(136 days)
The GE Lunar Visceral Fat Software option used on GE Lunar DEXA bone densitometer total body scans estimates the visceral adipose tissue (visceral fat) content within the android region in a male or female population between the age of 18 and 90 with a BMI between 18.5 and 40, excluding pregnant women. The content that is estimated is the Visceral Fat Mass and Visceral Fat Volume. The values can be displayed in user defined statistical formats and trends.
The estimated visceral fat content is useful to health care professionals in their management of diseases/conditions where the disease/condition itself, or its treatment, can affect the relative amounts of visceral fat content in the android region. The GE Lunar Visceral Fat Body Composition Software option does not diagnose disease, or recommend treatment regimens, or quantify treatment effectiveness. Only the health care professional can make these judgments. Some of the diseases/conditions for which visceral fat estimation is useful include hypertension, impaired fasting glucose, impaired glucose tolerance, diabetes mellitus, dyslipidemia, and metabolic syndrome.
The Visceral Fat Software option is part of a GE Lunar DXA bone densitometer. It uses data from a DXA total body scan to estimate the visceral adipose tissue (visceral fat), mass and volume within the android region in a male or female population between the age of 18 and 90 with a BMI between 18.5 and 40.
Here's a summary of the acceptance criteria and the study details for the GE Lunar Visceral Fat Software, based on the provided 510(k) Premarket Notification:
Acceptance Criteria and Device Performance
The provided document implies the acceptance criteria were based on demonstrating substantial equivalence to a predicate device (GE Lunar Body Composition Software option, K071570) and showing comparable performance to the accepted research standard for VAT imaging, Computed Tomography (CT), in a clinical trial. However, specific quantitative acceptance criteria (e.g., accuracy thresholds, correlation coefficients) are not explicitly stated in the provided text.
The document states: "A clinical trial was performed to compare dual-energy x-ray absorptiometry (DXA) estimated visceral fat adipose tissue (VAT) with the accepted research standard for VAT imaging, computed tomography (CT), using the Advantage Windows Tissue Volume option (K963345)."
Therefore, the reported device performance is that the "GE Lunar Visceral Fat Software... performs as well as the devices currently on the market," which is supported by the comparison to the CT standard in the clinical trial. Without explicit quantitative results from the trial in this summary, a detailed table with specific reported device performance cannot be generated.
| Acceptance Criteria (Implied) | Reported Device Performance (Implied) |
|---|---|
| Substantial equivalence to predicate device (K071570) | Achieved |
| Comparable performance to CT as the research standard for VAT imaging | Achieved |
| Safe for intended use and conforms to applicable standards | Achieved |
Study Information
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated in the provided text. The document only mentions "A clinical trial was performed."
- Data Provenance: Not explicitly stated in the provided text (e.g., country of origin). The study is described as a "clinical trial," implying prospective data collection, but this is not explicitly confirmed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not explicitly stated.
- Qualifications of Experts: Not explicitly stated.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication Method: Not explicitly stated.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Multi-reader multi-case (MRMC) comparative effectiveness study: No, this type of study was not described. The clinical trial compared the DXA estimated VAT to CT, not human readers with and without AI assistance. The device is a software option that estimates visceral fat, not an assistive AI for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Standalone Study: Yes, a standalone study was done. The clinical trial directly compared the "dual-energy x-ray absorptiometry (DXA) estimated visceral fat adipose tissue (VAT) with the accepted research standard for VAT imaging, computed tomography (CT)." This implies the algorithm's output was compared against the ground truth without human intervention in the estimation process.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Ground Truth Type: The accepted research standard for VAT imaging, Computed Tomography (CT), using the Advantage Windows Tissue Volume option (K963345).
8. The sample size for the training set:
- Training Set Sample Size: Not explicitly stated in the provided text.
9. How the ground truth for the training set was established:
- Ground Truth Establishment for Training Set: Not explicitly stated. The document focuses on the clinical trial used for validation, not the development or training phase of the algorithm.
{0}------------------------------------------------
K103730 Page 1 of 3
GE Medical Systems Lunar 510(k) Premarket Notification GE Lunar Visceral Fat Software
510(K) Summary of Safety and Effectiveness
MAY - 6 2011
Prepared in accordance with 21 CFR Part 807.92.
| Date Prepared: | December 20, 2010 |
|---|---|
| Submitter: | GE Healthcare, (GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC doing business as GE Medical Systems Lunar)3030 Ohmeda DriveMadison, WI 53718 |
| Primary Contact: | Chris PaulikRegulatory Affairs Leader, GE Medical Systems LunarTelephone: 262-548-2010; Fax: 262-546-0704e-mail: Christopher.Paulik@med.ge.com |
| Secondary Contact: | David BlonskiRegulatory Affairs Director, X-RayGE Healthcare, (GE Medical Systems, LLC)Telephone: 262-513-4072; Fax: 262-546-0704e-mail: David.Blonski@ge.com |
DEVICE INDENTIFICATION
| Trade Name: | GE Lunar Visceral Fat Software |
|---|---|
| Common/Usual Name: | Bone Densitometer |
| Classification Name: | Bone Densitometer |
| Class II, KGI, 21CFR 892.1170 |
DEVICE DESCRIPTION:
The Visceral Fat Software option is part of a GE Lunar DXA bone densitometer. It uses data from a DXA total body scan to estimate the visceral adipose tissue (visceral fat), mass and volume within the android region in a male or female population between the age of 18 and 90 with a BMI between 18.5 and 40.
ન્ડન્ટ
Page 14-04-03
{1}------------------------------------------------
K103730 Page 2 of 3
Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular border. The logo is in black and white, with the letters and border appearing in black against a white background. The design is simple and iconic, representing the well-known multinational conglomerate.
GE Medical Systems Lunar 510(k) Premarket Notification GE Lunar Visceral Fat Software
INDICATIONS FOR USE:
The GE Lunar Visceral Fat Software option used on GE Lunar DEXA bone densitometer total body scans estimates the visceral adipose tissue (visceral fat) content within the android region in a male or female population between the age of 18 and 90 with a BMI between 18.5 and 40, excluding pregnant women. The content that is estimated is the Visceral Fat Mass and Visceral Fat Volume. The values can be displayed in user defined statistical formats and trends.
The estimated visceral fat content is useful to health care professionals in their management of diseases/conditions where the disease/condition itself, or its treatment, can affect the relative amounts of visceral fat content in the android region. The GE Lunar Visceral Fat Body Composition Software option does not diagnose disease, or recommend treatment regimens, or quantify treatment effectiveness. Only the health care professional can make these judgments. Some of the diseases/conditions for which visceral fat estimation is useful include hypertension, impaired fasting glucose, impaired glucose tolerance, diabetes mellitus, dyslipidemia, and metabolic syndrome.
TECHNOLOGY
GE Lunar Bone Densitometers use a technique called Dual-energy X-ray Absorptionetry or DXA. DXA measures the attenuation of an x-ray beam comprised of two energy levels after passing the beam through the body of a subject. When the dual-energy x-ray beam passes through soft tissue, without any bone mineral, the tissue composition for the pixel can be measured by calculating the relative attenuation by the fat and lean issue of the two energy levels in the x-ray beam.
The Visceral Fat software algorithm combines direct measurements of fat and lean tissue with modeling information about the distribution of fat in the android region.
A clinical trial was performed to compare dual-energy x-ray absorptiometry (DXA) estimated visceral fat adipose tissue (VAT) with the accepted research standard for VAT imaging, computed tomography (CT), using the Advantage Windows Tissue Volume option (K963345).
COMPARISON WITH PREDICATE DEVICES:
The GE Lunar Visceral Fat Software option for GE Lunar DEXA Bone Densitometers is of comparable type and substantially equivalent to the GE Lunar Body Composition Software option (K071570). It has the same technological characteristics, is comparable in key safety and effectiveness features, it utilizes similar design, construction and materials, and has similar intended uses as the GE Lunar Body Composition Software option.
Image /page/1/Picture/12 description: The image shows the numbers 5 and 3 connected by a horizontal line. The number 5 is on the left and the number 3 is on the right. The numbers are written in a simple, sans-serif font. The image is black and white.
Page 12 or 95 -
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The logo is in black and white, with the letters and the outline of the circle in black and the background in white.
GE Medical Systems Lunar 510(k) Premarket Notification GE Lunar Visceral Fat Software
ADVERSE EFFECTS ON HEALTH:
The device has been evaluated for electrical, mechanical, and radiation safety, and conforms to applicable medical device safety standards, as confirmed by a Nationally Recognized Test Laboratory,
The potential hazards of electrical and mechanical are identified in a risk management summary (hazard analysis) and controlled by:
- System verification and validation to ensure performance to specifications, ● Federal Regulations, and user requirements.
- Adherence and certification to industry and international standards. .
- Compliance to applicable CDRH 21 CFR subchapter J requirements. .
The device is designed and manufactured under the Quality System Regulations of 21 CFR 820.
CONCLUSION:
The GE Lunar Visceral Fat Software is an extension of the currently cleared GE Lunar Body Composition Software (K071570). It does not result in any new potential safety risks, has the same technological characteristics, and performs as well as the devices currently on the market.
After analyzing performance testing on the bench and in a clinical trial it is the conclusion of GE Healthcare that the GE Lunar Visceral Fat Software is substantially equivalent to other marketed devices with similar indications for use and meeting the same standards.
5-4
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines above them, creating a sense of movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
MAY 1 0 2011
Mr. Chris Paulik Regulatory Affairs Leader GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC Doing business as GE Medical Systems Lunar 3030 Ohmeda Drive MADISON WI 53718
Re: K103730
Trade/Device Name: GE Lunar Visceral Fat Software Regulation Number: 21 CFR 892.1170 Regulation Name: Bone densitometer Regulatory Class: II Product Code: KGI Dated: December 20, 2010 Received: December 21, 2010
Dear Mr. Paulik:
This letter corrects our substantially equivalent letter of May 6, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Paulik
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucr1115800.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Production 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters 'GE' intertwined within a circular border. The letters are stylized and appear to be hand-drawn or calligraphic. The logo is black and white, with the letters and border in black and the background in white.
SION -510(k) Premarket Notification GE Lunar Viscoral Fat Software
Indications for Use
510(k) Number (if Known): Device Name: GE Lunar Visceral Fat Software
Indications for Use:
The GB Lunar Visceral Pat Software option used on GE Lunar DEXA bone densitometer
s to al body scans estimates the visceral adjnose tissue (risessed friences of the lotal body scans estimates the visceral adjose tissue (visceral fat) content within the and of the on outhers are visers adjose the age of 18 and 90thin was.
between 18.5 and 40, cxcluding between the age of 18 and 90with a BMI between 18.5 and 40, excluding pregnant women. The content that is estimated is the election is estimated is the Viscently and Visceral Fat Volume. The volume. The content that is estimated is their
statistical formats and trends, statistical formats and trends.
The estimated visceral fat content is useful to health care professionals in their management of discases/conditions where the disease/oondition itself, or its treatment, oan affect the relative amounts of visce)al at content in the android region. The Greatment,
Visceral Fat Body Composition Software ontion dess not discussion. The GI Lunar Visceial Pat Body Composition Software option does not diagnose disease, CD-Library, SB-L
Visceial Pat Body Composition Software option does not diagnose disease, or recommon cannent regimens, or quantify treatment effectiveness. Only the health ears professional can make these judgments. Some of the diseases/conditions for which visceral fat estimation is useful include bypertension, impaired fasting glucose, impaired glucose, impaired glucose tolerance, diabetes mellitus, dyperiension, imparicd fasting glucose,
tolerance, diabetes mellitus, dyslipidemia, and metabolic syndrome.
Prescription Use_ X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CPR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
5/6/11
Division Sign-Off
Office of In Vitro Diagnostic Device Bvaluation and Safety 510(k) K103
Page 1 of 1
March 29, 2011
. . . . . .
§ 892.1170 Bone densitometer.
(a)
Identification. A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.