC Medical piston type syringes are single use syringes, intended for injecting fluids into or withdrawing fluids from the body.

{1}------------------------------------------------

Technological Characteristics and Substantial Equivalence Table:

Component:ICU Medical, Inc.DMC Medical, Ltd
Syringe Barrel:PolycarbonatePolycarbonate
Plunger Tip:PolyisopreneElastomer
Silicone OilMedical Grade OilMedical Grade Oil
Calibrated Barrel Volume:YESYES
Sterilization method:EtOEtO
51 O(k) ApprovalK070856This submission

The operational characteristics, when compared to the predicate device, are identical. The user connects the syringe via the threaded luer, then manually advancing or withdrawing the plunger internal to the barrel, is able to express or withdraw fluids are measured via the printed external housing of the barrel; measurements are indicated in ml (milliliters). Operation is similar for most all piston syringes whether fitted with a threaded luer end or a slip-fit only end.

Safety and Performance:

DMC Medical syringes conform to the requirements offSO 7886-1, an FDA recognized standard. Additionally, the Sterility Assurance Level, (SAL) has been established to meet the 10- level. The single use syringes are packaged in a way as to ensure conformity with ISO 10993-1, including minimizing residual gases as well as discourage re-use.

Non-clinical Testing:

DMC Medical syringes are tested to ISO 7886-1, as well as pressure testing to failure. All syringes conform to ISO 10993-1, including minimizing residual gases. This testing ensures that the devices meet the needs of the healthcare professional using them. Additionally, the testing demonstrates that the device is as safe, as effective, and performs as well as or better than the predicate device.

Conclusion:

The materials, performance, and operational features of both the submitted device and the predicate device are substantially equivalent and are safe and effective for their intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DMC Medical, Limited C/O Ms. Tracy Best Tracy Best Consulting 944 North Main Street Bountiful, Utah 84010

JUN 1 0 2511

Re: K103736

Trade/Device Name: DMC Medical Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: May 26, 2011 Received: June 2, 2011

Dear Ms. Best:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Best

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for.Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K103736

. .. ..

Device Name: DMC Medical Syringe

Indications For Use: DMC Medical piston type syringes are single use syringes intended for injecting fluids into or withdrawing fluids from the body.

2 Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M. C. A 6/8/4

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

<103736

Page 1 of __ 1__

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).